K112118 · Bisco, Inc. · KLE · Nov 17, 2011 · Dental
Device Facts
Record ID
K112118
Device Name
DREAMBOND
Applicant
Bisco, Inc.
Product Code
KLE · Dental
Decision Date
Nov 17, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
The indications of use of DREAMBOND are: 1. all direct restorations 2. all indirect restorations 3. intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, metals, glass ceramics, tooth structure, and composites) 4. desensitizing/sealing of tooth structure 5. protective varnish for glass ionomer fillings 6. priming of enamel for orthodontic use.
Device Story
DREAMBOND is a single-component, ethanol/water-based dental adhesive; combines etching, priming, and bonding in one bottle. Used by dental professionals for direct/indirect restorations, intra-oral repairs, desensitizing, and orthodontic priming. Compatible with light-cured, self-cured, and dual-cured composites; can be used with or without phosphoric acid etchant. Applied to dentin and enamel to facilitate bonding of restorative materials. Benefits include simplified clinical workflow and versatility across multiple dental applications.
Clinical Evidence
Bench testing only. Physical and mechanical properties were evaluated using ISO, ADA/ANSI, and internal Bisco protocols. Biocompatibility was assessed via oral toxicity testing in mice, showing no mortality or toxicity.
Technological Characteristics
Single-component, ethanol/water-based dental adhesive. Contains unfilled, multifunctional methacrylate resin. Low viscosity. Compatible with light, self, and dual-cured composites. Designed for both self-etch and total-etch techniques.
Indications for Use
Indicated for patients requiring direct or indirect dental restorations, intra-oral repairs of fixed prostheses (zirconia, alumina, metal, glass ceramic, tooth structure, composites), desensitizing/sealing of tooth structure, protective varnishing for glass ionomer fillings, and enamel priming for orthodontic use.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
K112118
NOV 1 7 2011
l
Image /page/0/Picture/2 description: The image shows the logo for BISCO. The logo consists of the letters "B," "I," "S," "C," and "O" inside of hexagons. Below the logo is the text "Bringing Science to the Art of Dentistry."
## .510 (k) SUMMARY
| Applicant: | Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193 | | | | | | | | | | | | | | |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-----------|--------------|--------------------------------------------------|---------------|-----|----------------------|---------------------------|--------------------|-----------------|-----------------|----|-----------------------|--------|
| Contact Person: | Michelle Schiltz-Taing Tel: 847-534-6146 Fax: 847-534-6146 | | | | | | | | | | | | | | |
| Date Prepared: | 08 November 2011 | | | | | | | | | | | | | | |
| Device Name: | <table><tr><td>Trade Name:</td><td>DREAMBOND</td></tr><tr><td>Common Name:</td><td>Adhesive for Both Self and Total Etch Techniques</td></tr><tr><td>Product Code:</td><td>KLE</td></tr><tr><td>Classification Name:</td><td>Resin tooth bonding Agent</td></tr><tr><td>Regulation Number:</td><td>21 CFR 872.3200</td></tr><tr><td>Classification:</td><td>II</td></tr><tr><td>Classification Panel:</td><td>Dental</td></tr></table> | Trade Name: | DREAMBOND | Common Name: | Adhesive for Both Self and Total Etch Techniques | Product Code: | KLE | Classification Name: | Resin tooth bonding Agent | Regulation Number: | 21 CFR 872.3200 | Classification: | II | Classification Panel: | Dental |
| Trade Name: | DREAMBOND | | | | | | | | | | | | | | |
| Common Name: | Adhesive for Both Self and Total Etch Techniques | | | | | | | | | | | | | | |
| Product Code: | KLE | | | | | | | | | | | | | | |
| Classification Name: | Resin tooth bonding Agent | | | | | | | | | | | | | | |
| Regulation Number: | 21 CFR 872.3200 | | | | | | | | | | | | | | |
| Classification: | II | | | | | | | | | | | | | | |
| Classification Panel: | Dental | | | | | | | | | | | | | | |
### Predicate Device:
DREAMBOND is substantially equivalent to Ace Bond SE from Bisco, Inc. Schaumburg IL (K063780) and to Adhesive EXL-759 from 3M ESPE AG Seefeld, Bavaria (K110302).
#### Indications for Use:
The indications of use of DREAMBOND are:
- 1. all direct restorations
- all indirect restorations 2.
- intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, metals, 3. glass ceramics, tooth structure and composites)
- 4. desensitizing/sealing of tooth structure
- 5. protective varnish for glass ionomer fillings
- priming of enamel for orthodontic use 6.
### Description of Applicant Device:
Bisco's DREAMBOND is a single-component dental bonding agent that combines etching, priming and bonding all in one bottle. DREAMBOND is an ethanol/water-based dental adhesive which bonds to dentin and to cut and uncut enamel. As a universal adhesive, DREAMBOND has been designed to work with light-cured, self-cured and dual-cured composite materials. DREAMBOND can be used for the bonding of both direct and indirect restorations. DREAMBOND may be used with or without phosphoric acid etchant. .
An ISO 13485 Certified Company
BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for BISCO, a dental supply company. The logo consists of the letters "B I S C O" arranged in a hexagonal pattern. Below the logo is the text "Bringing Science to the Art of Dentistry."
## Technological Characteristics
The chemical components of DREAMBOND are based upon industry standard monomer chemistry and are found in legally marketed predicate device ACE Bond SE (K063780). A comparison of the chemical composition of DREAMBOND to the predicate device is provided.
| Chemical Composition | ACE Bond SE<br>(K063780) | DREAMBOND |
|----------------------------------------------|--------------------------|-----------|
| Light-cured | X | X |
| Unfilled, multifunctional methacrylate resin | X | X |
| Ethanol and water based adhesive | X | X |
| Two component adhesive | X | |
| Single component adhesive | | X |
### Performance Data:
They physical/mechanical properties of DREAMBOND were tested in the lab using ISO, ADA/ANSI, and Bisco Inc. QA/QC and R&D test protocols. The information provided in this 510(k) of DREAMBOND compared to ACE Bond SE demonstrates that DREAMBOND performs as well as or better than the predicate. A comparison of the physical and mechanical properties to the predicate device is included.
| Physical / Mechanical Property<br>Comparison | ACE Bond SE<br>(K063780) | DREAMBOND |
|----------------------------------------------|--------------------------|-----------|
| Low viscosity | X | X |
| Self-etching dental adhesive | X | X |
| Total-etch dental adhesive | | X |
### Biocompatibility:
A biocompatibility evaluation was conducted to determine the safety of DREAMBOND using Oral Toxicity in Mice. No mortality or toxicity was observed for the tested time period. The conclusion of the safety evaluation and subsequent oral toxicity testing is that DREAMBOND is safe for its intended use.
### Conclusion:
The information provided in this 510(k) submission demonstrates that DREAMBOND is substantially equivalent to the predicate devices Ace Bond SE (K063780) and to Adhesive EXL-759 (K110302) in terms of intended use, indications for use, chemical composition and physical properties.
It is concluded that the information supplied in this submission has proven the safety and efficacy of this product.
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
800-247-3368 or 847-534-6000 Fax: 847-891-5049 www.bisco.com
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Michelle Schiltz-Taing Regulatory Affairs Coordinator BISCO, Incorporated 1100 West Irving Park Road Schaumburg, Illinois 60193
NOV 1 7 2011
Re: K112118
Trade/Device Name: DREAMBOND Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: November 8, 2011 Received: November 9, 2011
Dear Ms. Schiltz-Taing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
{4}------------------------------------------------
# Page 2- Ms. Schiltz-Taing
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510 (k) Number (if known): ___________________________________________________________________________________________________________________________________________________
Device Name: __DREAMBOND
Indications for Use:
The indications of use of DREAMBOND are:
- 1. all direct restorations
- 2. all indirect restorations
- 3. intra-oral repairs (i.e. repair of any fixed dental prosthesis containing zirconia, alumina, metals, glass ceramics, tooth structure, and composites)
- desensitizing/sealing of tooth structure র্ব
- protective varnish for glass ionomer fillings 5.
- priming of enamel for orthodontic use 6.
Prescription Use √ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Ruse
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K112118
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.