K252785 · Rizhao Huge Biomaterials Company, Ltd. · EMA · Dec 12, 2025 · Dental
Device Facts
Record ID
K252785
Device Name
TopCEM Vigor SA Self-Adhesive Resin Cement
Applicant
Rizhao Huge Biomaterials Company, Ltd.
Product Code
EMA · Dental
Decision Date
Dec 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Cementation of all-ceramic, composite, and metal crowns, bridges, inlays, onlays, dental posts; 2-3-unite Maryland bridges and 3-unit inlay/onlay bridges (contraindicated for patients with bruxism or periodontitis). Cementation of fiber posts. Cementation of all-ceramic, composite, and metal restorations on implant abutments.
Device Story
TopCEM Vigor SA is a dual-curing (light/chemical) self-adhesive resin cement used by dentists for permanent cementation of dental restorations. Supplied in double-barrel syringes, the device consists of base and catalyst pastes containing methacrylate resins, MDP, and inorganic fillers. Upon mixing, polymerization occurs via photo and chemical initiators. The cement is radiopaque for radiographic identification. It is applied in a clinical setting to secure crowns, bridges, inlays, onlays, posts, and implant abutments. The material provides mechanical retention and bonding to tooth structure and restorative materials, benefiting patients by stabilizing dental prosthetics.
Clinical Evidence
No clinical data. Bench testing only. Physical properties (working/setting time, film thickness, flexural/compressive strength, bond strength, water sorption, solubility, radio-opacity) tested per ISO 4049, ISO 29022, and ISO/TS 11405. Biocompatibility testing performed per ISO 7405 and ISO 10993.
Technological Characteristics
Dual-cure resin cement; methacrylate resins, phosphorylated methacrylate (MDP), silane-treated glass/silica fillers. Particle size 0.01-5 microns. 40% inorganic filler volume. Radiopaque. ISO 4049 Type 2 Class 3. Delivered via double-barrel syringe. Polymerization via photo/chemical initiator system.
Indications for Use
Indicated for cementation of crowns, bridges, inlays, onlays, dental posts, Maryland bridges, and implant restorations using all-ceramic, composite, or metal materials. Contraindicated for patients with bruxism or periodontitis.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
K242366 — OverCEM SA Universal; OverCEM SA Translucent; OverCEM SA Opaque; OverCEM Ti-Abutment; OverCEM Cer-Abutment · Overfibers S.R.L. · Jan 10, 2025
K120378 — PANAVIA SA CEMENT HANDMIX VALUE PACK (UNIVERSAL (A2), WHITE) PANAVIA SA CEMENT HANDMIX (UNIVERSAL (A2), WHITE) PANAVIA S · Kuraray Medical, Inc. · Mar 30, 2012
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
Rizhao HuGe Biomaterials Company, Ltd.
Maggie Zheng
Regulatory Affairs Manager
No.2 North Zhaoyang Road, District of Donggang
Rizhao, 276800
CHINA
Re: K252785
Trade/Device Name: TopCEM Vigor SA Self-Adhesive Resin Cement
Regulation Number: 21 CFR 872.3275
Regulation Name: Dental Cement
Regulatory Class: Class II
Product Code: EMA
Dated: November 10, 2025
Received: November 12, 2025
Dear Maggie Zheng:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252785 - Maggie Zheng
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252785 - Maggie Zheng
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252785 | ? |
| Please provide the device trade name(s). | | ? |
| TopCEM Vigor SA Self-Adhesive Resin Cement | | |
| Please provide your Indications for Use below. | | ? |
| -Cementation of all-ceramic, composite, and metal crowns, bridges, inlays, onlays, dental posts; 2-3-unite Maryland bridges and 3-unit inlay/onlay bridges (contraindicated for patients with bruxism or periodontitis).
-Cementation of fiber posts.
-Cementation of all-ceramic, composite, and metal restorations on implant abutments. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
HUGE
K252785
# 510 (k) Summary
This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
1. Date Summary Prepared: December 11, 2025
2. Submitter Information:
Owner's Name: Rizhao HuGe Biomaterials Company, Ltd.
Address: No.2 North Zhaoyang Road, District of Donggang, Rizhao City, Shandong Province, China 276800
Telephone: 0086 633 2277268
Fax: 0086 633 2277298
Contact Person: Ms. Maggie Zheng
Contact Title: Regulatory Affairs Manager
E-mail: zhengxy@hugedent.com
3. Device Name
Trade name: TopCEM Vigor SA Self-Adhesive Resin Cement
Common name: Dental Cement
Classification name: Dental Cement (21 CFR 872.3275)
Regulatory Class: II
Product Code: EMA
4. Predicate Device Information
| Owner/Operator | Device Trade Name | 510 (k) No. | Product Code |
| --- | --- | --- | --- |
| 3M ESPE AG | Malta | K100756 | EMA |
No reference device were used in this submission.
5. Description of Device
TopCEM Vigor SA Self-Adhesive Resin Cement is a dual-curing (chemical and/or light cure) luting cement containing paste-paste of Base and Catalyst. The cement contains inorganic fillers, and the inorganic filler loading is about 40% by volume having particle size range of
18-1
{5}
HUGE
about 0.01 to 5 microns. The mixing ratio, based on volume, is 1 part base paste: 1 part catalyst paste.
TopCEM Vigor SA Self-Adhesive Resin Cement is delivered in double-barrel syringes.
TopCEM Vigor SA Self-Adhesive Resin Cement is radiopaque, allowing for easy identification on radiographs. It belongs to Type 2 Class 3 dual cure materials in accordance with ISO 4049. The curing mechanism of the predicate devices and subject device is polymerization of uncured methacrylate ester monomers. This reaction is caused by photo initiator and chemical polymerization initiator systems.
## 6. Indications for Use
- Cementation of all-ceramic, composite, and metal crowns, bridges, inlays, onlays, dental posts; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (contraindicated for patients with bruxism or periodontitis).
- Cementation of fiber posts.
- Cementation of all-ceramic, composite, and metal restorations on implant abutments.
## 7. Technological Characteristics Comparison
All components of the subject device are based upon industry well-known chemistry. The curing mechanism of the subject device and predicate device are all of polymerization of uncured methacrylate ester monomers. The reaction is caused by photo initiator and chemical polymerization initiator systems. The following table shows the significant technological characteristics for the subject device and indicates the following similarities and differences with the predicate device:
| Technological Characteristics Comparison Table | | |
| --- | --- | --- |
| Technological Characteristics | Subject device K252785 | Predicate device K100756 |
| Composition of Materials | Methacrylate resins, phosphorylated methacrylate (MDP), silane treated glass powder, silane treated silica, pigments, initiator and stabilizer | Methacrylate resins, phosphorylated methacrylate (MDP), silane treated glass powder, silane treated silica, pigments, initiator and stabilizer |
| Physical Form | Pastes of Catalyst and Base | Pastes of Catalyst and Base |
| Indications of Use | ● Cementation of all-ceramic, composite, and metal crowns, bridges, inlays, | ● Final cementing of all-ceramic, composite, or metal inlays, onlays, |
18-2
{6}
HUGE
| Technological Characteristics Comparison Table | | |
| --- | --- | --- |
| Technological Characteristics | Subject device K252785 | Predicate device K100756 |
| | onlays, dental posts; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (contraindicated for patients with bruxism or periodontitis). • Cementation of fiber posts. • Cementation of all-ceramic, composite, and metal restorations on implant abutments. | crowns and bridges; 2-3-unit Maryland bridges and 3-unit inlay/onlay bridges (excluded for patients with bruxism or periodontitis); • Final cementing of posts and screws; • Final cementation of all-ceramic, composite, or metal restorations on implant abutments. |
| Prescription/over-the-counter use | Prescription | Prescription |
| Curing method | Dual cure | Dual cure |
| Delivery form | Syringe | Syringe |
| Radiographic Appearance | Radiopaque | Radiopaque |
| Physical Properties | The subject device and the predicate device have substantially equivalent physical properties as they all conform to the ISO 4049, ISO 29022, ISO/TS 11405, and HUGE SOP. | |
All compositions of the subject device are based upon industry well-known chemistry. The technological characteristics of the subject device are very similar to those of the predicate device. The subject device is a similar product, manufactured with similar materials and used in the same way by the same types of users and patient populations. The subject device and predicate device have minor difference in Composition of Materials. However, the difference does not affect the intended use or substantial equivalence. Besides, other comparison items such as physical form, curing method, delivery form and physical properties, etc. are the same or very similar. And both of the subject device and the predicate device are supplied for prescription use.
# 8. Summary of Non-clinical testing
The physical properties of the subject device, including Working time, Setting time, Film thickness, Flexural strength, Water sorption, Solubility, Compressive strength, Elastic modulus, Surface hardness, Shear bond strength, Bonding durability and Radio-opacity, were determined and tested according to ISO 4049, ISO 29022, ISO/TS 11405, and HUGE SOP, and the test results demonstrated the qualification and substantial equivalence when compared to the predicate device.
{7}
HUGE
Additionally, the subject device is substantially equivalent to the predicate device that has been legally marketed already and with no clinical adverse events. The compositions of the subject device do not contain any non-conventional chemicals compared to the legally marketed predicate device. Biocompatibility tests were performed fully following the ISO 7405 and ISO 10993 standards, and test results are sufficient to prove the safety of the subject device.
## 9. Summary of Clinical testing
Clinical test is not applicable.
## 10. Conclusions
Based on the indications for use, technological characteristics, performance testing and comparison to the predicate device, the subject device has been shown to be safe and effective for its intended use and the minor differences in composition of the predicate device affecting neither the general intended use nor substantial equivalence. Rizhao HuGe Biomaterials Company, Ltd. concludes that the subject device is substantially equivalent to the predicate device described herein.
18-4
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.