K023842 · Kuraray Medical, Inc. · KLE · Feb 11, 2003 · Dental
Device Facts
Record ID
K023842
Device Name
CLEARFIL SE BOND PLUS
Applicant
Kuraray Medical, Inc.
Product Code
KLE · Dental
Decision Date
Feb 11, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
1) Direct restorations using light-cured composite resin or compomer 2) Cavity sealing as a pretreatment for indirect restorations 3) Treatment of hypersensitive and/or exposed root surfaces 4) Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics or composite resin using light-cured composite resin 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics and cured composite resin 6) Core build-ups using light- or dual-cured composite resin 7) Cavity sealing under amalgam restorations
Device Story
CLEARFIL SE BOND PLUS is a resin-based dental adhesive system composed of dimethacrylate monomers. It is applied by a clinician in a dental office to prepared tooth cavities or prosthetic surfaces. The device functions as a bonding agent to improve retention of restorative materials (composite resin, compomer, or amalgam). It is used for direct restorations, cavity sealing, treatment of root hypersensitivity, and intraoral repair of fractured crowns or bridges. The system includes a primer and a bond component. By facilitating adhesion between the tooth structure or prosthetic material and the restorative material, it aids in structural restoration and sealing, potentially reducing sensitivity and improving the longevity of dental repairs.
Clinical Evidence
Bench testing only. No clinical data provided. Evidence consists of comparative laboratory evaluations of bond strength (to bovine tooth, composite resin, and amalgam), marginal sealing performance, and fluoride ion release/mechanical stability compared to predicate devices.
Technological Characteristics
Resin-based dental adhesive system containing dimethacrylate monomers and MDPB. Components include primer and bond. Biocompatibility tested per ISO 7405:1997 and ISO 10993-1:1997. Fluoride-releasing properties evaluated. No electronic or software components.
Indications for Use
Indicated for dental patients requiring direct composite/compomer restorations, cavity sealing for indirect restorations, treatment of hypersensitive/exposed root surfaces, intraoral repair of fractured dental prosthetics (porcelain, hybrid ceramic, composite), surface treatment of prosthetic appliances, core build-ups, and cavity sealing under amalgam.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
K023842
Image /page/0/Picture/2 description: The image is a black and white logo for Kuraray. The logo features a stylized letter "K" design. The word "KURARAY" is printed in a bold, sans-serif font below the symbol.
].
# KURARAY MEDICAL INC.
Dental Material Department 3-1-6 Nihonbashi, Chuo-ku, Tokyo 103-8254, Japan Tel : +81.3.3277.6949 Fax : +81.3.3277.6956
# 510(k) SUMMARY
| Submitter | |
|-----------------------------|------------------------------------------------------|
| 1) Name | KURARAY MEDICAL INC. |
| 2) Address | 1621 Sakazu, Kurashiki, Okayama 710-8622, Japan |
| 3) Contact person | Koji Nishida |
| | Dental Material Department |
| 4) Date | November 11, 2002 |
| 5) Contact person in U.S.A. | Masaya Sasaki |
| | Kuraray America, Inc. |
| | 101 East 52nd Street, 26th Floor, New York, NY 10022 |
| | Telephone: (212)-986-2230 (Ext.115) |
| | Facsimile: (212)-867-3543 |
### : !. Name of Device
| 1) Proprietary Name | CLEARFIL SE BOND PLUS |
|------------------------|--------------------------------------------|
| 2) Classification Name | Resin tooth bonding agent (21CFR 872.3200) |
| 3) Common/Usual Name | Resin-based dental adhesive system |
- :). Predicate device:
The predicate products are;
- a) Resin Tooth Bonding Agent
| 1. | CLEARFIL SE BOND | by Kuraray Medical Inc. | (K012442) |
|----|---------------------------------------------------------------|-------------------------------------|-----------|
| 2. | CLEARFIL LINER BOND 2V | by Kuraray Medical Inc. | (K012440) |
| 3. | TOKUYAMA ONE-UP-BOND F | by Tokuyama America Inc. | (K993917) |
| 4. | IMPERVA FLUORO BOND | by Shofu Dental Corp. | (K953612) |
| 5. | PRIME & BOND NT DUAL CURE<br>UNIVERSAL DENTAL ADHESIVE SYSTEM | by Dentsply Intl. | (K982394) |
| 6. | OPTIBOND SOLO PLUS 4 | by KERR DENTAL<br>MATERIALS CENTER. | (K014027) |
| b) | Other dental materials | | |
| 1. | PANAVIA F | by Kuraray Medical Inc. | (K012441) |
| 2. | SEAL & PROTECT PROTECTIVE SEALANT | by Dentsply Intl. | (K021805) |
| 3. | PANAVIA EX | by Kuraray Co., Ltd. | (K855211) |
- 4. ESTENIA
(K012707) by Kuraray Medical Inc.
- 1. Description for the premarket notification
CLEARFIL SE BOND PLUS is classified into the resin tooth bonding agent, CFR 21 Section 872.3200, because it is a device composed of materials such as dimethacrylate monomers intended to painted on the interior of a prepared cavity of a tooth to improve retention of restorative materials.
{1}------------------------------------------------
### 5 Staterient of the intended use
This device has the following indications for use which are substantially equivalent to the legally marketed predicate devices.
| Indications for use | Predicate device |
|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| 1) Direct restorations using light-cured composite resin or<br>compomer | CLEARFIL SE BOND<br>(K012442) |
| 2) Cavity sealing as a pretreatment for indirect restorations | |
| 3) Treatment of hypersensitive and/or exposed root surfaces | |
| 4) Intraoral repairs of fractured crowns/bridges made of<br>porcelain, hybrid ceramics or composite resin using<br>light-cured composite resin | |
| 5) Surface treatment of prosthetic appliances made of<br>porcelain, hybrid ceramics and cured composite resin | |
| 6) Core build-ups using light- or dual-cured composite resin | OPTIBOND SOLO PLUS 4<br>(K014027) |
| 7) Cavity sealing under amalgam restorations | PRIME & BOND NT DUAL<br>CURE UNIVERSAL DENTAL<br>ADHESIVE SYSTEM<br>(K982394) |
#### 3. Statement of the technological characteristics and safety
### 3-1 Design/Components
This device consists of the primer, the bond and the accessories, and is substantially equivalent in design to CLEARFIL SE BOND.
### 6-2 Material and chemical ingredients
The ingredients other than MDPB are used in the legally marketed predicate devices, C. EARFIL SE BOND etc. The biocompatibilities of the new ingredient, MDPB, and this device were evaluated according to ISO 7405: 1997 and ISO 10993-1: 1997. As to the results, this device is judged that it is substantially equivalent in safety to the legally marketed predicate devices.
#### 6-3 Mechanical properties
The bond strengths between bovine tooth and a composite resin and a fresh amalgam were evaluated in comparison with the legally marketed predicate devices; CLEARFIL SE BOND etc. The bond strengths between composite resin and various substances were evaluated when used as intra oral repairs of fractured crowns/bridges. The marginal sealing was also evaluated when used as direct filling and cavity sealing under amalgam restoration. As to the result, this device is substantially equivalent to the legally marketed predicate devices in effectiveness.
#### 6-4 Physical properties
The amount of fluorine ion from the cured material immersed in water was evaluated in comparison with the legally marketed predicate devices; TOKUYAMA ONE UP-BOND F IMPERVA FLUORO BOND and PANAVIA F. The changes in mechanical strength due to fluorine releasing were also evaluated. As to the result, this device is substantially equivalent in the fluoride releasing property including the change of mechanical strength to the legally
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three lines representing its head and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter of the circle.
FEB 11 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Kuraray Medical, Incorporated C/O Ms. Masaya Sasaki Kuraray America, Incorporated 101 East 52nd Street, 26th Floor New York, New York 10022
Re: K023842
Trade/Device Name: CLEARFIL SE BOND PLUS Regulation Number: 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: November 12, 2002 Received: November 18, 2002
Dear Ms. Sasaki:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Sasaki
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Penner
Susan Runner, DDS, MA Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
[CLEARFIL SE BOND PLUS, Kuraray Medical Inc.]
## 3842 162
# Indications for Use
('LEAFIL SE BOND PLUS is indicated for the following applications'
- 1) :Direct restorations using light-cured composite resin or compomer
- 2) Cavity sealing as a pretreatment for indirect restorations
- 3) 'Treatment of hypersensitive and/or exposed root surfaces
- 4) Intraoral repairs of fractured crowns/bridges made of porcelain, hybrid ceramics or composite resin using light cured composite resin
- 5) Surface treatment of prosthetic appliances made of porcelain, hybrid ceramics and cured composite resin
- 6) Core build-ups using light- or dual-cured composite resin
- 7) Cavity sealing under amalgam restorations
"PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Part: 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Kein Mulay for MSR
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.