1. all direct restorations 2. all indirect restorations 3. intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, metals, glass ceramics, tooth structure, and composites) 4. desensitizing/sealing of tooth structure 5. protective varnish for glass ionomer fillings 6. priming of enamel for orthodontic use
Device Story
All-Bond Universal w/BAC is a light-cured, single-component dental adhesive; ethanol/water-based; contains benzalkonium chloride (BAC). Used by dentists in clinical settings for direct/indirect restorations, intra-oral repairs, desensitizing, and orthodontic priming. Applied to dentin and enamel; compatible with light-cured and dual-cured composites; can be used with or without phosphoric acid etchant. Simplifies bonding by combining etching, priming, and bonding in one bottle. Provides a stable interface between tooth structure and restorative materials; BAC acts as a bond enhancer. Clinical benefit includes improved bond strength durability and reduced sensitivity for patients.
Clinical Evidence
Bench testing only. Evaluated per ISO 29022:2013 (bond strength, moisture tolerance, immediate bond strength), ISO 4049:2009 (film thickness), ADA/ANSI Specification No. 27 (water sorption/solubility), and pH testing. Bond strength durability testing showed a 4% increase for the subject device compared to 23-27% reductions in predicates. Biocompatibility evaluated per ISO 7405:2008 and ISO 10993-1, including an oral toxicity study in 10 rats over 7 days, showing no toxicity.
Technological Characteristics
Single-component, light-cured, unfilled multifunctional methacrylate resin adhesive. Ethanol/water-based solvent. Contains benzalkonium chloride (BAC). Film thickness 3.4μm. Compatible with self-etch and total-etch techniques. Standards: ISO 29022:2013, ISO 4049:2009, ADA/ANSI Specification No. 27, ISO 7405:2008, ISO 10993-1.
Indications for Use
Indicated for patients requiring dental restorations, intra-oral repairs of fixed prostheses (zirconia, alumina, metals, glass ceramics, tooth structure, composites), tooth desensitization/sealing, protective varnishing for glass ionomer fillings, and enamel priming for orthodontic procedures.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
Bisco, Inc. Ryan Hobson Regulatory Affairs Product Registration Coordinator 1100 West Irving Park Road Schaumburg, Illinois 60193
Re: K161051
Trade/Device Name: All-Bond Universal w/ BAC Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: August 8, 2016 Received: August 10, 2016
Dear Ryan Hobson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael D. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K161051
Device Name All-Bond Unviersal w/BAC
Indications for Use (Describe)
1. all direct restorations
2. all indirect restorations
3. intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, metals, glass ceramics, tooth structure, and composites)
4. desensitizing/sealing of tooth structure
5. protective varnish for glass ionomer fillings
6. priming of enamel for orthodontic use
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the alphanumeric string "K161051" in a bold, sans-serif font. The string appears to be a code or identifier. The characters are uniformly sized and spaced, making the string easily readable. The background is plain and white.
Image /page/3/Picture/1 description: The image shows the BISCO logo. The logo consists of a series of connected hexagons in red, white, and blue. Below the hexagons is the word "BISCO" in a sans-serif font, with the top half of the letters in gray and the bottom half in black.
## 510 (k) SUMMARY
Applicant:
Bisco. Inc. 1100 W. Irving Park Road Schaumburg IL, 60193
Contact Person:
Rvan Hobson RA Product Registration Coordinator Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193 Tel: 847-534-6143 Fax: 847-534-6143
Date Prepared:
8 September 2016
Trade Name: Common Name: Product Code: Classification/Name:
All-Bond Universal w/BAC Light Curable Dental Adhesive w/BAC KLE Resin Tooth Bonding Agent Class II per 21 CFR 872.3200
## Predicate Devices:
## Primary Predicate:
DREAMBOND by Bisco, Inc. K112118, Agent, Tooth Bonding Resin - KLE
## Reference Predicate:
PEAK UNIVERSAL BOND (Light Cure Resin Adhesive and Chlorhexidine), by Ultradent, K100752, Agent, Tooth Bonding Resin - KLE
## Indications for Use:
## ALL-BOND UNIVERSAL w/BAC adhesive is used for:
- 1. all direct restorations
- 2. all indirect restorations
- intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, 3. metals, glass ceramics, tooth structure, and composites)
- 4. desensitizing/sealing of tooth structure
- 5. protective varnish for glass ionomer fillings
- 6. priming of enamel for orthodontic use
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the BISCO logo, which consists of three hexagons in red and blue, followed by the word "BISCO" in gray and black. The text "510 (k) SUMMARY (continued)" is located to the right of the logo. The background of the image features a pattern of light blue hexagons.
The indications for use of AII-Bond Universal w/BAC are the same as those for its predicates and are summarized in the table below:
| | DREAMBOND<br>(K112118) | All-Bond Universal w/BAC | Peak Universal Bond<br>(K100752) |
|----|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| 1. | all direct restorations | 7. all direct restorations | Peak™ Universal<br>Bond bonds to the<br>following materials:<br>• Dentin and<br>enamel<br>• Porcelain<br>• Metal<br>• Composite |
| 2. | all indirect restorations | 8. all indirect restorations | |
| 3. | intra-oral repairs (i.e.<br>repair of any fixed dental<br>prosthesis containing<br>zirconia, alumina, metals,<br>glass ceramics, tooth<br>structure, and composites) | 9. intra-oral repairs (i.e.<br>repair of any fixed dental<br>prosthesis containing<br>zirconia, alumina, metals,<br>glass ceramics, tooth<br>structure, and composites) | |
| 4. | desensitizing/sealing of<br>tooth structure | 10. desensitizing/sealing of<br>tooth structure | |
| 5. | protective varnish for glass<br>ionomer fillings | 11. protective varnish for glass<br>ionomer fillings | |
| 6. | priming of enamel for<br>orthodontic use | 12. priming of enamel for<br>orthodontic use | |
#### Description of Applicant Device:
BISCO's ALL-BOND UNIVERSAL w/BAC is a light-cured single-component dental bonding agent that combines etching, priming, and bonding with benzalkonium chloride (BAC) in one bottle. ALL-BOND UNIVERSAL w/BAC is an ethanol/water-based dental adhesive which bonds to dentin and cut and un-cut enamel. ALL-BOND UNIVERSAL w/BAC has been designed to be fully compatible with light-cured and dual-cured composite materials. ALL-BOND UNIVERSAL w/BAC can be used for the bonding of both direct and indirect restorations. ALL-BOND UNIVERSAL w/BAC may be used with or without phosphoric acid etchant.
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the BISCO company logo. The logo consists of five hexagons arranged in a row, with the two outer hexagons being red and the three inner hexagons being blue. Below the hexagons is the word "BISCO" in gray, with a gradient effect that makes the bottom of the letters darker than the top.
# 510 (k) SUMMARY (continued)
#### Technological Characteristics:
Comparisons of the chemical composition of ALL-BOND UNIVERSAL w/BAC to the predicates is provided in the following table:
| Chemical Composition | DREAMBOND<br>(K112118) | Peak Universal<br>Bond (K100752) | All-Bond Universal<br>w/BAC |
|--------------------------|-------------------------------------------------|-----------------------------------------------|-------------------------------------------------|
| Method of polymerization | Light Cured | Light Cured | Light Cured |
| Resin composition | Unfilled, multifunctional<br>methacrylate resin | Filled, multifunctional<br>methacrylate resin | Unfilled, multifunctional<br>methacrylate resin |
| Solvent | Ethanol based | Ethanol based | Ethanol based |
| Method of application | Single component<br>adhesive | Multi-component,<br>primer required | Single component<br>adhesive |
| Bond Enhancer | Not applicable | Chlorohexidine | Benzalkonium<br>Chloride |
| Physical / Mechanical Property<br>Comparison | DREAMBOND<br>(K112118) | Peak Universal Bond<br>(K100752) | All-Bond Universal<br>w/BAC |
|----------------------------------------------|--------------------------|----------------------------------|-----------------------------|
| Film thickness | 2.6μm | 17.0μm | 3.4μm |
| Etch Methods | Self-etch and total-etch | Self-etch and total-etch | Self-etch and total-etch |
The difference in chemical composition as compared to the DREAMBOND is the addition of Benzalkonium. Both DREAMBOND and ALL-BOND UNIVERSAL w/BAC have the same technological characteristics before the addition of Benzalkonium Chloride. All-Bond Universal w/BAC uses Benzalkonium Chloride while Peak Universal Bond (K100752) uses Chlorohexidine however; both are cationic surface-acting agents and both have a long history of use in dental medical devices. Peak Universal Bond (K100752) is a filled resin, All-Bond Universal w/BAC and DREAMBOND are both unfilled resins, the lack of filler does not affect intended use and performance data shows the materials are equivalent. All-Bond Universal w/BAC is a single component system, requiring no additional primer in self-etch mode and fewer steps in its use compared to Peak Universal Bond (K100752).
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the BISCO company logo. The logo consists of five connected hexagons, with the two outer hexagons colored red, the middle hexagon colored blue, and the two inner hexagons colored light blue. Below the hexagons is the word "BISCO" in gray letters, with the bottom half of the letters colored black.
# 510 (k) SUMMARY (continued)
#### Performance Data:
The following physical/mechanical properties of All-Bond Universal w/BAC were tested:
| Physical / Mechanical Property | All-Bond Universal w/BAC |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Bond Strength<br>(Modified ISO 29022:2013) | All-Bond Universal w/BAC is equivalent to the primary and reference predicate. |
| Moisture Tolerance<br>(Modified ISO 29022:2013) | All-Bond Universal w/BAC is equivalent to the primary and reference predicate. |
| Film Thickness<br>(ISO 4049:2009) | All-Bond Universal w/BAC meets the requirements of ISO 4049:2009 for Film Thickness. |
| Water Sorption / Solubility<br>(ADA/ANSI Specification No. 27) | All-Bond Universal w/BAC is equivalent to the primary predicate. |
| Immediate Bond Strength<br>(Modified ISO 29022:2013) | All-Bond Universal w/BAC is equivalent to the primary and reference predicate. |
| Bond Strength Durability | Reference predicate showed a 23% reduction in bond strength, primary predicate showed a 27% reduction in bond strength, and All-Bond Universal w/BAC showed a 4% increase in bond strength on completion of the test. |
| Solvent Evaporation | All-Bond Universal w/BAC is equivalent to the primary and reference predicate. |
| pH | All-Bond Universal w/BAC is equivalent to the primary predicate. |
#### Biocompatibility:
An evaluation of biocompatibility was conducted using ISO 7405:2008 and ISO 10993-1 to determine the biocompatibility of All-Bond Universal w/BAC. It is concluded from the evaluation and the results of the Oral Toxicity Study (10 rats, 7 days) that All-Bond Universal w/BAC was not toxic in this test.
#### Conclusion:
It is concluded from review of the predicate device indications, chemical composition, biocompatibility, and physical properties that All-Bond Universal w/BAC to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.