The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.
Device Story
The LOCATOR Angled Abutment is a titanium alloy (Ti-6Al-4V ELI) dental abutment used to connect endosseous implants to prosthetic restorations. It features a 15-degree angle and a TiN coating. The device is used by dental professionals in a clinical setting to restore masticatory function in patients with full or partial, fixed or removable restorations. The device is provided non-sterile and requires end-user steam sterilization. Compatibility with specific implant systems from BioHorizons and Implant Direct is achieved by replicating the OEM's design specifications for the implant-to-abutment connection and fixation screws. The clinician selects the appropriate abutment based on the implant system, attaches it to the implant, and uses it to support the final prosthetic restoration. The device benefits patients by providing a standardized, reliable interface for implant-retained prosthetics.
Clinical Evidence
Bench testing only. No clinical data was provided. Equivalence was established through engineering analysis of OEM design specifications, leveraging previous fatigue testing (ISO 14801), biocompatibility (ISO 10993), and MR compatibility testing (ASTM F2052, F2213, F2182, F2119).
Technological Characteristics
Material: Ti-6Al-4V ELI (ASTM F136) with TiN coating. Design: 15° angled abutment. Connection: Conical/Internal Hex (system-specific). Sterilization: Moist heat (end-user). Connectivity: None (mechanical interface).
Indications for Use
Indicated for patients requiring attachment of full or partial, fixed or removable, dental restorations to endosseous implants to restore masticatory function.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
K071838 — NEOSS VARIOUS TITANIUM ABUTMENTS · Neoss, Ltd. · Oct 19, 2007
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Zest Anchors, LLC
Maleata Hall
Director Regulatory Affairs
2875 Loker Ave E
Carlsbad, California 92010
October 15, 2025
Re: K252944
Trade/Device Name: LOCATOR® Angled Abutment
Regulation Number: 21 CFR 872.3630
Regulation Name: Endosseous Dental Implant Abutment
Regulatory Class: Class II
Product Code: NHA
Dated: September 5, 2025
Received: September 15, 2025
Dear Maleata Hall:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252944 - Maleata Hall
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K252944 - Maleata Hall
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252944
Device Name
LOCATOR® Angled Abutment
Indications for Use (Describe)
The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable restorations retained by endosseous implants to restore masticatory function for the patient.
IMPLANT COMPATIBILITY
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| | | 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| | Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| | | 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| | Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| | | 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| | | 3.5, 4.3, 5.0 | Drive GM | N/A | | Conical / Grand Morse |
| | | 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| | | 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| | Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| | | 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 | |
| | | 4.7 | Legacy 1 | N/A | 4.5 | |
| | BioHorizons | 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| | | 4.6 | Tapered Plus | | | |
| | | 3.8 | Tapered Internal | | | |
| | | 3.0, 3.8 | Tapered Tissue Level | | | |
| | | 4.6 | Tapered Short | | | |
| | | 4.2 | Tapered PTG | | | |
| | | 5.2 | Tapered Pro | Green | 4.5 | |
| | | 5.8 | Tapered Plus | | | |
| | | 4.6 | Tapered Internal | | | |
| | | 4.6 | Tapered Tissue Level | | | |
| | | 5.8 | Tapered Short | | | |
FORM FDA 3881 (7/17)
PISC Publishing Services (301) 443-6740
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6° Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6° Morse Taper |
| LAA 15°, BioHorizons Internal Hex | BioHorizons | 3.0, 3.4, 3.8 | Tapered Plus, Tapered 3.0, Mount Free Tapered Internal | 3.0 (Gray) | 3.0 | Internal Hex |
| LAA 15°, BioHorizons Conical | BioHorizons (Camlog) | 3.3 | CONELOG | Narrow (Gray) | N/A | Conical |
| | | 3.3, 3.8 | Tapered Pro Conical | | | |
| | | 3.8 | CONELOG | Regular (Yellow) | N/A | |
| | | 3.8, 4.2, 4.6, 5.2 | Tapered Pro Conical | | | |
| | | 4.2, 4.6, 5.2 | Tapered Short Conical | | | |
| LAA 15°, Implant Direct Legacy 3.0 | Implant Direct | 3.2 | Legacy3 (Blue) | Internal Hex Connection (Blue) 3.0 | 3.0 | Internal Hex |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Publishing Services (301) 443-6740
{5}
ZD ZEST DENTAL SOLUTIONS
Maleata Hall
Director of Regulatory Affairs
Email: maleata.hall@zestdent.com
Telephone: +1 (442) 244-4835 Ext. 5650
K252944
# 510(K) Summary – K252944
# LOCATOR Angled Abutment
## i. General Information on Submitter
Applicant: Zest Anchors, LLC
Address: 2875 Loker Avenue East
Carlsbad, CA 92010 USA
Telephone: (800) 262-2310
Contact Person: David Lin
Contact Title: Sr. Regulatory Affairs Specialist
Email: regulatoryaffairs@zestdent.com
Date Prepared: October 09, 2025
## ii. General Information on Device
Proprietary Name: LOCATOR Angled Abutment
Common Name: Dental Implant Abutment
Regulation Number: 21 CFR 872.3630
Classification Name: Endosseous dental implant abutment
Regulatory Class: Class II
Product Code: NHA (Abutment, Implant, Dental, Endosseous)
## iii. Predicate Device
| Predicate Device | 510(k) Number |
| --- | --- |
| LOCATOR Angled Abutment: Zest Anchors, LLC | K250721 |
| Reference Devices | 510(k) Number |
| BioHorizons CAD/CAM Abutments, BioHorizons Implant Systems, Inc. | K151621 |
| Tapered Pro Conical Implant System, BioHorizons Implant Systems, Inc. | K240187 |
| Legacy and Custom InterActive Titanium Abutments, Implant Direct LLC. | K192218 |
## iv. Description of Device
The purpose of this submission is to expand the indications of the LOCATOR® Angled Abutment product line to include use with new dental implant systems. This was achieved through collaboration with the implant manufacturers BioHorizons and Implant Direct to prove compatibility of the design of the implant to abutment connection, and equivalence of design specification and performance characteristics.
Previously, the LOCATOR Angled Abutment has been cleared by FDA in applications K233587, K243272, and K250721 for use with various dental implant systems by the manufacturers BioHorizons, Implant Direct, Implant Logistics, Neodent, Nobel Biocare, Straumann, and ZimVie (Zimmer). These LOCATOR Angled Abutments are also designed and
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{6}
ZD ZEST DENTAL SOLUTIONS
K252944
cleared for use with removable LOCATOR® Attachment Systems (K072878) and LOCATOR FIXED® (K213391), intended for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants in the mandible or maxilla. This 510(k) submission expands the indications of the LOCATOR Angled Abutment product line with new connections compatible with the following implant systems from BioHorizons (Camlog) and Implant Direct.
- BioHorizons Tapered 3.0 (Internal Hex)
- BioHorizons Tapered Pro Conical Narrow platform (CONELOG)
- BioHorizons Tapered Pro Conical Regular platform (CONELOG)
- BioHorizons Tapered Short Conical Regular platform (CONELOG)
- Implant Direct Legacy 3, 3.0mmD
Compatibility of the new LOCATOR Angled Abutments and screws with either the BioHorizons or Implant Direct implants has been demonstrated through documented collaboration with the OEM. This was achieved by utilizing the OEM's design specifications, which were shared with Zest to create a LOCATOR Angled Abutment and screw that is equivalent to the implant manufacturer's own abutment device in how it interfaces with their respective implant system(s). Additionally, the LOCATOR Angled Abutment parameters deemed critical to ISO 14801 fatigue testing are also equivalent to the manufacturer's cleared abutment specifications, and therefore do not create a new worst case construct or condition. An equivalent specification is deemed as being identical to or within the OEM's defined specification.
## Indication for Use
The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient.
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{7}
ZD
ZEST DENTAL
SOLUTIONS
K252944
IMPLANT COMPATIBILITY
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15°, Straumann, BLX | Straumann | 3.5, 3.75, 4.0, 4.5 | BLX | Regular Base | 2.9 | Bone Level |
| 5.0, 5.5, 6.5 | BLX | Wide Base | 2.9 | Bone Level |
| LAA 15°, Nobel (NP) | Nobel | 3.5 | NobelActive, NobelParallel CC, NobelReplace CC | Narrow Platform | 3.0 | Conical |
| Implant Direct | 3.2, 3.7 | InterActive | N/A | 3.0 | Conical |
| 3.2, 3.7, 4.2, 4.7 | Simply Iconic | N/A | 3.0 | Conical |
| LAA 15°, Nobel (RP) | Nobel | 4.3 | NobelActive, NobelParallel CC, NobelReplace CC | Regular Platform | 3.5 | Conical |
| Implant Direct | 4.3, 5.0 | InterActive | N/A | 3.4 | Conical |
| 4.7, 5.2, 5.7 | Simply Iconic | N/A | 3.4 | Conical |
| LAA 15°, Neodent (GM) | Neodent | 3.5, 3.75, 4.0, 5.0, 6.0 | Helix GM | N/A | 3.0 | Conical / Grand Morse |
| 3.5, 4.3, 5.0 | Drive GM | N/A | Conical / Grand Morse |
| 3.5, 3.75, 4.0, 5.0 | Titamax GM | N/A | Conical / Grand Morse |
| LAA 15°, TSV | ZimVie | 3.7 & 4.1 Green | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 3.5 | Internal Hex |
| 4.7 Purple | Trabecular Metal, Tapered Screw-Vent, Screw-Vent, Advent | N/A | 4.5 | Internal Hex |
| Implant Direct | 3.7, 4.2 | Legacy 1, 2, 3, 4 | N/A | 3.5 | Internal Hex |
| 4.7, 5.2 | Legacy 2, 3, 4 | N/A | 4.5 |
| 4.7 | Legacy 1 | N/A | 4.5 |
| BioHorizons | 4.2, 4.6 | Tapered Pro | Yellow | 3.5 | Internal Hex |
| 4.6 | Tapered Plus |
| 3.8 | Tapered Internal |
| 3.0, 3.8 | Tapered Tissue Level |
| 4.6 | Tapered Short |
| 4.2 | Tapered PTG |
| 5.2 | Tapered Pro | Green | 4.5 |
| 5.8 | Tapered Plus |
| 4.6 | Tapered Internal |
| 4.6 | Tapered Tissue Level |
| 5.8 | Tapered Short |
| LAA 15°, Implant Logistics, Implant-One 300 Series | Implant Logistics | 3.5, 4.1, 4.5 | Implant One | 300 Series | 2.75 | Internal Conical 6" Morse Taper |
| LAA 15°, Implant Logistics, Implant-One 400 Series | 4.0, 4.5, 5.5 | Implant One | 400 Series | 3.25 | Internal Conical 6" Morse Taper |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{8}
ZD
ZEST DENTAL
SOLUTIONS
K252944
| Zest Model | Implant Mfg | Implant Diameters (Ø) mm | Implant System Name | Implant Platform Name | Platform Diameter (Ø) mm | Connection Type |
| --- | --- | --- | --- | --- | --- | --- |
| LAA 15", BioHorizons Internal Hex | BioHorizons | 3.0, 3.4, 3.8 | Tapered Plus, Tapered 3.0, Mount Free Tapered Internal | 3.0 (Gray) | 3.0 | Internal Hex |
| LAA 15", BioHorizons Conical | BioHorizons (Camlog) | 3.3 | CONELOG | Narrow (Gray) | N/A | Conical |
| | | 3.3, 3.8 | Tapered Pro Conical | | | |
| | | 3.8 | CONELOG | Regular (Yellow) | N/A | |
| | | 3.8, 4.2, 4.6, 5.2 | Tapered Pro Conical | | | |
| | | 4.2, 4.6, 5.2 | Tapered Short Conical | | | |
| LAA 15", Implant Direct Legacy 3.0 | Implant Direct | 3.2 | Legacy3 (Blue) | Internal Hex Connection (Blue) 3.0 | 3.0 | Internal Hex |
Page 4 of 9
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{9}
ZD
ZEST DENTAL
SOLUTIONS
K252944
# v. Predicate Device Comparison
The following table compares the Indications for Use and key technological characteristics of the subject and predicate device:
Device Comparison Table
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc. LOCATOR Angled Abutment K252944 | Zest Anchors, Inc. LOCATOR Angled Abutment K250721 | BioHorizons Implant Systems, INC. BioHorizons CAD/CAM Abutments K151621 | BioHorizons Implant Systems, INC. Tapered Pro Conical Implant System K240187 | Implant Direct LLC. Custom Legacy and Custom Interactive Titanium Abutments, K192218 |
| Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | BioHorizons connection is compatible with Tapered Plus, Tapered 3.0, Mount Free Tapered Internal, CONELOG, and Tapered Pro Conical connection | BioHorizons connection is compatible with CONELOG, Tapered Short and Tapered Pro Conical connection | Implant Direct connection is compatible with legacy3, 3.0mmD connection |
| Indications for Use | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. A complete Implant Manufacturer's compatibility table is provided in the Instructions For Use. | The LOCATOR Angled Abutment is indicated for the attachment of full or partial, fixed and removable, restorations retained by endosseous implants to restore masticatory function for the patient. The complete Indications for Use Statement with OEM implant compatibilities is provided in the 510(k) Summary for K250721. | BioHorizons CAD/CAM Abutments are dental abutments placed onto a dental implant to provide support for dental prosthetic restorations. The abutments include: 1) Titanium abutment blanks with a pre-machined implant connection where the upper portion may be custom-milled in accordance with a patient-specific design using CAD/CAM techniques; and 2) Titanium bases with a pre-machined implant connection upon which a CAD/CAM designed superstructure may be fitted to complete a two-piece dental abutment. The abutments include an abutment screw for fixation to the underlying implant. The abutments may be used for single-unit (single-tooth) or multiple-unit (bridges and bars) restorations and are compatible for use with BioHorizons Internal and Tapered Internal implant systems and Zimmer® Dental Screw-Vent® and Tapered Screw-Vent® implants with 3.5mm, 4.5mm and 5.7mm internal hex-connection mating platform diameters. All digitally designed abutments and/or copings for use with BioHorizons CAD/CAM Abutments are intended to | BioHorizons Tapered Pro Conical dental implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. These dental implants may be restored immediately (1) with a temporary prosthesis that is not in functional occlusion or (2) when splinted together for multiple tooth replacement or when stabilized with an overdenture supported by multiple implants. BioHorizons Tapered Short Conical dental implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. These dental implants must be restored using delayed loading, for single tooth replacement, or may be used with a terminal or intermediate abutment for fixed or removable bridgework or for overdentures. Tapered Short Conical implants should be used only when there is not enough space for a longer implant. If the ratio of crown length to implant length is | Custom Titanium Abutments are customizable devices intended for use with dental implants in the maxillary and/or mandibular arches to provide support for crowns or bridges for edentulous or partially edentulous patients. • Custom Titanium Abutment for narrow (3.2mmD, 3.3mmD) implants: Indicated for single-tooth replacement of mandibular central and lateral incisors and maxillary lateral incisors. Also indicated for multiple tooth replacements. • Custom Titanium Abutment for short (8mm) 3.7mmD implants: Indicated for tooth replacement of mandibular and maxillary central and lateral incisors. |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{10}
ZD
ZEST DENTAL
SOLUTIONS
K252944
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc. LOCATOR Angled Abutment K252944 | Zest Anchors, Inc. LOCATOR Angled Abutment K250721 | BioHorizons Implant Systems, INC. BioHorizons CAD/CAM Abutments K151621 | BioHorizons Implant Systems, INC. Tapered Pro Conical Implant System K240187 | Implant Direct LLC. Custom Legacy and Custom InterActive Titanium Abutments, K192218 |
| Reason for Predicate/Reference | n/a | Design of Locator Angled Abutment | BioHorizons connection is compatible with Tapered Plus, Tapered 3.0, Mount Free Tapered Internal, CONELOG, and Tapered Pro Conical connection | BioHorizons connection is compatible with CONELOG, Tapered Short and Tapered Pro Conical connection | Implant Direct connection is compatible with legacy3, 3.0mmD connection |
| | | | be sent to a BioHorizons-validated milling center for manufacture. BioHorizons abutments designed using CAD/CAM techniques must fulfill the BioHorizons allowable range of design parameters. | unfavorable, the biomechanical risk factors have to be considered and appropriate measures have to be taken by the dental professional. BioHorizons conical dental prosthetic components connected to the endosseous dental implants are intended for use as an aid in prosthetic rehabilitations of the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed abutments for use with Conical CAD/CAM Ti Blanks and Ti Bases are to be sent to a BioHorizons validated milling center for manufacture. | |
| Design | | | | | |
| Abutment Cuff Height | 2.5 - 7.5 mm | 2.5 - 7.5 mm | 0-12.1mm | 0.5mm Min. | - |
| Abutment Type | Angled | Angled | Straight, Angled | Straight, Angled | Straight, Angled |
| Abutment connection | Conical Internal Hex | TorcFit (Hexalobe) Internal Hex Internal Tri-Channel Conical Grand Morse Taper Conical Conical Taper | Internal hex | Conical | Internal hex |
| Abutment Angle | 15° | 15° | 0°-30° | 0°-30° | Maximum angle of 30° from the axis of the implant |
| Sterilization Method | Moist heat end user sterilization | Moist heat end user sterilization | N/A; provided non-sterile, steam sterilization in accordance with qualified cycles as specified in the Instructions for Use | Non-sterile | Sold non-sterile, Steam sterilization by end user |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{11}
ZD
ZEST DENTAL
SOLUTIONS
K252944
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- | --- |
| | Zest Anchors, Inc.
LOCATOR Angled Abutment
K252944 | Zest Anchors, Inc.
LOCATOR Angled Abutment
K250721 | BioHorizons Implant Systems, INC.
BioHorizons CAD/CAM Abutments
K151621 | BioHorizons Implant Systems, INC.
Tapered Pro Conical Implant System
K240187 | Implant Direct LLC. Custom Legacy
and Custom InterActive Titanium
Abutments,
K192218 |
| Reason for
Predicate/Reference | n/a | Design of Locator Angled Abutment | BioHorizons connection is compatible
with Tapered Plus, Tapered 3.0, Mount
Free Tapered Internal, CONELOG, and
Tapered Pro Conical connection | BioHorizons connection is
compatible with CONELOG,
Tapered Short and Tapered Pro
Conical connection | Implant Direct connection is compatible
with legacy3, 3.0mmD connection |
| Compatibility | BioHorizons
(Tapered Plus, Tapered 3.0,
Mount Free Tapered Internal,
CONELOG Narrow
CONELOG Regular
Tapered Pro Conical Narrow,
Tapered Pro Conical Regular,
Tapered Short Conical Regular)
Implant Direct Legacy3, 3.0mmD | Straumann BLX
Zimmer (ZimVie) TSV 3.5
Zimmer (ZimVie)TSV 4.5
Nobel RP
Nobel NP
Neodent Grand Morse
Implant Direct Legacy
Implant Direct InterActive
Implant Direct Simply
BioHorizons
(Tapered Internal Implant System &
LASER-LOK)
Iconic Implant-One system,
300&400 series | BioHorizons
Tapered Plus,
Tapered 3.0,
Mount Free Tapered Internal,
CONELOG Narrow,
CONELOG Regular,
Tapered Pro Conical Narrow,
Tapered Pro Conical Regular,
Tapered Short Conical Regular | BioHorizons
CONELOG Narrow
CONELOG Regular
Tapered Pro Conical Narrow,
Tapered Pro Conical Regular,
Tapered Short Conical Regular | Implant Direct
Legacy3, 3.0mmD |
| Materials | | | | | |
| Abutment | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI |
| Abutment
Coating | TiN | TiN | Anodized, None | Anodized, None | - |
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{12}
ZD ZEST DENTAL SOLUTIONS
K252944
## vi. Summary of Non-Clinical Performance Testing
Confirmation of the compatibility of the new Locator Angled Abutments with their corresponding compatible implant systems from BioHorizons and Implant Direct was based on the requirements provided in the implant manufacturer's drawing specifications, as cleared in their respective 510(k) applications. The critical features were identified on the OEM component specifications (abutment-to-implant connection specifications and abutment fixation screws specifications), which are required for proper function. Using engineering analysis, critical areas for abutment-to-implant engagement were evaluated using the abutment and abutment screw drawings, provided by the manufacturers, and all critical tolerances functionally verified with the OEM implant.
No new V&V is required and performance is deemed equivalent to the OEM's FDA cleared device through the replication of critical specifications provided by the implant manufacturer and referenced 510k clearances.
The LOCATOR Angled Abutments are made of titanium alloy Ti-6Al-4V ELI, conforming to ASTM F136, and have a TiN (Titanium Nitride) coating, identical to the predicate device K250721. TiN coating performance was tested per ASTM F1044 and ASTM F1147 in K233587 and being leveraged in the current submission.
The packaging of the LOCATOR Angled Abutments is similar to the packaging of the predicate device, consisting of the LOCATOR angled abutment placed in a vial and sealed in a polybag, along with a parallel post (class I device) used to visually confirm the desired abutment orientation. Additionally, some packaged configuration will include the LOCATOR Angled Abutment Screw packaged in a separate vial and sealed in the same polybag as the abutment and parallel post. Packaging and shipping validation testing was completed previously where the LOCATOR Angled Abutment worst case device and packaging were undamaged after the test, as desired. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
The cleaning and sterilization are identical to the predicate device cleared under K250721. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
MR compatibility testing was conducted previously per ASTM F2052-21, ASTM F2213-17, ASTM F2182-19, ASTM F2119-07, and FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" on abutment and implant components made of Ti-6Al-4V and designed with similar features as the LOCATOR Angled Abutments of this 510(K) submission. The tests that were conducted are Force: static magnetic field induced displacement force, Torque: static magnetic field induced torque, Heating: Radiofrequency field (RF) induced heating, Image Quality: susceptibility induced image artifacts, Heating: Gradient field induced heating, and Vibration: Gradient field induced vibration. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
{13}
ZD ZEST DENTAL SOLUTIONS
K252944
An assessment for biocompatibility per ISO 10993-1 was conducted previously using testing from K072878 and additional cytotoxicity testing per ISO 10993-5 cleared under K233587. The results have been leveraged for the LOCATOR Angled Abutment where engineering analysis established that the subject device does not create a new worst-case scenario.
No other new testing was performed as a part of this submission for the determination of substantial equivalence.
## vii. Substantial Equivalence
As this is a modification to the manufacturer's own cleared and marketed device, the risk based analysis and results of the design control activities performed provide reasonable assurance that the subject devices have demonstrated substantial equivalence to the predicate devices in that they share the same intended use and principles of operation, use the same materials and manufacturing processes, and utilize the same fundamental design including identical prosthetic attachment features; thus the indications for use has been expanded to include compatibility with additional implant systems from Implant Direct and BioHorizons.
Zest Anchors, LLC. 2875 Loker Avenue East, Carlsbad, CA 92010, USA
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.