K150367 · Jjgc Industria E Comercio DE Materiais Dentarios S · NHA · Jul 13, 2015 · Dental
Device Facts
Record ID
K150367
Device Name
Neodent Implant System
Applicant
Jjgc Industria E Comercio DE Materiais Dentarios S
Product Code
NHA · Dental
Decision Date
Jul 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations. PreFace Abutment is a titanium abutment to be used in fabricating a full custom abutment and placed onto Neodent dental implants to provide support for customized prosthetic restorations. PreFace Abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.
Device Story
Neodent Implant System components (Titanium Base and PreFace Abutments) serve as interfaces between Neodent dental implants and customized prosthetic restorations (crowns/bridges). Titanium Base Abutment acts as a base for zirconia superstructures; PreFace Abutment serves as a blank for milling custom titanium alloy abutments. Used in dental clinics; operated by dentists/prosthodontists. Workflow involves scanning and designing via Straumann CARES System; final fabrication performed at Straumann Milling Center. Output is a patient-specific abutment providing structural support for dental prosthetics. Benefits include restoration of dental function and aesthetics.
Clinical Evidence
No clinical data submitted. Substantial equivalence supported by bench testing, including biocompatibility, engineering analysis, dimensional analysis, and dynamic fatigue testing per ISO 14801. Sterilization validated per ISO 17665-1 and ISO 17665-2.
Technological Characteristics
Materials: ASTM F136 Ti-6Al-4V ELI titanium alloy. Design: Morse taper implant interface; Titanium Base (3.5/4.5mm diameter) and PreFace (11.5/15.8mm milling diameter). Energy: N/A (mechanical). Connectivity: Compatible with Straumann CARES CAD/CAM system. Sterilization: Moist heat.
Indications for Use
Indicated for patients requiring dental implant-supported prosthetic restorations, including single or multi-unit cement-retained restorations or single-unit screw-retained restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 13, 2015
Jjgc Industria E Comercio De Materiais Dentarios S c/o Mr. Kevin Thomas Paxmed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K150367
Trade/Device Name: Neodent Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 27, 2015 Received: May 28, 2015
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
lina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K150367
Device Name
Neodent Implant System
Indications for Use (Describe)
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.
PreFace Abutment is a titanium abutment to be used in fabricating a full custom abutment and placed onto Neodent dental implants to provide support for customized prosthetic restorations. PreFace Abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# JJGC Indústria e Comércio de Materiais Dentários SA
# Neodent Implant System
July 13, 2015
## ADMINISTRATIVE INFORMATION
| Manufacturer Name | JJGC Indústria e Comércio de Materiais Dentários S.<br>Av. Juscelino Kubitschek de Oliveira, 3291 - CIC<br>Curitiba, Paraná, 81270-200, Brazil | | |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--|
| | Telephone: | +55 41 2169 4058 | |
| | Fax: | +55 41 2169 1043 | |
| Official Contact | Jacson Cambruzzi<br>Head of Quality and Regulatory Affairs | | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | |
| | | Telephone: +1 (858) 792-1235 | |
| | Fax: | +1 (858) 792-1236 | |
| | Email: | kthomas@paxmed.com | |
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Neodent Implant System |
|---------------------------|------------------------------------|
| Common Name | Endosseous dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulation | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
{4}------------------------------------------------
### INTENDED USE
Titanium Base Abutment is a titanium base placed onto Neodent dental implants to provide support for customized prosthetic restorations. It is used with a coping and crown, or crown alone, and is indicated for cement-retained single or multi-unit restorations, or screw-retained single restorations.
PreFace Abutment is a titanium abutment to be used in fabricating a full custom abutment and placed onto Neodent dental implants to provide support for customized prosthetic restorations. PreFace Abutments are indicated for screw-retained single restorations or cement-retained single or multi-unit restorations.
### DEVICE DESCRIPTION
The purpose of this submission is to expand the Neodent Implant System components cleared under K101945, K123022, and K133592; these submissions included dental implants with a Morse taper abutment interface, mating abutments, abutment screws, and other associated components. This submission includes the Titanium Base Abutment, provided in two prosthetic platform diameters (3.5 and 4.5 mm), each in five gingival heights (0.8, 1.5, 2.5, 3.5 and 4.5 mm). Titanium Base is an abutment to be used as a base when fabricating a zirconia superstructure. This submission also includes the PreFace Abutment in two overall (milling) diameters of 11.5 and 15.8 mm. The PreFace Abutment with overall (milling) diameter of 15.8 mm is only used to mill an angled abutment.
PreFace is an abutment to be used in fabricating a full custom abutment in titanium alloy. After milling the minimum wall thickness for the PreFace must be no less than 0.4 mm, and the emergence profile diameter must be at least 3.3 mm. The maximum limit for angulation is 30°, and the maximum gingival height is 6 mm from the implant platform. In the case of an angled abutment, the cementable height from the emergence profile (prosthetic height) must not exceed 6 mm. The total height of the abutment (above the implant-abutment connection platform) must be no greater than 14.2 mm and no shorter than 4 mm. The range for the abutment body diameter (above the implant-abutment connection platform) is 3.3 mm. The implant interface must remain intact and cannot be modified.
The subject abutments are made of titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401). The subject abutments are compatible with Neodent dental implants having the Morse taper interface cleared in K101945, K123022 and K133592.
### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility, engineering analysis, and dimensional analysis. Mechanical testing was performed on the Titanium Base and PreFace abutments according to ISO 14801 Dentistry -Implants - Dynamic fatigue test for endosseous dental implants. Sterilization testing was performed according to ISO 17665-1 Sterilization of health care products - Moist heat -Part 1: Requirements for the development, validation and routine control of a sterilization
{5}------------------------------------------------
process for medical devices and ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1. Clinical data were not submitted in this premarket notification.
#### EQUIVALENCE TO MARKETED DEVICE
Neodent Implant System is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K120822, Straumann® CARES® Variobase™ Abutments, Institut Straumann AG;
K052272, Straumann C.A.R.E.S. Titanium Abutment, Institut Straumann AG;
K133421, Straumann® Magellan™ Abutment System, Institut Straumann AG;
K133592, Neodent Implant System, Neodent USA, Inc .;
K 123022, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA , and
K101945, Neodent Implant System, JJGC Indústria e Comércio de Materiais Dentários SA.
The primary predicate device for the Titanium Base abutment is K120822. The primary predicate device for the PreFace abutment is K052272. The reference predicate devices are K133421, K133592, K123022, and K101945. A comparison of the technological characteristics of the subject device and the primary predicate devices K120822 and K052272 is provided in the following table.
| | Subject Device | Primary Predicate Devices | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | JJGC Indústria e Comércio de<br>Materiais Dentários SA<br>Neodent Implant System<br>K150367 | Institut Straumann AG<br>Straumann® CARES® Variobase™<br>Abutments<br>K120822 | Institut Straumann AG<br>Straumann C.A.R.E.S. Titanium<br>Abutment<br>K052272 |
| Indications for Use | Titanium Base Abutment is a<br>titanium base placed onto Neodent<br>dental implants to provide support<br>for customized prosthetic<br>restorations. It is indicated for<br>cement-retained single or multi-unit<br>restorations, or screw-retained<br>single restorations.<br>PreFace Abutment is a titanium<br>abutment to be used in fabricating a<br>full custom abutment and placed<br>onto Neodent dental implants to<br>provide support for customized<br>prosthetic restorations. PreFace<br>Abutments are indicated for screw-<br>retained single restorations or<br>cement-retained single or multi-unit<br>restorations. | The Straumann® CARES®<br>Variobase™ Abutment is a two-<br>piece dental abutment consisting of<br>the Straumann® Variobase™<br>Abutment and the Straumann®<br>CARES® Variobase™ Coping<br>which is intended to be placed onto<br>Straumann dental implants to<br>provide support for prosthetic<br>reconstruction such as crowns and<br>bridges. Straumann® CARES®<br>Variobase™ Abutments are<br>indicated for screw-retained single<br>tooth or cement-retained single<br>tooth and bridge restorations.<br>The Straumann® CARES®<br>Variobase™ Coping polycon® ae in<br>combination with the Straumann®<br>Variobase™ Abutment is indicated<br>for temporary (up to 180 days)<br>dental restoration of a Straumann<br>dental implant. | Abutments are intended to be<br>placed into dental implants to<br>provide support for prosthetic<br>reconstructions such as crowns or<br>bridges. The Straumann C.A.R.E.S.<br>Titanium Abutment is indicated for<br>cemented restorations. The<br>abutment can be used in single<br>tooth replacements and multiple<br>tooth restorations. |
| Abutment Characteristics | | | |
| Design | Titanium Base and<br>CAD/CAM Blank | Titanium Base | CAD/CAM Blank |
| Prosthesis Attachment | Screw-retained<br>Cement-retained | Cement-retained | Cement-retained |
{6}------------------------------------------------
| | Subject Device | Primary Predicate Devices | |
|-----------------------------|---------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Comparison | JJGC Indústria e Comércio de<br>Materiais Dentários SA<br>Neodent Implant System<br>K150367 | Institut Straumann AG<br>Straumann® CARES® Variobase™<br>Abutments<br>K120822 | Institut Straumann AG<br>Straumann C.A.R.E.S. Titanium<br>Abutment<br>K052272 |
| Restoration | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit | Single-unit<br>Multi-unit |
| Abutment Diameter | Titanium Base: 3.5 and 4.5 mm<br>PreFace: 11.5 and 15.8 mm<br>(milling diameter) | 3.3 - 6.5 mm | 3.3 - 6.5 mm |
| Abutment Angle | Up to 30° | Up to 30° | Up to 30° |
| Abutment/ Implant Interface | Indexed | Indexed, non-indexed | Indexed, non-indexed |
| Abutment Material | Ti-6Al-4V | Ti-6Al-7Nb | Ti-6Al-7Nb |
| Abutment Screw Material | Ti-6Al-4V | Ti-6Al-7Nb | Ti-6Al-7Nb |
The subject device and the primary predicate devices have substantially equivalent indications for use, including providing support for customized prosthetic restorations.
The subject Titanium Base abutments are substantially equivalent in design, function, size, and material to the Straumann CARES Variobase Abutments cleared in K120822. All are titanium alloy bonding bases for CAD/CAM zirconia superstructures. The scanning and designing software used for the zirconia superstructure is the Straumann CARES System (K120822). Final fabrication of the abutment will be done at the Straumann Milling Center. The subject PreFace abutments are substantially equivalent in design, function, size, and material to the Straumann C.A.R.E.S. Titanium Abutment cleared in K052272. Both are abutment blanks for CAD/CAM fabrication of a custom titanium alloy abutment. Each has a precision implant-abutment interface compatible with specific implant systems. The scanning and designing software used for the subject PreFace abutment is the Straumann CARES System (K052272). Final fabrication of the abutment will be done at the Straumann Milling Center.
Minor differences in the technological characteristics of the subject device and primary predicate devices include the option for screw-retained prosthesis attachment, the range of (finished) abutment diameters, the provision of only indexed implant interfaces for the subject devices, and slight differences in the device materials (Ti-6Al-4V for the subject devices and Ti-6A1-7Nb for the primary predicate devices).
Mechanical test results demonstrated that the strengths of subject Titanium Base and PreFace abutments are equal to or greater than that of predicate abutments in K120822 and K133421, respectively.
{7}------------------------------------------------
#### 510(k) Summary K150367
## CONCLUSION
Overall, the subject device has the following similarities to the predicate devices:
- has the same intended use,
- uses the same operating principle,
- incorporates the same basic design, ●
- incorporates the same or very similar materials, and ●
- has similar packaging and is to be sterilized using the same processes. .
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.