Post-market complaint records; Clinical literature (18 publications)
The sponsor provided a review of 13.5 years of global post-market complaint data and a review of 18 clinical publications to demonstrate successful product use and safety across all indicated clinical applications.
Post-market surveillance; Complaint history; Clinical literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Review of clinical literature and post-market complaint history; Retrospective review of clinical literature and post-market surveillance data; Follow-up/Duration: 13.5 years; Study Period: 13.5 years
Patients treated with Straumann BoneCeramic in clinical practice; Sample Size: Not specified (18 publications reviewed); Number of Sites: Not applicable
Not applicable for this study
Product safety and successful clinical use
Indications for Use
Straumann® BoneCeramic is intended to be used for filling and/or augmenting intraoral/maxillofacial osseous defects.
Device Story
Straumann® BoneCeramic is a synthetic, osteoconductive bone graft substitute composed of biphasic calcium phosphate (60% hydroxyapatite, 40% beta-tricalcium phosphate). It features 90% porosity with interconnected pores (100-500 microns). Supplied as sterile granules in two size ranges (400-700 µm and 500-1000 µm), it is used by dental clinicians to fill or augment osseous defects in the oral/maxillofacial region. The material acts as a scaffold, gradually resorbing and being replaced by the patient's vital bone during the natural remodeling process. It is intended for surgical implantation to support bone regeneration in periodontal, alveolar, and sinus grafting procedures.
Clinical Evidence
Bench testing only. Biocompatibility evaluated per ISO 10993. Packaging validated per ISO 11607 and ASTM F1980. Sterilization validated per ISO 11137. Pyrogenicity confirmed via LAL Endotoxin Analysis (limit 20 EU/device). Clinical performance supported by a review of 18 clinical publications and a 13.5-year post-market complaint rate of <0.05%.
Indicated for filling and/or augmenting intraoral/maxillofacial osseous defects, including intrabony periodontal osseous and furcation defects, augmentation of alveolar ridge bony defects, filling of tooth extraction sites, and sinus elevation grafting.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 24, 2020
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01801
Re: K201051
Trade/Device Name: Straumann® BoneCeramic Regulation Number: 21 CFR 872.3930 Regulation Name: Bone grafting material Regulatory Class: Class II Product Code: LYC Dated: August 23, 2020 Received: August 25, 2020
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K201051
### Device Name:
Straumann® BoneCeramic
Indications for Use (Describe)
Straumann® BoneCeramic is indicated for filling and/or augmenting the following intraoral/maxillofacial osseous defects:
- Intrabony periodontal osseous and furcation defects
- Augmentation of bony defects of the alveolar ridge ●
- Filling tooth extraction sites ●
- Sinus elevation grafting .
Type of Use (Select one or both, as applicable) 図Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{4}------------------------------------------------
# Straumann® BoneCeramic
510(k) Summary
#### 510(k) Summary 5
#### 5.1 Submitter's Contact Information
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 900505-<br>On the behalf of:<br>Institut Straumann AG<br>Peter Merian-Weg 12<br>CH-4002 Basel, Switzerland |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 |
| Prepared By: | Christelle Gerspach-Gasser<br>Senior Lead RA Biomaterials<br>Institut Straumann AG<br>Phone number: +41 61 965 1666 |
| Date Prepared: | September 23, 2020 |
#### Name of the Device 5.2
| Trade Names: | Straumann® BoneCeramic | |
|------------------------|-----------------------------------|--|
| Common Name: | Bone Grafting Material | |
| Classification Name: | Bone Grafting Material, Synthetic | |
| Regulation Number: | §872.3930 | |
| Device Classification: | II | |
| Product Code(s): | LYC | |
| Classification Panel: | Dental | |
{5}------------------------------------------------
### Straumann® BoneCeramic
510(k) Summary
#### Predicate Device(s) 5.3
Primary Predicate:
- K040646 Straumann Granules .
#### 5.4 -Device Description
Straumann® BoneCeramic is a fully synthetic bone graft substitute of medical grade purity in particulate form composed of biphasic calcium phosphate. It consists of a mixture of 60% hydroxyapatite (HA), which is 100% crystalline, and of 40% of the beta form of tricalcium phosphate (beta-TCP). BoneCeramic is 90% porous with interconnected pores of 100-500 microns in diameter. It is osteoconductive and gradually resorbed and replaced by vital bone during bone remodeling.
BoneCeramic is available in two granule sizes: 400-700 µm diameter and 500-1000 µm diameter and in three different filling volumes: 0.25g, 0.5g and 1.0g. It is delivered sterile.
#### Intended Use 5.5
Straumann® BoneCeramic is intended to be used for filling and/or augmenting intraoral/maxillofacial osseous defects.
#### 5.6 Indications for Use
Straumann® BoneCeramic is indicated for filling and/or augmenting the following intraoral/maxillofacial osseous defects:
- . Intrabony periodontal osseous and furcation defects
- . Augmentation of bony defects of the alveolar ridge
- Filling tooth extraction sites .
- . Sinus elevation grafting
#### Technological Characteristics 5.7 -
The chemical composition, structure and indications for use of the subject device are identical to those of the predicate device. The only difference between the two relies in the different primary packaging utilized for the subject device. A comparison of the technological characteristics between the subject device and the primary predicate is provided in Table 1.
{6}------------------------------------------------
### Straumann® BoneCeramic
### 510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K201051 | K040646 |
| Indications for<br>use | Straumann® BoneCeramic is indicated for<br>filling and/or augmenting the following<br>intraoral/maxillofacial osseous defects:<br>Intrabony periodontal osseous and<br>furcation defects Augmentation of bony defects of the<br>alveolar ridge Filling tooth extraction sites Sinus elevation grafting | Straumann Granules are indicated for filling<br>and/or augmenting intraoral /maxillofacial<br>osseous defects, such as:<br>Intrabony periodontal osseous defects Furcation defects Augmentation of bony defects of the<br>alveolar ridge Filling of tooth extraction sites and<br>sinus elevation grafting. |
| Material<br>(Chemical<br>composition) | Hydroxyapatite (60%) and beta-tricalcium<br>phosphate (40%) | Hydroxyapatite (60%) and beta-tricalcium<br>phosphate (40%) |
| Structure /<br>Porosity | Matrix consisting of interconnected pores<br>(90% porosity), the size of the pores lies<br>within a range of approx. 100 – 500 microns<br>in diameter. | Matrix consisting of interconnected pores<br>(90% porosity), the size of the pores lies<br>within a range of approx. 100 – 500 microns<br>in diameter. |
| Particle sizes | 400 - 700 µm and 500 – 1000 µm | 400 - 700 µm and 500 – 1000 µm |
| Packaging | Sterile barrier system and protective<br>packaging:<br>1) Primary packaging: polystyrene jar closed<br>with a polypropylene cap<br>2) Secondary packaging: PETG blister<br>sealed with a Tyvek lid (sterile barrier)<br>3) Tertiary packaging: cardboard box | Sterile barrier system and protective<br>packaging:<br>1) Primary packaging: PET blister with PE<br>sealing layer<br>2) Secondary packaging: PET blister sealed<br>with a Tyvek lid (sterile barrier)<br>3) Tertiary packaging: cardboard box |
| Sterilization | Gamma irradiation | Gamma irradiation |
| Biological<br>characteristics | Biocompatible | Biocompatible |
#### Performance Testing 5.8
Chemical characterization and biological testing were conducted according to the ISO standard for the biological evaluation of medical devices, ISO 10993, and to the FDA guidance document for this standard (FDA 2016, Use of International Standard ISO-10993). The new packaging material has been assessed in accordance to the above-mentioned standard and the biocompatibility of the device was confirmed. Furthermore, the packaging process and shelf life of the device have been validated per ISO 11607 and ASTM F1980.
The sterilization process for the subject device as indicated in the labeling has been validated according to ISO 11137 "Sterilization of health care products – Radiation", part 1 and 2. The
{7}------------------------------------------------
### Straumann® BoneCeramic
### 510(k) Summary
devices will not be marketed as non-pyrogenic. Pyrogenicity information provided is based on FDA Guidance on "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submission for Devices Labeled as Sterile, issued on 21 January 2016." The method used to determine the device meets pyrogen limit specifications is LAL Endotoxin Analysis on every batch with a testing limit of 20 EU/device, based on a blood contacting and implanted device.
Device marketing and complaint history and review of the relevant clinical literature was provided. Over a period of 13.5 years, the overall complaint rate based on individual units sold globally for the subject devices was <0.05%. The review of the relevant clinical literature included 18 clinical publications across the relevant indications for use: intrabony osseous and furcation defects, augmentation of bony defects of the alveolar ridge, filling of tooth extraction sites, and sinus elevation grafting. The device marketing and complaint history and review of the relevant clinical literature provided demonstrates successful product use.
#### 5.9 Conclusion
The documentation submitted in this premarket notification demonstrates that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.