Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
Device Story
Dynamic TiBase is a two-piece titanium base abutment system for dental implants. The device consists of a titanium alloy base (Ti-6Al-4V) designed to be bonded to a patient-specific CAD-CAM zirconia superstructure. The system is used by clinicians to restore chewing function in partially or fully edentulous patients. The clinician prescribes the custom superstructure, which is manufactured at a Thommen Medical validated milling center. The base features a prosthetic post with a cutout to facilitate angled screw access. The final two-piece abutment is secured to the implant using an abutment screw. The device is compatible with specific Thommen SPI CONTACT dental implants. The system benefits patients by providing a customized, esthetic, and functional prosthetic restoration.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: biocompatibility (ISO 10993-12), moist heat sterilization validation (ISO 17665-1, ISO 11737-2), and static compression/compression fatigue testing (ISO 14801) on worst-case constructs.
Technological Characteristics
Two-piece abutment system: titanium alloy (ASTM F136, ISO 5832-3) base and CAD-CAM zirconia (ISO 13356) superstructure. Internal connection. Gingival height 0.7 mm. Prosthetic post heights 4.0 mm/9.0 mm. Compatible with Thommen SPI CONTACT implants. Non-sterile; requires moist heat sterilization. Cemented using G-CEM LinkAce (K120243).
Indications for Use
Indicated for partially or fully edentulous patients requiring single-unit or multi-unit dental implant-supported prostheses in the maxillary or mandibular arch.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 24, 2021
Talladium España. SL % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K212108
Trade/Device Name: Dynamic TiBase Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 2, 2021 Received: July 6, 2021
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K212108
Device Name
Dynamic TiBase
Indications for Use (Describe)
Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|-----------------------------|---------------------------|----------------------|
| SPI® CONTACT Dental Implant | 2.7 | 3.5 |
| | 3.5 | 4.0 |
| | 3.5 | 4.5 |
| | 4.2 | 5.0 |
All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Thommen Medical validated milling center for manufacture.
| Type of Use (Select one or both, as applicable) | <span> <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary K212108 Talladium España, SL Dynamic TiBase
September 23, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Talladium España, SL<br>C/ Virginia Woolf, 17<br>Lleida, Lleida, ES 25005<br>Telephone +34 973-289-580 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact | Xavier Soca Filella, General Manager |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Dynamic TiBase |
|------------------------|------------------------------------------------------------------------------------------------|
| Common Names | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory,<br>ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K202026, Blue Sky Bio CAD-CAM Abutments, Blue Sky Bio, LLC Reference Devices K182246, MIST IC, Imagine Milling Technologies, LLC K034014, SPI® CONTACT Dental Implant, Thommen Medical AG K072933, SPI® CONTACT Platform Ø 4.0 mm, Thommen Medical AG K191919, Elos Accurate® Hybrid Base, Elos Medtech Pinol A/S K162021, 3.0 Dynamic TiBase, Talladium España, SL
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandibular arch of a partially or fully edentulous patient.
| Implant Compatibility | Implant Body Diameter, mm | Implant Platform, mm |
|-----------------------------|---------------------------|----------------------|
| SPI® CONTACT Dental Implant | 2.7 | 3.5 |
| | 3.5 | 4.0 |
| | 3.5 | 4.5 |
| | 4.2 | 5.0 |
All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Thommen Medical validated milling center for manufacture.
#### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to obtain marketing clearance for Dynamic TiBase, a series of twopiece titanium base abutments that require the fabrication of patient-specific custom superstructures using CAD-CAM technology. In final, finished form, the subject device abutments are intended to be used as a two-piece abutment composed of the base bottom-half (titanium base) with a bonded CAD-CAM zirconia top-half. Each patient-specific zirconia superstructure is individually prescribed by the clinician and manufactured by an authorized milling center.
The subject device abutments are compatible with SPT® CONTACT dental implant fixtures manufactured by Thommen Medical AG, and cleared in K034014 and K072933, in the following sizes:
- 2.7 mm body diameter/3.5 mm platform;
- 3.5 mm body diameter/4.0 mm platform;
- 3.5 mm body diameter/4.5 mm platform; and
- 4.2 mm body diameter/5.0 mm platform.
The compatibility between the subject device and the Thommen SPI® CONTACT dental implant fixtures was established by a business agreement between Talladium España, SL and Thommen Medical AG.
All subject device abutments and abutment screws are made of titanium alloy (Ti-6Al-4V) conforming to ASTM F136 and ISO 5832-3.
All Dynamic TiBase abutments are provided in a straight design (no angulation in the TiBase portion), and with implant connections for crowns (engaging) or bridges (non-engaging). All Dynamic TiBase abutments are provided with a gingival height of 0.7 mm; additional gingival height may be provided in the zirconia superstructure as described below. All Dynamic TiBase abutments are provided with a cutout in the prosthetic post to accommodate a restoration with an angled screw channel when clinically necessary. The prosthetic post heights are 4.0 mm (maximum height) / 2.3 mm (cutout height), and 9.0 mm/3.5 mm. Dynamic TiBase abutments with a 9.0 mm post height may be shortened to no less than 4 mm for a single-unit restoration.
{5}------------------------------------------------
All zirconia copings (superstructures) for use with the subject device Dynamic TiBase will be made at a Thommen Medical validated milling center under FDA quality system regulations, and the material will conform to ISO 13356.
The design parameters for the CAD-CAM zirconia superstructure for Dynamic TiBases are: Minimum wall thickness - 0.43 mm Minimum post height for single-unit restorations – 4.0 mm Maximum gingival height - 5.83 mm Minimum gingival height - 0.7 mm (in the TiBase, not the zirconia superstructure) Maximum angulation - 30°
The recommended cement for bonding the zirconia superstructure to the Dynamic TiBases to create the final two-piece abutment is G-CEM LinkAce™ cleared in K120243.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: biocompatibility testing according to ISO 10993-12; moist heat sterilization validation according to ISO 17665-1 and ISO 11737-2; and static compression and compression fatigue testing according to ISO 14801 of worst-case constructs comprising the subject device abutments, zirconia superstructures made to the limits listed above, subject device abutment screws, and compatible Thommen SPI® CONTACT dental implant fixtures. No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and reference devices.
The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate device K202026 and the reference devices K182246, K191919, and K162021; differences in language of the Indications for Use statements do not affect the intended use as an endosseous dental implant abutment for support of a prosthesis to restore chewing function.
The minor differences among the IFUS for the subject device, the primary predicate K202026, and the reference devices K182246, K191919, and K162021 include differences in the specific wording regarding compatible dental implants, and differences in the specific wording the required validated milling centers. K162021 is indicated only for multi-unit prostheses. Similarities among the IFUS for the subject device, the primary predicate K202026, and the reference devices K182246, K191919, and K162021 include wording single-unit and multi-unit restorations (K162021 is indicated only for multi-unit) and the requirement for use of validated milling centers.
The reference devices K072933 and K034014 are for substantial equivalence of the identical Thommen Medical AG SPI® CONTACT implant interface connections and platforms. Differences among the IFUS for the subject device and these reference devices are wording the implant components included
{6}------------------------------------------------
in K072933 and K034014; the relevant similarities include wording concerning prosthetic support of single-unit and multi-unit restorations.
The reference device K191919 is for support of substantial equivalence of the performance testing data. Similarities between the IFUS for the subject device and K191919 include wording on the intended use for single-unit and multi-unit restorations and the requirement for use of validated milling centers. Minor differences between the IFUS for the subject device and K191919 include specific wording about base abutments and the listing of compatible implant systems. None of these minor differences impact substantial equivalence because all IFUS express an equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording.
The reference device K162021 is for identical manufacturing and technological characteristics for the proposed titanium alloy abutments and abutment screws, including the anodization process that is identical to that used for the subject devices.
The subject device abutments have interface connections and platforms that are identical to the compatible Thommen Medical AG SPI® CONTACT implant interface connections and platforms cleared in K072933 and K034014. The subject device includes designs for compatible Thommen Medical AG SPI® CONTACT implant platforms ranging from 3.5 mm to 5.0 mm.
The subject device abutments are substantially equivalent in material and design to the corresponding abutments in the primary predicate device K202026. The final finished device for both the subject device abutments and the titanium base abutments cleared in K202026 is intended to be used as a two-piece abutment composed of the base bottom-half (titanium base) bonded to a CAD-CAM zirconia top-half.
The subject device abutments are provided with abutment-implant platform diameters and prosthetic platform diameters that are substantially equivalent to those of the corresponding abutments in K202026. Differences in the zirconia superstructure design parameters between the subject device and the primary predicate device K202026 include slight differences in the minimum wall thickness and gingival height; these differences are mitigated by the mechanical testing described below.
The reference device K182246 is for support of substantial equivalence of the subject device prosthetic post design, which has a cut-out to allow a screwdriver to access the abutment screw at an angle (if necessary) to tighten the screw and secure the abutment to the implant fixture. The MIST IC S-LINK abutment cleared in K182246 is a titanium base abutment manufactured from titanium alloy to be used with a CAD-CAM zirconia superstructure intended for cement-retained, single-unit or multi-unit restorations. Prior to bonding of the zirconia superstructure, MIST IC S-LINK has a post height of 3.8 mm with a cut-down channel to allow a screwdriver to access the abutment screw at an angle (if necessary) to tighten the screw. The subject device abutments have a prosthetic post with a similar cutdown channel with a minimum post height of 4 mm prior to bonding of the zirconia superstructure.
Minor differences in the design parameters for zirconia copings to be used with the subject device base abutments and for zirconia copings to be used with the predicate device base abutments (minimum wall thickness, maximum gingival height, maximum angulation) are mitigated by mechanical testing of the subject device performed in conformance with ISO 14801.
{7}------------------------------------------------
The subject device includes abutments made of titanium alloy conforming to ASTM F136 and ISO 5832-3. The titanium alloy subject device components are manufactured from identical materials, in the identical facilities using the identical manufacturing processes as used for Talladium España. SL products cleared previously in the reference device K162021. In addition, subject device titanium alloy abutments and abutment screws are anodized using a process that is identical to the anodization process used for Talladium España, SL products cleared previously in the reference device K162021. The subject device abutments are to be used with superstructures fabricated from zirconia conforming to ISO 13356; this is the same material used for superstructures in the primary predicate device K202026 and in the reference devices K182246 and K191919. Confirmatory biocompatibility testing for finished subject devices made from titanium alloy with cemented zirconia superstructures was performed according to ISO 10993-5 and ISO 10993-12.
Mechanical performance testing of the subject device was performed in conformance to ISO 14801. The fatigue limit data demonstrated that constructs of the subject device abutments fabricated to the limits stated in the proposed labeling, in combination with previously-cleared compatible Thommen Medical AG SPI® CONTACT implants, have sufficient strength for their intended use.
All subject device abutments and all abutments cleared in K202026, K182246, and K191919 are provided non-sterile and are to be moist heat sterilized by the end user.
Minor differences in the designs, dimensions, sizes, or compatible implant lines among the subject device, the primary predicate device, and the reference devices do not affect substantial equivalence. These minor differences do not impact safety or effectiveness because these differences are related to the compatible implant designs and are mitigated by the mechanical performance testing.
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device, the primary predicate, and the reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate device listed above.
{8}------------------------------------------------
## Table of Substantial Equivalence – Indications for Use Statements
| | Indications for Use Statements | | | | | | | | | | | | | | | |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Subject Device<br>K212108<br>Dynamic TiBase<br>Talladium España, SL | Dynamic TiBase abutments are intended for use with dental implants as a support for single-unit or multi-unit prostheses in the maxillary or mandible. Implant Compatibility Implant Body Diameter, mm Implant Platform, mm 2.7 3.5 3.5 4.0 SPI® CONTACT Dental Implant 3.5 4.5 4.2 5.0 All digitally designed custom abutments for use with Dynamic TiBase abutments are to be sent to a Thommen Medical validated milling center. | | | | | | | | | | | | | | | |
| Primary Predicate Device | | | | | | | | | | | | | | | | |
| K202026<br>Blue Sky Bio CAD-CAM Abutments<br>Blue Sky Bio, LLC | Blue Sky Bio CAD-CAM Abutments are intended to be used in conjunction with Blue Sky Bio endosseous dental implants in the maxillary or mandibular prosthetic restorations.<br>All digitally designed abutments for use with Blue Sky Bio CAD-CAM Abutments are intended to be sent to a Blue Sky Bio validated milling center. | | | | | | | | | | | | | | | |
| Reference Devices | | | | | | | | | | | | | | | | |
| K182246<br>MIST IC<br>Imagine Milling Technologies, LLC | MIST IC abutments are intended for use to support a prosthetic device in a partially or completely edentulous patient. They are intended to support the mandible or maxilla. MIST IC abutments are compatible for use with the following implants: Keystone Dental Implant Line Platform Diameter (mm) Body Diameter (mm) Genesis 3.8, 4.5, 5.5, 6.5 3.8, 4.5, 5.5, 6.5 PrimaConnex® 1.0 (Straight) 3.5, 4.1, 5.0 3.3, 4.0, 5.0 PrimaConnex® 1.0 (Tapered) 3.5, 4.1, 5.0 3.5, 4.1, 5.0 All digitally designed custom abutments for use with MIST IC abutments are to be sent to an Imagine Milling Technologies validated milling center. | | | | | | | | | | | | | | | |
| K072933<br>SPI® CONTACT Platform Ø 4.0 mm<br>Thommen Medical AG | SPI® CONTACT Platform Ø 4.0 mm is intended to be surgically placed, either immediately after extraction or following healing, in the bone of the mandible or maxilla, to support single or multiple-unit bridges or overdentures. SPI® Dental Implants can be loaded immediately if they are splinted with a bar on four implants in the mandibular arch or six in the maxilla. | | | | | | | | | | | | | | | |
| K034014<br>SPI® CONTACT Dental Implant<br>Thommen Medical AG | The Thommen SPI® CONTACT Dental Implant is intended to be surgically placed, either immediately or delayed, in the bone of the maxillary or mandibular arch, to support single or multiple-unit bridges or overdentures. SPI® CONTACT implants can be loaded immediately if they are splinted with a bar on four implants in the mandibular arch or six in the maxilla.<br>Contraindications for the use of SPI® CONT ACT implants Ø 3.5 mm:<br>These implants are not suitable for applications in areas where pronounced rotation and translation movements occur, causing the implant to be subjected to excessive stress.<br>- Restoration of posterior teeth in the upper or lower jaw<br>- Single-tooth restoration of canines and central incisors in the upper jaw<br>- Any application involving retentive anchors. | | | | | | | | | | | | | | | |
libular arch of a partially or fully edentulous patient.
for manufacture.
andibular arch to provide support for single-unit or multi-unit
nter for manufacture.
rt a single-unit or multi-unit, cement-retained prosthesis in the
ter for manufacture.
he maxillary and/or mandibular arch to provide support for crowns, r six implants in the maxillary arch.
r mandibular arch to provide support for crowns, bridges or implants in the maxillary arch .
ibjected to large bending moments.
{9}------------------------------------------------
## Table of Substantial Equivalence – Indications for Use Statements
| Table 1. | | |
|--------------------------------|------------------------|----------------------------|
| Implant Platform compatibility | Platform diameter [mm] | Implant Body diameter [mm] |
| Nobel Replace NP | 3.5 | 3.5 |
| Nobel Replace RP | 4.3 | 4.3 |
| Nobel Replace WP | 5 | 5 |
| Nobel Replace 6.0 | 6 | 6 |
| Nobel CC 3.0 | 3 | 3 |
| Nobel CC NP | 3.5 | 3.5 & 3.75 |
| Nobel CC RP | 3.9 | 4.3 & 5 |
| Nobel CC WP | 5.1 | 5.5 |
| Straumann Bone Level NC | 3.3 | 3.3 |
| Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
| Astra Tech 3.0 | 3 | 3 |
| Astra Tech 3.5/4.0 | 3.5 & 4 | 3.5 & 4 |
| Astra Tech 4.5/5.0 | 4.5 & 5 | 4.5 & 5 |
| Astra Tech EV 3.0 | 3 | 3 |
| Astra Tech EV 3.6 | 3.6 | 3.6 |
| Astra Tech EV 4.2 | 4.2 | 3.6 & 4.2 |
| Astra Tech EV 4.8 | 4.8 | 4.2 & 4.8 |
| Astra Tech EV 5.4 | 5.4 | 5.4 |
orations.
Iuded prosthetic screw and attached to the zirconia superstructure
n FDA registered Elos Medtech approved milling facility.
r fully edentulous patient.
Image /page/9/Figure/7 description: The image shows a table with two rows. The first row contains the text 'Abutment Connection'. The second and third rows contain the text 'INTERNAL HEXAGON'. The table appears to be describing a type of abutment connection.
{10}------------------------------------------------
# Table of Substantial Equivalence – Technological Characteristics
| Comparison | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device | Reference Device |
|---------------------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| | K212108<br>Dynamic TiBase | K202026<br>Blue Sky Bio CAD-CAM<br>Abutments | K182246<br>MIST IC | K072933<br>SPI® CONTACT<br>Platform Ø 4.0 mm | K034014<br>SPI® CONTACT Dental Implant | K191919<br>Elos Accurate® Hybrid Base |
| | Talladium España, SL | Blue Sky Bio, LLC | Imagine Milling Technologies, LLC | Thommen Medical AG | Thommen Medical AG | Elos Medtech Pinol A/S |
| Reason for Predicate Device / Reference<br>Device | Not applicable | Indications for Use Statement<br>Abutment designs | Prosthetic post design, with<br>angled screw channel feature | Compatible identical implant interface<br>and platforms | Compatible identical implant interface<br>and platforms | Performance testing data |
| Product Codes | NHA | NHA | NHA | DZE, NHA | DZE | NHA |
| Intended Use | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla | Functional and esthetic rehabilitation of<br>the edentulous mandible or maxilla |
| Designs/Features | | | | | | |
| Abutment Design | CAD-CAM Titanium Base Abutments | CAD-CAM Titanium Base Abutments<br>CAD-CAM Cobalt Base Abutments<br>CAD-CAM Titanium Blank Abutments | CAD-CAM Titanium Base Abutments<br>CAD-CAM Titanium Blank Abutments | | | CAD-CAM Titanium Base Abutments |
| Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | | | Single-unit<br>Multi-unit |
| Prosthesis Attachment | Cement-retained<br>Screw-retained | Cement-retained<br>Screw-retained | Cement-retained | | | Cement-retained<br>Screw-retained |
| Abutment-Implant Interface | Internal | Internal | Internal | Internal | Internal | Internal |
| | | | | | | |
| Abutment-Implant Platform<br>Diameter, mm | 3.5, 4.0, 4.5, 5.0 | 2.8 – 5.0 | 3.5 - 5.0 | 4.0 | 3.5, 4.5, 5.0, 6.0 | 3.0 - 6.0 |
| Prosthetic Platform Diameter, mm | 4.1, 4.3, 4.75, 5.25 | 3.9—6.5 | Not stated in 510(k) Summary | | | Not stated in 510(k) Summary |
| Superstructure (Coping)<br>Design Parameters | | | | | | |
| Minimum wall thickness, mm | 0.43 | 0.5 | 0.5 or 0.7 | | | Not stated in 510(k) Summary |
| Minimum post height for single-unit<br>restorations, mm | 4.0 | 4.0 | 4.0 | | | Not stated in 510(k) Summary |
| Minimum gingival height<br>(of the coping), mm | 0<br>All bases have GH of 0.7 mm | 0.5 | Not stated in 510(k) Summary | | | Not stated in 510(k) Summary |
| Maximum gingival height<br>(of the coping), mm | 5.83 | 6.7 | 5.0 | | | Not stated in 510(k) Summary |
| Angulation of Finished Abutment | Up to 30° | Up to 30° | Up to 20° | | | Up to 20° |
| Cement to bond coping to base | G-CEM LinkAce™<br>GC America Inc. | Multilink Hybrid Abutment Cement<br>Ivoclar Vivadent AG | Not stated in 510(k) Summary | | | Not stated in 510(k) Summary |
| Materials | | | | | | |
| Abutment Materials | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136<br>Cobalt-chromium alloy, ASTM F1537 | Titanium alloy, ASTM F136 | | | Titanium alloy, ASTM F136 |
| Coping | Zirconia, ISO 13356 | Zirconia, ISO 13356 | Zirconia, ISO 13356 | | | Zirconia, ISO 13356 |
| Abutment Screws | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | | | Titanium alloy, ASTM F136 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.