K143448 · Sonendo, Inc. · ELC · May 1, 2015 · Dental
Device Facts
Record ID
K143448
Device Name
Sonendo GentleWave System
Applicant
Sonendo, Inc.
Product Code
ELC · Dental
Decision Date
May 1, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sonendo GentleWave™ System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave™ Molar Handpiece, the System is indicated for 1 and 200 molar teeth. When used with the Sonendo GentleWave™ Anterior/Premolar Handpiece, the System is indicated for anterior and premolar teeth.
Device Story
System delivers degassed treatment fluids (3% NaOCl, 8% EDTA) into tooth via sterile, single-use handpiece; deflection plate inside handpiece redirects fluid stream to create hydroacoustic and hydrodynamic effects; removes organic/inorganic debris from root canals; waste fluid collected into console canister. Operated by dental professionals in clinical setting. Output allows clinician to perform root canal therapy; benefits include effective cleaning of complex root canal anatomy.
Clinical Evidence
Bench testing only. Evaluated sterility (ISO 11137), biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601), and software validation. Internal testing confirmed cleaning efficacy, hydroacoustic characteristics, cavitation, apical pressure, and apical extrusion volume on extracted teeth.
Technological Characteristics
Console-based system with sterile, single-use handpieces. Uses hydroacoustic/hydrodynamic fluid delivery. Materials: Medical grade. Electrical: 100-240V, 50-60Hz. Frequency: 0.5-14 kHz (Molar) or 0.5-32 kHz (Anterior/Premolar). Flow rate: 35-45 ml/min. Max output: 21W. Sterilization: Radiation (handpieces).
Indications for Use
Indicated for patients requiring root canal therapy. Used for preparation, cleaning, and irrigation of 1st and 2nd molar teeth (with Molar Handpiece) or anterior and premolar teeth (with Anterior/Premolar Handpiece).
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, suggesting a sense of community and care. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 1, 2015
Sonendo, Inc. Ms. Jenny Fam Senior Director of Regulatory Affairs 26061 Merit Circle, Suite 101 Laguna Hills, California 92653
Re: K143448
Trade/Device Name: Sonendo GentleWave™ System Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: March 30, 2015 Received: March 31, 2015
Dear Ms. Fam:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Jenny Fam
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 5.0 Indications for Use Statement
### INDICATIONS FOR USE STATEMENT
510(k) Number (if known): __K143448
Device Name: Sonendo GentleWave System
Indications for Use:
The Sonendo GentleWave™ System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave™ Molar Handpiece, the System is indicated for 1 and 200 molar teeth. When used with the Sonendo GentleWave™ Anterior/Premolar Handpiece, the System is indicated for anterior and premolar teeth.
AND/OR
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page _1_ of _1__
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for SONENDO. The logo consists of a stylized blue wave-like symbol on the left, followed by the word "SONENDO" in a bold, sans-serif font. The wave symbol is made up of several parallel lines that curve to create a sense of motion. The word "SONENDO" is in black, and there is a registered trademark symbol to the right of the word.
# 510(k) Summary K143448
This 510(k) summary information is being submitted in accordance with 21 CFR 807.92
| Applicant | Sonendo, Inc. |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | April 28, 2015 |
| Contact | Jenny Fam<br>Senior Director, Regulatory Affairs<br>26061 Merit Circle, Suite 101<br>Laguna Hills, CA 92653<br>949-766-3636 |
| Trade Name | Sonendo GentleWave™ System |
| Common Name | Sonic Cleaning and Irrigation System<br>Ultrasonic Scaler |
| Classification | Ultrasonic Scaler |
| Device Classification | Class II per 21 CFR 872.4850 |
| Product Code | ELC |
| Predicate Devices | Primary: Sonendo OmniClean Endotherapy System (K133752)<br>Reference: EMS Piezon Master 700 (K093000) |
#### Substantial Equivalence:
The Sonendo GentleWave System is substantially equivalent in intended use, principle of operation and technological characteristics to the Sonendo OmniClean Endotherapy System (K133752) and the EMS Piezon Master 700 (K093000).
### Description of the Device Subject to Premarket Notification:
The Sonendo GentleWave System (Console and Handpiece) delivers a stream of degassed treatment fluids into the tooth. The treatment fluids delivered to the tooth include a 3% Sodium Hypochlorite (NaOCI) solution and an 8% Ethylenediaminetetraacetic (EDTA) solution which are commonly utilized in traditional endodontic treatments. The stream of solutions delivered into the tooth come in contact with accumulated fluid and are redirected by a deflection plate, creating hydroacoustic and hydrodynamic effects which result in removal of organic and inorganic debris from the root canals. The Console collects the treatment fluid from the tooth into a waste collection canister.
The GentleWave System requires use with a Sonendo Handpiece which is used to treat 1* and 20 molar teeth or the Anterior/Premolar Handpiece which is used to treat anterior and premolar teeth. The Molar and Anterior/Premolar Handpieces are hand held dental instruments which are connected to the Console and provide treatment fluid delivery and waste fluid recovery. The Molar Handpiece is used with various accessories provided with the System. Both the Molar and Anterior/Premolar Handpieces are provided sterile and intended for single use only.
#### Indications for Use:
The Sonendo GentleWave System is intended to prepare, clean, and irrigate teeth indicated for root canal therapy. When used with the Sonendo GentleWave Molar Handpiece, the System is indicated for 1st and
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for SONENDO. The logo consists of a stylized blue wave-like design on the left, followed by the word "SONENDO" in bold, black capital letters. The wave design is made up of several parallel curved lines, and the word "SONENDO" is in a sans-serif font, with a registered trademark symbol next to the "O".
2nd molar teeth. When used with the Sonendo GentleWave Anterior/Premolar Handpiece, the System is indicated for anterior and premolar teeth.
#### Technological Characteristics:
The Sonendo GentleWave System has similar physical and technological characteristics to the predicate devices. The differences in physical (weight, dimensions) and electrical (power usage) characteristics when compared to the EMS Piezon Master do not affect substantial equivalence.
| Characteristics | Sonendo GentleWave<br>System (K143448) | Primary Predicate:<br>Sonendo OmniClean<br>Endotherapy System<br>(K133752) | Reference Predicate:<br>EMS Piezon Master 700<br>(K093000) |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Sonendo GentleWave<br>System is intended to<br>prepare, clean and irrigate<br>teeth indicated for root<br>canal therapy. When<br>used with the Molar<br>Handpiece, the System is<br>intended to prepare, clean<br>and irrigate 1st and 2nd<br>molar teeth indicated for<br>root canal therapy. When<br>used with the<br>Anterior/Premolar<br>Handpiece, the System is<br>intended to prepare, clean<br>and irrigate anterior and<br>premolar teeth indicated<br>for root canal therapy. | The Sonendo OmniClean<br>System is intended to<br>prepare, clean and irrigate<br>1st and 2nd molar teeth<br>indicated for root canal<br>therapy. | Intended for use in the<br>following dental and<br>periodontal applications:<br>● Preparing, cleaning and<br>irrigating root canals<br>● Retrograde preparation<br>of root canals<br>● Removing supra and<br>sub-gingival calculus<br>deposits and stains from<br>teeth<br>● Periodontal pocket<br>lavage with simultaneous<br>ultrasonic tip movement<br>● Scaling and root planning<br>● Releasing crowns,<br>bridges, inlays and posts<br>as well as condensing<br>gutta percha<br>● Plugging for amalgam<br>condensation<br>● Cavity preparation,<br>cementing inlays and<br>onlays |
| Principle of Operation | Generation of<br>hydroacoustic waves and<br>fluid motion. The tip of the<br>device is placed in or on<br>top of the tooth during<br>cleaning. Hydroacoustics<br>are created by the<br>treatment fluid stream<br>flowing through the<br>enclosure and coming into<br>contact with the fluid<br>inside the tooth at the<br>distal tip. The fluid stream<br>is dispersed and deflected<br>by the deflection plate of<br>the enclosure creating<br>hydrodynamics (fluid<br>motion) within the tooth. | Generation of<br>hydroacoustic waves and<br>fluid motion. The tip of the<br>device is placed in or on<br>top of the tooth during<br>cleaning. Hydroacoustics<br>are created by the water<br>treatment fluid stream<br>flowing through the<br>enclosure and coming into<br>contact with the fluid<br>inside the tooth at the<br>distal tip. The fluid stream<br>is dispersed and deflected<br>by the deflection plate of<br>the enclosure creating<br>hydrodynamics (fluid<br>motion) within the tooth. | Generation of hydroacoustic<br>waves and fluid motion |
| Characteristics | Sonendo GentleWave<br>System (K143448) | Primary Predicate:<br>Sonendo OmniClean<br>Endotherapy System<br>(K133752) | Reference Predicate:<br>EMS Piezon Master 700<br>(K093000) |
| Treatment Site | Root canal | Root canal | Various, including Root<br>canal |
| Components | Console (Control Unit),<br>Irrigation Fluid Reservoirs,<br>Foot Pedal, Handpiece<br>and Accessories | Console (Control Unit),<br>Irrigation Fluid Reservoirs,<br>Foot Pedal, Handpiece<br>and Accessories | Control Unit, Irrigation Fluid<br>Reservoirs, Foot Pedal,<br>Handpiece and Instruments |
| Irrigation Fluids | 3% Sodium Hypochlorite<br>(NaOCI) solution and 8%<br>EDTA solution | 3% Sodium Hypochlorite<br>(NaOCI) solution and 8%<br>EDTA solution | Various, including 0.3%<br>Sodium Hypochlorite<br>(NaOCI) solution and low<br>grade acids |
| Handpiece Type | Two (2) Offered: Molar<br>Handpiece and<br>Anterior/Premolar<br>Handpiece | One (1) Offered: Molar<br>Handpiece | Universal Handpiece with<br>various attachments. |
| Sterility: Handpiece | Sterile, Single Use Only | Sterile, Single Use Only | User sterilized, Reusable |
| Sterility: Console | Non-Sterile, User<br>Disinfected, Reusable | Non-Sterile, User<br>Disinfected, Reusable | Non-Sterile, User<br>Disinfected, Reusable |
| Materials | Medical Grade Materials | Medical Grade Materials | Medical Grade Materials |
| Flow rate | With Molar Handpiece:<br>45±10 ml/min<br>With Anterior/Premolar<br>Handpiece: 35±10 ml/min | With Molar Handpiece:<br>50±10 ml/min | 50 ml/min |
| Maximum Ultrasonic<br>Output | 21 watts | 21 watts | 12 watts |
| Frequency range | With Molar Handpiece:<br>0.5-14 kHz<br>With Anterior/Premolar<br>Handpiece: 0.5-32 kHz | Molar Handpiece: 0.5-14<br>kHz | 24-32 kHz |
| Supply voltage | 100-240V, 50-60Hz | 100-120V, 50-60Hz<br>120-240V, 50-60Hz | 100-240 VAC/ 50-60 Hz |
| Power usage | 6 VA | 6 VA | 80 VA |
| EN 60601-1<br>classification | Class I Applied Part, Type<br>B, IP 20, unit | Class I Applied Part, Type<br>B, IP 20, unit | Class I Applied Part, Type<br>BF, IP 20, unit |
| Weight | 60 kg | 59 kg | 2.4 kg |
| Unit Dimensions (mm) | 305 x 584 x 1066 | 305 x 584 x 1066 | 225 x 280 x 295 |
| Operating conditions | 10°-40°C, 30-75% Rh | 10°-40°C, 30-75% Rh | 10°-40°C, 30-75% Rh |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for SONENDO. The logo consists of a stylized blue wave-like design on the left, followed by the word "SONENDO" in bold, black letters. The wave design is made up of several curved lines that create a sense of movement. The word "SONENDO" is in a sans-serif font, and there is a small circle to the right of the word.
The differences do not impact substantial equivalence as the differences in the indications and technical specifications are within the range of the identified predicates.
### Performance Data:
All necessary performance testing has been conducted on the Sonendo GentleWave System to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. The device passed the following tests, which were conducted in accordance with noted standards:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for SONENDO. The logo consists of a stylized "S" made up of several curved lines on the left, followed by the word "SONENDO" in a bold, sans-serif font. A small circle, likely a registration mark, is present to the upper right of the "O" in "SONENDO". The logo is simple and modern.
| Test | Consensus Standard/FDA Guidance/Description |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterility | Sterile per ISO 11137-1,-2, Sterilization of health care products - Radiation Sterilization |
| Biocompatibility | Biocompatible per ISO 10993-1, -5, -10, -11, Biological Evaluation of Medical Devices |
| Electromagnetic<br>Compliance and<br>Electrical Safety | Certified per IEC 60601-1 General Requirements for Basic Safety and Essential<br>Performance and IEC 60601-1-2 Medical Electrical Equipment: Safety, Radiofrequency<br>Emissions and Electromagnetic Immunity. |
| Software | Validated per FDA Guidance: Guidance for the Content of Premarket Submissions for<br>Software Contained in Medical Devices, 2005. |
| Thermal Test | Internal Test Method: Determined the temperature generated on the external surface of<br>the tooth when receiving treatment with the device. |
| Cavitation Test | Internal Test Method: Evaluated transient cavitation bubbles which clean root canals. |
| Hydroacoustics Test | Internal Test Method: Demonstrated hydroacoustic characteristics of the device. |
| Cleaning Test | Internal Test Method: Demonstrated cleaning efficacy of the device on extracted teeth. |
| Apical Pressure Test | Internal Test Method: Determined the hydraulic pressure exerted at the apical terminus<br>of the root canal during use of the device. |
| Apical Extrusion Test | Internal Test Method: Measured the relative volume of treatment fluid extruded through<br>the apical terminus during use of the device. |
# Basis for Determination of Substantial Equivalence:
Upon reviewing the information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the Sonendo GentleWave System is substantially equivalent to existing legally marketed devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.