ATLANTIS ABUTMENT IN ZIRCONIA FOR 3I CERTAIN MICROAMINI IMPLANTS, GEMINI ABUTMENT IN ZIRCONIA FOR ASTRA IMPLANTS
K072483 · Atlantis Components, Inc. · NHA · Nov 16, 2007 · Dental
Device Facts
| Record ID | K072483 |
| Device Name | ATLANTIS ABUTMENT IN ZIRCONIA FOR 3I CERTAIN MICROAMINI IMPLANTS, GEMINI ABUTMENT IN ZIRCONIA FOR ASTRA IMPLANTS |
| Applicant | Atlantis Components, Inc. |
| Product Code | NHA · Dental |
| Decision Date | Nov 16, 2007 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 872.3630 |
| Device Class | Class 2 |
Intended Use
The devices covered by this submission are abutments which are placed into a dental implant to provide support for a prosthetic reconstruction. The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant. Please note, highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to the limited strength of the implant fixture.
Device Story
Atlantis Abutment in Zirconia is a dental implant abutment; provides support for cement-retained prosthetic restorations. Device consists of yttria-stabilized tetragonal zirconia polycrystals (Y-TZP) abutment and titanium alloy (Ti-6Al-4V ELI) screw. Placed over dental implant shoulder; secured via screw. Used by dental professionals in clinical settings. Supports single or multiple tooth prostheses in mandible or maxilla. Highly angled abutments (30 degrees) on implants <4mm diameter restricted to anterior region due to fixture strength limitations. Benefits patient by providing stable foundation for prosthetic reconstruction in edentulous sites.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Materials: Yttria-stabilized tetragonal zirconia polycrystals (Y-TZP) per ISO 6972 & 13356; Titanium grade Ti-6Al-4V ELI screw per ASTM F-136. Form factor: Dental abutment for 3i Certain MicroMini implants. Mechanical fixation via screw.
Indications for Use
Indicated for partially or completely edentulous patients requiring single or multiple tooth prosthetic support in the mandible or maxilla. Compatible with 3i Osseotite NT Certain and 3i Osseotite Certain (parallel walled) 3.25mm diameter implants. Contraindicated for posterior use with highly angled (30 degree) abutments on implants <4mm diameter.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
- Atlantis Abutment for 3i Certain MicroMini (K062069)
- Atlantis Abutment in Zirconia for 3i MicroMini (K063734)
- 3i Certain Implant System (K014235)
Related Devices
- K071946 — ATLANTIS, GEMINI AND GEMINI+ ABUTMENTS IN ZIRCONIA FOR ASTRA IMPLANTS · Atlantis Components, Inc. · Oct 5, 2007
- K073540 — ATLANTIS ABUTMENT IN ZIRCONIA FOR BIOHORIZONS IMPLANT · Atlantis Components, Inc. · Feb 29, 2008
- K103759 — ATLANTIS ABUTMENT IN ZIRCONIA FOR OSSEOSPEED TX PROFILE IMPLANT · Astra Tech, Inc. · Apr 8, 2011
- K052070 — ATLANTIS ABUTMENT IN ZIRCONIA, ATLANTIS GEMINI ABUTMENT IN ZIRCONIA, ATLANTIS GEMEINI + ABUTMENT IN ZIRCONIA · Atlantis Components, Inc. · Oct 14, 2005
- K062277 — ATLANTIS ABUTMENT IN ZIRCONIA FOR NOBEL BIOCARE REPLACE · Atlantis Components, Inc. · Dec 20, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Pre-market Notification Page - 00009 - - - - - - - - - - - - - - -
NOV 1 6 2007
## VII. SECTION 10 - 510(K) SUMMARY
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
#### l. Applicant's Name and Address
Atlantis Components Inc. 25 First Street Cambridge, Massachusetts 02141 Telephone Number: 617-661-9799 Fax Number: 617-661-9063 Contact Person: Franklin Uyleman Manager of Quality and Regulatory Affairs
#### 2. Name of Device
| Trade Name: | Atlantis™ Abutment in Zirconia for 3i Certain<br>MicroMini Implant |
|----------------------|------------------------------------------------------------------------|
| Common Name: | Endosseous dental implant abutment |
| Classification Name: | Endosseous dental implant abutment<br>21 CFR 872.3630 Product code NHA |
### Legally Marketed Device to which Equivalence is claimed (Predicate Device) 3.
| Manufacturer | Device | 510(k)<br>Number |
|-----------------------------|---------------------------------------------------|------------------|
| Atlantis Components<br>Inc. | Atlantis Abutment for 3i Certain<br>MicroMini | K062069 |
| Atlantis Components<br>Inc. | Atlantis Abutment in Zirconia for 3i<br>MicroMini | K063734 |
| 3i Implant Innovations | 3i Certain Implant System | K014235 |
#### 4. Description of the Device
The devices covered in this submission are abutments which are placed into the dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for cemented restorations.
{1}------------------------------------------------
Pre-market Page - 00010 - - - - - - - - - - - - - - - -
#### 4. Description of the Device (continued)
The Atlantis™ Abutments in Zirconia for 3i Certain MicroMini Implant is made of biocompatible material, yttria-stabilized tetragonal zirconia polycrystals (Y-TZP) and meets ISO Standards 6972 & 13356). The abutment screws are made from titanium grade Ti-6A1-V ELI and meets ASTM Standard F-136. The abutment is placed over the implant shoulder and is mounted into the implant with a screw. The abutments are compatible with the 3i Osseotite NT® Certain™ and 3i Osseotite® Certain™ (parallel Walled) 3.25mm diameter implants.
#### 5. Intended Use of the Device
The devices covered by this submission are abutments which are placed into a dental implant to provide support for a prosthetic reconstruction. The Atlantis Abutment is intended for use as an accessory to an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
Please note, highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to the limited strength of the implant fixture.
#### 6. Basis for Substantial Equivalence
# The Atlantis™ Abutments in Zirconia for 3i Certain MicroMini Implant are
substantially equivalent in intended use, material, design and performance to the Atlantis Abutments for 3i Certain MicroMini Implant cleared under K062069, the Atlantis Abutment in Zirconia 3i Certain MicroMini Implant cleared under K063734 and for the 3i Certain Implant System cleared under K014235.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal for the Department of Health and Human Services (HHS). The seal features an abstract eagle design with three stylized lines representing the eagle's body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle emblem.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 6 2007
Atlantis Components, Incorporated C/O Ms. Betsy A. Brown Consultant B.A. Brown & Associates 8944 Tamaroa Terrace Skokie, Illinois 60076
Re: K072483
Trade/Device Name: Atlantis™ Abutment in Zirconia for 3i Certain MicroMini Implants Regulation Number: 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: August 22, 2007 Received: September 4, 2007
Dear Ms. Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Neil R. Oglin
Chiu Lin, Ph.D.
Chiu Lin, Ph.D Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Pre-market Notification Page - 000227
## Indications for Use
510(k) Number (if Known) KD72483
Device Name: Atlantis ™ Abutment in Zirconia for 3i Certain MicroMini Implants
Indication for Use:
The Atlantis Abutment is intended for use with an endosseous implant to support a prosthetic device in a partially or completely edentulous patient. It is intended for use to support single and multiple tooth prosthesis, in the mandible or maxilla. The prosthesis can be cement retained to the abutment. The abutment screw is intended to secure the abutment to the endosseous implant.
This device is compatible with the following manufacturers' implant systems: 3i Osseotite NT® Certain™ and 3i Osseotite® Certain™ (parallel Walled) 3.25mm diameter implants.
Please note: This device may be used in an early load situation, but is dependent on the specific implant system and protocol used by the dental professional.
Also, highly angled abutments (i.e. 30 degrees) on implants with diameters less than 4 mm are intended for the anterior region of the mouth and are not intended for the posterior region due to limited strength of the implant.
Susan Roars
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Prescription Use X (Part 21 CFR 801 SubpartD)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)