OsteoPlan System

K212570 · Osteomed, LLC · DZJ · Feb 11, 2022 · Dental

Device Facts

Record IDK212570
Device NameOsteoPlan System
ApplicantOsteomed, LLC
Product CodeDZJ · Dental
Decision DateFeb 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4120
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

The OsteoPlan™ System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the OsteoPlan™ System and the result is an output data file that may then be provided as digital models or used as input to the additive manufacturing portion of the system that produces physical outputs including anatomic models and splints for use in maxillofacial surgery. The OsteoPlan System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options.

Device Story

System processes CT/CBCT imaging data to assist maxillofacial surgeons in pre-operative planning and surgical execution. Uses off-the-shelf (OTS) software for image segmentation and surgical simulation. Physician provides input to define surgical plan; system generates digital models, anatomic models (maxilla, mandible, skull), and orthognathic occlusal splints. Physical outputs produced via additive manufacturing (SLS/SLA). Used in clinical settings by physicians to visualize complex anatomy and translate surgical plans into physical guides. Benefits include improved surgical accuracy and planning for complex maxillofacial procedures.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing: IQ/OQ/PQ, software validation (moderate level of concern), cleaning/sterilization validation (ANSI/AAMI/ISO 17665-1), biocompatibility (ISO 10993-1), packaging, shelf-life, and a cadaver study for simulated use.

Technological Characteristics

Additive manufacturing (SLS/SLA) using photopolymers and Polyamide 12. Software-based image segmentation and surgical planning. Sterilization via steam (ANSI/AAMI/ISO 17665-1). Biocompatibility per ISO 10993-1. Software level of concern: moderate.

Indications for Use

Indicated for patients undergoing maxillofacial surgery requiring pre-operative simulation, anatomical modeling, or surgical splints.

Regulatory Classification

Identification

A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square are the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" below it. February 11, 2022 OsteoMed Diane Rutherford Regulatory Affairs Manager 3885 Arapaho Road Addison, Texas 75001 Re: K212570 Trade/Device Name: OsteoPlan System Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument And Accessories Regulatory Class: Class II Product Code: DZJ, LLZ Dated: November 12, 2021 Received: November 15, 2021 Dear Diane Rutherford: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K212570 Device Name OsteoPlan System ### Indications for Use (Describe) The OsteoPlan™ System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the OsteoPlan™ System and the result is an output data file that may then be provided as digital models or used as input to the additive manufacturing portion of the system that produces physical outputs including anatomic models and splints for use in maxillofacial surgery. The OsteoPlan System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | <span style="font-size:14px"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:14px"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY K212570 #### I. SUBMITTER OsteoMed 3885 Arapaho Rd. Addison, TX 75001 USA Phone: 972-677-4795 Fax: 972-677-4601 Email: diane.rutherford(@acumed.net Contact Person: Diane Rutherford Date Prepared: February 9, 2022 #### II. DEVICE | Name of the Device: | OsteoPlan™ System | |-----------------------|-------------------------------------------------------------------------------| | Common or Usual Name: | Patient specific maxillofacial anatomical models, splints, and surgical plans | | Classification Name: | Bone Cutting Instrument and Accessories | | Classification: | Class II, 21 CFR 872.4120 | | Product Code: | DZJ, LLZ | #### III. PREDICATE DEVICES Predicate Device: VSP® System, Medical Modeling, Inc (K133907, K120956) Reference Device: Surgicase Guides, Materialise (K103136) #### IV. DEVICE DESCRIPTION OsteoMed uses computer aided modeling to assist the physician with planning complex maxillofacial surgeries. Specifically, the OsteoPlan™ System provides patient-specific anatomical models, splints, and patient-specific surgical plans and digital files of the surgical plan to assist physicians with maxillofacial surgeries. Outputs of the OsteoPlan™ System are designed with physician input and reviewed by the physician prior to finalization and distribution. All outputs are manufactured by OsteoMed using additive manufacturing (SLS and SLA), only with direct physician involvement to reduce the criticality of the outputs. The system uses electronic medical images of the patient anatomy (CT and CBCT) with input from the physician to create the plan and splints for executing surgery. Off-the-shelf (OTS) software is used for surgical planning. The outputs of the system include Orthognathic Occlusal Splints, Case Reports, and Anatomic models. The splints are offered in commonly used forms, in both intermediate and final positioning, and some are available with ligature holes. Case reports are digital and physical documents created to lay out the surgical plan, dictated by the surgeon, and show outputs of the OsteoPlan™ system that will be used to translate the plan during surgery. Anatomic models are tools provided to physicians for complex anatomy visualization or to preplan surgery with an accurate physical representation of patient anatomy. Anatomic models may include maxilla, mandible, or skull models. {4}------------------------------------------------ #### V. INDICATIONS FOR USE The OsteoPlan™ System is intended for use as a software system and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the OsteoPlan System and the result is an output data file that may then be provided as digital models or used as input to the additive manufacturing portion of the system that produces physical outputs including anatomic models and splints for use in maxillofacial surgery. The OsteoPlan System is also intended as a pre-operative software tool for simulating / evaluating surgical treatment options. #### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The principles of operation and technological characteristics are substantially equivalent between the subject device and the primary predicate, VSP System. Specifically, the predicate device includes patient specific anatomic models and splints produced through 3D printing. The only difference in the Technological Characteristics is found in the materials of the outputs. The OsteoPlan System uses photopolymers for splints while the VSP System uses photopolymers (with stainless-steel inserts) in the cutting guides of the system. The OsteoPlan System uses Polyamide anatomic models (Polyamide 12) produced through an SLS printing process equivalent to the reference device. SurgiCase Guides (K103136), which includes outputs for the same anatomic region, made out of identical material (Polyamide 12) and printing process (SLS). The subject device contains a subset of the outputs of the predicate and reference device. | | New Device | Primary Predicate | Reference Predicate | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | OsteoMed OsteoPlan™<br>System K212570 | Medical Modeling VSP<br>System K133907, K120956 | Materialise Surgicase<br>Guides K103136 | | Indication for Use | The OsteoPlan™ System is<br>intended for use as a software<br>system and image segmentation<br>system for the transfer of imaging<br>information from a medical scanner<br>such as a CT based system. The<br>input data file is processed by the<br>OsteoPlan System and the result is<br>an output data file that may then be<br>provided as digital models or used<br>as input to the additive<br>manufacturing portion of the system<br>that produces physical outputs<br>including anatomical models and<br>splints for use in maxillofacial<br>surgery. The OsteoPlan System is<br>also intended as a pre-operative<br>software tool for simulating /<br>evaluating surgical treatment<br>options. | The Medical Modeling VSP<br>System is intended for use as a<br>software system and image<br>segmentation system for the<br>transfer of imaging information<br>from a medical scanner such as a<br>CT based system. The input data<br>file is processed by the VSP<br>System and the result is an output<br>data file that may then be<br>provided as digital models or used<br>as input to a rapid prototyping<br>portion of the system that<br>produces physical outputs<br>including anatomical models,<br>templates, and surgical guides for<br>use in maxillofacial surgery. The<br>VSP System is also intended as a<br>pre-operative software tool for<br>simulating / evaluating surgical<br>treatment options. | SurgiCase Guides are<br>intended to be used as<br>surgical tools to transfer a<br>pre-operative plan to<br>the surgery. The devices are<br>intended to guide the<br>marking of bone and/or<br>guide surgical<br>instruments in mandibular<br>and maxillofacial surgical<br>procedures.<br>SurgiCase Guides are<br>intended for single use only. | | Outputs | Anatomic Models<br>Surgical Positioning Templates<br>(splints)<br>Patient Specific Case Reports | Anatomical Models<br>Surgical Positioning<br>Templates/Guides<br>Osteotomy Templates/Guides<br>Patient Specific Case Reports<br>Plate Bending Templates/Guides<br>Stainless Steel Culling and Drill<br>Inserts | Surgical Positioning<br>Templates/Guides<br>Osteotomy Templates/Guides<br>Patient Specific Case Reports<br>Plate Bending<br>Templates/Guides | {5}------------------------------------------------ ## OsteoMed OsteoPlan™ System Traditional 510(k) | | New Device | Primary Predicate | Reference Predicate | |-----------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------------------------------|-----------------------------------------| | Trade Name | OsteoMed OsteoPlanTM<br>System K212570 | Medical Modeling VSP<br>System K133907, K120956 | Materialise Surgicase<br>Guides K103136 | | Additive<br>Manufacturing | Yes | Yes | Yes | | Biocompatible<br>Materials | Yes | Yes | Yes | | Patient-Contacting<br>External<br>Communicating<br>Tissue/Bone/Dentin | Limited (<24 hours):<br>Models<br>Prolonged (24 hours -30 Days):<br>Splints | Limited and Prolonged | Limited | #### VII. SUMMARY OF NON-CLINICAL TESTING The following performance data were provided in support of the substantial equivalence determination. - Equipment/Process Qualification (IQ/OQ/PQ) was performed in accordance with the OsteoMed quality management system with all validations meeting acceptance criteria. - Software Validation and documentation for software of moderate level of concern was provided per the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Off-The-Shelf Software Use in Medical Devices." All software verification/validation passed. - Cleaning Validations were performed in accordance with FDA Guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling. All testing passed. - Steam Sterilization Validation was conducted using both pouches and rigid sterilization containers in accordance with ANSI/AAMI/ISO 17665-1 "Sterilization of health care products-Moist heat-Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices - Biocompatibility testing for the worst-case splint and anatomical model was conducted per its contact classification according to the FDA Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" Testing included: cytotoxicity, sensitization, irritation, acute toxicity, pyrogenicity, subchronic toxicity, and implantation. All biocompatibility testing passed. - Packaging Validation was conducted to confirm that the packaging for the OsteoPlan System protects and retains the splints and models during normal shipping and handling. All testing passed. - Shelf Life was established with functional testing to confirm the specified requirements were met. All testing passed. - A Cadaver Study was conducted to verify the functionality of the design of the OsteoPlan system via simulated use with all acceptance criteria being met. Performance equivalence was shown through the verification comparison to the predicate device. Clinical testing is not required to support substantial equivalence. #### CONCLUSION VIII. The conclusions drawn from the nonclinical tests demonstrate that the OsteoMed OsteoPlan™ System is substantially equivalent to the legally marketed device predicates.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...