K103234 · GC America, Inc. · NHA · Jun 30, 2011 · Dental
Device Facts
Record ID
K103234
Device Name
GC AADVA BO AC ABUTMENTS
Applicant
GC America, Inc.
Product Code
NHA · Dental
Decision Date
Jun 30, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The "GC Aadva Ti Abutments – Bo Ac Sv Bh" is intended for use with an endossous dental implant to support a prosthetic device in a partially or completely endentulous patient. It is intended for use to support a tooth prosthesis, in the mandible or maxilla. The abutment screw is intended to secure the abutment to the endosseous dental implant.
Device Story
GC Aadva Ti Abutments are dental implant components designed to support prosthetic restorations. The device consists of a titanium abutment and a titanium screw used to secure the abutment to a compatible endosseous dental implant fixture. These components are placed by dental professionals in a clinical setting. The abutment provides the interface between the implant fixture and the final tooth prosthesis, enabling the restoration of chewing function in patients with missing teeth. The device is available in various geometries (internal square or hexagon anti-rotational features) to ensure compatibility with specific implant systems from manufacturers including Institut Straumann AG, Nobel Biocare AB, BioHorizons, and Zimmer. By providing a stable connection to the implant, the abutment allows for the secure attachment of a prosthetic device, benefiting the patient through restored oral function and aesthetics.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing only. Static and fatigue testing were performed on worst-case implant-abutment assemblies in accordance with FDA's Guidance Document for Dental Implants and ISO 14801, demonstrating compatibility and structural integrity.
Technological Characteristics
Materials: Titanium alloy Ti-6Al-4V (ISO 5832-3). Design: Abutments feature internal square or hexagon anti-rotational interfaces for specific implant compatibility. Dimensions: Collar 2.5-6.0mm, angulation 0-20°, length 2.5-12.0mm, width 4.1-11.6mm. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for partially or fully edentulous patients requiring restoration of chewing function via attachment to dental implant fixtures placed in the maxilla or mandible.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
K103234 JUN 3 0 2011
## 510(k) Summary
# GC AADVA TI ABUTMENTS – BO AC SV BH
1994 : 1999 : 1999 : 1999 : 1999 : 1999 : 1999 : 1999 : 1999 : 1999 :
Date Prepared: June 10, 2011
| Submitter/Contact<br>Person | H. Carl Jenkins<br>The Wood Burditt Group<br>FDA Regulatory Counseling<br>10 E. Scranton Avenue, Suite 201<br>Lake Bluff, IL 60044<br>(ph) (847) 234-7500 x 205<br>(fax) (847) 574-0728<br>(email) hcjenkins@woodburditt.com |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | GC America, Inc.<br>3737 W. 127th Street<br>Alsip, IL 60803<br>800.323.3386 x4042<br>708.897.4042<br>708.897.4031 (fax) |
| Manufacturer | GC CORPORATION.<br>76-1 HASUNUMA-CHO,<br>ITABASHI-KU<br>TOKYO 174-8585 JAPAN |
| Device Name | GC AADVA TI ABUTMENTS - BO AC SV BH |
| Common Name | Endosseous Dental Implant Abutment |
| Classification | Class II<br>Procode NHA<br>Regulation: 21 CFR 872.3630 |
{1}------------------------------------------------
## Device Description
The "GC Aadva Ti Abutments – Bo Ac Sv Bh" are abutments which are placed into a dental implant to provide support for a prosthetic restoration. The subject abutments are indicated for use with the following implant fixtures:
. . .
| Types | Material | Antirotational<br>Features | Engaged Fixture<br>(Manufacturer) |
|------------------|-----------|----------------------------|----------------------------------------------------------------------------|
| IN-Bo-3.3mm #354 | Ti-6Al-4V | Internal Square | $\Phi$ 3.3<br>Bone Level Implant NC<br>(Institut straumann AG) |
| IN-Bo-4.1mm #355 | Ti-6Al-4V | Internal Square | $\Phi$ 4.1, $\Phi$ 4.8<br>Bone Level Implant RC<br>(Institut straumann AG) |
| IN-Ac-3.5mm #356 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 3.5<br>NobelActive NP<br>(Nobel Biocare AB) |
| IN-Ac-4.3mm #357 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 4.3, $\Phi$ 5.0<br>NobelActive RP<br>(Nobel Biocare AB) |
| IN-Bh-3.8mm #358 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 3.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Bh-4.6mm #359 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 4.6mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Bh-5.8mm #360 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 5.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Sv-3.5mm #351 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 3.7, $\Phi$ 4.1<br>Tapered Screw-Vent<br>(Zimmmer) |
| IN-Sv-4.5mm #352 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 4.7<br>Tapered Screw-Vent<br>(Zimmmer) |
| IN-Sv-5.7mm #353 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 6.0<br>Tapered Screw-Vent<br>(Zimmer) |
The "GC Aadva Ti Abutments – Bo Ac Sv Bh" components are made of Titanium grade Ti-6Al-4V and meets the requirements of ISO : 5832-3. The abutments are mounted into the implant
{2}------------------------------------------------
| Collar | Angulation | Length | Width | Head space |
|------------------------|-------------------|------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------|
| Max.6.0mm<br>Min.2.5mm | Max.20°<br>Min.0° | Max.12.0mm<br>Min.2.5mm -<br>3.5mm<br>(depending<br>on abutment<br>selected) | Max. $\Phi$ 11.6mm<br>Min. $\Phi$ 4.1mm -<br>$\Phi$ 6.1mm<br>(depending on<br>abutment<br>selected) | Max.11.8mm<br>Min.1.0mm |
with a screw also made of Titanium grade Ti-6A1-4V, which meets the requirements of ISO : 5832-3. Each of these abutments have the following design limitations:
## Intended Use:
The "GC Aadva Ti Abutments – Bo Ac Sv Bh" is intended for use with an endossous dental implant to support a prosthetic device in a partially or completely endentulous patient. It is intended for use to support a tooth prosthesis, in the mandible or maxilla. The abutment screw is intended to secure the abutment to the endosseous dental implant.
### Indications for use:
Indications for Use: "GC AADVA TI ABUTMENTS – BO AC SV BH" are dental implant abutments for use with partially or fully edentulous patients to restore chewing function by attachment to a dental implant fixture placed in the maxilla or mandible. Each abutment is accompanied by a screw in order to engage corresponding dental implant fixture.
| Abutment | Material | Antirotational<br>Features | Implant -- Engaged Fixture<br>(Manufacturer) |
|------------------|-----------|----------------------------|----------------------------------------------------------------|
| IN-Bo-3.3mm #354 | Ti-6A1-4V | Internal Square | Ф3.3<br>Bone Level Implant NC<br>(Institut straumann AG) |
| IN-Bo-4.1mm #355 | Ti-6Al-4V | Internal Square | Ф4.1, Ф4.8<br>Bone Level Implant RC<br>(Institut straumann AG) |
| IN-Ac-3.5mm #356 | Ti-6Al-4V | Internal Hexagon | Ф3.5<br>NobelActive NP<br>(Nobel Biocare AB) |
| IN-Ac-4.3mm #357 | Ti-6Al-4V | Internal Hexagon | Ф4.3, Ф5.0<br>NobelActive RP<br>(Nobel Biocare AB) |
#### Abutment - Implant Compatibility Tahle
{3}------------------------------------------------
| IN-Bh-3.8mm #358 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 3.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
|------------------|-----------|------------------|-----------------------------------------------------------|
| IN-Bh-4.6mm #359 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 4.6mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Bh-5.8mm #360 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 5.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Sv-3.5mm #351 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 3.7, $\Phi$ 4.1<br>Tapered Screw-Vent<br>(Zimmmer) |
| IN-Sv-4.5mm #352 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 4.7<br>Tapered Screw-Vent<br>(Zimmmer) |
| IN-Sv-5.7mm #353 | Ti-6Al-4V | Internal Hexagon | $\Phi$ 6.0<br>Tapered Screw-Vent<br>(Zimmmer) |
Prescription Use Only.
## Substantial Equivalence:
The applicant device is substantially equivalent to the predicate devices in its intended use, indications for use, and design, as described below:
| Product | Manufacturer | K Number |
|----------------------------------------------|-----------------------------------------------------------------------------------|----------|
| Atlantis Straumann Bone Level<br>Abutment | Astra Tech Inc.<br>25 First Street Cambridge, Massachusetts<br>02141 USA | K083871 |
| NobelActive Internal Connection | Nobel Biocare USA, LLC.<br>22715 Savi Ranch Parkway, Yorba Londa<br>CA 92887 USA | K071370 |
| GC Aadva Abutment | GC Corporation<br>76-1 Hasunuma-Cho, Itabashi-Ku<br>Tokyo, Japan | K072100 |
| Atlantis Abutment for BioHorizons<br>Implant | Atlantis Components Inc.<br>25 First Street Cambridge, Massachusetts<br>02141 USA | K073258 |
| ATLANTIS ABUTMENT FOR ZIMMER<br>INTERFACE | Atlantis Components, INC.<br>8944 Tamaroa Terrace Skokie, IL 60076 | K053373 |
{4}------------------------------------------------
| Component | Applicant Device Materials | Predicate Devices that Use Same Materials | |
|-----------|------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Block | Ti-6Al-4V<br>ISO : 5832-3<br><br>Ti : Bal.<br>Al : 5.5~6.75<br>V : 3.5~4.5<br>Fe<0.3<br>O<0.2<br>C<0.08<br>N<0.05<br>H<0.015 | Atlantis Straumann Bone Level Abutment<br>(K083871)<br>NobelActive Internal Connection (K071370)<br>GC AADVA ABUTMENT (K072100)<br>Atlantis Abutment for BioHorizons Implant<br>(K073258)<br>ATLANTIS ABUTMENT FOR ZIMMER INTERFACE<br>(K053373) | |
| | Screw | Ti-6Al-4V<br>ISO : 5832-3<br><br>Ti : Bal.<br>Al : 5.5~6.75<br>V : 3.5~4.5<br>Fe<0.3<br>O<0.2<br>C<0.08<br>N<0.05<br>H<0.015 | Atlantis Straumann Bone Level Abutment<br>(K083871)<br>NobelActive Internal Connection (K071370)<br>NobelActive Zirconia Abutment (K072129)<br>GC AADVA ABUTMENT (K072100)<br>Atlantis Abutment for BioHorizons Implant<br>(K073258)<br>ATLANTIS ABUTMENT FOR ZIMMER INTERFACE<br>(K053373) |
| Technological Characteristic | Applicant Device | Predicate Devices |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Material | Ti-6Al-4V<br>ISO : 5832-3 | Ti-6Al-4V<br>ISO : 5832-3 |
| Performance Characteristics | Allows the prosthesis to be<br>retained to the abutment;<br>abutment screw is intended to<br>secure the abutment to the<br>endosseous dental implant. | Allows the prosthesis to be<br>retained to the abutment;<br>abutment screw is intended to<br>secure the abutment to the<br>endosseous dental implant. |
| Intended Use | Intended for use with an<br>endosseous dental implant to<br>support a prosthetic device in<br>a partially or fully endentulous<br>patient to restore chewing | Intended for use with an<br>endosseous dental implant to<br>support a prosthetic device in<br>a partially or fully endentulous<br>patient to restore chewing |
Materials used in blocks and screws
{5}------------------------------------------------
| | function. Intended to be<br>attached to a dental implant<br>fixture placed in the maxilla or<br>mandible. | function. Intended to be<br>attached to a dental implant<br>fixture placed in the maxilla or<br>mandible. |
|--|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
## Summary of Non-Clinical Performance Testing:
Static and fatigue testing was conducted on the "worst case scenario" implant-abutment combination assemblies in accordance with FDA's Guidance Document for Dental Implants and ISO 14801. Test results demonstrated that the "GC Aadva Ti Abutments – Bo Ac Sv Bh" are compatible with the referenced implant fixtures and the implant-abutment assemblies support adequate static and fatigue test loads. Performance testing demonstrates that the device performs as intended and is as safe and effective as the cited predicates.
## Substantial Equivalence Conclusion Statement:
Based on noted similarities and comparative traits in the indications for use, manufacturing materials, design and performance characteristics, and the fact that the applicant device and the predicate devices have demonstrated acceptable static and fatigue test loads during performance testing conducted in accordance with FDA's Guidance Document for Dental Implants and ISO 14801, the "GC Aadva Ti Abutments – Bo Ac Sv Bh" is substantially equivalent to the following predicate devices:
| Product | Manufacturer | K Number |
|----------------------------------------------|-----------------------------------------------------------------------------------|----------|
| Atlantis Straumann Bone Level<br>Abutment | Astra Tech Inc.<br>25 First Street Cambridge, Massachusetts<br>02141 USA | K083871 |
| NobelActive Internal Connection | Nobel Biocare USA, LLC.<br>22715 Savi Ranch Parkway, Yorba Londa<br>CA 92887 USA | K071370 |
| GC Aadva Abutment | GC Corporation<br>76-1 Hasunuma-Cho, Itabashi-Ku<br>Tokyo, Japan | K072100 |
| Atlantis Abutment for BioHorizons<br>Implant | Atlantis Components Inc.<br>25 First Street Cambridge, Massachusetts<br>02141 USA | K073258 |
| ATLANTIS ABUTMENT FOR ZIMMER<br>INTERFACE | Atlantis Components, INC.<br>8944 Tamaroa Terrace Skokie, IL 60076 | K053373 |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -- WO66-G609 Silver Spring, MD 20993-0002
GC America, Incorporated C/O Mr. H. Carl Jenkins Wood Burditt Group 10 East Scranton Avenue. Suite 201 Lake Bluff. Illinois 60044
JUN 3 0 2011
Re: K103234
Trade/Device Name: GC AADVA TI ABUTMENTS - BO AC SV BH Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: June 10, 2011 Received: June 13, 2011
Dear Mr. Jenkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 – Mr. Jenkins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{8}------------------------------------------------
Section 5 - Indications for Use Statement
Indications for Use
510(k) Number (if known):
Device Name: GC AADVA TI ABUTMENTS - BO AC SV BH
Indications for Use: "GC AADVA TI ABUTMENTS – BO AC SV BH" are dental implant abutments for use with partially or fully edentulous patients to restore chewing function by attachment to a dental implant fixture placed in the maxilla or mandible. Each abutment is accompanied by a screw in order to engage corresponding dental implant fixture.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|--|
|--------------------------------------------------------|--|
Susa Luise
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 5.1.of 5.2
510(k) Number: 1403234
{9}------------------------------------------------
Abutment – Implant Compatibility Table
| Abutment | Material | Antirotational Features | Implant -- Engaged Fixture (Manufacturer) |
|------------------|-----------|-------------------------|----------------------------------------------------------------------------|
| IN-Bo-3.3mm #354 | Ti-6Al-4V | Internal Square | $\phi$ 3.3<br>Bone Level Implant NC<br>(Institut straumann AG) |
| IN-Bo-4.1mm #355 | Ti-6Al-4V | Internal Square | $\phi$ 4.1, $\phi$ 4.8<br>Bone Level Implant RC<br>(Institut straumann AG) |
| IN-Ac-3.5mm #356 | Ti-6Al-4V | Internal Hexagon | $\phi$ 3.5<br>NobelActive NP<br>(Nobel Biocare AB) |
| IN-Ac-4.3mm #357 | Ti-6Al-4V | Internal Hexagon | $\phi$ 4.3, $\phi$ 5.0<br>NobelActive RP<br>(Nobel Biocare AB) |
| IN-Bh-3.8mm #358 | Ti-6Al-4V | Internal Hexagon | $\phi$ 3.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Bh-4.6mm #359 | Ti-6Al-4V | Internal Hexagon | $\phi$ 4.6mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Bh-5.8mm #360 | Ti-6Al-4V | Internal Hexagon | $\phi$ 5.8mm<br>Tapered Internal Implant<br>(BioHorizons) |
| IN-Sv-3.5mm #351 | Ti-6Al-4V | Internal Hexagon | $\phi$ 3.7, $\phi$ 4.1<br>Tapered Screw-Vent<br>(Zimmer) |
| IN-Sv-4.5mm #352 | Ti-6Al-4V | Internal Hexagon | $\phi$ 4.7<br>Tapered Screw-Vent<br>(Zimmer) |
| IN-Sv-5.7mm #353 | Ti-6Al-4V | Internal Hexagon | $\phi$ 6.0<br>Tapered Screw-Vent<br>(Zimmer) |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 5.2 of 5.2
... ... ... ... ..............................................................................................................................................................
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.