KAIFENG/CRYSTAL, KAILI, KAIJING, KAIYUE, MAIST AND KAITONG/ROYAL SYNTHETIC POLYMER TEETH
K101029 · Huge Dental Material Co., Ltd. · ELM · Aug 16, 2010 · Dental
Device Facts
Record ID
K101029
Device Name
KAIFENG/CRYSTAL, KAILI, KAIJING, KAIYUE, MAIST AND KAITONG/ROYAL SYNTHETIC POLYMER TEETH
Applicant
Huge Dental Material Co., Ltd.
Product Code
ELM · Dental
Decision Date
Aug 16, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3590
Device Class
Class 2
Indications for Use
It is intended for use as a tooth in a denture. These teeth are used for making full or partial dentures. The artificial resin teeth are mounted on an appliance made of dental acrylic resin. The appliance is made to fit the patient's mouth and is a removable. It is not implanted.
Device Story
Synthetic polymer teeth (Kaifeng/Crystal, Kaili, Kaijing, Kaiyue, MAIST, Kaitong/Royal) used for fabrication of full or partial removable dentures. Produced via dough molding process using polymethylmethacrylate (PMMA), methylmethacrylate monomer, and crosslinking agent. Post-polymerization, teeth are trimmed of burrs and polished. Intended for mounting on dental acrylic resin appliances by dental professionals. Device is non-implanted; provides functional and aesthetic tooth replacement for patients.
Clinical Evidence
No clinical data. Performance and safety established via bench testing and biocompatibility studies.
Indicated for use as artificial teeth in full or partial removable dentures for patients requiring tooth replacement.
Regulatory Classification
Identification
A preformed plastic denture tooth is a prefabricated device, composed of materials such as methyl methacrylate, that is intended for use as a tooth in a denture.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
DENTAL VIPI LTDA ACRY PAN, VIPI DENT PLUS, BIOLUX, BIOLUX V, NEW DENT, DENTOLUXX, VIPI DENT N.H., VIPI DENT V PERFORMED PLASTIC DENTURE TEETH (K022300)
Submission Summary (Full Text)
{0}------------------------------------------------
# Summary Prepared Date: 03/26/2010
# Submission Sponsor:
:
AUG 1 6 2010
HUGE DENTAL MATERIAL CO., LTD. No.5069 Chuannanfeng Rd, Zhuqiao, Nanhui Shanghai, China 201323 Phone: +86-21-61629296 Fax: +86-21-61629298 www.hugedental.com
#### Submission Correspondent:
Mr. Leon Lu Director of Regulatory Affairs
MEDevice Services, LLC 3500 South Dupont Highway Dover, Delaware 19901 USA Phone: 1-877-202-1588 Fax: 1-888-202-8884 Email: info@medeviceservices.com www.medeviceservices.com
Trade/Device Name: SYNTHETIC POLYMER TEETH (Kaifeng/Crystal, Kaili, Kaijing, Kaiyue, MAIST and Kaitong/Royal)
Common or Usual Name: Preformed Plastic Denture Teeth Device Class: II Classification Name: Preformed Plastic Denture Tooth Regulation Number: 21 CFR 872.3590 Product Code: ELM Review Panel: Dental
#### Predicate Device:
- K061337, XPDENT CORPORATION ● WIEDENT ESTHETIC TEETH
- . K070591, UNION DENTAL S.A./UNIDESA.ODI REPLICA, ORTOLUX TOP, ODIPAL, ODILUX, ODIDENT, VITACRILIC, NATURE'S BEST, DENTAL ACRYLIC TEETH
- K022300, DENTAL VIPI LTDA . ACRY PAN, VIPI DENT PLUS, BIOLUX, BIOLUX V, NEW DENT, DENTOLUXX, VIPI DENT N.H., VIPI DENT V PERFORMED PLASTIC DENTURE TEETH
{1}------------------------------------------------
The device is similar in size, shape, color, chemical composition and usage as the above products.
#### Device Description:
- D
ﺎ
The Synthetic Polymer Teeth (Kaifeng/Crystal, Kaili, Kaijing, Kaiyue, MAIST and Kaitong/Royal) used in the production of dentures are produced through a dough molding process. Polymethylmethacrylate (PMMA) is mixed with methylmethacrylate monomer and a crosslinking agent.
Once the polymerization process has been carried out, the teeth are obtained and the next step is to remove any burrs and polish them.
#### Intended Use:
It is intended for use as a tooth in a denture.
These teeth are used for making full or partial dentures. The artificial resin teeth are mounted on an appliance made of dental acrylic resin. The appliance is made to fit the patient's mouth and is a removable. It is not implanted.
#### Comparison to Predicate Devices:
The device has comparable chemical composition as the predicate devices.
The device is similar in size, shape, color and usage as the predicate devices.
#### Discussion of Non-Clinical Tests Performed:
Performance and safety testing activities were conducted against recognized standards to establish the reliability characteristics of the new devices. Testing involved bench studies, and biocompatibility tests.
### Discussion of Clinical Tests Performed:
None
#### Conclusion:
The proposed device is as safe and effective as the predicate devices. The proposed device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the proposed device and its predicate devices raise no new issues of safety or effectiveness. Thus, the Synthetic Polymer Teeth (Kaifeng/Crystal, Kaili, Kaijing, Kaiyue, MAIST and Kaitong/Royal) is substantially equivalent to its predicate devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Huge Dental Material Company, Limited C/O Mr. Jing Li MEDevice Services, LLC 3500 South Dupont Highway Dover, Delaware 19901
AUG 1 6 2010
Re: K101029
Trade/Device Name: SYNTHETIC POLYMER TEETH (Kaifeng/ Crystal, Kaili, Kaijing, Kaiyue, MAIST and Kaitong/ Royal) Regulation Number: 21 CFR 872.3590 Regulation Name: Preformed Plastic Denture Tooth Regulatory Class: II Product Code: ELM Dated: July 12, 2010 Received: July 22, 2010
Dear Mr. Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2- Mr. Li
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
## 510(k) Number (if known):
Device Name: SYNTHETIC POLYMER TEETH (Kaifeng/Crystal, Kaili, Kaijing, Kaiyue, MAIST and Kaitong/Royal)
#### Indications for Use:
It is intended for use as a tooth in a denture.
These teeth are used for making full or partial dentures. The artificial resin teeth are mounted on an appliance made of dental acrylic resin. The appliance is made to fit the patient's mouth and is a removable. It is not implanted.
Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Kei Mulyani MSR
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K101029
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.