NC Straumann Variobase Abutment for Bars/Bridges, RC Straumann Variobase Abutment for Bars/Bridges, NNC Straumann Variobase Abutment for Bars/Bridges, RN Straumann Variobase Abutment for Bars/Bridges, WN Straumann Variobase Abutment for Bars/Bridges
K151157 · Straumann USA, LLC · NHA · Aug 5, 2015 · Dental
Device Facts
Record ID
K151157
Device Name
NC Straumann Variobase Abutment for Bars/Bridges, RC Straumann Variobase Abutment for Bars/Bridges, NNC Straumann Variobase Abutment for Bars/Bridges, RN Straumann Variobase Abutment for Bars/Bridges, WN Straumann Variobase Abutment for Bars/Bridges
Applicant
Straumann USA, LLC
Product Code
NHA · Dental
Decision Date
Aug 5, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Straumann® Variobase® abutments are intended to be placed into Straumann dental implants to provide support for prosthetic reconstructions such as bridges and over-dentures.
Device Story
Straumann Variobase Abutments for Bridges/Bars are non-engaging titanium-6aluminum-7niobium (TAN) alloy components; they connect to endosseous dental implants to support multi-unit prosthetic reconstructions (bridges or over-dentures). Dental laboratory technicians design and manufacture the bridge/bar framework using traditional or CAD/CAM workflows. The abutment is secured to the implant via a basal screw. The device facilitates prosthetic rehabilitation by providing a stable interface for the restoration. Used in dental clinics by clinicians; the device benefits patients by enabling functional and aesthetic tooth replacement. The non-engaging design prevents mechanical constraints in multi-unit frameworks, ensuring proper fit.
Clinical Evidence
Bench testing only. Dynamic fatigue testing was performed in accordance with FDA guidance for Root-form Endosseous Dental Implants and Endosseous Dental Abutments and ISO 14801 standards to verify performance of the abutment design, including taller variants.
Technological Characteristics
Material: Titanium-6aluminum-7niobium (TAN) alloy. Design: Non-engaging Morse taper interface; 10° or 15° coronal conical taper. Dimensions: 4.2–7.0 mm diameter; 3.5–5.2 mm height. Connectivity: Mechanical interface to dental implant. Workflow: Compatible with traditional casting/pressing or CAD/CAM milling. Sterilization: Not specified (standard dental component).
Indications for Use
Indicated for use as an aid in prosthetic rehabilitations for patients requiring bridge or over-denture restorations on endosseous dental implants. Temporary restorations may be used during the healing phase to maintain, stabilize, and form soft tissue; these must not be placed into occlusion. Final abutments and restorations may be placed into occlusion once the implant is fully osseointegrated.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, but instead of snakes, it features three abstract human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 5, 2015
Straumann USA, LLC Mr. Christopher Klaczyk Director of Regulatory Affairs and Clinical Research 60 Minuteman Road Andover, Massachusetts 01810
Re: K151157
Trade/Device Name: Straumann® Variobase® Abutment for Bars/bridges Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: July 2, 2015 Received: July 6, 2015
Dear Mr. Klaczyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang-S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151157
Device Name
Straumann® Variobase Abutments for Bridges/Bars
#### Indications for Use (Describe)
Straumann® Variobase™ prosthetic components directly connected to the endosseous dental implants are indicated for use as an aid in prosthetic rehabilitations. The prosthetic restoration (bridge or over-denture) can be cemented on the Straumann® Variobase™ prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and form the soft tissue during the healing phase. They may not be placed into occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K151157
# 510(k) Summary
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315<br>Owner/Operator No.: 9005052 | | |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact Person: | Christopher Klaczyk<br>Director of Regulatory Affairs and Clinical Research | | |
| Date Prepared: | August 4, 2015 | | |
| Product Code(s): | NHA (21 CFR 872.3630) | | |
| Device Class: | II<br>(21 CFR 872.3630) | | |
| Classification Panel: | Dental | | |
| Classification Name: | Endosseous dental abutment (21 CFR 872.3630) | | |
| Proprietary Name: | Straumann® Variobase® Abutment for Bridges/Bars | | |
| Predicate Device(s): | • K142890, Straumann Variobase Abutments | | |
| Reference Device(s): | • K132219, Straumann Variobase Abutments<br>• K120822, Straumann CARES Variobase Abutment NNC,<br>RN, WN, NC, RC<br>• K080239, Straumann Bone Level Multibase Abutments<br>• K990342, Straumann synOcta Prosthetics<br>• K140737, Straumann Screw-Retained Bridge and Bar | | |
| Device Description: | The Straumann Variobase Abutments for Bars & Bridges are<br>non-engaging abutments made from Ti-6Al-7Nb (TAN) that<br>support a bridge or bar reconstruction (framework or full<br>contour) on two or more dental implants. The corresponding<br>basal screw is delivered with the abutment for connecting the<br>abutment to the implant. A dental laboratory technician designs<br>and manufactures the bridge/bar reconstruction via their<br>preferred workflow using traditional or CAD/CAM methods. | | |
{4}------------------------------------------------
| Indications For Use: | Straumann® Variobase® prosthetic components directly<br>connected to the endosseous dental implants are indicated for<br>use as an aid in prosthetic rehabilitations. The prosthetic<br>restoration (bridge or overdenture) can be cemented on the<br>Straumann® Variobase® prosthetic components. A temporary<br>restoration can be used prior to the insertion of the final<br>components to maintain, stabilize and form the soft tissue<br>during the healing phase. They may not be placed into<br>occlusion. Final abutments and restorations may be placed into<br>occlusion when the implant is fully osseointegrated. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Straumann® Variobase® abutments are intended to be placed<br>into Straumann dental implants to provide support for prosthetic<br>reconstructions such as bridges and over-dentures. |
| Materials: | The subject devices are produced from titanium-6aluminum-<br>7niobium alloy (TAN). This is the same material as for the<br>predicate devices cleared to market per premarket notification<br>submissions K120822, K132219 and K142890. |
| Technological<br>Characteristics: | The subject Straumann Variobase Abutments for Bridge/Bar<br>devices have the same Indications For Use, Implant-to-<br>Abutment platforms, materials, packaging and dynamic fatigue<br>performance as the identified Straumann Variobase Abutment<br>predicate devices. |
| | The subject devices differ in that they do not include the<br>engaging features of the implant-to-abutment interface design<br>allowing full axial rotation and the coronal aspect takes a form<br>better suited to multi-unit restorations. |
{5}------------------------------------------------
| FEATURE | Straumann® Variobase®<br>Abutments for Bridges/Bars<br>Subject Submission | Primary Predicate<br>Straumann® Variobase®<br>Abutments<br>(K142890) | Reference Devices<br>Straumann® Multibase<br>(K080239) and synOcta<br>(K990342) Abutments | Equivalence<br>Discussion |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Straumann® Variobase®<br>prosthetic components directly<br>connected to the endosseous<br>dental implants are indicated for<br>use as an aid in prosthetic<br>rehabilitations. The prosthetic<br>restoration (bridge or over-<br>denture) can be cemented on the<br>Straumann® Variobase®<br>prosthetic components. A<br>temporary restoration can be<br>used prior to the insertion of the<br>final components to maintain,<br>stabilize and form the soft tissue<br>during the healing phase. They<br>may not be placed into<br>occlusion. Final abutments and<br>restorations may be placed into<br>occlusion when the implant is<br>fully osseointegrated. | The Straumann® Variobase®<br>Abutment is a titanium base<br>placed onto Straumann dental<br>implants to provide support for<br>customized prosthetic<br>restorations. Straumann®<br>Variobase® Abutments are<br>indicated for screw-retained<br>single tooth or cement-retained<br>single tooth and bridge<br>restorations. | Not applicable | Equivalent<br>The basic indication of<br>providing support for prostheses<br>is identical. Where the predicate<br>is indicated for both single-unit<br>and multi-unit restorations, the<br>subject devices are only suitable<br>for multi-unit restorations. This<br>limited indication is within the<br>scope of the indication of the<br>predicate devices. |
| Abutment<br>Material | Titanium-Aluminum-Niobium<br>alloy (Ti-6Al-7Nb) | Titanium-Aluminum-Niobium<br>alloy (Ti-6Al-7Nb) | Not applicable | Identical |
| Abutment<br>Diameter | 4.2 - 7.0 mm | 3.8 - 7.0 mm | Not applicable | Equivalent<br>Subject device diameters are<br>within the range of diameters for<br>the predicate device. |
| FEATURE | Straumann® Variobase®<br>Abutments for Bridges/Bars<br>Subject Submission | Primary Predicate<br>Straumann® Variobase®<br>Abutments<br>(K142890) | Reference Devices<br>Straumann® Multibase<br>(K080239) and synOcta<br>(K990342) Abutments | Equivalence<br>Discussion |
| Abutment<br>Height | 3.5 – 5.2 mm | 3.5 – 4.5 mm | Not applicable | Equivalent<br>The subject devices have the<br>same minimum height as the<br>Variobase predicates.<br>Performance of the taller subject<br>device is addressed via bench<br>testing. |
| Abutment<br>Apical<br>Design | Non-engaging Morse taper | Engaging CrossFit® and Morse<br>taper | Not applicable | Equivalent<br>The Morse taper portion of both<br>the subject and Variobase<br>predicate devices are identical.<br>Non-engaging abutments are<br>necessary so as not to introduce<br>mechanical constraints in the<br>framework that would affect fit. |
| Abutment<br>Coronal<br>Design | 10° or15° conical taper | Straight wall with detents | 6° conical taper (K990342)<br>30° conical taper (K080239) | Equivalent<br>The conic taper of the subject<br>devices is within the range<br>established by the synOcta and<br>Multibase predicates. |
| FEATURE | Straumann® Variobase®<br>Abutments for Bridges/Bars<br>Subject Submission | Primary Predicate<br>Straumann® Variobase®<br>Abutments<br>(K142890) | Reference Devices<br>Straumann® Multibase<br>(K080239) and synOcta<br>(K990342) Abutments | Equivalence<br>Discussion |
| Restoration<br>Material | Materials cleared by the FDA<br>under 21 CFR 872.6660 or<br>exempt materials as described<br>under 21 CFR 872.3060 (Noble<br>metal alloys) and 21 CFR<br>872.3710 (Base metal alloys) | Traditional Workflow:<br>Type 4 Metals (ISO 22674)<br>IPS e.max® Press Ceramic<br>Digital Workflow:<br>polycon® ae (temporary)<br>zerion® (permanent)<br>IPS e.max® CAD Ceramic<br>(permanent)<br>coron® (permanent) | Not applicable | Equivalent<br>The range of materials cleared<br>for use with the predicate<br>devices is within the scope of<br>materials indicated for the<br>subject devices. |
| Design<br>Workflow | Wax-up or Open CAD<br>Exempt from pre-market<br>notification according to 21<br>CFR 807.85(a)(1). | Wax-up or Open CAD | Not applicable | Identical |
| Manufacturing<br>Workflow | Traditional casting, pressing or<br>Straumann Milling Center<br>Exempt from pre-market<br>notification according to 21<br>CFR 807.85(a)(1). | Traditional casting, pressing or<br>Straumann Milling Center | Not applicable | Identical |
| Mode of<br>Attachment | Screw-retained or cement<br>retained | Screw-retained or cement<br>retained | Not applicable | Identical |
| Reusable | No | No | Not applicable | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
| Discussion of<br>Substantial<br>Equivalence: | The basic indication of providing support for prostheses is<br>identical. Where the predicate is indicated for both single-unit<br>and multi-unit restorations, the subject devices are only<br>suitable for multi-unit restorations. This limited indication is<br>within the scope of the indication of the predicate devices.<br>Subject device diameters are within the range of diameters for<br>the predicate devices. The subject devices have the same<br>minimum height as the Variobase predicates. |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Performance of the taller subject device is addressed via bench<br>testing. The Morse taper portion of both the subject and<br>Variobase predicate devices are identical. Non-engaging<br>abutments are necessary so as not to introduce mechanical<br>constraints in the framework that would affect fit. The conic<br>taper of the subject devices is within the range established by<br>the synOcta and Multibase predicates. The bonding surface<br>areas for the subject devices are consistent with those of the<br>predicate. The range of materials cleared for use with the<br>predicate devices is within the scope of materials indicated for<br>the subject devices. |
| Performance Data: | Per Guidance for Industry and FDA Staff - Class II Special<br>Controls Guidance Document: Root-form Endosseous Dental<br>Implants and Endosseous Dental Abutments dated May 12,<br>2004, the substantial equivalence of the subject device(s) are<br>satisfactorily addressed via bench studies. Dynamic fatigue test<br>data consistent with FDA guidance and ISO 14801 have been<br>referenced in support of this submission. |
| Conclusions: | Based upon our assessment of the design and applicable<br>performance data, the subject devices have been determined to<br>be substantially equivalent to the identified predicate devices. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.