SD Abutment is intended for use in partially or fully edentulous maxilla, in support of single or multiple-unit prosthetic restorations. SD Abutment is for single stage and two stage surgical procedures. This system is intended for delayed loading.
Device Story
SD Abutment system comprises dental abutments (healing, solid, dual, dual long, angled, temporary) and screws; interfaces with specific third-party endosseous dental implants. Device fabricated from Ti-6Al-4V ELI (ASTM F136) or Pure Titanium Gr4 (ASTM F67). Used by dental clinicians in clinical settings to support prosthetic restorations. Clinician selects appropriate abutment based on implant platform/body diameter and length; secures to implant fixture. Provides structural support for dental prosthetics; facilitates soft tissue healing or temporary restoration. Benefits patient by enabling functional and aesthetic tooth replacement in edentulous sites.
Clinical Evidence
Bench testing only. Performance data includes 30° compressive loads, fatigue testing, adaptation accuracy, torsional breaking force, and removal torque force, all demonstrating equivalence to predicate devices. Biocompatibility evaluated per ISO 10993-1 (cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, implantation).
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F136) and Pure Titanium Gr4 (ASTM F67). Components: Healing, solid, dual, angled (15°, 25°), and temporary abutments with screws. Interface: Internal hex/non-hex. Supplied non-sterile; requires steam sterilization (ISO 17665-1/2).
Indications for Use
Indicated for partially or fully edentulous maxilla and mandible patients requiring single or multiple-unit prosthetic restorations via single or two-stage surgical procedures with delayed loading.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2019
Surgident Co., Ltd. % Peter Chung Representative Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213
Re: K173570
Trade/Device Name: SD Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: April 17, 2019 Received: April 19, 2019
Dear Peter Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173570
Device Name SD Abutment
Indications for Use (Describe)
SD Abutment is intended for use in partially or fully edentulous maxilla, in support of single or multipleunit prosthetic restorations. SD Abutment is for single stage and two stage surgical procedures. This system is intended for delayed loading.
The SD Abutment is compatible with the following devices:
| No. | Trade Name | Platform<br>Diameters<br>(mm) | Body<br>Diameters<br>(mm) | Lengths<br>(mm) |
|-----|------------------------------------------|-------------------------------|---------------------------|---------------------|
| 1 | s-Clean OneQ-SL Narrow<br>Implant System | 03.0 | 03.0 | 10, 12, |
| | | 03.3 | 03.3 | 14 |
| 2 | OneQ-SL s-Clean Implant<br>System | 03.7 | 03.5 | |
| | | 03.9 | 03.6 | |
| | | 04.2 | 03.7 | 7, 8, 10,<br>12, 14 |
| | | 04.7 | 04.2 | |
| | | 05.2 | 04.7 | |
| | | 06.0 | 04.8 | 7, 8, 10, |
| | | 07.0 | 05.8 | 12 |
| | | 08.0 | 06.8 | |
Type of Use (Select one or both, as applicable)
| <span> <span style="text-decoration:overline">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------------------|
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K173570 - 510(k) Summary [as required by 807.92(c)]
## 1. Applicant
- a) Company: Surgident Co., Ltd.
- Address: #209~210, Woolim Lion's Valley, 27, Dunchon-daero 457beon-gil, b) Jungwon-gu, Seongnam-si, Gyeonggi-do, 13219, Korea
- +82-31-737-2875 c) Tel :
- d) Fax : +82-31-737-2876
- e) President of Company: Mr. Duck Su Hur
- f) Contact Person : Peter Chung
- Contact Person Telephone: 412-687-3976 g)
- Contact Person Address: 300, Atwood Street, Pittsburgh, PA, 15213, USA h)
- i) Preparation Date: May 17, 2019
#### 2. Device Information
- Trade Name : SD Abutment a)
- Common Name : Endosseous Dental Implant Abutment b)
- Classification Name : c) abutment, implant, dental, endosseous
- d) Product Code : NHA
- e) Regulation Number : 872.3630
- f) Class of device : Class II
- g) Panel : Dental
## 3. Primary Predicate
K131682, UD Implant System / MEDIMECCA Co., Ltd.
#### Reference Device
K161244, s-Clean OneQ-SL Narrow Implant System, Dentis Co., Ltd. K153639, OneQ-SL s-Clean Implant System, Dentis Co., Ltd.
## 4. Device Description and Technological Characteristics
SD Abutment made of Ti-6A1-4V ELI alloy (ASTM F136) and/or Pure Titanium Gr4 (ASTM F67) is intended for use as an aid in single or multiple-unit prosthetic restorations. It consists of Healing Abutment, Solid Abutment, Dual Abutment, Dual Long Abutment, Angled Abutment (15° and 25°), Temporary Abutment and Abutment Screws. All abutments are supplied non-sterile and autoclaved by the end user. The abutments are to be used with the as followed fixtures:
| No. | 510k No. | Trade Name | Platform<br>Diameters<br>(mm) | Body<br>Diameters<br>(mm) | Lengths<br>(mm) | Manufacturer |
|-----|----------|------------------------------------------|-------------------------------|---------------------------|-----------------|---------------------|
| 1 | K161244 | s-Clean OneQ-SL Narrow<br>Implant System | 03.0<br>03.3 | 03.0<br>03.3 | 10, 12,<br>14 | Dentis Co.,<br>Ltd. |
| 2 | K153639 | OneQ-SL s-Clean<br>Implant System | 03.7<br>03.9<br>04.2 | 03.5<br>03.6<br>03.7 | 7, 8, 10, | |
{4}------------------------------------------------
| | 04.7 | 04.2 12, 14 |
|--|------|----------------|
| | 05.2 | 04.7 |
| | 06.0 | 04.8 7, 8, 10, |
| | 07.0 | 05.8 12 |
| | 08.0 | 06.8 |
#### 5. Indication for Use
SD Abutment is intended for use in partially or fully edentulous mandibles and maxilla, in support of single or multiple-unit prosthetic restorations. SD Abutment is for single stage and two stage surgical procedures. This system is intended for delayed loading.
| No. | Trade Name | Platform<br>Diameters<br>(mm) | Body<br>Diameters<br>(mm) | Lengths<br>(mm) |
|-----|------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|----------------------------------------|
| 1 | s-Clean OneQ-SL Narrow<br>Implant System | 03.0<br>03.3 | 03.0<br>03.3 | 10, 12,<br>14 |
| 2 | OneQ-SL s-Clean Implant<br>System | 03.7<br>03.9<br>04.2<br>04.7<br>05.2<br>06.0<br>07.0<br>08.0 | 03.5<br>03.6<br>03.7<br>04.2<br>04.7<br>04.8<br>05.8<br>06.8 | 7, 8, 10,<br>12, 14<br>7, 8, 10,<br>12 |
# The SD Abutment is compatible with the following devices:
#### 6. Performance Data
- a) Physical Tests
| No. | Test items | Results |
|-----|--------------------------|---------------------------------|
| 1. | 30° Compressive Loads | |
| 2. | Fatigue | |
| 3. | Adaptation Accuracy | |
| 4. | Torsional Breaking Force | |
| 5. | Removal Torque Force | |
| 6. | Adaptation Accuracy | |
| 7. | Torsional Breaking Force | Equivalent to predicate devices |
| 8. | Removal Torque Force | |
| 9. | 30° Compressive Loads | |
| 10. | Fatigue | |
| 11. | Visual | |
| 12. | Dimension | |
| 13. | Package | |
Reverse Engineering tolerance analysis was conducted on the original OEM 3rd party implant body, abutment, and abutment screw to ensure compatibility of the SD Abutment.
{5}------------------------------------------------
## 7. Predicate Device Comparison Table
The SD Abutment is substantially equivalent to previously marketed devices. The design features and sizing of the components were compared and the SD Abutment found to be substantially equivalent as the predicate device. The primary predicate includes a list of examples for single and multiple unit restorations whereas the subject device contains this information in the labeling for each specific abutment. Additionally, the primary predicate is dental abutments for the predicates own implant bodies whereas the subject device is a 30 party abutment compatible. The inclusion of the specific compatible implant bodies, while different as compared to the primary predicate's Indications for Use, is necessary to adequately identify the indications for the subject device and the appropriate patient population. There are no significant differences between the SD Abutment and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to other devices in design, function, material and intended use.
| | | SUBJECT Device | Primary PREDICATE Device<br>K131682 | | | | | | | | | | | | | | | | | | | | | |
|---------------------|------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | | Surgident Co., Ltd. | MEDIMECCA Co., Ltd. | | | | | | | | | | | | | | | | | | | | | |
| Common Name | | Endosseous Dental Implant<br>Abutment | Endosseous Dental<br>Implant | | | | | | | | | | | | | | | | | | | | | |
| Trade Name | | SD Abutment | UD Implant System | | | | | | | | | | | | | | | | | | | | | |
| Indications for Use | | SD Abutment is intended for use in<br>partially or fully edentulous mandibles and<br>maxilla, in support of single or multiple-<br>unit prosthetic restorations. SD Abutment<br>is for single stage and two stage surgical<br>procedures. This system is intended for<br>delayed loading. The SD Abutment is<br>compatible with the following devices:<br><br>No. 510k No. Trade Name Platform Diameters<br>(mm) Body Diameters<br>(mm) Lengths<br>(mm) Manufacturer 1 K161244 s-Clean OneQ-SL<br>Narrow Implant System 03.0<br>03.3 03.0<br>03.3 10, 12,<br>14 Dentis Co.,<br>Ltd. 2 K153639 OneQ-SL s-Clean<br>Implant System 03.7<br>03.9<br>04.2<br>04.7<br>05.2<br>06.0<br>07.0<br>08.0 03.5<br>03.6<br>03.7<br>04.2<br>04.7<br>04.8<br>05.8<br>06.8 7, 8, 10,<br>12, 14 | | | | | | | | | | | | | | | | | | | | | | UD Implant System<br>intended for use in<br>partially or fully<br>edentulous mandibles and<br>maxilla, in support of<br>single of multiple-unit<br>restorations<br>including: cemented<br>retained, screw retained,<br>or overdenture<br>restorations, and<br>terminal or<br>intermediate abutment<br>support for fixed<br>bridgework. UD<br>Implant System is for<br>single stage and two<br>stage surgical<br>procedures.<br>This<br>system is<br>intended for delayed<br>lading. |
| Components | | Abutments | Fixtures and abutments | | | | | | | | | | | | | | | | | | | | | |
| Healing | Type | Internal | Internal | | | | | | | | | | | | | | | | | | | | | |
{6}------------------------------------------------
| Abutment | | Diameter (mm) | | | |
|----------------------------|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------|--|-----------------------------------------------------------|--|
| | | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5 mm, 7.5 mm, 8.5 mm | | 4.0 mm, 4.5 mm, 5.5 mm, 6.5 mm | |
| Solid Abutment | Type | Internal | | Internal | |
| | Diameter (mm) | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5 mm | | 4.0 mm, 4.5 mm, 5.5 mm, 6.5 mm | |
| Dual Abutment | Type | Internal Hex, Internal Non-Hex | | Internal Hex, Internal Non-Hex | |
| | Diameter (mm) | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5 mm | | 4.0 mm, 4.5 mm, 5.5 mm, 6.5 mm | |
| Dual Long Abutment | Type | Internal Hex, Internal Non-Hex | | Internal Hex, Internal Non-Hex | |
| | Diameter (mm) | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5 mm | | 4.0 mm, 4.5 mm, 5.5 mm, 6.5 mm | |
| Angled Abutment | Type | Internal Hex, Internal Non-Hex | | Internal Hex, Internal Non-Hex | |
| | Diameter (mm) | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5mm | | 4.5 mm, 5.5 mm | |
| | Angulation | 15°, 25° | | 15°, 25° | |
| Temporary Abutment | Type | Internal Hex, Internal Non-Hex | | Internal Hex, Internal Non-Hex | |
| | Diameter (mm) | 4.0 mm, 4.5 mm, 5.0 mm, 5.5 mm, 6.0 mm, 6.5 mm | | 4.0 mm, 4.5 mm | |
| Material | | Ti-6Al-4V ELI (ASTM F136), or Titanium Grade 4 (ASTM F67) | | Ti-6Al-4V ELI (ASTM F136), or Titanium Grade 4 (ASTM F67) | |
| Physical testing performed | | Test according to ISO 14801 | | Test according to ISO 14801 | |
| Biocompatibility | | Biocompatible according to ISO 10993-1 (evaluated cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, implantation) | | Biocompatible according to ISO 10993-1 | |
| Single Use | | yes | | yes | |
| Sterilization method | | The abutment is supplied non-sterile (Steam sterilization prior to use). Validation via ISO 17665-1 & ISO 17665-2 | | Gamma radiation | |
## 8. Conclusion
The SD Abutment is similar in fundamental scientific technology to the predicate device in that they all have been designed, manufactured and tested in compliance with "FDA's Class II special controls guidance document root-form endosseous dental implants and endosseous dental implant abutments".
The subject device and the predicate device have the same intended use and have the same technological characteristics. The subject and predicate implants are all made of the titanium
{7}------------------------------------------------
alloys and have not the surface treatments. The subject and the predicate devices encompass the similar range of physical dimensions, including diameter and length of the implants. and diameter of the abutments.
Overall, the SD Abutment has the following similarities to the predicate device:
- . has the same intended use.
- uses the same operating principle. .
- . incorporates the same basic design.
- . incorporates the same material.
Based on the similarities, we conclude that the SD Abutment is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.