K212577 · Terrats Medical SL · NHA · Dec 10, 2021 · Dental
Device Facts
Record ID
K212577
Device Name
DESS Dental Smart Solutions
Applicant
Terrats Medical SL
Product Code
NHA · Dental
Decision Date
Dec 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations. All digitally designed custom abutments for use with Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture.
Device Story
DESS Dental Smart Solutions are pre-milled blanks used to fabricate patient-specific custom dental abutments via CAD/CAM processes. Clinicians order custom abutments; designs are sent to a Terrats Medical validated milling center for manufacturing. Blanks are made of titanium alloy (Ti-6Al-4V) or cobalt-chromium alloy (Co-Cr-Mo). The device supports prosthetic restorations (cement or screw-retained) for single or multi-unit cases. This submission updates labeling to allow custom abutment angulation up to 30° and adds 14 mm diameter blanks. The device is used in dental clinics; operation involves CAD/CAM software for design and professional milling for fabrication. The output is a custom abutment that connects the dental implant to the prosthetic restoration, providing functional and esthetic rehabilitation for the patient.
Clinical Evidence
Bench testing only. Static and dynamic testing performed according to ISO 14801. Biocompatibility and sterilization information leveraged from prior clearances (K170588, K173908). No clinical data included.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) and Co-Cr-Mo (ASTM F1537). Design: Pre-milled blanks for CAD/CAM fabrication. Angulation: up to 30°. Minimum wall thickness: 0.45 mm. Minimum post height: 4.0 mm. Gingival height: 0.3 mm to 6.0 mm. Sterilization: End-user sterilized.
Indications for Use
Indicated for patients requiring prosthetic restoration support in the maxillary or mandibular arch using endosseous dental implants. Compatible with specific NobelActive, NobelParallel Conical, Straumann Bone Level, and Zimmer Screw-Vent/Tapered Screw-Vent implant systems.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Terrats Medical SL % Melissa Burbage Senior Regulatory Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K212577
Trade/Device Name: DESS Dental Smart Solutions Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 8, 2021 Received: November 9, 2021
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K212577
Device Name
DESS Dental Smart Solutions Indications for Use (Describe)
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant Systems | | |
|--------------------------------------------|------------------------------|------------------|
| Compatible Implant System | Implant Body Diameter,<br>mm | Implant Platform |
| NobelActive®, NobelParallel<br>Conical | 3.5 | NP |
| | 4.3, 5.0 | RP |
| | 5.5 | WP |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1/4.8 | RC |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="text-decoration: underline;"><b> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information
unless it displays a currently valid OMB number."*
{3}------------------------------------------------
## 510(k) Summary
## Terrats Medical SL DESS® Dental Smart Solutions K212577
### December 10, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Terrats Medical SL<br>Carrer Mogoda, 75-99<br>08210 Barberà del Vallès<br>Barcelona, Spain |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone | +34 935 646 006 |
| Official Contact | Roger Terrats, CEO |
| Representative/Consultant | Melissa Burbage, BS<br>Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email mburbage@paxmed.com<br>flarson@paxmed.com<br>kthomas@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | DESS Dental Smart Solutions |
|------------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Division | DHT1B: Division of Dental Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K191986, DESS Dental Smart Solutions, Terrats Medical SL
Additional Predicate Devices K170588, DESS Dental Smart Solutions, Terrats Medical SL K173908, DESS Dental Smart Solutions, Terrats Medical SL
Reference Devices K193425, Pre-Milled Blank, ARUM Dentistry Co. Ltd K183518, Preat Abutments, Preat Corporation
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for prosthetic restorations.
All digitally designed custom abutments for use with Pre-milled Blanks are to be sent to a Terrats Medical validated milling center for manufacture.
| Compatible Implant System | Implant Body<br>Diameter, mm | Implant Platform |
|--------------------------------------------|------------------------------|------------------|
| NobelActive®, NobelParallel<br>Conical | 3.5 | NP |
| | 4.3, 5.0 | RP |
| | 5.5 | WP |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1/4.8 | RC |
| Zimmer Screw-Vent®/ Tapered<br>Screw-Vent® | 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
#### Compatible Implant Systems
### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to implement a labeling change for certain Pre-milled Blanks of the DESS Dental Smart Solutions abutment system, cleared under K170588, K173908 and K191986, to permit the manufacture of custom abutments with angulation up to 30° and to add three (3) Pre-milled Blanks that also will include such labeling. Design parameters cleared in K170588. K173908 and K191986 were limited to straight abutments only. This change in labeling is for Pre-milled Blanks compatible with three (3) systems: NobelParallel Conical Connection, Straumann Bone Level, and Zimmer Screw-Vent/Tapered Screw-Vent implants. Note that, because NobelActive and Nobel Parallel Conical Connection share the same implant/abutment interface, they are considered one system for purposes of this submission, as they were in K170588, K173908 and K191986. No new compatibilities are added in this submission.
Pre-milled Blanks are designed for fabrication of a custom abutment by a CAD/CAM process. All patient-specific custom abutment fabrication is by prescription on the order of the clinician. Nobelcompatible Pre-milled Blanks are made of titanium alloy (Ti-6A1-4V) conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELL (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401) or from cobalt chromium alloy (Co-Cr-Mo) conforming to ASTM F1537 Standard Specification for Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants (UNS R31537, UNS R31538, and UNS R31539. Straumann-compatible and Zimmercompatible Pre-milled Blanks are made of titanium alloy (Ti-6Al-4V). They are available in engaging designs and are compatible with the implant systems shown the table below Compatible Implant Systems and Platforms. Except for the maximum angulation of the final abutment and the addition of three (3) Pre-milled Blanks of 14 mm diameter to the previously cleared Pre-milled Blanks of 10 mm and 14 mm diameter, subject device Pre-milled Blanks are identical to Pre-milled Blanks cleared (with slight variations on the name) in K170588, K173908 and K191986. Compatibility with the implant platform was demonstrated in K170588 and K191986.
The design parameters for the CAD/CAM fabrication of custom abutments from subject device Premilled Blanks are:
Minimum wall thickness - 0.45 mm Minimum post height – 4.0 mm
{5}------------------------------------------------
Maximum gingival height - 6.0 mm Minimum gingival height - 0.3 mm Maximum angulation of the final abutment - 30°
Manufacture of CAD/CAM custom abutments from Pre-milled Blanks is to be performed at a Terrats Medical validated milling center.
| Compatible Implant<br>System | DESS Abutment<br>System | Implant Body<br>Diameter, mm | Implant Platform | Connection |
|--------------------------------------------|-------------------------|------------------------------|------------------|------------|
| NobelActive®,<br>NobelParallel Conical | Active Hex | 3.5 | NP | Internal |
| | | 4.3, 5.0 | RP | Internal |
| | | 5.5 | WP | Internal |
| Straumann® Bone<br>Level | Conical BL | 3.3 | NC | Internal |
| | | 4.1/4.8 | RC | Internal |
| Zimmer Screw-Vent®/<br>Tapered Screw-Vent® | Internal Hex USA | 3.7, 4.1 | 3.5 | Internal |
| | | 4.7 | 4.5 | Internal |
| | | 6.0 | 5.7 | Internal |
Compatible Implant Systems and Platforms
## PERFORMANCE DATA
Non-clinical data submitted to demonstrate substantial equivalence included: static and dynamic testing according to ISO 14801. Information was leveraged from prior clearance (K170588 and K173908) to demonstrate substantial equivalence with regards to sterilization and biocompatibility. No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the additional predicate devices listed above. Provided at the end of this summary are tables comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, the additional predicate devices, and the reference devices.
Subject device abutments are substantially equivalent in intended use to the primary predicate device cleared in K191986, the additional predicate devices K173908 and the reference device K 193425. All are intended for use with endosseous dental implants to provide functional and esthetic rehabilitation of the edentulous maxilla and mandible. The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K191986 and additional predicate devices K170588 and K173908.
All subject device abutments are identical in design, materials, and technological characteristics to those of the primary predicate K191986 and additional predicate devices K170588 and K173908. The only change to the subject device is a labeling change to permit custom abutments to be manufactured from the Pre-milled Blanks with an angulation up to 30°, which is substantially equivalent to that of the reference device K193425, and the addition of three new Pre-milled Blanks of 14 mm diameter to the Pre-milled Blanks of 10 mm and 14 mm diameter previously cleared for the predicate devices. No new compatibilities are included in this submission.
Substantial equivalence of the subject device with design parameters that permit angulation of the final abutment is supported by dynamic testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants and by the reference device K183518.
The subject device is to be sterilized by the end-user, the same as primary predicate device K191986 and additional predicate devices K170588 and K173908.
{6}------------------------------------------------
### CONCLUSION
The subject device, the primary predicate device, additional predicate devices, and the reference devices have the same intended use, have similar technological characteristics, and are made of the same materials. The subject device, the primary predicate, and the additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{7}------------------------------------------------
| Subject Device | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in<br>the maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with Pre-milled Blanks are to be sent to a Terrats Medical<br>validated milling center for manufacture. | |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|
| DESS Dental Smart Solutions | | |
| K212577 | | |
| Terrats Medical SL | | |
| Compatible Implant Systems | | |
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform |
| NobelActive®, NobelParallel Conical | 3.5 | NP |
| | 4.3, 5.0 | RP |
| | 5.5 | WP |
| Straumann® Bone Level | 3.3 | NC |
| | 4.1/4.8 | RC |
| Zimmer Screw Vent®/ Tapered<br>Screw-Vent® | 3.7, 4.1 | 3.5 |
| | 4.7 | 4.5 |
| | 6.0 | 5.7 |
| Primary Predicate Device | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in<br>the maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with Ti Base abutments or Pre-milled (Blank) abutments are to be<br>sent to a Terrats Medical validated milling center for manufacture. | |
| K191986 | | |
| DESS Dental Smart Solutions | | |
| Terrats Medical SL | | |
| Compatible Implant Systems | | |
| Compatible Implant System | Implant Body Diameter, mm | Implant Platform |
| Ankylos C/X | 3.5, 4.5, 5.5 | 2.52 mm |
| | 3.6 | 2.9 mm |
| Astra Tech EV | 4.2 | 3.5 mm |
| | 4.8 | 4.1 mm |
| | 3.0 | 3.0 mm |
| Astra Tech OsseoSpeed™ | 3.5/4.0 | 3.5/4.0 mm |
| | 4.5/5.0 | 4.5/5.0 mm |
| | 3.25 | 3.45 mm |
| Biomet 3i Certain® | 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| | 3.25 | 3.4 mm |
| Biomet 3i OSSEOTITE® | 3.75, 4.0 | 4.1 mm |
| | 5.0 | 5.0 mm |
| | 3.3 | 3.3 mm |
| | 3.8 | 3.8 mm |
| Camlog | 4.3 | 4.3 mm |
| | 5.0 | 5.0 mm |
| | 3.4 | 3.4 mm |
| FRIADENT XiVE® | 3.8 | 3.8 mm |
| | 4.5 | 4.5 mm |
# Table of Substantial Equivalence – Indications for Use Statement
{8}------------------------------------------------
| | | | 5.5 | 5.5 mm | | |
|-----------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------|------------------------|--|--|
| | | MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 mm | | |
| | | | 3.0 | 3.0 (3.0 mm) | | |
| | | NobelActive®, | 3.5 | NP (3.5 mm) | | |
| | | NobelReplace/NobelParallel Conical | 4.3, 5.0 | RP (3.9 mm) | | |
| | | | 5.5 | WP (5.1 mm) | | |
| | | | 3.5 | NP (3.5 mm) | | |
| | | NobelReplace® Trilobe | 4.3 | RP (4.3 mm) | | |
| | | | 5.0 | WP (5.0 mm) | | |
| | | | 6.0 | 6.0 (6.0 mm) | | |
| | | | 3.3 | NP (3.5 mm) | | |
| | | Nobel Brånemark System® | 3.75, 4.0 | RP (4.1 mm) | | |
| | | | 5.0 | WP (5.1 mm) | | |
| | | Osstem TS | 3.5 | Mini (2.8 mm) | | |
| | | | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular (3.35 mm) | | |
| | | Straumann® Bone Level | 3.3 | NC (3.3 mm) | | |
| | | | 4.1/4.8 | RC (4.1/4.8 mm) | | |
| | | | 3.3 | NNC (3.5 mm) | | |
| | | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN (4.8 mm) | | |
| | | | 4.8 | WN (6.5 mm) | | |
| | | Zimmer Screw Vent®/ Tapered Screw-<br>Vent® | 3.3, 3.7, 4.1 | 3.5 mm | | |
| | | | 4.7 | 4.5 mm | | |
| | | | 6.0 | 5.7 mm | | |
| | | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in | | | | |
| Additional Predicate Device | | the maxillary or mandibular arch to provide support for prosthetic restorations. | | | | |
| | | | | | | |
| | All digitally designed custom abutments for use with Ti Base or Pre-milled (Blank) abutments are to be sent to a | | | | | |
| K170588 | | Terrats Medical validated milling center for manufacture. | | | | |
| | Compatible Implant Systems | | | | | |
| DESS Dental Smart Solutions | | | | | | |
| Terrats Medical SL | | Implant System Compatibility | Implant Diameter (mm) | Platform Diameter (mm) | | |
| | | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | |
| | | 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | |
| | | OsseoSpeed™ | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 | | |
| | | FRIADENT XiVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 | | |
| | | NobelActive® | 3.5, 4.3, 5.0 | NP, RP | | |
| | | NobelReplace Conical | 3.5, 4.3, 5.0 | NP, RP | | |
| | | Nobel Replace Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP | | |
| | | Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP | | |
| | | Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC | | |
| | | | | | | |
| | | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN, WN | | |
| | | Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | |
{9}------------------------------------------------
| Additional Predicate Device | DESS Dental Smart Solutions abutments are intended to be used in conjunction with endosseous dental implants in<br>the maxillary or mandibular arch to provide support for prosthetic restorations.<br>All digitally designed custom abutments for use with Aurum TM Abutment or Pre-milled Blank are to be sent to a<br>Terrats Medical validated milling center for manufacture. | | | | | |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-----------------------------------|------------------------------------------|-------------------------------|----------------|
| K173908 | Compatible Implant Systems | | | | | |
| DESS Dental Smart Solutions<br>Terrats Medical SL | Implant System Compatibility | Implant Body | Implant Platform | | | |
| | 3i Certain® | 3.25, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | |
| | 3i OSSEOTITE® | 3.25, 3.75, 4.0, 5.0 | 3.4, 4.1, 5.0 | | | |
| | OsseoSpeedTM | 3.5, 4.0, 5.0 | 3.5/4.0, 4.5/5.0 | | | |
| | FRIADENT XiVE | 3.4, 3.8, 4.5 | 3.4, 3.8, 4.5 | | | |
| | NobelActive® | 3.5, 4.3, 5.0 | NP, RP | | | |
| | NobelReplace® Conical | 3.5, 4.3, 5.0 | NP, RP | | | |
| | NobelReplace® Trilobe | 3.5, 4.3, 5.0 | NP, RP, WP | | | |
| | Brånemark | 3.5, 3.75/4.0, 5.0 | NP, RP, WP | | | |
| | Straumann® Bone Level | 3.3, 4.1, 4.8 | NC, RC | | | |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RP, WP | | | |
| | Tapered Screw-Vent® | 3.7, 4.1, 4.7, 6.0 | 3.5, 4.5, 5.7 | | | |
| Reference Device | ARUM Dentistry's Pre-Milled Blank abutments are intended for attachment to dental implants in order to provide<br>support for customized prosthetic restorations. Pre-Milled Blank abutments are indicated for screw-retained single<br>restorations or cement-retained single or multi-unit restorations. The customized Pre-Milled Blank abutment will<br>be attached to a dental implant using the included ARUM Dentistry prosthetic screw. | | | | | |
| K193425 | | | | | | |
| Pre-Milled Blank<br>ARUM Dentistry Co. Ltd | ARUM Dentistry's Pre-Milled Blanks are compatible with the implant systems listed in the Compatibility Table: | | | | | |
| | Compatibility Table | | | | | |
| | ARUM Pre-Milled Blank | | Implant Platform<br>compatibility | Restorative<br>Platform<br>Diameter (mm) | Implant Body<br>diameter (mm) | Abutment Screw |
| | 10 mm | 14 mm | | | | |
| | CIHE037 | CIHE038 | NobelActive NP | 3.5 | 3.5 | CSTO001 |
| | CIHE039 | CIHE040 | NobelActive NP | 3.9 | 4.3/5.0 | CSTO002 |
| | CIHE135 | CIHE136 | NobelActive WP | 5.1 | 5.5 | |
| | All digitally-designed Pre-Milled Blank abutments are intended to be sent to an ARUM Dentistry-validated milling<br>center for manufacture. | | | | | |
{10}------------------------------------------------
| Compatible Implant Systems | | | |
|----------------------------|----------------------------------------|----------------------------|--------------------------------|
| | | | |
| | Compatible Implant System | Implant Body Diameter (mm) | Implant Platform Diameter (mm) |
| | 3i OSSEOTITE® Certain® | 3.25 | 3.4 |
| | | 4.0 | 4.1 |
| | | 5.0 | 5.0 |
| | | 6.0 | 6.0 |
| | Astra Tech OsseoSpeed™ | 3.0 | 3.0 |
| | | 3.5, 4.0 | 3.5/4.0 |
| | | 4.5, 5.0 | 4.5/5.0 |
| | BioHorizons Tapered Internal | 3.0 | 3.0 |
| | | 3.5 | 3.5 |
| | | 4.0 | 4.5 |
| | HIOSSEN ET III | 3.5 | Mini |
| | | 4.0, 4.5, 5.0, 6.0, 7.0 | Regular |
| | Implant Direct Legacy | 3.2 | 3.0 |
| | | 3.7, 4.2 | 3.5 |
| | | 4.7, 5.2 | 4.5 |
| | | 5.7, 7.0 | 5.7 |
| | MegaGen AnyRidge | 3.5, 4.0, 4.5, 5.0, 5.5 | 3.5 |
| | Neoss | 3.5, 4.0, 4.5, 5.0, 5.5 | 4.1 |
| | NobelActive® | 3.5 | NP |
| | | 4.3, 5.0 | RP |
| | Nobel Replace™ | 3.5 | NP |
| | | 4.0, 4.3, 5.0 | RP |
| | | 5.0 | WP |
| | | 6.0 | 6.0 |
| | Straumann® Bone Level | 3.3 | NC |
| | | 4.1, 4.8 | RC |
| | Straumann® Tissue Level | 3.3, 4.1, 4.8 | RN |
| | | 4.8, 6.5 | WN |
| | Zimmer Screw-Vent®/Tapered Screw-Vent® | 3.3, 3.7, 4.1 | 3.5 |
| | | 4.7 | 4.5 |
| | | 6.0 | 5.7 |
{11}------------------------------------------------
| Comparison | Subject Device | Primary Predicate<br>Device | Additional Predicate Devices | Reference Devices | | |
|----------------------------------------------|-----------------------------------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|----------------------------------------------------------|-------------------------------------------------------------------------------------|
| | K212577<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K191986<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K170588<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K173908<br>DESS Dental Smart<br>Solutions<br>Terrats Medical SL | K193425<br>Pre-Milled Blank<br>ARUM Dentistry<br>Co. Ltd | K183518<br>Preat Abutments<br>Preat Corporation |
| Reason for<br>Predicate | n/a | Design | Design | Design | 30° angle | Performance<br>Testing |
| Design | | | | | | |
| Abutment Designs | CAD/CAM Blanks | Healing,<br>Temporary,<br>Straight, Multi-unit,<br>Locator-type,<br>CAD/CAM Bases,<br>CAD/CAM Blanks | Healing, Temporary,<br>Straight, Multi-unit,<br>Locator-type,<br>CAD/CAM Bases,<br>CAD/CAM Blanks | CAD/CAM Bases,<br>CAD/CAM Blanks | CAD/CAM<br>Blanks | Healing,<br>Temporary,<br>Straight, Multi-unit,<br>CAD/CAM Bases,<br>CAD/CAM Blanks |
| Prosthesis<br>Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained | Cement-retained<br>Screw-retained |
| Restoration | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit | Single-unit,<br>Multi-unit |
| Abutment/Implant<br>Platform<br>Diameter, mm | 2.52 - 6.0 | 2.52 - 6.0 | 3.0-6.0 | 3.6 - 5.0 | 3.5 - 5.5 | 3.0 - 6.5 |
| Prosthetic Platform<br>Diameter, mm | 4.5 – 6.5 | 4.5-6.5 | 4.5 | 4.0 - 6.5 | 3.5 - 5.1 | 3.0 - 6.5 |
| Blank Abutment<br>Angulation | 0° to 30° | 0° | 0° | 0° | 0° to 30° | 0° to 30° |
| Abutment/ Implant<br>Interface | Internal | Internal | Internal | Internal | Internal | Internal |
| Material | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy | Ti-6Al-4V ELI | Ti-6Al-4V ELI<br>Co-Cr-Mo Alloy…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.