BIOACTYS

K082286 · Graftys · LYC · Feb 24, 2009 · Dental

Device Facts

Record IDK082286
Device NameBIOACTYS
ApplicantGraftys
Product CodeLYC · Dental
Decision DateFeb 24, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3930
Device ClassClass 2
AttributesTherapeutic

Indications for Use

BIOACTYS® is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region. These defects may be surgically created osseous or defects created from traumatic injury to bone.

Device Story

BIOACTYS® is a sterile, single-use, bioresorbable bone void filler provided as granules. Composed of a two-phase ceramic (60% hydroxyapatite, 40% beta-tricalcium phosphate), it features a microporous and macroporous structure (70% total volume ratio). Used by clinicians in oral and maxillofacial surgery, the granules are placed into bony voids or gaps. The material acts as an osteoconductive scaffold, promoting bone ingrowth while gradually resorbing and being replaced by natural bone during the healing process. It is intended for non-structural support of osseous defects.

Clinical Evidence

Bench testing only. In vitro tests were performed in accordance with the Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices (April 2005) to demonstrate substantial equivalence.

Technological Characteristics

Sterile, single-use, bioresorbable ceramic granules. Composition: 60% Hydroxyapatite / 40% beta-Tricalcium Phosphate. Porous structure with 70% total volume ratio. Osteoconductive mechanism. No electronic or software components.

Indications for Use

Indicated for patients requiring filling, augmentation, or reconstruction of periodontal or bony defects in the oral and maxillofacial region, including surgically created osseous defects or those resulting from traumatic injury.

Regulatory Classification

Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a stylized letter G with an arrow pointing upwards and to the right. The letter G is drawn in a thick, bold font, and the arrow is positioned within the curve of the G. The image has a grainy texture, suggesting it may be a scanned or low-resolution image. K082286 # Graftys ## BIOACTYS® 510(k) Summary FEB 2 4 2009 #### Prepared: July 29, 2008 #### 1. Submitter Information | Name : | GRAFTYS | |------------|----------------------------------------------------------------------------------------------------| | Address : | Eiffel Park – Bât D<br>415, rue Claude Nicolas Ledoux-<br>13 854 AIX EN PROVENCE Cedex 3<br>FRANCE | | Telephone: | + 33 (0) 4 42 60 30 00 | | Facsimile: | + 33 (0) 4 42 60 30 11 | | Contact: | Anthony LE NAOUR - Regulatory Manager | #### Name of Device 2. | Trade Name: | BIOACTYS® | |----------------------|--------------------------------------------------------------------------| | Common Name: | Resorbable calcium salt bone void filler device | | Classification name: | Dental Bone Grafting Material Device (CFR 872.3920 ; Product Code : LYC) | | Device panel | Oral/Dental | #### 3. Legally Marketed Predicate Device MBCP ™ - Resorbable calcium salt bone void filler device [K051885] Predicate GRAFTYS®BCP- Resorbable calcium salt bone void filler device [K073064] #### Device Description 4. BIOACTYS® is a sterile single use bioresorbable bone void filling substitute. BIOACTYS® is a microporous and macroporous two-phase calcium phosphate ceramic made of 60% Hydroxyapatite and 40% beta-tricalcium phosphate. Following placement in the bony voids or gap, BIOACTYS® resorbs and is replaced with bone during the healing process. BIOACTYS® is available in the form of granules. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows a handwritten letter 'G' in a bold, sans-serif font. The letter is slightly tilted to the left. The image quality is somewhat grainy, with visible noise and artifacts around the edges of the letter. ### 5. Intended Use BIOACTYS® is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region #### 6. Technological characteristics BIOACTYS® and the predicate devices have the same technological characteristics. They are sterile single use devices made of 60% Hydroxyapatite/40% β-Tricalcium Phosphate and have a similar porous structure (total volume ratio of 70%) that promotes bone ingrowth by osteoconductivity. BIOACTYS® and predicates are provided sterile for single-use. #### 7. Non clinical performance data In vitro tests, performed according to the Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices; Guidance for Industry and FDA Staff issued April 28 2005, support the substantial equivalence between BIOACTYS® and the predicate devices. #### 8. Conclusion BIOACTYS® is claimed to be substantially equivalent in terms of safety and effectiveness to the predicate devices as a non structural osteoconductive bone void filler for osseous defect. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Anthony Le Naour Regulatory Affairs Manager Eiffel Park-Bât D 415, rue Claude Nicolas Ledoux FEB 2 4 2009 FRANCE F-13 854 Aix en Provence Cedex 3 Re: K082286 Graftys Trade/Device Name: Bioactys® Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: February 12, 2009 Received: February 17, 2009 Dear Mr. Le Naour: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Le Naour Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Saytte Y. Michael Dons. · Ginette Y. Michaud, M.D. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health #### Enclosure {4}------------------------------------------------ 510(k) Premarket Notification Graftys Bioactys- Resorbable Bone Void Filler Confidential # Statement of Indications for Use 510(k) Number (if known): Device Name: Bioactys® Indications For use: Bioactys® is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region. These defects may be surgically created osseous or defects created from traumatic injury to bone. x Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart C) (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...