K130025 · Sonendo, Inc. · ELC · Jun 11, 2013 · Dental
Device Facts
Record ID
K130025
Device Name
SONENDO ENDOTHERAPY SYSTEM
Applicant
Sonendo, Inc.
Product Code
ELC · Dental
Decision Date
Jun 11, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Sonendo Endotherapy System is intended to prepare, clean and irrigate 1st and 2nd molar teeth indicated for root canal therapy.
Device Story
The Sonendo Endotherapy System is a dental device used by clinicians to prepare, clean, and irrigate root canals in 1st and 2nd molars. The system consists of a console, foot pedal, and a handpiece with a molar procedure kit. The device operates by generating hydroacoustic waves and fluid motion within the tooth. A water stream flows through a guide tube and contacts fluid inside the tooth at the distal tip; the fluid stream is dispersed and deflected by the distal end plate of the tube, creating hydrodynamics that clean the canal. This process replaces traditional mechanical instrumentation methods. The clinician uses the device to facilitate root canal therapy, potentially improving cleaning efficacy. The system is intended for prescription use in a clinical setting.
Clinical Evidence
Bench testing only. Performance data included simulated use, EMC and electrical safety, thermal safety, hydroacoustics, apical extrusion and pressure, and cleaning efficacy testing on units representative of finished devices.
Technological Characteristics
System components: console, foot pedal, handpiece, and irrigation reservoirs. Principle of operation: hydroacoustic wave generation via fluid stream deflection at a distal end plate. Energy source: electrical (console). Connectivity: not specified. Sterilization: not specified. Software: not specified. The device is a Class II ultrasonic scaler (21 CFR 872.4850).
Indications for Use
Indicated for patients requiring root canal therapy on 1st and 2nd molar teeth.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
K133752 — SONENDO OMNICLEAN ENDOTHERPAY SYSTEM CONSOLE, SONENDO OMNICLEAN MOLAR PROCEDURE KIT · Sonendo, Inc. · May 15, 2014
K153157 — Sonendo Gentle Wave System · Sonendo, Inc. · Jan 29, 2016
K190359 — Sonendo GentleWave System · Sonendo, Inc. · Mar 16, 2019
K160905 — Sonendo GentleWave System · Sonendo, Inc. · Jun 3, 2016
K203302 — GentleWave X · Sonendo, Inc. · Feb 5, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
Section 6
K130025
## 510(k) Summary
# JUN 1 1 2013
#### 6. 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Sonendo, Inc. | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------|--|
| DATE PREPARED: | December 29, 2012 | |
| CONTACT PERSON: | Dan W. Miller<br>Sonendo, Inc.<br>26061 Merit Circle, Suite 101<br>Laguna Hills, CA 92653<br>Phone: (949) 766.3636 x 544 | |
| TRADE NAME: | Sonendo Endotherapy System | |
| COMMON NAME: | Sonic Cleaning and Irrigation System | |
| CLASSIFICATION<br>NAME: | Ultrasonic Scaler | |
| DEVICE<br>CLASSIFICATION: | Class 2, per 21 CFR 872.4850 | |
| PRODUCT CODE | ELC | |
| PREDICATE DEVICES: | EMS Piezon Master 700 (K093000)<br>Sonic Air MM 1500+ (MID) (K081268) | |
#### Substantially Equivalent To:
The Sonendo Endotherapy System is substantially equivalent in intended use, principal of operation and technological characteristics to the EMS Piezon Master 700 (K093000) and the Sonic Air MM 1500+ (MID) (K081268).
#### Description of the Device Subject to Premarket Notification:
The Sonendo Endotherapy System is a medical device intended to prepare, clean and irrigate root canals. The Sonendo Endotherapy System is comprised of a Console, Foot Pedal and Molar Procedure Kit with a Handpiece.
#### Indication for Use:
The Sonendo Endotherapy System is intended to prepare, clean and irrigate 1st and 2nd molar teeth indicated for root canal therapy.
### Technical Characteristics:
The Sonendo Endotherapy System has similar physical and technical characteristics to the predicate devices. These characteristics are tabulated below:
Sonendo, Inc. Sonendo Endotherapy System
Page 11 of 85 Deficiencies Response 130610
{1}------------------------------------------------
| Characteristics | Sonendo Endotherapy<br>System | EMS Piezon Master 700<br>(K093000) | Sonic Air MM 1500 +<br>(MID) (K081268) |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Function | Preparation cleaning and<br>irrigation of root canals | Various, including<br>preparation, cleaning and<br>irrigation of root canals | Preparation cleaning and<br>irrigation of root canals |
| Principle of<br>Operation | Generation of<br>hydroacoustic waves and<br>fluid motion.<br>The tip of the device is<br>placed inside the tooth<br>during cleaning.<br>Hydroacoustics is created<br>by the water stream<br>flowing through the guide<br>tube and coming into<br>contact with the fluid<br>inside the tooth at the distal<br>tip.<br>The fluid stream is<br>dispersed and deflected by<br>the distal end plate of the<br>tube creating<br>hydrodynamics (fluid<br>motion) within the tooth. | Generation of<br>hydroacoustic waves and<br>fluid motion.<br>The tip of the device is<br>placed inside the tooth<br>during cleaning.<br>A piezoelectric element is<br>used to create vibration in<br>the tip of the device<br>creating hydrodynamics<br>(fluid motion) within the<br>tooth. | Generation of<br>hydroacoustic waves and<br>fluid motion.<br>The tip of the device is<br>placed inside the tooth<br>during cleaning.<br>Air is used to create<br>vibration in the tip of the<br>device creating<br>hydrodynamics (fluid<br>motion) within the tooth.<br>The tip of the device<br>discontinues moving if it<br>comes in contact with the<br>walls of a canal. |
| Treatment Site | Root canal | Various, including Root<br>canal | Root canal |
| Components | Control Unit<br>Irrigation reservoirs<br>Foot pedal<br>Handpiece<br>Accessories | Control Unit<br>Irrigation reservoirs<br>Foot pedal<br>Handpiece<br>Instruments | Handpiece<br>Instruments |
## Performance Data:
All necessary performance testing has been conducted for the Sonendo Endotherapy System to assure substantial equivalence to the predicate devices and demonstrate the devices perform as intended. All testing was performed on test units representative of finished devices. Testing included:
- . Simulated Use
- EMC and Electrical Safety .
- Thermal Safety .
- Hydroacoustics .
- Apical Extrusion and Pressure ●
- Cleaning .
{2}------------------------------------------------
## Section 6
## Basis for Determination of Substantial Equivalence:
Upon reviewing the safety and efficacy information provided in this submission and comparing intended use, principle of operation and overall technological characteristics, the Sonendo Endotherapy System is determined by Sonendo, Inc., to be substantially equivalent to existing legally marketed devices.
Sonendo, Inc. Sonendo Endotherapy System
Page 13 of 85 K130025 Deficiencies Response 130610
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## June 11, 2013
Mr. Dan W. Miller Sonendo, Incorporated 26061 Merit Circle, Suite 101 LAGUNA HILLS CA 92653
Re: K130025
Trade/Device Name: Sonendo Endotherapy System Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: May 7, 2013 Received: May 16, 2013
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic "Act"(Act) that do not require approval of a premarket approval application (PMA): You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Miller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. '
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
5. Indications for Use Statement
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): _ K |30 025
Device Name: Sonendo Endotherapy System
Indications for Use:
The Sonendo Endotherapy System is intended to prepare, clean and irrigate 15 and 2nd molar teeth indicated for root canal therapy.
AND/OR
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew I. Steen
2013.06.11 11:50:37 -04'00'
Page _________________________________________________________________________________________________________________________________________________________________________
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division of Anesthesions 37.
Infection Control, Dental Devices
K130525 510(k) Number:_
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.