DENTAVIT, MONARCH, ACROTONE, SENATOR, DENTORIUM

K122955 · Wright Health Group , Ltd. · ELM · Mar 1, 2013 · Dental

Device Facts

Record IDK122955
Device NameDENTAVIT, MONARCH, ACROTONE, SENATOR, DENTORIUM
ApplicantWright Health Group , Ltd.
Product CodeELM · Dental
Decision DateMar 1, 2013
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3590
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use for these plastic teeth is to make partial or full dentures. These dentures are made to fit the patient's mouth as a removable device, not implanted. These plastic tech may also be attached to dental implants. Federal law restricts this device to sale by or on the order of a physician. It is intended to be used only by appropriately trained and qualified healthcare professionals and servicing staff in clinical environments.

Device Story

Synthetic plastic teeth (models: Dentavit, Monarch, Acrotone, Senator, Dentorium) used for denture fabrication. Produced via polymerization of polymethylmethacrylate (PMMA) with methylmethacrylate monomer, cross-linking agents, and pigments under heat and pressure. Finished teeth are checked for flash and polished. Used by trained dental professionals in clinical environments to construct removable dentures or implant-supported prosthetics. Provides patients with functional and aesthetic tooth replacement. No software or automated processing involved.

Clinical Evidence

Bench testing only. Compliance with ADA specification No. 15:2008 (Artificial Teeth), BS EN ISO 22112:2006, and biocompatibility standards ISO 10993-1 and ISO 7405:2008.

Technological Characteristics

Materials: Polymethylmethacrylate (PMMA), methylmethacrylate monomer, cross-linking agents, non-toxic pigments. Standards: ADA specification No. 15:2008, BS EN ISO 22112:2006, ISO 10993-1, ISO 7405:2008. Manufacturing: Heat and pressure polymerization. Form factor: Preformed plastic teeth for denture integration.

Indications for Use

Indicated for use as components in the fabrication of removable partial or full dentures, or for attachment to dental implants, for patients requiring prosthetic tooth replacement.

Regulatory Classification

Identification

A preformed plastic denture tooth is a prefabricated device, composed of materials such as methyl methacrylate, that is intended for use as a tooth in a denture.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(K) SUMMARY August 10, 2012 ### SUBMITTER INFORMATION: (3003103749) Wright Health Group Ltd. Dunsinane Industrial Estate Kingsway West Dundee, Scotland DD2 3QD UK ## FDA AGENT/CORRESPONDING OFFICIAL INFORMATION: North American Technical Services (NATS) Corp. 30 Northport Rd Sound Beach, NY 11789 Tel: 631-744-0059 Fax: 631-744-0192 Email: natscorp@aol.com Contact: Stephen T. Micoch ### DEVICE NAME: Name: Denture Preformed Plastic Teeth Proprietary Models: Dentavit, Monarch, Acrotone, Senator, Dentorium Classification: II Product Code: ELM Regulation: 21 CFR 872.3590 ### PREDICATE DEVICES: Union Dental S.A./ Unidesa Odi K070591 - model Replica Dental Vipi Ltd K022300 - model Vipi Dent Plus Vident K914476 - model Vitapan Huge Dental Material Co., Ltd. K101029 -- model Kaili ### DESCRIPTION: The synthetic plastic teeth are used in the production of dentures. They are produced from a I he synuletic process. Polymethylmethacrylate or PMMA is mixed with methylmethacrylate monomer, a cross linking agent and non-toxic pigments under heat and pressure. Once the monomer, a cross iniking agoit and non teen are checked for flash and polished according to the relevant model characteristics. # SUBSTANCIAL EQUIVALENCE AND TECHNOLOGY: The plastic teeth have comparable chemical composition as the predicate devices. The models I he piastic tecur have comparable chesage to the predicate devices. Also, preamendment are similal in size, shape, color, and asage to the productive. The added models use the same materials and controls established for over 20 years. MAR 0 1 2013 {1}------------------------------------------------ The intended use for these plastic teeth is to make partial or full dentures. These dentures are made to fit the patient's mouth as a removable device, not implanted. These plastic tech may also be attached to dental implants. Federal law restricts this device to sale by or on the order of a physician. It is intended to be used only by appropriately trained and qualified healthcare professionals and servicing staff in clinical environments. # SUMMARY OF NONCLINICAL TESTS AND DESIGN CONTROL ACTIVITIES: The performance and safety testing activities were conducted on acceptable dental materials to establish the reliability characteristics of these teeth devices. Validation and verification of compliance with the following mandatory and voluntary standards has been made - CE Medical Directive 93/42/EEC 2007/47/EC t - ISO9001/GMP/ISO13485: Quality System Certification . Factory control procedures are established for production QA and QC with use of ADA specification No. 15:2008 Artificial Teeth, and BS EN ISO22112:2006, with Biocompatibility Standards ISO10993-1, ISO7405:2008 This control activity shows that there are no new questions of safety and effectiveness for the plastic teeth models made by Wright. ### CONCLUSION: For and and the transity The plastic teeth are substantially equivalent to the predicate teeth and have the same intended use. They are suitable for equivalent use on dentures according to the specifications stated based on the same materials, production process and technology. Wright has been a manufacturer for about 100 years and has been in the market for over 40 years and has met preamendment rules. There are no significant health incidents in that time. These models have the same intended use and similar indications, technical characteristics and function as the predicate devices. Suitable dental industry materials have been used by Wight and predicate companies that solve biocompatibility, durability and esthetic concerns. The minor differences raise no new issues of safety or effectiveness. These models are substantially equivalent to the predicate devices. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines representing the staff and two snakes intertwined around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 1, 2013 Wright Health Group, Limited C/O Mr. Stephen Mlcoch Chief Executive Officer North American Technical Services (NATS) Corporation 30 Northport Road Sound Beach NY 11764 Re: K122955 Trade/Device Name: Plastic Teeth Models Dentavit, Monarch, Acrotone, Senator, and Dentorium Regulation Number: 21 CFR 872.3590 Regulation Name: Prosthetic Devices Regulatory Class: II Product Code: ELM Dated: January 15, 2013 Received: February 13, 2013 Dear Mr. Mlcoch: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Mlcoch Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. Kwame O. Ulmer for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K122955 #### INDICATIONS FOR USE 510(K) Number (if known): Device Name: Plastic teeth models Dentavit, Monarch, Acrotone, Senator, and Dentorium. Indications for Use: These dental plastic teeth are parts used for denture fabrication. PMMA is mixed with monomer in a polymenzation process. The plastic teeth are part of dentures made to fit a patient's mouth which are removable or may be attached to dental implant devices. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mary S. Runner -S 2013.02.27 15:20:12 -05'00' (Division Sign-Off) . Division of Anesthesiology, General Hospital ntection Control, Dental Devices 510(k) Number: K1208955 Page 1 of 1
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...