INION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM

K052444 · Inion , Ltd. · JEY · Sep 20, 2005 · Dental

Device Facts

Record IDK052444
Device NameINION CPS BABY 1.5 BIOABSORBABLE FIXATION SYSTEM
ApplicantInion , Ltd.
Product CodeJEY · Dental
Decision DateSep 20, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.4760
Device ClassClass 2
AttributesTherapeutic, Pediatric

Intended Use

The INION CPS™ BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, mid-face and maxilla. Specific indications: Fractures of the cranium, mid-face and maxilla; Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations); LeFort (I, II, III) osteotomies; Pediatric reconstructive procedures; Orthognathic or reconstructive procedures of the cranium, mid-face, or maxilla; Craniotomy flap fixation.

Device Story

Inion CPS Baby 1.5 Bioabsorbable Fixation System is a line extension adding a large X plate to an existing system. Device consists of bioabsorbable bone plates used to maintain alignment of bone fractures and osteotomies in craniofacial, mid-face, and maxillary regions. Operated by surgeons in clinical settings. Provides mechanical stabilization of bone segments during healing; material degrades over time. Benefits include avoiding secondary removal surgery due to bioabsorbable nature. Device is substantially equivalent to predicate systems in material composition, degradation profile, and mechanical properties.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on performance data and specifications demonstrating that the device's material composition, degradation profile, and mechanical properties are equivalent to the predicate devices.

Technological Characteristics

Bioabsorbable bone fixation plate. Material: Copolymer (composition identical to K010351). Principle: Mechanical stabilization of bone fractures/osteotomies via bioabsorbable fixation. Form factor: Large X plate. Sterilization: Not specified (identical to K010351).

Indications for Use

Indicated for trauma and reconstructive procedures of the craniofacial skeleton, mid-face, and maxilla in pediatric and infant populations, including craniosynostosis, congenital malformations, LeFort osteotomies, and craniotomy flap fixation. Contraindicated for mandibular tumor resection, active/potential infection, limited blood supply, insufficient bone quality/quantity, non-compliant patients, and mandibular use.

Regulatory Classification

Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K052 444 Page 1 of 2 Image /page/0/Picture/2 description: The image shows a logo with the word "INION" in large, bold white letters against a black background. Below the word "INION", there is the phrase "Intelligent Solution" in a smaller, white font. The logo appears to be for a company or product named INION, emphasizing its focus on intelligent solutions. SEP 2 0 2005 SPECIAL 510(k) SUMMARY for the line extension of INION CPS ™ Baby 1.5 Bioabsorbable Fixation System (K010351) MANUFACTURER Inion Ltd., Lääkärinkatu 2, FIN-33520 Tampere, FINLAND #### Contact Person Hanna Marttila, Regulatory Affairs Director Lääkärinkatu 2, FIN-33520 Tampere Phone: +358 3 2306 600 Fax: +358 3 2306 691 hanna.marttila@inion.com ## DEVICE NAME Trade name: Large X plate Common/Usual Name: Bone plate #### ESTABLISHMENT REGISTRATION NUMBER 9710629 ## DEVICE CLASSIFICATION AND PRODUCT CODE Classification Panel: Dental Regulation Number: 21 CFR 872.4760 Regulation Name: Bone plate Regulatory Class: Class II Product Code: JEY #### PREDICATE DEVICE Howmedica Leibinger Resorbable Fixation System (K982531) #### CONFORMANCE WITH PERFORMANCE STANDARDS No applicable mandatory performance standards exist for this device. Compliance to voluntary consensus standards is listed in the application. #### THE REASON FOR SCPECIAL 510(k) Additional plate design to Inion CPS™ Baby 1.5 Bioabsorbable Fixation System (K010351). {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for INION Intelligent Solution. The logo is white text on a black background. The word INION is in large, bold letters, and the words Intelligent Solution are in smaller letters below. # DEVICE DESCRIPTION AND PRINCIPLES OF OPERATION The Inion CPS™ Baby 1.5 Bioabsorbable Fixation System is indicated for use in trauma and reconstructive procedures in the craniofacial skeleton, mid-face and maxilla. Specific indications: - . Fractures of the cranium, mid-face and maxilla - Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations) . - LeFort (I, II, III) osteotomies . - . Pediatric reconstructive procedures - Orthognathic or reconstructive procedures of the cranium, mid-face, or maxilla . - . Craniotomy flap fixation. The new large X plate is identical with the other implants in the previously 510(k) cleared Inion CPSTM Baby 1.5 Bioabsorbable Fixation System (K010351) in terms of copolymer composition, intended use and indications for use as described in the device labelling, manufacturing method, sterilization method and packaging solution. Similar design from other manufacturers is currently commercially available and in clinical use. # EQUIVALENCE TO MARKETED PRODUCTS Based on the performance data and specifications presented, it can be concluded that the intended use, material composition and scientific technology, degradation profile and mechanical properties of Inion CPS™ Baby 1.5 Bioabsorbable Fixation system large X plate are substantially equivalent with the other Inion CPS™ Baby 1.5 Bioabsorbable Fixation system (K010351) implants as well as the predicate device Howmedica Leibinger Resorbable Fixation System (K982531). Inion CPS™ Baby 1.5 Bioabsorbable Fixation system large X plate is substantially equivalent to predicate Class II devices used in maintaining accurate alignment of bone fractures and osteotomies in the cranium and mid-face because the differences between Inion CPS™ Baby 1.5 Bioabsorbable Fixation system large X plate and the predicate devices do not raise new questions of safety and effectiveness. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. The text is in all capital letters and is evenly spaced around the logo. SEP 2 0 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Hanna Marttila Regulatory Affairs Director Inion Limited Laakarinkatu 2 Tampere, FINLAND 33520 Re: K052444 R022444 Trade/Device Name: Inion CPS™ Baby 1.5 Bioabsorbable Fixation System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: JEY Dated: August 25, 2005 Received: September 6, 2005 Dear Ms. Marttila: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becales or child the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interstate continered prior to that have been reclassified in accordance with the provisions of Amendinents, or to do rooms and tic Act (Act) that do not require approval of a premarket the roderal I ood, Drug, and Ocentraly, therefore, market the device. subject to the general approval uppressions of the Act. The general controls provisions of the Act include confors providens of an annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 MA), it may of subject to tach adala f Federal Regulations, Title 21, Parts 800 to 898. In your device can oc found in the beach nouncements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 -- Ms. Marttila Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements modifine and Federal statutes and regulations administered by other Federal agencies. or the Act of all , eath all the Act's requirements, including, but not limited to: registration r ou intist contpry wart 807); labeling (21 CFR Part 801); good manufacturing practice alle fishing (21 CF rate of 1), and 10 the quality systems (QS) regulation (21 CFR Part 820); and if requirements as set form in the quarity sion control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will are n Jourse of Substantial equivalence of your device to a premarket notification - 11 - 11 - 11 - 11 - 11 - 12 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), 11 you desire specific at Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Clive Shia-Lian, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K052444 # Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Inion CPS1M Baby 1.5 Bioabsorbable Fixation System Indications for Use: - General indications: The INION CPS™ BABY 1.5 BIOABSORBABLE FIXATION A. SYSTEM is intended for use in trauma and reconstructive procedures in the craniofacial skeleton, mid-face and maxilla. - в. Specific indications: - · Fractures of the cranium, mid-face and maxilla - · Infant craniofacial surgery (i.e. craniosynostosis, congenital malformations) - · LeFort (I, II, III) osteotomies - · Pediatric reconstructive procedures - · Orthognathic or reconstructive procedures of the cranium, mid-face, or maxilla - · Craniotomy flap fixation Contraindications: The INION CPSIM BABY 1.5 BIOABSORBABLE FIXATION SYSTEM is not intended for use in and is contraindicated for mandibular tumor resection; active or potential infection; patient conditions including limited blood supply, insufficient quantity or quality of bone; and where patient cooperation cannot be guaranteed (e.g., alcoholism, drug abuse). DO NOT use in the mandible. V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runne n of Anesthesiology, General Hospital Infection Control, Dental D 510(k) Number Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...