K160905 · Sonendo, Inc. · ELC · Jun 3, 2016 · Dental
Device Facts
Record ID
K160905
Device Name
Sonendo GentleWave System
Applicant
Sonendo, Inc.
Product Code
ELC · Dental
Decision Date
Jun 3, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sonendo GentleWave(R) System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave(R) Molar Handpiece, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo GentleWave(R) Anterior/Premolar Handpiece, the System is indicated for and premolar teeth.
Device Story
System prepares, cleans, and irrigates root canals; utilizes hydroacoustic waves and fluid motion. Comprised of console, irrigation reservoirs, foot pedal, and disposable single-use handpiece. Clinician places handpiece tip inside tooth; water stream flows through guide tube, contacts fluid at distal tip, and is deflected by end plate to create hydrodynamics within tooth. Software-controlled interface allows clinician to select treatment fluid concentrations (default or three decreased settings). Used in dental clinic by dental professionals. Output is cleaning/irrigation of root canal; enables clinician-determined fluid concentration based on clinical evaluation of treatment site; benefits patient by facilitating root canal therapy.
Clinical Evidence
Bench testing only. Software validation performed in accordance with ANSI/AAMI/IEC 62304:2006. Risk analysis conducted per ISO 14971:2007. No clinical data presented.
Technological Characteristics
Hydroacoustic wave generation; fluid motion via distal tip deflection. Components: console, irrigation reservoirs, foot pedal, disposable handpiece. Software-controlled user interface for fluid concentration adjustment. Design controls per 21 CFR 820.30.
Indications for Use
Indicated for patients requiring root canal therapy. Compatible with 1st and 2nd molars (using Molar Handpiece) and anterior/premolar teeth (using Anterior/Premolar Handpiece).
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2016
Sonenedo, Inc. Mr. Dan W. Miller Sr. Vice President of Regulatory Affairs, Clinical Affairs, and Quality Assurance 26061 Merit Circle, Suite 102 Laguna Hills, Ca 92653
Re: K160905
Trade/Device Name: Sonendi Gentle Wave System Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: May 2, 2016 Received: May 4, 2016
Dear Mr. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Dan W. Miller
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160905
Device Name Sonendo GentleWave(R) System
Indications for Use (Describe)
The Sonendo GentleWave(R) System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave(R) Molar Handpiece, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo GentleWave(R) Anterior/Premolar Handpiece, the System is indicated for and premolar teeth.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 6. 510(k) K160905 Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Sonendo, Inc. |
|---------------------------|----------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | March 31, 2016 |
| CONTACT PERSON: | Dan W. Miller<br>26061 Merit Circle, Suite 102<br>Laguna Hills, CA 92653<br>Phone: 949.766.3636<br>Fax: 949.305.5201 |
| TRADE NAME: | Sonendo GentleWave® System |
| COMMON NAME: | Sonic Cleaning and Irrigation System |
| CLASSIFICATION<br>NAME: | Ultrasonic Scaler |
| DEVICE<br>CLASSIFICATION: | Class 2, per 21 CFR 872.4850 |
| PRODUCT CODE | ELC |
| PREDICATE DEVICES: | Sonendo GentleWave® System (K153157 |
### Description of the Device Subject to Premarket Notification:
The Sonendo GentleWave® System is a medical device intended to prepare, clean and irrigate root canals. The Sonendo GentleWave® System is comprised of a Console, and a disposable single-use Handpiece. The Handpiece is offered in two versions: a Molar Handpiece which is intended to be used on 1st and 2nd molar teeth and an Anterior/Premolar Handpiece which is intended to be used on anterior and pre-molar teeth.
### Indication for Use:
The Sonendo GentleWave® System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWaye® Molar Handpiece, the System is indicated for 1st and 2nd molar teeth. When used with the Sonendo GentleWave® Anterior/Premolar Handpiece, the System is indicated for anterior and premolar teeth.
### Substantially Equivalent To:
The Sonendo GentleWave® System is a modified device of the existing Sonendo GentleWave® System. The modified Sonendo GentleWave® System is substantially equivalent in intended use, principal of operation and technological characteristics to the Sonendo GentleWave® System cleared under premarket notification K153157.
{4}------------------------------------------------
# Technical Characteristics:
The Sonendo GentleWave® System has similar physical and technical characteristics to the predicate device. The software controlled user interface has been modified to allow for clinician-determined treatment fluid concentrations with the default fluid concentration setting identical to the concentrations for the predicate K153157 with three setting options to decrease the concentration of the fluids.
| Technical Characteristics | Sonendo GentleWave®<br>System (modified) | Sonendo GentleWave®<br>System (K153157) |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| Function | Preparation, cleaning and irrigation<br>or root canal | SAME |
| Principle of Operation | Generation of hydroacoustic<br>waves and fluid motion. The tip<br>of the device is placed inside the<br>tooth during cleaning.<br>Hydroacoustics are created by the<br>water stream flowing through the<br>guide tube and coming into<br>contact with the fluid inside the<br>tooth at the distal tip. The fluid<br>stream is dispersed and deflected<br>by the distal end plate of the tube<br>creating hydrodynamics (fluid<br>motion) within the tooth. | SAME |
| Treatment Site | Root canal | SAME |
| Components | Control Unit<br>Irrigation reservoirs<br>Foot pedal<br>Handpiece<br>Accessories | SAME |
| Treatment times | Fixed or User selected | SAME |
| Treatment fluid concentration | Default mode at the concentration<br>value identical to the predicate or<br>3 setting options to decrease the<br>concentrations of the fluids | Fixed |
Each of the technical attributes are present in the predicate device. The modification to use treatment fluids at lower concentrations does not affect the substantial equivalent nature of the modified device, as this change merely allows for ability to select the treatment fluid concentration based on clinical evaluation of the treatment site. This change is consistent with traditional root canal procedures in that it enables the User to utilize treatment fluids with varying concentrations.
# Performance Data:
Design control activities and risk analysis were performed in accordance with 21 CFR Part 820.30 and ISO14971:2007, respectively. Based on the changes being made and the risks associated with those changes, the following testing was determined to be necessary:
{5}------------------------------------------------
- Software validation in accordance with ANSI/AAMI/IEC 62304:2006, Medical . device software - Software life cycle processes
Based on the results of this testing, the modified GentleWave System performs as intended and is substantially equivalent to the predicate device.
### Basis for Determination of Substantial Equivalence:
The indications for use and the fundamental scientific technology of the modified device have not been changed and are the same as those described in the unmodified predicate device. The Sonendo GentleWave® System is similar to the predicate device and is determined by Sonendo, Inc. to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.