Kerecis Oral

K213904 · Kerecis · NPL · Sep 30, 2022 · Dental

Device Facts

Record IDK213904
Device NameKerecis Oral
ApplicantKerecis
Product CodeNPL · Dental
Decision DateSep 30, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3930
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants; Covering of implants placed in immediate extraction sockets; Covering of implants placed in delayed extraction sockets; Covering of bone defects after root resection and removal of retained teeth; and Guided tissue regeneration procedures in periodontal and recession defects.

Device Story

Kerecis Oral is an acellular, resorbable fish dermal matrix derived from cod fish skin. It serves as a surgical graft for periodontal and oral procedures. The device is supplied as a solid, single-layer sheet in various sizes. It is intended for use by dental surgeons in clinical settings to aid in soft tissue and bone regeneration. The matrix is biocompatible, non-cross-linked, and flexible, allowing for fixation via sutures or pins. During surgery, the clinician cuts and shapes the graft to cover defects or implants. The porous structure of the fish-derived collagen facilitates cellular infiltration, tissue adherence, and eventual resorption as the patient's own tissue regenerates. By providing a scaffold for cellular ingrowth, the device supports the healing of periodontal defects and the augmentation of keratinized tissue, potentially improving clinical outcomes for patients undergoing dental implant or periodontal surgery.

Clinical Evidence

No human clinical trials were performed. Evidence consists of bench testing and animal studies. Bench testing verified morphology (SEM/H&E), tensile strength (ASTM D638), suture/pin pull-out strength (ISO 7198/ASTM F-1839-08), compression (ASTM F-1306), and stability in simulated oral environments. Animal study (canine bilateral mandibular bony defect model) compared the subject device to the predicate at 30, 60, and 90 days, showing no significant differences in new bone formation, membrane resorption, or soft tissue infiltration via histology and micro-CT volumetric analysis.

Technological Characteristics

Acellular fish dermal matrix (cod skin). Single-layer sheet. Biocompatible, non-cross-linked, resorbable. Sterilized via Ethylene Oxide (ISO 11135, ISO 11737-1). Mechanical properties: Tensile strength 14.3 MPa. Complies with ISO 10993 (biocompatibility), ASTM D638 (tensile), ASTM F-1839-08 (pin pull-out), ASTM F-1306 (compression), and ICH Q3D (heavy metals).

Indications for Use

Indicated for patients requiring localized gingival augmentation to increase keratinized tissue around teeth or implants, covering of implants in extraction sockets, covering of bone defects after root resection/tooth removal, and guided tissue regeneration in periodontal and recession defects.

Regulatory Classification

Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Kerecis Gudmundur Sigurjonsson CEO Eyrargata 2 Isafjordur, 400 Iceland 9/30/22 Re: K213904 Trade/Device Name: Kerecis® Oral Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPL Dated: September 6, 2022 Received: September 7, 2022 Dear Gudmundur Sigurjonsson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K213904 Device Name Kerecis® Oral #### Indications for Use (Describe) - · Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants; - · Covering of implants placed in immediate extraction sockets; - · Covering of implants placed in delayed extraction sockets; - · Covering of bone defects after root resection and removal of retained teeth; and - · Guided tissue regeneration procedures in periodontal and recession defects. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------| | <span style="text-decoration: underline;"> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Ke ## 510(k) Summary Traditional Premarket Notification Submission (510(k)) Summary) Prepared in accordance with 21 CFR § 807.92 #### 1. Submitter Information | Sponsor Name: | Kerecis Limited | |-----------------------------|---------------------------------------------------| | Sponsor Address: | Eyrargata 2 – PO Box 151, 400 Isafjordur, Iceland | | Sponsor Telephone: | +011 354-419-8000 | | Establishment Registration: | 301060025 | | Primary Contact Person: | Gudmundur Fertram Sigurjonsson | | Contact Title: | Founder, Chairman, President & CEO | | Email Direct: | gfs@kerecis.com | | Secondary Contact Person: | Daniel L. Mooradian, Ph.D. | | Contact Title: | Senior VP for Research & Development | | Email Direct: | dmooradan@kerecis.com | | Additional Contact Person: | Jennifer Michelle Chambers, MPA, PMP, RAC | | Contact Title: | Senior Regulatory Specialist | | Email Direct: | jchambers@kerecis.com | | Date Summary Prepared: | September 30, 2022 | #### 2. Device Information | Trade Name (Proprietary): | Kerecis® Oral | |---------------------------|-----------------------------------| | Common (Usual): | Intraoral Surgical Graft | | Classification Name: | Barrier, Animal Source, Intraoral | | Device Classification: | Class II, 21 CFR § 872.3930 | | FDA Device Code: | NPL | #### 3. Predicate Device and Reference Devices #### Predicate Device Company Name: Ed. Geistlich Söhne Ag für Chemische Industrie MUCOGRAFT® Collagen Matrix Device Name (proprietary): K073711 Device 510(k): {4}------------------------------------------------ Reference Device Company Name: Device Name (proprietary): Device 510(k): Kerecis Kerecis Gingiva Graft K192612 #### 4. Device Description The subject device is an acellular resorbable fish dermal matrix, intended for use in periodontal surgical procedures to aid in soft tissue and bone regeneration. It is obtained from cod fish skin by a standardized controlled manufacturing process, and supplied in terminally sterile peel-pouch packaging in the following solid sizes: - 15mm x 20mm - 20mm x 30mm - . 30mm x 40mm It is biocompatible, non-cross linked, and therefore resorbable, strong, flexible, and supports fixation by sutures and pins. #### 5. Intended Use The subject device is indicated for: - Localized gingival augmentation to increase keratinized tissue (KT) around teeth or implants; - . Covering of implants placed in immediate extraction sockets; - Covering of implants placed in delayed extraction sockets; - Covering of bone defects after root resection and removal of retained teeth; and - Guided tissue regeneration procedures in periodontal and recession defects. ### 6. Technological Characteristics and Substantial Equivalence Comparison with the predicate device (K073711), and the reference device (K192612), demonstrate that it is substantially equivalent with regards to: intended use, materials, design, and operational principle. See Table 6.1. Kerecis Oral is comparison with predicate and reference devices. Table 6.1. Kerecis Oral is comparison with predicate and reference devices. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the word "Kerecis" in a stylized font. The letters are primarily dark blue, except for the "c" in the middle, which is orange. There is a registered trademark symbol to the right of the word. | | Subject<br>Device | Predicate<br>Device | Reference<br>Device | Discussion | |------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | Kerecis Oral | Geistlich<br>MUCOGRAFT<br>Collagen<br>Matrix | Kerecis<br>Gingiva Graft | N/A | | 510(k) | TBD | K073711 | K192612 | N/A | | Regulation | 21 CFR 872.3930 | 21 CFR 872.3930 | 21 CFR 872.3930 | Same regulation | | Device Class | II | II | II | Same device class | | Product Code | NPL | NPL | NPL | Same product code | | Device<br>Classification | Barrier, Animal<br>Source, Intraoral | Barrier, Animal<br>Source, Intraoral | Barrier, Animal<br>Source, Intraoral | Same device classification | | Intended Use | indicated for:<br>Localized<br>gingival<br>augmentation<br>to increase<br>keratinized<br>tissue (KT)<br>around teeth<br>or implants;<br>-Covering of<br>implants<br>placed in<br>immediate<br>extraction<br>sockets;<br>-Covering of<br>implants<br>placed in<br>delayed<br>extraction<br>sockets; | indicated for:<br>-Simultaneous<br>use of GBR-<br>membrane and<br>implants<br>-Covering of<br>implants<br>placed in<br>immediate<br>extraction<br>sockets;<br>-Covering of<br>implants<br>placed in<br>delayed<br>extraction<br>sockets;<br>-Localized ridge<br>augmentation | indicated for:<br>-Localized<br>gingival<br>augmentation<br>to increase<br>keratinized<br>tissue (KT)<br>around teeth<br>or implants. | Intended use includes<br>elements of the predicate and<br>reference devices, supported by<br>performance testing in canines,<br>intended to demonstrate<br>equivalence in the oral<br>environment | | | | | | | | | -Covering of<br>bone defects<br>after root<br>resection and<br>removal of<br>retained teeth;<br>-Guided tissue<br>regeneration<br>procedures in<br>periodontal<br>and recession<br>defects. | for later<br>implantation;<br>-Alveolar ridge<br>reconstruction<br>for prosthetic<br>treatment;<br>-Covering of<br>bone defects<br>after root<br>resection,<br>cystectomy,<br>removal of<br>retained teeth;<br>-Guided bone<br>regeneration in<br>dehiscence<br>defects; and<br>-Guided tissue<br>regeneration<br>procedures in<br>periodontal<br>and recession<br>defects. | | | | Animal Origin<br>Material | Fish: skin<br>tissue, single<br>layer sheet | Porcine: skin<br>and connective<br>tissue, double<br>layer sheet | Fish: skin<br>tissue, single<br>layer sheet | Same animal species as<br>reference | | Biocompatibility | Yes | Yes | Yes | Biocompatibility testing<br>performed per ISO 10993 series<br>standards | | NON-Pyrogenic | Yes | Yes | Yes | Materials-mediated<br>pyrogenicity safe per ISO 10993<br>series standard | | Tensile Strength | 14.3 MPa | 4.6 MPa | 14.3 MPa | Meets tensile strength value of<br>the predicate device per ASTM<br>D638 | | Resorbable | Yes | Yes | Yes | All devices are resorbable per<br>comparative performance data | | Sizes | 15x20 mm | 15x20 mm | 15x20 mm | Equivalent sizes by dimensional<br>analysis | | | 20x30 mm | 20x30 mm | 20x30 mm | | | | 30x40 mm | 30x40 mm | 30x40 mm | | | Sterilization | Ethylene Oxide | Gamma<br>Irradiation | Ethylene Oxide | Traditional Sterilization Methods<br>per ISO 11135 and ISO 11737-1.<br>EO residual testing per<br>ISO 10993-7 | | Sterility<br>Assurance Level | 10-6 | 10-6 | 10-6 | Equivalent SAL per ISO 11137<br>and ISO 11737 | | Shelf Life | 3 years | 3 years | 3 years | Equivalent shelf life per ASTM<br>F1980 and Q5C (R2)[ICH] | | Mode of Action | Fixation | Fixation | Fixation | Equivalent mode of action per<br>ANSI/AAMI/ISO 7198 and ASTM<br>F-1839-08 | {6}------------------------------------------------ Image /page/6/Picture/5 description: The image shows the word "Kerecis" in a bold, sans-serif font. The letters are primarily a dark blue color, except for the "c" in the middle of the word, which is a bright orange color. There is a registered trademark symbol to the right of the word. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for Kerecis. The logo is in blue, except for the "c" in "cis", which is in orange. There is a registered trademark symbol to the right of the logo. #### 7. Performance Testing - BenchBench testing was performed following FDA guidance, "Preparation of a Premarket Notification Application for a Surgical Mesh" (1999). The following performance studies were conducted on representative product samples to verify that material properties remain unchanged and support substantial equivalence: #### 7.1 Morphology Observation The subject device and the predicate device are based on the collagen rich animal tissue, piscine and porcine, respectively. Based on H&E staining, both materials are rich in collagen and porous, therefore favoring cellular infiltration. Scanning Electron Microscope (SEM) shows equivalent preserved collagen structure of the animal origin tissues used for both devices. Cross section of both devices showed that the porous surfaces in the skin derived collagen structure of both materials allows tissue adherence and promotes tissue regeneration by favoring cellular ingrowth when applied to soft tissue defect areas. #### 7.2. Cellular ingrowth comparison Both materials were tested for cellular ingrowth capability by fibroblast seeding onto the materials in vitro cellular modes. Both materials showed favorable cellular infiltration of fibroblasts after 14 days {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the word "Kerecis" in a stylized font. The letters are primarily dark blue, except for the "c" in the middle of the word, which is orange. A small registered trademark symbol is located to the upper right of the word. which is a key component for tissue augmentation and re-epithelization of defected keratin tissue in the oral cavity. #### 7.3. Tensile Strength The tensile strength of the subject device was determined to be comparable to the predicate device measured by ultimate tensile strength. #### 7.4. Heavy Metal Analysis A heavy metal analysis was evaluated to show that the limits of cadmium (Cd), lead (Pb), arsenic (As) and mercury (Hg) contained within the subject device were acceptable under the ICH guidelines: Q3D Elemental Impurities-Guidance for Industry. #### 7.5. Dimensional Validation Manufacturing dimensional validation was performed on three lots of the subject device where its dimension and thickness were tested. Kerecis has set the quality limits for a capability index to be > 1.33. All tested parameters met or exceeded the set goal. #### 7.6. Stability in a simulated physiological environment A stability test was done in a simulated physiological oral environment (artificial saliva buffer) to investigate the dissolution of both material over time and to compare the effects that the products have on the pH levels and conductivity of the buffer over 24 hours. The subject device is structurally more stable than the predicate device since it dissolved slower than the predicate device at neutral pH 7. #### 7.7. Suture Pull-Out Strength Consistent with recognized standard ANSI/AAMI/ISO 7198, The suture pull-out strength of the subject device meets or exceeds that of the predicate with a confidence of greater than 95%. The products are equivalent. #### 7.8. Pin Pull-Out Strength Consistent with recognized standard ASTM F-1839-08, The pin pull-out strength of the subject device exceeds that of the predicate at a confidence level of greater than 95%. The products are equivalent. #### 7.9. Compression In accordance with ASTM-F-1306, "Standard Test Method for Slow Rate Penetration Resistance of Flexible Barrier Films and Laminates", the mechanical test for compressive strength was measured for the subject and predicate device. The compressive Peak-Load, Load-at-Break, Probe Penetrationat-Break, and Energy-to-Break of the subject device meet or exceed those of the predicate device, with a confidence of greater than 95%. The products are equivalent. #### 7.10. User Evaluation of Device for Cutting and Shaping The subject device was evaluated in comparison to the predicate for use in the oral environment using a questionnaire. The results showed favorable usability that was substantially equivalent to the predicate for cutting and shaping the device for use as a dental barrier membrane. #### 7.11.Biocompatibility, Sterilization, Shelf-life and Animal origin. {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the word "Kerecis" in a bold, sans-serif font. The letters are primarily dark blue, except for the "C" in the middle, which is a bright orange color. A small registered trademark symbol is located in the upper right corner of the image. Testing from the applicant's own predicate device (K190528 and K153364) was used to s substantial equivalence. Biocompatibility per ISO 10993 series - · Cytotoxicity - Sensitization - Irritation or Intracutaneous reactivity - · Acute systemic toxicity - · Subacute/sub-chronic toxicity - · Genotoxicity - Implantation - Materials-Mediated Pyrogenicity - Chronic Toxicity - Carcinogenicity • Sterilization validation per ISO 11737-1, Ethylene Oxide residual test following ISO 10993-7 • Endotoxin validation (<20 EU/device) of sterilization method per LAL turbidimetric kinetic method following ISO 10993-11 - Shelf life per ASTM F1980 and Q5C (R2)[ICH] using accelerated and real-time aged samples - Packaging per ISO 11607-series, ASTM F88 and ASTM F1886 - Animal Origin and Viral inactivation per ISO 22422 series The performance data demonstrates that the subject device is substantially equivalent to the predicate device. #### 8. Performance Testing - Animal Comparative performance testing in an animal canine (beagle bilateral mandibular bony defect model was used to support equivalence with the predicate device for the stated indications for use. On the 30, 60 and 90 day post-surgery, animal tissues were evaluated for gross pathology and histological, and quantitative micro-CT volumetric analysis of new bone (%) and xenograft (%). In a side-by-side comparison between the Test (Subject device) and Control (Predicate device), no qualitative or significant differences was observed for new bone formation, xenograft membrane resorption kinetics and soft tissue infiltration. Histologic and micro-CT volumetric analysis revealed no noticeable or significant differences between the two membranes at any time point. Soft tissue endpoints were part of the applicant's previous canine testing as described in the reference device application K192612. Studies were performed in AAALC accredited facilities in compliance with the applicable requirements of Good Laboratory Practices (GLP) 21 CFR §58. #### 9. Performance Testing - Clinical No human clinical trials were performed in support of this submission. {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for Kerecis. The logo is in dark blue, except for the letter 'c' in the middle, which is in orange. There is a registered trademark symbol in the upper right corner of the logo. ### 10. Conclusion The data provided within this submission support substantial equivalence of the subject device to the predicate device with regards to intended use, materials, design, and technological characteristics, including principles of operation, performance characteristics, and device safety. For use in oral surgical applications, Kerecis Oral is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...