GC BLUE SILICONE

K120521 · GC America, Inc. · ELW · May 10, 2012 · Dental

Device Facts

Record IDK120521
Device NameGC BLUE SILICONE
ApplicantGC America, Inc.
Product CodeELW · Dental
Decision DateMay 10, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

GC BLUE SILICONE is a silicone impression material for fit check by detecting the high spots and pressure points of ceramic crowns, ceramic bridges, ceramic inlays and the occlusal surface.

Device Story

GC BLUE SILICONE is a two-part silicone impression material (base and catalyst) supplied in a cartridge; used by dental professionals for fit checking of ceramic restorations (crowns, bridges, inlays) and occlusal surfaces. The material is applied to the restoration or tooth surface to identify high spots and pressure points; the clinician evaluates the resulting impression to adjust the restoration for proper fit. Benefits include improved accuracy of dental prosthetic seating and reduced patient discomfort from ill-fitting restorations.

Clinical Evidence

Bench testing only. Performance evaluated against ISO 4823:2000(E) standards for light-bodied consistency materials. Results: detail reproduction ≤20µm; elastic recovery ≥96.5%; strain in compression 2.0-20%.

Technological Characteristics

Two-part silicone impression material (base and catalyst). Light-bodied consistency. Complies with ISO 4823:2000(E) standards. Supplied in 62g (48mL) cartridges with mixing tips. Shelf life: 2 years; storage 15-25°C.

Indications for Use

Indicated for use as a dental impression material to detect high spots and pressure points on ceramic crowns, bridges, inlays, and occlusal surfaces.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K120521. Image /page/0/Picture/1 description: The image shows the letters 'GCI' in a bold, sans-serif font. The letters are black and the background is white. There are quotation marks on either side of the letters. MAY 1 0 2012 GC AMERICA INC. 3737 WEST 127TH STREET ALSIP. ILLINOIS 60803 TEL (708) 597-0900 FAX (708) 371-5103 # Section 5 - 510(k) Summary - 1. Submitter Information: GC AMERICA INC. 3737 W. 127" Street Alsip, IL 60803 | Contact Person: | Mark Heiss, D.D.S. | |-----------------|--------------------| | Phone: | (708) 926-3090 | | Fax: | (708) 597-6222 | Date Prepared: February 17, 2012 - 2. Device Name: | Proprietary Name: | GC BLUE SILICONE | |------------------------|----------------------------| | Classification Name: | Dental Impression material | | Device Classification: | 872.3660 | | Product Code: | ELW | - 3. Predicate Devices: | Product | Applicant | 510(k) No. | Code No | |-----------------------|-----------------|------------|---------| | Fit Checker Advanced | GC AMERICA, INC | K110871 | ELW | | Fit Checker II | GC AMERICA, INC | K032289 | ELW | | GC Fusion (EXA'lence) | GC AMERICA, INC | K041398 | ELW | #### 4. Description of Device: GC BLUE SILICONE is available in a 2-part cartridge device in which part 1 is a base silicone paste and part 2 is a catalyst silicone. #### 5. Indications for Use: GC BLUE SILICONE is a silicone impression material for fit check by detecting the high spots and pressure points of ceramic crowns, ceramic bridges, ceramic inlays and the occlusal surface. - 6. Description of Safety and Substantial Equivalence: The applicant device, GC BLUE SILICONE is equivalent to predicate device, Fit Checker II, in its intended use. The applicant device, GC BLUE SILICONE, is equivalent to the predicate device, Fit Checker Advanced and GC Fusion (EXA'lence) in its chemical compositions. Image /page/0/Picture/19 description: The image contains two logos. The logo on the left is a diamond shape with text inside that reads "CHECKED TO AS 14181 BY POWER SAFE". The logo on the right has the letters "DTA" stacked vertically with the word "MEMBER" above it. {1}------------------------------------------------ #### 7. Test Methods | | Property | Standards | Test methods | Requirements | |---|---------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------------------|--------------| | 1 | Detail<br>reproduction<br>(µm) | ISO 4823: 2000(E)<br>Classification: Light bodied<br>consistency | Detail reproduction was measured by the detail<br>reproduction test method in ISO 4823 | 20µm | | 2 | Elastic<br>Recovery<br>(%) | ISO 4823: 2000(E)<br>Classification: Light bodied<br>consistency | Elastic recovery was measured by the Elastic<br>recovery test method in ISO4823 | ≥96.5 | | 3 | Strain in<br>compression<br>(%) | ISO 4823: 2000(E)<br>Classification: Light bodied<br>consistency | Strain in compression was measured by the<br>Strain in compression test method in ISO 4823 | 2.0-20 | ## Test standards and methods based on ISO standards - 8. Shelf Life Evaluation and Storage Conditions: - Shelf Life 2 years - - Store in a cool and dark place. 15-25°C (60-77.0°F) - - 9. Packaging Two cartridge package : Cartridge 62g (48mL) x 1, Mixing Tip II S x 2 Mixing Tip II SS x 2 ### Optional - · 1. CARTRIDGE DISPENSER II - 2. MIXING TIP II SS Refill Package Accessories necessary for proper functioning of the device: Mixing tip II S {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of several curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mark Heiss, DDS Director - New Business Development, Academic and Regulatory Affairs GC America, Incorporated 3737 West 127th Street Alsip, Illinois 60803 MAY 1 0 2012 Re: K120521 Trade/Device Name: GC Blue Silicone Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: February 17, 2012 Received: February 21, 2012 Dear Dr. Heiss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ ### Page 2 - Dr. Heiss Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. th for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health ### Enclosure {4}------------------------------------------------ Section 4 - Indications for Use Statement Indications for Use 12052 510(k) Number (if known): Device Name: _________________________________________________________________________________________________________________________________________________________________ GC BLUE SILICONE Indications for Use: GC BLUE SILICONE is a silicone impression material for fit check by detecting the high spots and pressure points of ceramic crowns, ceramic bridges, ceramic inlays and the occlusal surface. Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Page 4. Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: Page 4.1 of 4.1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...