OSSEOSPEED ANGLED ABUTMENT EV

K121810 · Astra Tech AB · NHA · Jan 23, 2013 · Dental

Device Facts

Record IDK121810
Device NameOSSEOSPEED ANGLED ABUTMENT EV
ApplicantAstra Tech AB
Product CodeNHA · Dental
Decision DateJan 23, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

OsseoSpeed™ Angled Abutment EV is intended to be used in conjunction with Astra Tech Implant System EV in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures. The AtlantisTM Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis™ Crown Abutment in Zirconia is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant. Atlantis™ Abutment and Atlantis™ Crown Abutment are compatible with Ø5.4 Astra Tech Implant System EV.

Device Story

OsseoSpeed Angled Abutment EV and Atlantis Zirconia abutments serve as prosthetic interfaces for Astra Tech Implant System EV. Components connect endosseous implants to dental prostheses (bridges, overdentures, or crowns). Abutments are provided in various diameters, angles (20°, 30°), and gingival heights to accommodate anatomical variations. Used in dental clinics by clinicians to secure restorations via screw-retention, cementation, or friction fit. Device provides structural support for final dental restorations, restoring masticatory function and aesthetics for edentulous or partially edentulous patients.

Clinical Evidence

Bench testing only. Mechanical strength validated via dynamic fatigue testing per ISO 14801.

Technological Characteristics

Dental implant abutments; available in 3.6, 4.2, 4.8, and 5.4 mm diameters; 20° and 30° angulations; indexed and non-indexed interfaces. Materials consistent with predicate systems. Sterilized components. Mechanical fixation via abutment screws.

Indications for Use

Indicated for partially or fully edentulous patients requiring maxillary or mandibular dental implant-supported restorations, including bridges, overdentures, or single/multiple tooth prostheses.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 510(k) Summary # JAN 2 3 2013 ## Astra Tech AB OsseoSpeed™ Angled Abutment EV K121810 January 23, 2013 #### ADMINISTRATIVE INFORMATION Manufacturer Name: Astra Tech AB Aminogatan 1, P.O. Box 14 Mölndal, Sweden SE-431-21 +46 31 776 30 00 Telephone: +46 31 776 30 10 Fax: Official Contact: Representative/Consultant: Christina Lewing Head of Regulatory Affairs Linda K. Schulz, BSDH, RDH Kevin A. Thomas, PhD PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 Telephone: +1 (858) 792-1235 Fax: +1 (858) 792-1236 email: lschulz@paxmed.com kthomas@paxmed.com #### DEVICE NAME AND CLASSIFICATION Trade/Proprietary Name: Common Name: Classification Name: Classification Regulation: Product Code: Classification Panel: Reviewing Branch: OsseoSpeed™ Angled Abutment EV Dental implant abutment Endosseous dental implant abutment Class II, 21 CFR 872.3630 NHA Dental Products Panel Dental Devices Branch {1}------------------------------------------------ ## INTENDED USE OsseoSpeed™ Angled Abutment EV is intended to be used in conjunction with Astra Tech Implant System EV in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures. The AtlantisTM Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis™ Crown Abutment in Zirconia is intended for use with an endosseous implant to function as a substructure that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant. Atlantis™ Abutment and Atlantis™ Crown Abutment are compatible with Ø5.4 Astra Tech Implant System EV. ### DEVICE DESCRIPTION OsseoSpeed Angled Abutment EV is designed for multi-unit, screw-retained restorations in a partially or fully edentulous situation. It is provided in three platform diameters (3,6, 4,2 and 4.8 mm) and two angles (20° and 30°). It is intended to fit Astra Tech implants of corresponding diameters cleared under K111287 and K120414. All abutments with a 20°angle are provided in two gingival heights (1.0 and 2.0 mm) with an indexed interface. All abutments with a 30°angle have a non-indexed interface and are provided in two gingival heights (1.0 and 3.0 mm). OsseoSpeed Angled Abutment EV is not compatible with Astra Tech 3.0 mm diameter implants. Atlantis Abutment in Zirconia for OsseoSpeed EV is a custom abutment for prosthesis attachment. Atlantis Crown Abutment in Zirconia for OsseoSpeed EV is fabricated as a custom single-tooth final restoration and is shaped in the design of the final prosthesis onto which porcelain is added. Both Atlantis abutments are provided in a platform diameter of 5,4 mm. #### EQUIVALENCE TO MARKETED DEVICE OsseoSpeed™ Angled Abutment EV is substantially equivalent in indications and design principles to the following legally marketed predicate devices: Astra Tech AB - K111287 Astra Tech, Inc. - K112138 Astra Tech, Inc. - K110356 {2}------------------------------------------------ The purpose of this submission is to expand the OsseoSpeed Plus implant system to include three new components, the OsseoSpeed Angled Abutment EV, the Atlantis Abutment in Zirconia for OsseoSpeed EV, and Atlantis Crown Abutment in Zirconia for OsseoSpeed EV. OsseoSpeed Angled Abutment EV is provided in two angles (20° and 30°). Platform sizes correspond to OsseoSpeed Plus implant diameters cleared in K111287. OsseoSpeed Angled Abutment EV is indicated for multi-unit restorations as are the predicate devices. The original Atlantis Abutment in Zirconia for OsseoSpeed Plus was cleared in K112138 and the Atlantis Crown Abutment in Zirconia for OsseoSpeed in K110356. The interface remains the same as in K112138. A new size of Ø5.4 mm has been added for both designs. Testing of OsseoSpeed™ Angled Abutment EV according to ISO 14801 Dentistry - Implants -Dynamic fatigue test for endosseous dental implants, was performed and the results show that OsseoSpeed™ Angled Abutment EV has sufficient mechanical strength for the intended clinical application and demonstrate that it is substantially equivalent to the listed predicate devices. Overall, OsseoSpeed Angled Abutment EV has the following similarities to the predicate devices: - has the same intended use, - uses the same operating principle, . ● - . incorporates the same basic design, - . incorporates the same or very similar materials, and - has similar packaging and is sterilized using the same materials and processes . {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" are arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 23, 2013 Astra Tech AB C/O Ms. Linda K. Schulz Regulatory Affairs PaxMed International, Limited Liability Company 12264 El Camino Real, Suite 400 SAN DIEGO CA 92130 Re: K121810 Trade/Device Name: OsseoSpeed 110 Angled Abutment EV Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: January 2, 2013 Received: January 3, 2013 Dear Ms. Schulz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The . general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Schulz Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Kwame O. Ulmer Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 510(k) Number: K)21810 OsseoSpeed™ Angled Abutment EV Device Name: Indications for Use: OsseoSpeed™ Angled Abutment EV is intended to be used in conjunction with Astra Tech Implant System EV in fully edentulous or partially edentulous maxillary and/or mandibular arches to provide support for bridges or overdentures. The Atlantis™ Abutment is intended for use with an endosseous implant to support a prosthetic I he Atiantis - I Duanon in intelly edentulous patient. It is intended for use to support single and multiple tooth prostheses, in the mandible or maxilla. The prosthesis can be cemented, screw retained or friction fit to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. The Atlantis™ Crown Abutment in Zirconia is intended for use with an endosseous implant to The Attamily - Crown Fre that also serves as the final restoration, in partially or completely edentulous patients. The prosthesis is screw retained. The abutment screw is intended to secure the crown abutment to the endosseous implant. Atlantis™ Abutment and Atlantis™ Crown Abutment are compatible with Ø5.4 Astra Tech Implant System EV. Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ## Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Sheena A. Green 2013.01.23 16:10:12 -05/00 Division Sign Off) (Division Sign-Off) Division of Anesthestology, General Hospital Division of Anton, Dental Devices 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...