KATANA Zirconia Block

K190436 · Kuraray Noritake Dental, Inc. · EIH · May 23, 2019 · Dental

Device Facts

Record IDK190436
Device NameKATANA Zirconia Block
ApplicantKuraray Noritake Dental, Inc.
Product CodeEIH · Dental
Decision DateMay 23, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

KATANA Zirconia Block is used for the fabrication of the ceramic restorations (frameworks, FCZ crowns, FCZ bridges (14Z L only), inlays, onlays and veneers).

Device Story

Pre-sintered yttrium-stabilized zirconia block; designed for CEREC milling systems. Used by dental professionals in clinical settings to fabricate dental restorations (crowns, bridges, inlays, onlays, veneers). Milling process follows specific system instructions; followed by end-user sintering at 1550°C-1560°C. Provides durable, aesthetic ceramic restorations; benefits patient through restoration of tooth function and appearance.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 6872:2015 standards. Subject device demonstrated compliance with requirements for uniformity, freedom from extraneous materials, flexural strength (>= 500MPa), chemical solubility (< 100µg/cm²), and coefficient of thermal expansion. No clinical data provided.

Technological Characteristics

Material: Yttrium-stabilized zirconia (ZrO2 + Y2O3 + pigments). Form: Pre-sintered block. Sizes: 12Z, 14Z, 14Z L. Sintering temperature: 1550°C-1560°C. Classification: Type II, Class 4a per ISO 6872:2015. Non-sterile. Compatible with CEREC milling systems.

Indications for Use

Indicated for fabrication of ceramic restorations including frameworks, full-contour zirconia (FCZ) crowns, FCZ bridges (14Z L size only), inlays, onlays, and veneers in dental patients.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. May 23, 2019 Kuraray Noritake Dental Inc. Yasujiro Ohara Manager, Quality Assurance Department Ote Center Bldg. 7F Chiyoda-ku, 100-0004 JAPAN Re: K190436 Trade/Device Name: KATANA Zirconia Block Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: February 20, 2019 Received: February 22, 2019 Dear Yasujiro Ohara: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Malvina B Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known): K190436 Device Name: KATANA Zirconia Block Indications for Use: KATANA Zirconia Block is used for the fabrication of the ceramic restorations (frameworks, FCZ crowns, FCZ bridges (14Z L only), inlays, onlays and veneers). > Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {4}------------------------------------------------ Date: May 13, 2019 #### 510(k) number: K190436 # 510(k) Summary #### 3-1. 510(k) owner (submitter) - 1) Name - 2) Address - 3) Contact person 4) Contact person in US Kuraray Noritake Dental Inc. 1621 Sakazu, Kurashiki, Okayama 710-0801, Japan Yasujiro Ohara Manager Quality Assurance Department Manabu Suzuki Director KURARAY AMERICA, INC. 33 Maiden Lane, 6th Floor, New York, NY 10038 Tel: (212)-986-2230 (Ext. 115) or (800)-879-1676 Fax: (212)-867-3543 #### 3-2. Name of Device - 1) Trade / Proprietary name - 2) Classification name - 3) Common name #### 3-3. Predicate device KATANA Zirconia Block #### 3-4. Reference device KATANA Zirconia #### KATANA Zirconia Block Porcelain powder for clinical use (21 CFR section 872.6660. Product code: EIH) Dental ceramics 510(k) Number Classification: Product Code: 21 CFR Section: Applicant: K170986 Porcelain powder for clinical use EIH 872.6660 Kuraray Noritake Dental Inc. 510(k) Number Classification: Product Code: 21 CFR Section: Applicant: K143439 Porcelain powder for clinical use EIH 872.6660 Kuraray Noritake Dental Inc. {5}------------------------------------------------ #### 3-4. Device Description In this submission, we add a new size [14Z L (a: 17.8×b: 19.2×c: 40.0 mm)] to KATANA Zirconia Block (K170986) and expand to the same indications as KATANA Zirconia (K143439). Device description of KATANA Zirconia Block is modified as follows. The subject device, KATANA Zirconia Block is a pre-sintered zirconia block composed of yttrium stabilized zirconia, which is designed for CEREC milling systems. The subject device is meant to be end-user sintered after milling for the fabrication of dental restorations. KATANA Zirconia Block STML has 3 sizes (12Z, 14Z and 14Z L) and has 14 shade variations: (A1, A2, A3, A3.5, A4, B1, B2, B3, C1, C2, C3, D2, D3, NW). KATANA Zirconia Block ST has 1 size: (12Z) and has 1 shade variation: (CL). STML consists of 4 gradated shade layers and ST is mono color. They are classified as a Type II and Class 4a) material by ISO 6872. #### 3-5. Statement of Indications for use KATANA Zirconia Block is used for the fabrication of the ceramic restorations (frameworks, FCZ crowns, FCZ bridges (14Z L only), inlays, onlays and veneers). #### 3-6. Substantial Equivalence Discussion 1) Indications for use The indications of the subject device, the predicate device and the reference device are as listed on the following table. | | Subject device<br>(a modification to the<br>predicate device) | Predicate device<br>(K170986) | Reference device<br>(K143439) | |-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade<br>name | KATANA Zirconia Block | KATANA Zirconia Block | KATANA Zirconia | | Size | 12Z: 15.3×19.2×20.2<br>(mm)<br>14Z: 17.8×19.2×20.2<br>(mm)<br>14Z L: 17.8×19.2×40.0<br>(mm) | 12Z: 15.3×19.2×20.2<br>(mm)<br>14Z: 17.8×19.2×20.2<br>(mm) | $\phi$ 98.5×14 to $\phi$ 98.5×<br>22 (mm) | | Indication<br>s for use | KATANA Zirconia Block is<br>used for the fabrication of the<br>ceramic restorations<br>(frameworks, FCZ crowns,<br>FCZ bridges (14Z L), inlays,<br>onlays and veneers). | KATANA Zirconia Block is<br>used for the fabrication of the<br>ceramic restorations (FCZ<br>crowns, inlays, onlays and<br>veneers). | "KATANA Zirconia" is used<br>for the fabrication of the<br>ceramic restorations<br>(frameworks, FCZ crowns,<br>FCZ bridges, inlays, onlays<br>and veneers). | | Table 3-6-1. Indications for useof the subject device, the predicate device and the reference device | | | | | |------------------------------------------------------------------------------------------------------|--|--|--|--| |------------------------------------------------------------------------------------------------------|--|--|--|--| The indications for useof the subject device was written up based on the predicate device (KATANA Zirconia Block) and the reference device (KATANA Zirconia). This new size is larger than predicate device, and smaller than reference device (K143439), whose size is 98.5×14 to 22 (mm).Therefore, the indications for useof the subject device is substantially equivalent to the predicate device and the reference device. ### 2) Chemical ingredients All ingredients in the subject device have been used in the predicate device and the reference device. Regarding the above predicate device and the reference device, there have not been any reported problems or recalls according to the post-market adverse event reporting requirements in the US. Accordingly, it was concluded that the subject device is substantially equivalent in biocompatibility to the predicate device and the reference device. 3) Technological characteristics Table 3-6-2 compares the subject device to the predicate device and the reference device with respect to technological characteristics. {6}------------------------------------------------ | | Subject device<br>(a modification to the<br>predicate device) | predicate device<br>(K170986) | Remark | |----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade<br>name | KATANA Zirconia Block | KATANA Zirconia Block | Subject device and<br>predicate device are the<br>same except for new size.<br>This new size is larger than<br>predicate device, and<br>smaller than reference<br>device (K143439), whose<br>size is $\phi$ 98.5 × 14 to 22<br>(mm), and comply with<br>ISO 6872: 2015 as shown<br>in 3-7. From the above,<br>Subject device is<br>substantially equivalent to | | Type | ST&STML | ST&STML | | | Shapes | Block | Block | | | Sizes | 12Z: 15.3×19.2×20.2<br>(mm)<br>14Z: 17.8×19.2×20.2<br>(mm)<br>14Z L: 17.8×19.2×40.0<br>(mm) | 12Z: 15.3×19.2×20.2<br>(mm)<br>14Z: 17.8×19.2×20.2<br>(mm) | | | Shades | A1, A2, A3, A4, B1, B2,<br>B3, C1, C2, C3, D2, D3,<br>NW, CL | A1, A2, A3, A4, B1, B2,<br>B3, C1, C2, C3, D2, D3,<br>NW, CL | in 3-7. From the above,<br>Subject device is<br>substantially equivalent to | | Materials | ZrO2+Y2O3+Pigments | ZrO2+Y2O3+Pigments | predicate device and | | Sintering<br>status | 1550°C-1560°C | 1550°C-1560°C | reference device. | | Sterility<br>status | Non-sterile | Non-sterile | | | Milling<br>technique | Place the block into the<br>milling machine; then<br>begin the milling process<br>following the milling<br>systems technical<br>instructions. | Place the block into the<br>milling machine; then<br>begin the milling process<br>following the milling<br>systems technical<br>instructions. | | Table 3-6-2. Comparison of Technological characteristics 4) Non clinical Perfomance Testing <Bench testing> Physical and mechanical properties of the subject device were evaluated according to ISO 6872: 2015. According to ISO 6872:2015, the subject device is classified into the following: - Type II: All other forms of ceramic products. - Class 4: a) Monolithic ceramic for three-unit prostheses involving molar restoration. The results of comparative study performed according to ISO 6872: 2015 were indicated Table 3-6-3. The results show the subject device, the predicate device and the reference device comply with ISO 6872: 2015. Therefore, the performance of the subject device is substantially equivalent to the predicate device and the reference device. {7}------------------------------------------------ | Section | | | | Subject device<br>(a modification to the<br>predicate device) | Predicate device<br>(K170986) | Reference device<br>(K143439) | |-------------------|--------------------------------------------|-----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|-------------------------------------|------------------------------------| | | | | Requirement | KATANA Zirconia<br>Block (14Z L) | KATANA Zirconia<br>Block (12Z, 14Z) | KATANA Zirconia<br>Disc (ST, STML) | | 5.1<br>Uniformity | | | The inorganic pigment(s) used to<br>produce the color of a fired dental<br>ceramic and any organic colorants (for<br>color coding) shall be uniformly<br>dispersed throughout the dental<br>ceramic material and in powdered<br>ceramic products, no segregation of<br>the pigment(s) shall take place when<br>the powder is mixed as in 7.1.3. Check<br>by visual inspection. | COMPLIES | COMPLIES | COMPLIES | | 5.2 | Freedom<br>from<br>extraneous<br>materials | 5.2.1 | Dental ceramic materials shall be free<br>from extraneous materials when<br>assessed by visual inspection. | COMPLIES | COMPLIES | COMPLIES | | | | 5.2.2 | Dental ceramic materials shall not<br>have an activity concentration of more<br>than 1.0 Bq·g-1 of 238U. Test in<br>accordance with 7.2.2. | COMPLIES | COMPLIES | COMPLIES | | 5.4 | Physical<br>and<br>chemical<br>properties | Flexural<br>strength | Equal to or greater than 500MPa. | COMPLIES | COMPLIES | COMPLIES | | | | Chemical<br>solubility | Less than 100µg/cm² | COMPLIES | COMPLIES | COMPLIES | | | | The<br>coefficient<br>of thermal<br>expansion | The coefficient of thermal expansion<br>of the ceramics shall not deviate by<br>more than 0.5 × 10-6 K-1 from the<br>value stated by the manufacturer.<br>(The value stated by the<br>manufacturer : 9.8×10-6K-1) | COMPLIES | COMPLIES | COMPLIES | ## Table 3-6-3. The results of comparative study performed according to ISO 6872:2015 "COMPLIES" means that test value was in the acceptable range to pass the testing in compliance with the requirements of ISO 6872: 2015. {8}------------------------------------------------ <Biocompatibility testing> The subject device is categorized into the external communicating device (dentin) and permanent contact device. We evaluated according to ISO 10993-1: 2009, the subject device have the same ingredients and manufacturing process as the predicate device and the reference device and therefore no further biocompatibility testing was necessary. #### 3-7. Conclusion The comparison for indications for use, chemical ingredients, technological characteristics shows that the subject device is substantially equivalent to the predicate device and the reference device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...