FIT TEST C&B

K110050 · Voco GmbH · ELW · Mar 28, 2011 · Dental

Device Facts

Record IDK110050
Device NameFIT TEST C&B
ApplicantVoco GmbH
Product CodeELW · Dental
Decision DateMar 28, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

Fit Test C&B is intended for the following applications: Control of precise fit, marginal tightness and marginal gap of - full or partial crowns - bridges - inlays/onlays - cast post cores/post crowns - metal bases - implant-supported crowns and bridges - attachments and other secondary parts of dentures

Device Story

Fit Test C&B is a dental impression material used by clinicians to verify the accuracy of fit for dental restorations. The material is applied to the internal surface of a restoration (e.g., crown, bridge, inlay) or the prepared tooth structure. Upon seating the restoration, the material records the contact points and marginal gaps. The clinician visually inspects the resulting impression to identify areas of interference or insufficient contact. This process allows for precise adjustments to the restoration before final cementation, ensuring optimal marginal integrity and patient comfort.

Clinical Evidence

Bench testing only.

Technological Characteristics

Impression material; chemical composition consistent with dental materials regulated under 21 CFR 872.3660. Formulated for high-precision surface replication to detect marginal gaps and fit discrepancies in dental prosthetics.

Indications for Use

Indicated for dental professionals to assess the fit, marginal tightness, and marginal gaps of various dental restorations including crowns, bridges, inlays/onlays, cast post cores, metal bases, implant-supported restorations, and denture attachments.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Dr. Thorsten Gerkensmeier Regulatory Affairs VOCO GmbH Anton-Flettner-Strasse 1-3 D-27472 Cuxhaven GERMANY . MAR 2 8 2011 Re: K110050 Trade/Device Name: Fit Test C&B Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: January 4, 2011 Received: January 7, 2011 Dear Dr. Gerkensmeier: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ ## Page 2 - Dr. Gerkensmeier Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH /CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Nh for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology . General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {2}------------------------------------------------ ## Indications for Use Statement | 510(k) Number: | K110050 | |----------------|--------------| | Device Name: | Fit Test C&B | Indications for Use: Fit Test C&B is intended for the following applications: Control of precise fit, marginal tightness and marginal gap of - full or partial crowns - bridges - inlays/onlays - cast post cores/post crowns - metal bases - implant-supported crowns and bridges - attachments and other secondary parts of dentures | Prescription Use | X | OR | Over-The-Counter Use | |------------------|---|----|----------------------| |------------------|---|----|----------------------| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sasa Runos (Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K110050 ...
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...