K233285 · Kuraray Noritake Dental, Inc. · NHA · May 24, 2024 · Dental
Device Facts
Record ID
K233285
Device Name
KATANA Zirconia ONE For IMPLANT
Applicant
Kuraray Noritake Dental, Inc.
Product Code
NHA · Dental
Decision Date
May 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
KATANA Zirconia ONE For IMPLANT is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations. For the SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible. KATANA Zirconia ONE For IMPLANT is used in combination with the TiBase and Sirona Dental CAD/CAM System. KATANA Zirconia ONE For IMPLANT cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Device Story
Device is pre-sintered zirconia block for milling into ceramic mesostructures; used as component of two-piece hybrid abutment or hybrid abutment crown. Input: digital design data via Sirona Dental CAD/CAM System (CEREC/inLab software). Transformation: milling of zirconia block into patient-specific shape; cementation to pre-manufactured titanium base (TiBase) using PANAVIA SA Cement Universal. Output: custom hybrid abutment/crown for endosseous dental implants. Used in dental clinics by trained professionals. Healthcare provider uses CAD/CAM software to design mesostructure based on patient anatomy; output provides functional and aesthetic restoration. Benefits: patient-specific fit, restoration of oral function/aesthetics, and compatibility with established implant systems.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including ISO 6872:2015 (ceramic materials) and ISO 14801:2016 (dynamic fatigue testing for endosseous dental implants). Biocompatibility evaluated per ISO 10993-1 and ISO 7405. MRI safety assessed via scientific rationale and literature review.
Technological Characteristics
Pre-sintered zirconia block (zirconium oxide ceramic per ISO 6872:2015). Dimensions: one block size with small/large access hole options. Connectivity: integrated into Sirona Dental CAD/CAM System libraries. Sterilization: validated per ISO 17665-1. Design parameters: abutment post height 4.0-16.7mm, angulation 0-20 degrees, wall thickness 0.8-9.5mm.
Indications for Use
Indicated for partially or fully edentulous patients requiring single cement-retained restorations. SBL 3.3 L titanium base use restricted to maxillary lateral incisors and mandibular lateral/central incisors.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Sirona Dental CAD/CAM System with CEREC Chairside software (K193408)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Kuraray Noritake Dental Inc. Yasujiro Ohara General Manager Tokiwabashi Tower, 2-6-4, Otemachi Chiyoda-ku, Tokyo 100-0004 JAPAN
May 24, 2024
Re: K233285
Trade/Device Name: KATANA Zirconia ONE For IMPLANT Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: April 26, 2024 Received: April 26, 2024
Dear Yasujiro Ohara:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{3}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233285
Device Name KATANA Zirconia ONE For IMPLANT
Indications for Use (Describe)
KATANA Zirconia ONE For IMPLANT is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations.
For the SBL 3.3 L titanium bases, the indication is replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible.
KATANA Zirconia ONE For IMPLANT is used in combination with the TiBase and Sirona Dental CAD/CAM System. KATANA Zirconia ONE For IMPLANT cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Compatible implant systems and TiBase are as follows.
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{5}------------------------------------------------
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{6}------------------------------------------------
24th May 2024 Date:
# 510(k) Summary
## 1. 510(k) owner (submitter)
| 1) Name | Kuraray Noritake Dental Inc. |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2) Address | 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan |
| 3) Contact person | Yasujiro Ohara<br>General Manager<br>Quality Assurance Department |
| 4) Contact person in US | Manabu Suzuki<br>Director<br>KURARAY AMERICA, INC.<br>32 Old Slip, 7th Floor, New York, NY 10005<br>Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676<br>Fax: (212)-867-3543 |
| Name of Device | |
| 1) Trade / Proprietary name | KATANA Zirconia ONE For IMPLANT |
|-----------------------------|-------------------------------------------------------------|
| 2) CFR Number | 21 CFR 872.3630 |
| 3) Device Class | Class II |
| 4) Product code | |
| • Primary Product code | NHA (Abutment, Implant, Dental, Endosseous) |
| • Secondary Product code | PNP (Dental Abutment Design Software for Dental Laboratory) |
5) Panel
2. N
Dental
K200191
Dentsply Sirona
#### 3. Predicate and Reference devices 3-1. Primary predicate device
1) Trade / Proprietary name
Sirona Dental CAD/CAM System with CEREC Chairside software K193408 Abutment, Implant, Dental, Endosseous (21 CFR section 872.3630. Product code: NHA) Dental Abutment Design Software For Dental Laboratory (21 CFR section 872.3630. Product code: PNP) Dentsply Sirona
4) Applicant name
2) 510(k) number
3) Classification name
#### 3-2. Reference devices Reference device 1
1) Trade / Proprietary name 2) 510(k) number 3) Classification name
## 4) Applicant name
## Reference device 2
1) Trade / Proprietary name 2) 510(k) number 3) Classification name
4) Applicant name
## Reference device 3
1) Trade / Proprietary name 2) 510(k) number
KATANA Zirconia Block K190436 Powder, Porcelain (21 CFR section 872.6660. Product code: EIH) Kuraray Noritake Dental Inc.
(21 CFR section 872.3630. Product code: PNP)
Sirona Dental CAD/CAM System with InLab Software
Dental Abutment Design Software For Dental Laboratory
IPS e.max CAD Abutment Solutions- extra systems K191382
{7}------------------------------------------------
| 3) Classification name | Abutment, Implant, Dental, Endosseous<br>(21 CFR section 872.3630. Product code: NHA |
|------------------------|--------------------------------------------------------------------------------------|
| 4) Applicant name | Ivoclar Vivadent, AG |
## Reference device 4
| 1) Trade / Proprietary name | PANAVIA SA Cement Universal |
|-----------------------------------------------------------|--------------------------------------------|
| 2) 510(k) number | K183537 |
| 3) Classification name | Cement, Dental/Agent, Tooth Bonding, Resin |
| (21 CFR section 872.3275/872.3200. Product code: EMA/KLE) | |
| 4) Applicant name | Kuraray Noritake Dental Inc. |
## Reference device 5
| 1) Trade / Proprietary name | Sirona Dental CAD/CAM System |
|----------------------------------------------|---------------------------------------|
| 2) 510(k) number | K181520 |
| 3) Classification name | abutment, implant, dental, endosseous |
| (21 CFR section 872.3630. Product code: NHA) | |
| 4) Applicant name | Dentsply Sirona |
{8}------------------------------------------------
## 4. Device Description
The KATANA Zirconia ONE For IMPLANT mesostructure material (conforming to ISO 6872:2015) is a pre-sintered zirconia block to be used as the ceramic mesostructure component of a two-piece titanium base abutment. KATANA Zirconia ONE For IMPLANT is compatible with the specified TiBase and the CAD/CAM component of the Sirona Dental CAD/CAM system (K193408, K200191) as identified in the Indications for Use. This device is further processed by the trained professional to make individually designed mesostructure that are milled into the desired shape of a hybrid abutment or hybrid abutment crown.
The Zirconia block has identical composition as our own previously cleared reference device, KATANA Zirconia Block (K190436) under product code "EIH".
The mesostructure material of KATANA Zirconia ONE For IMPLANT is available in shades, A1, A2, A3, A3.5, B1, B2, C1, C2, D2 and NW for flexibility and application variety to meet individual patient needs. The mesostructure material of KATANA Zirconia ONE For IMPLANT is available with one block size and two sizes of access hole, which is small (S) and large (L).
The abutment must be sterilized after the cementation of the CAD/CAM patient matched mesostructure on the pre-manufactured titanium base component. PANAVIA SA Cement Universal (K183537) is used as the cement to set the mesostructure material to TiBase when the mesostructure material of the subject device set to TiBase prior to sterilization.
Dentsply Sirona has issued a letter of authorization indicating that the subject mesostructure material of the subject device can be selected in combination with the CAD/CAM system. Kuraray Noritake Dental and Dentsply Sirona have a business agreement for adding this new material to the Sirona CAD/CAM system. Kuraray Noritake Dental has worked with Dentsply Sirona to implement their new material into the Sirona CAD/CAM system libraries.
KATANA Zirconia ONE For IMPLANT which is the subject of this premarket notification consists of:
- Sirona Dental CAD/CAM System with CEREC Chairside Software or with inLab Software
- Sirona TiBase
- . KATANA Zirconia ONE mesostructured blocks
Compatible titanium bases are shown in below table.
| Ti Base | | Compatible System | | | | |
|--------------------|------------------------|-------------------|--------------------|------------------------------------|----------|---------|
| Name | Dentsply<br>Sirona Ref | Interface<br>size | Manufacturer/ Line | Platform | Diameter | |
| TiBase NB RS 3.5 L | 6282474 | L | Nobel Biocare | NP | 3.5 | |
| TiBase NB RS 4.3 L | 6282482 | L | | RP | 4.3 | |
| TiBase NB RS 5.0 L | 6282490 | L | | WP | 5.0 | |
| TiBase NB RS 6.0 L | 6282508 | L | | 6.0 | 6.0 | |
| TiBase NB A 4.5 L | 6308188 | L | Nobel Biocare | Nobel Active, | NP | 3.5 |
| TiBase NB A 5.0 L | 6308253 | L | | NobelReplace<br>Conical Connection | RP | 4.3/5.0 |
| TiBase S BL 3.3 L | 6308154 | L | Straumann | Bone Level | NC | 3.3 |
| TiBase S BL 4.1 L | 6308337 | L | | Bone Level | RC | 4.1/4.8 |
| TiBase BC 3.4 S | 6308048 | S | Biomet 3i | Certain | 3.4 | 3.4 |
| TiBase B C 4.1 L | 6308097 | L | | Certain | 4.1 | 4.1 |
| TiBase B C 5.0 L | 6308121 | L | | Certain | 5.0 | 5.0 |
| TiBase Z TSV 3.5 L | 6282581 | L | Zimmer | Tapered Screw-Vent | 3.5 | 3.7/4.1 |
| TiBase Z TSV 4.5 L | 6282599 | L | | Tapered Screw-Vent | 4.5 | 4.7 |
| TiBase Z TSV 5.7 L | 6282607 | L | | Tapered Screw-Vent | 5.7 | 6 |
| | | | | | | Table 4: Compatibility of intended Sirona TiBase with KATANA Zirconia ONE For IMPLANT |
|--|--|--|--|--|--|---------------------------------------------------------------------------------------|
|--|--|--|--|--|--|---------------------------------------------------------------------------------------|
{9}------------------------------------------------
The subject device is procced within the following design parameter range.
Image /page/8/Figure/2 description: The image shows three diagrams comparing a hybrid abutment and a hybrid abutment crown. Each diagram includes measurements of less than 20 degrees and greater than or equal to 0.8mm. The diagrams illustrate the differences in the angle and size of the abutments and crowns.
OR
The minimum wall thickness of the restoration must be 0.8 mm.
The parameters of this product for fabricating prosthetics are as following:
| Item | Parameter | Remark |
|------------------------------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| a) Abutment post<br>height*1 | 4.0 mm ~ 16.7 mm | Minimum value is followed by FDA's recommendation.<br>Maximum value set by considering product's dimension and<br>shrinkage of block. |
| b) Angulation | 0° ~ 20° | Angulation over 20° is contraindication at TiBase. |
| c) Wall thickness | 0.8 mm ~ 9.5 mm | Range of wall thickness is limited by allowance of access hole,<br>the dimensions and shrinkage of block. |
| d) Diameter | 3.3 mm ~ 6.0 mm*2 | This value set by Dentsply Sirona. It was confirmed that the<br>mesostructure of subject device could applied within those<br>range of diameter. |
| e) Gingival height | 1.0 mm*2 | This value set by Dentsply Sirona. |
*1 The abutment post height is considered by FDA to be the "stump" portion above the gingival collar, to which the restorations attach.
*2 Parameters of diameter and gingival height are depended on a selected TiBase.
{10}------------------------------------------------
## 5. Substantial Equivalence Discussion
## 5-1. Indications for Use / Intended Use
It is described that the indications for use of the subject device set as followings:
KATANA Zirconia ONE For IMPLANT is intended for use in partially or fully edentulous mandibles and maxillae in support of single cement-retained restorations.
For the SBL 3.3 L titanium bases, the indication is restricted to the replacement of single lateral incisors in the maxilla and lateral and central incisors in the mandible.
KATANA Zirconia ONE For IMPLANT is used in conjunction with TiBase and Sirona Dental CAD/CAM System. KATANA Zirconia ONE For IMPLANT cemented to the TiBase is recommended for two-piece hybrid abutments for single tooth restorations and hybrid abutment crowns, used in conjunction with endosseous dental implants.
Compatible implant systems and TiBase are as follows:
| Implant system<br>(Manufacturer) | Platform | Sirona TiBase Name | Size |
|--------------------------------------------------------------------|----------------------|--------------------|------|
| Replace, Replace Select<br>(Nobel Biocare) | NP | TiBase NB RS 3.5 L | L |
| | RP | TiBase NB RS 4.3 L | L |
| | WP | TiBase NB RS 5.0 L | L |
| | 6.0 | TiBase NB RS 6.0 L | L |
| Nobel Active<br>NobelReplace Conical Connection<br>(Nobel Biocare) | NP | TiBase NB A 4.5 L | L |
| | RP | TiBase NB A 5.0 L | L |
| Bone Level<br>(Straumann) | NC (3.3 mm) | TiBase S BL 3.3 L | L |
| | RC (4.1 mm / 4.8 mm) | TiBase S BL 4.1 L | L |
| Certain<br>(Biomet 3i) | 3.4 | TiBase B C 3.4 S | S |
| | 4.1 | TiBase B C 4.1 L | L |
| | 5.0 | TiBase B C 5.0 L | L |
| Tapered Screw-Vent<br>(Zimmer) | 3.5 | TiBase Z TSV 3.5 L | L |
| | 4.5 | TiBase Z TSV 4.5 L | L |
| | 5.7 | TiBase Z TSV 5.7 L | L |
{11}------------------------------------------------
This Indications for Use is substantially equivalence device (IPS e.max CAD Abutnent Solutions- extra systems, K.191382 / the predicate device (Sirona Dental CAD/CAM System with CEREC Chairside software (K193408)) the reference device (Sirona Dental CAD/CAM System (K181520) ). It is the description as follows that we conducted the comparison between the subject device and the reference device.
| Subject device:<br>KATANA Zirconia ONE For IMPLANT | Primary predicate device:<br>Sirona Dental CAD/CAM System with CEREC<br>Chairside software (K193408) | Reference device:<br>Sirona Dental CAD/CAM System<br>(K181520) | Reference device:<br>IPS e.max CAD Abutment Solutions- extra<br>systems (K191382) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| KATANA Zirconia ONE For IMPLANT is<br>intended for use in partially or fully edentulous<br>mandibles and maxillae in support of single<br>cement-retained restorations. | The Sirona Dental CAD/CAM System with<br>CEREC Chairside Software is intended for use<br>in partially or fully edentulous mandibles and<br>maxillae in support of single or multiple-unit<br>cement retained restorations. | The Sirona Dental CAD/CAM System is<br>intended for use in partially or fully edentulous<br>mandibles and maxillae in support of single or<br>multiple-unit cement retained restorations. | IPS e.max CAD Abutment Solutions is<br>intended for use in partially or fully<br>edentulous mandibles and maxillae in support<br>of single cement-retained restorations. |
| For the SBL 3.3 L titanium bases, the indication<br>is restricted to the replacement of single lateral<br>incisors in the maxilla and lateral and central<br>incisors in the mandible.<br>KATANA Zirconia ONE For IMPLANT is used<br>in combination with the TiBase and Sirona<br>Dental CAD/CAM System. KATANA Zirconia<br>ONE For IMPLANT cemented to the TiBase is<br>recommended for two-piece hybrid abutments for<br>single tooth restorations and hybrid abutment<br>crowns, used in conjunction with endosseous<br>dental implants. | For the AT TX 3.0 S, BH 3.0 S, SSO 3.5 L, and<br>SBL 3.3 L titanium bases, the indication is<br>restricted to the replacement of single lateral<br>incisors in the maxilla and lateral and central<br>incisors in the mandible.<br>The system consists of three major parts:<br>TiBase, inCoris mesostructure and CAD/CAM<br>software. Specifically, the inCoris mesostructure<br>and TiBase components make up a two- piece<br>abutment which is used in conjunction with<br>endosseous dental implants to restore the<br>function and aesthetics in the oral cavity. The<br>inCoris mesostructure may also be used in<br>conjunction with the Camlog Titanium base<br>CAD/CAM (types K2244.XXXX) (K083496) in<br>the Camlog Implant System. The CAD/CAM<br>software is intended to design and fabricate the<br>inCoris mesostructure. | For the BH 3.0 S, SSO 3.5 L and S/L 3.3 L<br>titanium bases, the indication is restricted to the<br>replacement of single lateral incisors in the<br>maxilla and lateral and central incisors in the<br>mandible.<br>The system consists of three major parts: TiBase,<br>inCoris mesostructure, and CAD/CAM software.<br>Specifically, the inCoris mesostructure and<br>TiBase components make up a two-piece<br>abutment which is used in conjunction with<br>endosseous dental implants to restore the<br>function and aesthetics in the oral cavity. The<br>inCoris mesostructure may also be used in<br>conjunction with the Camlog Titanium base<br>CAD/CAM (types K2244.XXXX) (K083496) in<br>the Camlog Implant System. The CAD/CAM<br>software is intended to design and fabricate the<br>inCoris mesostructure. | The system comprises three parts:<br>- IPS e.max CAD ceramic structure<br>- Ti base<br>- CAD/CAM system.<br>The IPS e.max CAD ceramic structure<br>cemented to the Ti base is recommended for<br>two-piece hybrid abutments for single tooth<br>restorations and hybrid abutment crowns, used<br>in conjunction with endosseous dental<br>implants. |
{12}------------------------------------------------
| Subject device:<br>KATANA Zirconia ONE For IMPLANT | | | Primary predicate device:<br>Sirona Dental CAD/CAM System with CEREC<br>Chairside software (K193408) | Reference device:<br>Sirona Dental CAD/CAM System (K181520) | Reference device:<br>IPS e.max CAD Abutment Solutions- extra systems (K191382) | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------------------------------------------------------------|----------|----------|----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Compatible implant systems and Ti-bases are as follows. | | | The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems | The inCoris mesostructure and TiBase two-piece abutment is compatible with the following implant systems | The compatible Implant systems, titanium bases and CAD/CAM systems are shown below:<br>-Implant Systems:<br>- Dentsply Sirona: AstraTech OsseoSpeed,<br>Frialit/Xive (K130999, K013867)<br>- BioHorizons Implant System: Internal Connection (K143022, K071638, K093321, K042429),<br>- Osstem: TS Implant System (K121585)<br>- Straumann: Tissue Level RN/WN (K061176)<br>- Nobel Biocare: Branemark (K022562)<br>- Zimmer: Tapered Screw-Vent (K061410)<br>- Camlog: Camlog Screw-Line, Conelog Screw-Line, iSy (K083496, K113779, K133991)<br>- CAD/CAM Systems: Sirona Dental CAD/CAM System (K181520)<br>- Titanium Bases | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Implant system (Manufacturer) | Platform | Diameter | Sirona Ti-base Name | Size | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Replace,<br>Replace Select<br>(Nobel Biocare) | NP | 3.5 | TiBase<br>NB<br>RS 3.5 | | Manufacturer Name of Implant System Implant Size Platform Diameter Nobel Biocare Replace NP 3.5 RP 4.3 Active NP 3.5 RP 4.3/5.0 Branemark NP 3.3 RP 3.75/4.0 Synocta NN (3.5mm) 3.3 RN (4.8mm) 3.3/4.1/4.8 Straumann WN (6.5mm) 4.8 Bone Level NC (3.3mm) 3.3 RC (4.1mm/4.8mm 4.1/4.8 Osseospeed 3.5/4.0 3.5 S / 4.0 S 4.5/5.0 4.5/5.0/5.0 S Dentsply Sirona Implants 3.4 3.4 3.8 3.8 Xive 4.5 4.5 5.5 5.5 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | RP | 4.3 | TiBase<br>NB<br>RS 4.3 | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | WP | 5.0 | TiBase<br>NB<br>RS 5.0 | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | 6.0 | 6.0 | TiBase<br>NB<br>RS 6.0<br>L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Nobel Active,<br>NobelReplace<br>Conical<br>Connection<br>(Nobel Biocare) | NP | 3.5 | TiBase<br>NB A<br>4.5 L | L | Manufacturer Name of Implant System Implant Size Platform Diameter Nobel Biocare Replace NP 3.5 RP 4.3 Active NP 3.5 RP 4.3/5.0 Branemark NP 3.3 RP 3.75/4.0 Synocta NN (3.5mm) 3.3 RN (4.8mm) 3.3/4.1/4.8 Straumann WN (6.5mm) 4.8 Bone Level NC (3.3mm) 3.3 RC (4.1mm/4.8mm 4.1/4.8 Osseospeed 3.5/4.0 3.5 S / 4.0 S 4.5/5.0 4.5/5.0/5.0 S Dentsply Sirona Implants 3.4 3.4 3.8 3.8 Xive 4.5 4.5 5.5 5.5 | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | RP | 4.3/5.0 | TiBase<br>NB A<br>5.0 L | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Bone Level<br>(Straumann) | NC | 3.3 | TiBase<br>S BL<br>3.3 L | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | RC | 4.1/4.8 | TiBase<br>S BL<br>4.1 L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Certain<br>(Biomet 3i) | 3.4 | 3.4 | TiBase<br>B C<br>3.4 S | S | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | 4.1 | 4.1 | TiBase<br>B C<br>4.1 L | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | 5.0 | 5.0 | TiBase<br>B C<br>5.0 L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Tapered<br>Screw-Vent<br>(Zimmer) | 3.5 | 3.7/4.1 | TiBase<br>Z TSV<br>3.5 L | L | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
{13}------------------------------------------------
| Subject device:<br>KATANA Zirconia ONE For IMPLANT | | | | | Primary predicate device: | Sirona Dental CAD/CAM System with CEREC<br>Chairside software (K193408) | | Reference device:<br>Sirona Dental CAD/CAM System<br>(K181520) | | | | Reference device:<br>IPS e.max CAD Abutment Solutions- extra<br>systems (K191382) | | | | | | | | | |
|----------------------------------------------------|--|-----|----------------|-----------|---------------------------|-------------------------------------------------------------------------|--------------|----------------------------------------------------------------|------------------|-----------------------------------------------------|--------------|-----------------------------------------------------------------------------------|---------------|--------------------|----------------------|------|----------------------|------------|--|--|--|
| | | 4.5 | 4.7 | TiBase | | | mplant Siz | | Manufacture | | Implant Size | | Camlog TiBase | | | | | | | | |
| | | | | ZTSV | Manufacturer | | Platforn | Diameter | | | Platform | | | | | | | | | | |
| | | | | 4.5 L | | | 36 | ર 6 | | | 36 | | | | | | | | | | |
| | | 5.7 | 6 | TiBase | | | 4.2 | 4.2<br>4.8 | | | 4.2 | 4.2<br>4.8 | | | | | | | | | |
| | | | | ZTSV | | | 4.8<br>54 | | | | 4.8<br>5.4 | 54 | | | | | | C2244,3348 | | | |
| | | | | 5.7 L | Dentsply Sirona Implants | Ankylos | C/X | A. B. C. D | | Ankylos | C/X | A. B. C. D | | | | | | 7244.384 | | | |
| | | | | | | | 3.0 | 3.0 | | | 3.4 | 3.25 | | | | | | CD3AA 434 | | | |
| | | | | | | sseospeed TX | 3.5/4/ | 3.5/4/1 | | | | 3.75 | | | | | | 2244.504 | | | |
| | | | | | | | 4.5/5.1 | 4.5/5 0 | | | 4.1 | 4.1 | | | | | | | | | |
| | | | | | | | 3.4 | 3.25<br>3.75 | | Osscotite | | 3/4 | | | | | | | | | |
| | | | | | | | | | | | | 5.0 | | | | | | | | | |
| | | | | | | | 4 1 | 4.1 | | | 5.0 | 4/2 | | | | | | | | | |
| | | | | | | | | 3/4 | | | | 3.25 | | | | | | | | | |
| | | | | | | | 5.0 | 5.0 | Biomet 3i | | | | | | | | | | | | |
| | | | | | | | 4/5 | | | 3.4 | 4/3 | | | | | | 2242-382 | | | | |
| | | | | Biomet 3i | | | 3.25 | | | | 3/4/3 | | | | | | | | | | |
| | | | | | 3.4 | 4/3 | | Certain | | 4.0 | | | | | | | | | | | |
| | | | | | | 3/4/3 | | | | 4/5/4 | | | | | AD/CAM, for @ 4.3 mm | | | | | | |
| | | | | | Certain | | 4.0 | | | | 5/4 | | | | | | | | | | |
| | | | | | | | 4.1 | 4/5/4 | | | | 5.0 | | | | | | | | | |
| | | | | | | | ડાવ | | | 5.0 | 4/5 | | | | | | C2242,5020 | | | | |
| | | | | | | | | 5.0 | | | | 3.7/4.1 | | Sy 3.8 / 4.4 / 5.0 | | | | 92244.440 | | | |
| | | | | | | | 5.0 | 4/5 | | | 4.5 | 4.7 | | | | 1.8/ | 5 man GM 2.0 mes | P2244,442 | | | |
| | | | | | | | 35 | 3.7/4.1 | | PI FLEMENT SPI | | | | | | 44 | 2 miss. GM O.R micro | P2244.5008 | | | |
| | | | | | Zimmer | apered Screw-Ve | 4.7<br>4.5 | | | | | 3.5 | | | | | | P2244.5020 | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | | | | | | | 25 | 3.5 | Osstem / Hiossen | | 4.5 | 4 5 | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | | | | | Thommen Medic | | | 4.5 | | | | | | | | | | | | | |
| | | | | | | | | | | Osstem TS Implant Syste | Mini | 3.5<br>4.0/4.5/5.0/6.0/7.0 | | | | | | | | | |
| | | | | | | Osstem TS Implant System | More | 3.5 | | | | | | | | | | | | | |
| | | | | | | liossen Implant Syst | | 4.0/4.5/5.0/6.0/7.0 | | | | | | | | | | | | | |
| | | | | | | | | | Manufacturer | | | | | | | | | | | | |
| | | | | | | | | | | | | | | | | | | | | | |
| | | | | | | | | | | | | Diameter | | | | | | | | | |
| | | | | | Manufacturer | | Implant Size | | | Tapered 3.0.Tanered n | | 3.0/3.4/3.1 | | | | | | | | | |
| | | | | | | | | Diameter | | | | | | | | | | | | | |
| | | | | | | | | 3.0/3.4/3.8 | | | | | | | | | | | | | |
| | | | | | | | | 3.0 | | | | 3.0/3.8 | | | | | | | | | |
| | | | | | | | | 4.6 | | | | | | | | | | | | | |
| | | | | | | | | 3.0/3.8 | | | | 3.5/4.0 | | | | | | | | | |
| | | | | | | | | 3.5 | | | | | | | | | | | | | |
| | | | | | | 3.5/4.0 | | anered internal Taner | | | | | | | | | | | | | |
| | | | BioHorizon | | | 5.8 | | ternal dental im | | 4.0 | | | | | | | | | | | |
| | | | Internal Conne | | | 4.6 | | | | 4.0/5.0 | | | | | | | | | | | |
| | | | | | | | | Tapered internal, Tapere | | 58 | | | | | | | | | | | |
| | | | | | | | | 4.0<br>4.0/5.0 | | ternal tissue leve | | | | | | | | | | | |
| | | | | | | | | | | ternal dental implant<br>Single stage dental implar | | 5.0/6.0 | | | | | | | | | |
| | | | | | | | | 5.8 | | | | | | | | | | | | | |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.