INCLUSIVE ZIRCONIA ABUTMENT BLANKS

K083480 · Inclusive Dental Solutions · NHA · May 5, 2009 · Dental

Device Facts

Record IDK083480
Device NameINCLUSIVE ZIRCONIA ABUTMENT BLANKS
ApplicantInclusive Dental Solutions
Product CodeNHA · Dental
Decision DateMay 5, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3630
Device ClassClass 2

Intended Use

The Inclusive Zirconia Abutment Blank is intended to be used in conjunction with endosseous implants in the maxillary and/or mandibular arch to provide support for crowns, bridges or overdenture prostheses. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. Inclusive Zirconia Abutment Blanks for Zimmer Implants are compatible with Zimmer Screw Vent and Tapered Screw Vent internal hex implants. Inclusive Zirconia Abutment Blanks for 3i Implant are compatible with 3i Certain internal hex implants. Inclusive Zirconia Abutment Blanks for Nobel Biocare Implant are compatible with Nobel Biocare Replace straight and tapered internal-connection implants. Abutments with angulations greater than 20 degrees are contraindicated.

Device Story

Inclusive Zirconia Abutment Blanks are dental components used by clinicians to support crowns, bridges, or overdentures on endosseous implants. Blanks are compatible with specific internal hex implant systems (Zimmer Screw Vent/Tapered Screw Vent, 3i Certain, Nobel Biocare Replace). Clinicians use these blanks to fabricate custom abutments; the final prosthesis is cement-retained. The abutment screw secures the assembly to the implant. These devices provide a restorative interface for patients needing tooth replacement, facilitating functional and aesthetic dental rehabilitation.

Clinical Evidence

No clinical data. Substantial equivalence is based on design similarity and established use of predicate devices.

Technological Characteristics

Zirconia abutment blanks designed for compatibility with specific internal hex dental implant systems. Mechanical fixation via abutment screw. Intended for custom fabrication of dental restorations.

Indications for Use

Indicated for patients requiring dental restoration support via endosseous implants in maxillary or mandibular arches. Supports crowns, bridges, or overdentures. Contraindicated for abutment angulations > 20 degrees.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary (K083480) Inclusive Dental Solutions Submitter: 4141 MacArthur Blvd. Newport Beach, CA 92660 MAY - 5 2009 Contact Person(s): Keith D. Allred, 949-440-2683 (p) / 949-440-2787 (f) and consultants, Greg Minzenmayer & Grant Bullis ## Date of Application: November 14, 2008 Device Name, Type and Classification: Inclusive Zirconia Abutment Blanks, product code 'NHA,'' non-exempt Class II (Sec. 872.3630 - Endosseous dental implant abutment) ## Indications for Usc: The Inclusive Zirconia Abutment Blank is intended to be used in conjunction with endosseous implants in the maxillary and/or mandibular arch to provide support for crowns, bridges or overdenture prostheses. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. Inclusive Zirconia Abutment Blanks for Zimmer Implants are compatible with Zimmer Screw Vent and Tapered Screw Vent internal hex implants. Inclusive Zirconia Abutment Blanks for 3i Implant are compatible with 3i Certain internal hex implants. Inclusive Zirconia Abutment Blanks for Nobel Biocare Implant are compatible with Nobel Biocare Replace straight and tapered internal-connection implants. Abutments with angulations greater than 20 degrees are contraindicated. Substantial Equivalence: The device is substantially equivalent to other legally marketed devices in the United States. Substantially equivalent devices include the following: Biomet 3i Dental Abutments cleared under K072642; for Zimmer, Biomet3i and Nobel Biocare implants cleared under K011028, K063341, and K023113 respectively; Atlantis™ Abutment in Zircnia for Nobel Biocare Replace (K062277); and, Atlantis™ Abutment in Zirconia for 3i Certain (K072483). Safety and Efficacy: The device functions in a similar manner to other comparative devices and the intended use is the same. The differences between comparative devices are minor and do not raise new safety concerns. The effectiveness and suitability to the intended purpose of the device is assured through wide, general use of similar other predicate devices, and demonstrates the safe use of the device to construct dental restorations. {1}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES .. Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a series of interconnected shapes. Public Health Service MAY - 5 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Keith D. Allred Inclusive Dental Solutions 4141 MacArthur Boulevard Newport Beach, California 92660 Re: K083480 Trade/Device Name: Inclusive Zirconia Abutment Blanks Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: April 21, 2009 Received: April 22, 2009 Dear Mr. Allred: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adultcration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting {2}------------------------------------------------ Page 2- Mr. Allred (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Susan Runnes Susan Runner, D.D.S., MA Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Schedule B ### Indications for Use Statement 510(k) Number: K083480 Device Name: Inclusive Zirconia Abutment Blanks Indications for Use: The Inclusive Zirconia Abutment Blank is intended to be used in conjunction with endosseous implants in the maxillary and/or mandibular arch to provide support for crowns, bridges or overdenture prostheses. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. Inclusive Zirconia Abutment Blanks for Zimmer Implants are compatible with Zimmer Screw Vent and Tapered Screw Vent internal hex implants. Inclusive Zirconia Abutment Blanks for 3i Implant are compatible with 3i Certain internal hex implants. Inclusive Zirconia Abutment Blanks for Nobel Biocare Implant are compatible with Nobel Biocare Replace straight and tapered internal-connection implants. Abutments with angulations greater than 20 degrees are contraindicated. Prescription Use X (Part 21 CFR 801 SubpartD) AND/OR Over-The-Counter Use (21 CFR 801 SubpartC) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Rani Mielzynska MSK (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number:
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...