K033938 · Kuraray Medical, Inc. · KLE · Jan 29, 2004 · Dental
Device Facts
Record ID
K033938
Device Name
CLEARFIL PROTECT BOND
Applicant
Kuraray Medical, Inc.
Product Code
KLE · Dental
Decision Date
Jan 29, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CLEARFIL PROTECT BOND device is indicated for the following applications: 1) direct restorations using light-cured composite resin or compomer 2) cavity sealing as a pretreatment for indirect restorations 3) treatment of hypersensitive and/or exposed root surfaces 4) intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin 5) surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin 6) core build-ups using light- or dual-cured composite resin 7) cavity sealing under amalgam restorations ** For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
Device Story
CLEARFIL PROTECT BOND is a resin-based dental adhesive system containing the antibacterial monomer 12-methacryloyloxydodecylpyridinium bromide (MDPB). Used by dental professionals in clinical settings for restorative procedures, the adhesive is applied to tooth surfaces or prosthetic materials. The device functions as a bonding agent while providing an antibacterial cavity cleansing effect. It facilitates adhesion for direct/indirect restorations, core build-ups, and repairs, and treats hypersensitivity. The inclusion of MDPB provides the antibacterial property, which benefits the patient by reducing bacterial presence in the cavity preparation. The device is applied directly by the clinician during standard dental procedures.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing demonstrating biocompatibility and efficacy of the MDPB ingredient as an antibacterial cavity cleanser compared to control groups of dental resin material without MDPB.
Technological Characteristics
Resin-based dental adhesive system containing 12-methacryloyloxydodecylpyridinium bromide (MDPB) as an antibacterial agent. The device is a chemical bonding agent used for dental restorative applications. No changes to composition from predicate K023842.
Indications for Use
Indicated for direct restorations (composite/compomer), cavity sealing for indirect restorations, treatment of hypersensitive/exposed root surfaces, intraoral repair of fractured crowns/bridges, surface treatment of prosthetic appliances, core build-ups, and cavity sealing under amalgam. Provides antibacterial cavity cleansing effect for most applications.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
K033938
JAN 2 9 2004
## 510(k) SUMMARY Kuraray Medical Inc. CLEARFIL PROTECT BOND
### Name of Device
Trade or Proprietary Name: Common Name: Classification Name: Product Code:
CLEARFIL PROTECT BOND resin-based dental adhesive system resin tooth bonding agent 76KLE
#### Preparation Date
December 11, 2003
#### 510(k) Sponsor
Kuraray Medical Inc. 1621 Sakazu, Kurashiki, Okayama 710-8622 Japan
## 510(k) Sponsor Contact
Mr. Koji Nishida Kuraray America, Inc. 101 East 52"d Street 26th Floor New York, NY 10022 Phone: 800-879-1676 Fax: 888-700-5200
#### Intended Use
CLEARFIL PROTECT BOND is a resin-based dental adhesive system. The device includes the antibacterial ingredient, 12-methacryloyloxydodecylpyridinium bromide (MDPB), as fully disclosed in the original filing for the device K#023842 (under the name CLEARFIL SE BOND PLUS). No changes are being made to the composition of the device. The only purpose of this 510(k) filing is to add the claim that the device also has an "antibacterial cavity cleansing effect."
{1}------------------------------------------------
## Technological Characteristics and Substantial Equivalence
CLEARFIL PROTECT BOND is a resin-based dental adhesive system. The device includes the new antibacterial ingredient 12-methacryloyloxydodecylpyridinium bromide (MDPB), as fully disclosed in the original filing for the device K#023842 (under the name CLEARFIL SE BOND PLUS). No changes are being made to the composition of the device. The only purpose of this 510(k) filing is to add the claim that the device also has an "antibacterial cavity cleansing effect."
Substantial safety and efficacy testing has been conducted on the device, and specifically of the ingredient MDPB. Testing shows MDPB to be biocompatible, and also an effective antibacterial cavity cleanser compared to control groups of dental resin material without MDPB.
CLEARFIL PROTECT BOND is substantially equivalent for purposes of FDA market authorization to Pulpdent Cavity Cleanser (K#974202), Bisco Cavity Cleanser (K#915668), Den-Mat Cavity Cleanser (K#832497), and Ultradent Consepsis Cavity Varnish (K#925375 and K#924982). Although the CLEARFIL PROTECT BOND device has a different antibacterial ingredient (MDPB), the biocompatibility studies submitted with K#023842 demonstrate the safety of MDPB, and the efficacy studies submitted herewith demonstrate its effectiveness.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be three overlapping lines or strokes, possibly representing a human form or abstract design.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JAN 2 9 2004
Kuraray Medical, Incorporated C/O Mr. Keith A. Barritt, Esq. Attorney Fish & Richardson P.C. 1425 K Street, N.W. Washington, District of Columbia 20005
Re: K033938
Trade/Device Name: CLEARFIL PROTECT BOND Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Codes: KLE and LBH Dated: December 18, 2003 Received: December 22, 2003
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becasin a reasy for any for any and and and if the referenced above and have determined are to tegally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Interstate comments, or to that been reclassified in accordance with the for wreating Amendinens, of to devices that hart e est (Act) that do not require approval of a premarket the Federal Food, Drug, and Ocombile : 100 crefore, market the device, subject to the general approval application (1 Mr.). - The general controls provisions of the Act include controls provisions of the real - 110 gisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III if your device is classified (see above) into exactives. Existing major regulations affecting (PMA), it may be subject to such additions, Title 21, Parts 800 to 898. In your device can be found in the ecas nouncements concerning your device in the Ecderal Register.
{3}------------------------------------------------
Page 2 - Mr. Keith A. Barritt, Esq.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Tim S. Liao, Ph.D.
Chiu S. Lin, PhD Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
K033939 510(k) Number (if known): Device Name: CLEARFIL PROTECT BOND Indications For Use:
> The CLEARFIL PROTECT BOND device is indicated for the following applications:
l ) direct restorations using light-cured composite resin or compomer
2) cavity sealing as a pretreatment for indirect restorations
3) treatment of hypersensitive and/or exposed root surfaces
4) intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics, or composite resin using light-cured composite resin
5) surface treatment of prosthetic appliances made of porcelain, hybrid ceramics, or cured composite resin
6) core build-ups using light- or dual-cured composite resin
7) cavity sealing under amalgam restorations
** For indications 1, 2, 3, 6, and 7 the device exhibits an antibacterial cavity cleansing effect.
(DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use V
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
Susan Dunne
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number: k003938
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.