Geistlich Mucograft® and Geistlich Mucograft® Seal are indicated for: • covering of implants placed in immediate or delayed extraction sockets; • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants; • alveolar ridge reconstruction for prosthetic treatment; and • recession defects for root coverage. Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for: • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants • Alveolar ridge reconstruction for prosthetic treatment • Recession defects for root coverage
Device Story
Geistlich Mucograft/Mucograft Seal and Fibro-Gide are resorbable, porcine-derived collagen matrices for oral tissue regeneration and soft tissue augmentation. Mucograft features a dual-structure design: a smooth, compact collagen layer for suturing and an inner porous layer for cell invasion and tissue integration. Fibro-Gide is a porous, weakly cross-linked collagen scaffold providing volume stability and space for soft tissue ingrowth. Devices are surgically implanted by clinicians (dentists/periodontists) in clinical settings. They function as scaffolds to facilitate wound healing and tissue integration, degrading over time. Benefits include improved keratinized tissue volume and successful root coverage. No electronic components or algorithms are involved.
Clinical Evidence
No new clinical data provided. Substantial equivalence is supported by leveraging existing clinical performance data from predicate submissions (K102531, K171050). Non-clinical evidence includes viral inactivation studies per ISO 22442-3, real-time stability studies per ICH Q1A(R), and mechanical suturability testing for new size variants.
Technological Characteristics
Materials: Porcine collagen. Structure: Mucograft (dual-layer: smooth/porous); Fibro-Gide (porous scaffold). Sterilization: Gamma irradiation. Form factor: Sheets/discs of varying dimensions. Connectivity: None. Software: None. Mechanical principle: Resorbable scaffold for tissue ingrowth.
Indications for Use
Indicated for patients requiring soft tissue augmentation, including localized gingival augmentation to increase keratinized tissue around teeth and implants, alveolar ridge reconstruction for prosthetic treatment, recession defects for root coverage, and covering of implants in extraction sockets.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Geistlich Mucograft® and Mucograft® Seal (K210280)
Geistlich Mucograft® and Mucograft® Seal (K192042)
Submission Summary (Full Text)
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
Geistlich Pharma AG
% Roshana Ahmed
President
Quaras, LLC
2101 Camino Rey
Fullerton, California 92833
November 25, 2025
Re: K252253
Trade/Device Name: Geistlich Mucograft® /Geistlich Mucograft® Seal; Geistlich Fibro-Gide®
Regulation Number: 21 CFR 872.3930
Regulation Name: Bone Grafting Material
Regulatory Class: Class II
Product Code: NPL
Dated: October 20, 2025
Received: October 27, 2025
Dear Roshana Ahmed:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K252253 - Roshana Ahmed
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K252253 - Roshana Ahmed
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K252253 | |
| Device Name Geistlich Mucograft® /Geistlich Mucograft® Seal; Geistlich Fibro-Gide® | |
| Indications for Use (Describe) Geistlich Mucograft® and Geistlich Mucograft® Seal are indicated for: • covering of implants placed in immediate or delayed extraction sockets; • localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants; • alveolar ridge reconstruction for prosthetic treatment; and • recession defects for root coverage. | |
| Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for: • Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants • Alveolar ridge reconstruction for prosthetic treatment • Recession defects for root coverage | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
Geistlich Mucograft® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
# 510(k) Summaries
# Geistlich Mucograft® and Geistlich Mucograft® Seal
## I. Submitter
Geistlich Pharma AG
Bahnhofstrasse 40
CH-6110 Wolhusen
Switzerland
Phone: +41 41 492 55 55
Contact Person: Marco Steiner, Head Regulatory Affairs Management
Date Prepared: November 24, 2025
## II. Device
| Device Proprietary Name: | Geistlich Mucograft®
Geistlich Mucograft® Seal |
| --- | --- |
| Common or Usual Name: | Collagen Matrix |
| Classification Name: | Barrier, Animal Source, Intraoral |
| Regulation Number: | 872.3930 |
| Product Code: | NPL |
| Device Classification | II |
## III. Predicate Device
Substantial equivalence is claimed to the following devices:
| Product Name | 510(k) | Applicant |
| --- | --- | --- |
| Geistlich Mucograft® and Mucograft® Seal | K210280 | Geistlich Pharma AG |
The following devices are referenced within the submission:
- Mucograft, K102531, Ed. Geistlich Soehne AG Fuer Chemische Industrie
- Mucograft Collagen Matrix, K073711, Ed. Geistlich Soehne AG Fuer Chemische Industrie
Page 1 of 6
{6}
Geistlich Mucograft® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
## IV. Device Description
Geistlich Mucograft® and Geistlich Mucograft® Seal are surgically implanted, fully resorbable devices intended for oral tissue regeneration. The matrices are made of collagen without further cross-linking. All configurations of the product are sterilized in a double package by gamma irradiation. Geistlich Mucograft® and Geistlich Mucograft® Seal are composed of two structures: one smooth structure and one porous structure. The device allows tissue adherence as a prerequisite for favorable wound healing. The "outer" side (i.e., turned towards the soft tissue) with a smooth surface consists of compact collagen and has a smooth texture with the appropriate elastic properties to accommodate suturing. The "inner" porous structure consists of collagen fibers in a loose, porous arrangement to allow cell invasion for soft tissue ingrowth. This roughened surface is placed next to the host tissue to facilitate tissue integration.
The products are provided as follows:
- Geistlich Mucograft®: 15 x 20 mm, 20 x 30 mm, and 30 x 40 mm
- Geistlich Mucograft® Seal: 8 mm and 12 mm diameter
## V. Indications for Use
Geistlich Mucograft® and Geistlich Mucograft® Seal are indicated for:
- covering of implants placed in immediate or delayed extraction sockets;
- localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants;
- alveolar ridge reconstruction for prosthetic treatment; and
- recession defects for root coverage.
## VI. Comparison of Technological Characteristics
The Indications for Use Statement is identical to the predicate device.
The subject device is identical to the predicate device with respect to materials characteristics, manufacturing and sterilization methods, packaging, and size. Both the subject and predicate devices have identical final product specifications. A comparison of the subject and predicate device is provided in the table below.
Page 2 of 6
{7}
Geistlich Mucograft® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
| | Subject Device | Geistlich Mucograft® and Mucograft® Seal (K210280, K192042) |
| --- | --- | --- |
| Material | Porcine collagen | Same |
| Shape | Geistlich Mucograft®: Rectangle
Geistlich Mucograft® Seal: Cirle | Same |
| Sizes | Geistlich Mucograft®:
15 x 20 mm
20 x 30 mm
30 x 40 mm
Geistlich Mucograft® Seal:
8 mm diameter
12 mm diameter | Same |
| Single-Use | Yes | Same |
| Sterilization | Gamma | Same |
The purpose of this submission is to obtain clearance for the use of an alternate supplier of porcine raw material. This change does not raise different questions of safety and effectiveness and the slight technological differences are addressed by the information described in the Performance Data section below.
## VII. Performance Data
Mechanical testing, biocompatibility (K073711), sterilization (K102531), shelf-life (K102531), and clinical performance testing (K102531) from the applicant’s own predicate device was leveraged in support of substantial equivalence.
The following non-clinical data was provided within this submission to demonstrate substantial equivalence:
- results from a viral inactivation study performed in accordance with ISO 22442-3
- results from real-time stability studies performed in accordance with ICH Q1A (R) (extension of studies submitted in support of K171050)
## VIII. Conclusion
The subject devices are identical to the predicate device. The addition of a new raw material supplier does not raise different questions of substantial equivalence. Therefore, it is concluded that Geistlich Mucograft® and Geistlich Mucograft® Seal are substantially equivalent to the identified predicate device.
Page 3 of 6
{8}
Geistlich Mucograft® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
# Geistlich Fibro-Gide®
## I. Submitter
Geistlich Pharma AG
Bahnhofstrasse 40
CH-6110 Wolhusen
Switzerland
Phone: +41 41 492 55 55
Contact Person: Marco Steiner, Head Regulatory Affairs Management
Date Prepared: November 24, 2025
## II. Device
| Device Proprietary Name: | Geistlich Fibro-Gide® |
| --- | --- |
| Common or Usual Name: | Bone Grafting Material |
| Classification Name: | Barrier, Animal Source, Intraoral |
| Regulation Number: | 872.3930 |
| Product Code: | NPL |
| Device Classification | II |
## III. Predicate Device
Substantial equivalence is claimed to the following device:
| Product Name | 510(k) | Applicant |
| --- | --- | --- |
| Geistlich Fibro-Gide® | K171050 | Geistlich Pharma AG |
## IV. Device Description
Geistlich Fibro-Gide® is a fully resorbable, porous, collagen matrix of porcine origin of a spongious consistency. The collagen is extracted from veterinary certified pigs and is carefully purified to avoid antigenic reactions. The collagen scaffold is weakly crosslinked. Geistlich Fibro-Gide® is sterilized in double packaging by Gamma-irradiation.
Geistlich Fibro-Gide® is an implantable device intended for use in soft tissue augmentation procedures. As described in more detail below, the device is indicated specifically for insufficient tissue volume at the alveolar ridge and for soft tissue recession. It has mechanical properties appropriate to withstand the mechanical stresses that occur after wound closure in soft tissue augmentation procedures, i.e., it has good volume stability and it withstands early
Page 4 of 6
{9}
Geistlich Mucograft® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
resorption to allow the formation of new soft tissue and degrades over time. In addition, the matrix is designed with an appropriate thickness to provide sufficient space for the ingrowth of new soft tissue. Due to its good wettability, suturability and biological properties, the device becomes well integrated into the surrounding soft tissue.
The products are provided as follows:
- 15 x 20 x 6 mm
- 20 x 40 x 6 mm
- 15 x 20 x 3 mm
- 20 x 40 x 3 mm
## V. Indications for Use
Geistlich Fibro-Gide® is intended for soft tissue augmentation. It is indicated for:
- Localized gingival augmentation to increase keratinized tissue (KT) around teeth and implants
- Alveolar ridge reconstruction for prosthetic treatment
- Recession defects for root coverage
## VI. Comparison of Technological Characteristics
The Indications for Use Statement is identical to the predicate device.
The subject device is identical to the predicate device with respect to materials characteristics, manufacturing and sterilization methods, and packaging. Both the subject and predicate device have identical final product specifications. A comparison of the subject and predicate device is provided in the table below.
| | Subject Device | Geistlich Fibro-Gide® (K171050) |
| --- | --- | --- |
| Material | Porcine collagen | Same |
| Shape | Rectangle | Same |
| Sizes | 15 x 20 x 6 mm | Different |
| | 20 x 40 x 6 mm | |
| | 15 x 20 x 3 mm | |
| | 20 x 40 x 3 mm | |
| Single-Use | Yes | Same |
| Sterilization | Gamma | Same |
The purpose of this submission is to obtain clearance for the use of an alternate supplier of porcine raw material and notify the FDA of the introduction of the additional 3 mm size variants,
Page 5 of 6
{10}
Geistlich Mucografi® and Geistlich Fibro-Gide®
Traditional 510(k)
Geistlich Pharma AG
slight modifications to the device labels and instructions for use, changes to the manufacturing process, and extension to shelf life. These changes do not raise different questions of safety and effectiveness and are addressed by the information summarized within the Performance Data section below.
## VII. Performance Data
Mechanical testing, biocompatibility (K171050), sterilization (K171050), shelf-life (K171050), and non-clinical and clinical performance testing (K171050) from the applicant’s own predicate device was leveraged in support of substantial equivalence.
The following non-clinical data was provided within this submission to demonstrate substantial equivalence:
- results from a suturability study on the 3 mm variants according to methods used in K171050
- results from a viral inactivation study performed in accordance with ISO 22442-3
- results from real-time stability studies performed in accordance with ICH Q1A (R) (extension of studies submitted in support of K171050)
## VIII. Conclusion
The subject device is identical to the predicate device. The addition of a raw material supplier does not raise different questions of substantial equivalence. Therefore, it is concluded that Geistlich Fibro-Gide® is substantially equivalent to the identified predicate device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.