GC Temp Print

K193113 · GC America, Inc. · EBG · Jul 9, 2020 · Dental

Device Facts

Record IDK193113
Device NameGC Temp Print
ApplicantGC America, Inc.
Product CodeEBG · Dental
Decision DateJul 9, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3770
Device ClassClass 2

Intended Use

1. Fabrication of temporary crowns, bridges, inlays, onlays and veneers 2. Fabrication of long-term temporary restorations

Device Story

GC Temp PRINT is a 3D printable light-curing composite resin for temporary dental restorations. Input consists of digital design data processed by compatible DLP/SLA 3D printers. The device is a liquid paste that undergoes photopolymerization during the printing process and subsequent post-curing in a light-curing unit. Used in dental clinics/laboratories by dental professionals to create temporary crowns, bridges, inlays, onlays, and veneers. Output is a hardened dental restoration bonded to tooth structure using adhesive temporary cement. Benefits include the ability to fabricate long-term temporary restorations via additive manufacturing. Healthcare providers use the output to provide patients with functional temporary dental coverage during treatment.

Clinical Evidence

No clinical testing performed. Substantial equivalence supported by bench testing including depth of cure, flexural strength, water sorption, and solubility, all meeting ISO 10477:2018 and company specifications. Biocompatibility assessment per ISO 10993-1:2009 confirmed no cytotoxic effect.

Technological Characteristics

Light-cured composite resin containing Urethane dimethacrylate (UDMA), dimethacrylate, silicon dioxide, initiator, and pigment. Fabricated via DLP/SLA 3D printing (additive manufacturing) followed by post-curing in a light-curing unit. Complies with ISO 10477:2018 and company specification 1AB-1500-3-10696(3).

Indications for Use

Indicated for the fabrication of temporary dental restorations including crowns, bridges, inlays, onlays, and veneers, including long-term temporary restorations.

Regulatory Classification

Identification

A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it. July 9, 2020 GC America Inc. Mark Heiss Director, Regulatory & Academic Affairs 3737 W. 127th Street Alsip, Illinois 60803 Re: K193113 Trade/Device Name: GC Temp PRINT Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown And Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: June 11, 2020 Received: June 12, 2020 Dear Mark Heiss: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910 0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K193113 Device Name GC Temp PRINT Indications for Use (Describe) 1. Fabrication of temporary crowns, bridges, inlays, onlays and veneers 2. Fabrication of long-term temporary restorations Type of Use (Select one or both, as applicable) Z Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection. including suggestions for reducing this burden. to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond fo, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text 'GC' in a stylized, teal-colored font. The text is surrounded by single quotation marks on both sides. The font appears to be bold and slightly italicized, giving it a dynamic and modern look. The background is plain white, which makes the teal text stand out. Image /page/3/Picture/1 description: The image shows a stylized number 7 with a pattern of small, teal rectangles to the right of the number. The number 7 is drawn with a thin, light blue-gray line. The rectangles are arranged in a scattered, somewhat diagonal pattern, suggesting a design element rather than a representation of a specific object. ## 1 Section 5 - 510(k) Summary - 1. Submitter Information: GC America Inc. 3737 W. 127th Street Alsip, IL 60803 | Contact Person: | Mark Heiss, D.D.S. | |------------------------|----------------------------------| | Phone: | (708) 926-3090 | | Alternate Contact: | Lori Rietman | | Phone: | (708) 926-3092 | | Fax: | (708) 925-0373 | | Date Prepared: | July 9, 2020 | | Device Name: | | | Proprietary Name: | GC Temp PRINT | | Classification Name: | Temporary crown and bridge resin | | Device Classification: | Class II, 872.3770 | | Product Code: | EBG | | 510(k) Number: | K193113 | - 510(k) Number: 3. Predicate Devices: 2. | Product | Applicant | 510(k) No. | Code No | Predicate | Decision Date | |------------------------------------------|------------------|------------|---------|----------------------|---------------| | Resin for<br>Temporary<br>Crown & Bridge | Dentis Co., Ltd. | K180657 | EBG | Primary<br>Predicate | 12/07/2018 | | TEMPSMART<br>(MSN-006) | GC America Inc. | K141562 | EBG | Reference<br>Device | 09/05/2014 | #### Description of Device: 4. - న్. GC Temp PRINT is 3D printable light curing composite for temporary crown and bridge. The device consists of a paste delivered in a bottle. GC Temp PRINT is fabricated with certain Digital Light Processing/Stereolithography Apparatus (DLP/SLA)a` 3D printers and post cured with light curing device into dental restorations, which are then bonded to tooth structure with adhesive temporary cement. GC Temp PRINT Package: Bottle (500 g) -QTY: 1 Shades available: Light, Medium and Bleach Shelf Life and Storage Conditions: Shelf Life 2 years Recommended for optimal performance, store in a cool and dark place (4-25°C / 39.2-77.0°F) away from high temperatures or direct sunlight. {4}------------------------------------------------ #### 6. Performance Bench Tests It is confirmed that the device conforms to the required specifications and is suitable for its intended use. Performance testing includes: - . Depth of cure - Flexural strength . - . Water sorption - o Solubility The applicant device complies with some requirements of ISO 10477 and Company specification (1AB-1500-3-10696(3)). (see table below). | | Property | Standards | Requirements in ISO 10477:<br>2018 and/or Company<br>specification<br>(1AB-1500-3-10696(3)) | |---|-------------------|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | 1 | Depth of cure | Company specification (1AB-1500-3-10696(3)) | 0.098 – 0.130 mm | | 2 | Flexural strength | ISO 10477: 2018<br>5.4 Flexural strength (Fabricated by 3D Printer) | > 50 MPa | | 3 | Water sorption | ISO 10477: 2018<br>5.6 Water sorption (Fabricated by 3D Printer) | < 40 µg/mm3 | | 4 | Solubility | ISO 10477: 2018<br>5.7 solubility (Fabricated by 3D Printer) | < 7.5 µg/mm3 | #### 7. Non-Clinical Performance Testing A biocompatibility assessment was completed according to ISO 10993-1:2009, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process. Based on the criteria of the protocol and the ISO 10993-5 guidelines, the test article meets the requirements of the test and is not considered to have a cytotoxic effect. #### 8. Clinical Performance Testing No clinical testing has been performed on this device. #### 9. Substantial equivalence Both applicant and primary predicate device are categorized in same Code No. EBG and both curing mechanism are polymerization of uncured methacrylate oligomers. The products are a liquid photocurable material that is polymerized by UV laser. Though primary predicate device reacts with UV light 405nm, the applicant device can be polymerized by UV light which is 405nm or shorter wavelength. After printing, both applicant and primary predicate device are treated by extra light curing in UV light box to make final polymerization. Therefore, the new and predicate devices are the same in clinical function, and similar in composition and intended use. This supports that the compatibility and safety of the applicant device are substantially equivalent to the predicate device. ## 10. Differences The following differences may be noted between the predicate device and GC Temp PRINT. - Though primary predicate device indicates "Duration is less than 30 days in oral environment", applicant device is for "long-term temporary restorations". - The chemical formula is not exactly the same between applicant device and predicate devices. - 2 {5}------------------------------------------------ | | Applicant device | Primary Predicate Device | Reference Device | |------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product<br>category | Light cured temporary crown and bridge<br>resin,<br>Class II<br>872.3770; EBG | Resin for temporary crown and bridge<br>Class II<br>872.3770; EBG | Dual cured temporary crown and bridge resin<br>Class II<br>872.3770; EBG | | Trade name | GC Temp PRINT | Resin for Temporary Crown & Bridge<br>(K180657) | TEMPSMART<br>(MSN-006) K141562 | | Manufacturer | GC Manufacturing Europe N.V. | Dentis Co., Ltd. | GC Corporation | | Indication for<br>use | 1. Fabrication of Temporary crowns,<br>bridges, inlays, onlays and veneers.<br>2. Fabrication of long-term temporary<br>restorations. | Resin for Temporary Crown & Bridge is indicated<br>for the fabrication of temporary dental<br>restorations in conjunction with extra-oral curing<br>light equipment. Duration is less than 30 days in<br>oral environment. | 1. Fabrication of temporary crowns, bridges,<br>inlays, onlays and veneers<br>2. Fabrication of long-term temporary<br>restorations. | | Product<br>description | GC Temp PRINT is 3D printable light<br>curing composite for temporary crown<br>and bridge. The device consists of a<br>paste delivered in a bottle. GC Temp<br>PRINT is fabricated with certain Digital<br>Light Processing/Stereolithography<br>Apparatus (DLP/SLA)a` 3D printers and<br>post cured with light curing device into<br>dental restorations, which are then<br>bonded to tooth structure with adhesive<br>temporary cement. | The Resin for Temporary Crown & Bridge is<br>made by Methacrylate Oligomer based on the<br>Urethane Acrylate Oligomer with 0.01~0.1wt%<br>inorganic filler. It has stored in a brown 1000ml<br>of HDPE bottle. It contains materials with colors<br>of A2 based on the shade guide. This Product<br>is a liquid photo-curable material that is<br>polymerized by UV laser at 405nm. It can be<br>used to make a tooth model with a photo-<br>curable polymer that is cured by ultraviolet light.<br>The liquid UV curing resin is cured at a specific<br>wavelength (405nm) by the photo-initiator<br>contained in the resin. Curing in a 3D printer is<br>related to the conditions of the printer<br>equipment, and is typically 0.1 to 0.010mm in<br>thickness, and is output at a resolution of 0.1 to<br>0.03 mm on the x, y axis. This device should<br>use ZENITH 3D Printer equipment using UV<br>light source of Dentis Co., Ltd., and it is<br>possible to produce three dimensional printed<br>matter by curing lamination step by step a<br>thicknesses of 100, 50 and 16μm. | TEMPSMART is a dual-cured, temporary crown<br>and bridge resin. The components consist of<br>Paste A and B, which are filled in a cartridge.<br>Both pastes are automixed with a mixing tip. | | Components | *Urethane dimethacrylate (UDMA)<br>*Dimethacrylate<br>*Silicon dioxide<br>*Initiator<br>*Pigment | The Resin for Temporary Crown & Bridge is<br>made by Methacrylate Oligomer based on the<br>Urethane Acrylate Oligomer with 0.01~0.1wt%<br>inorganic filler. | <Paste A><br>* Composite filler<br>* Urethane dimethacrylate (UDMA)<br>* Dimethacrylate<br>* Silicon dioxide<br>* Initiator<br>* Pigment | {6}------------------------------------------------ | | | | <Paste B><br>* Urethane dimethacrylate (UDMA)<br>* Dimethacrylate<br>* Silicon dioxide<br>* Initiator | |-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Instructions<br>for use | Manufacturing process dental<br>prosthetics<br>Please refer to respective instructions of<br>the additive manufacturing systems. List<br>of additive manufacturing systems<br>which support GC Temp PRINT can be<br>downloaded from GC website -<br>www.gceurope.com<br>1. Digital Data Designing<br>Use the instructed software related to<br>the additive manufacturing system to<br>modify and distribute the printed object<br>and support structures on the virtual<br>building platform. Create slice data with<br>50 µm and send it to the additive<br>manufacturing device.<br>Note:<br>· Select the proper print program for<br>GC Temp PRINT including all relevant<br>process parameters.<br>2. Print processing<br>Before pouring GC Temp PRINT into<br>the reservoir of the additive<br>manufacturing device, shake the GC<br>Temp PRINT in the original bottle for<br>approximately 2 minutes. After shaking,<br>directly pour it into the reservoir and set<br>both the reservoir and the platform in<br>accordance with the additive<br>manufacturing system instruction.<br>Start printing.<br>Note:<br>Make sure your printing platform,<br>reservoir, working space and device are<br>cleaned properly according to the<br>additive manufacturing system<br>instructions. Dirty tools can cause<br>defect and therefore failure of the<br>printed restoration(s). | Indication for Use The Resin for Temporary<br>Crown & Bridge and the predicate device have<br>similar indications for use statements. Limitation<br>on the duration of the subject device was<br>intended to ensure the performance of subject<br>device by narrowing down the allowed terms of<br>its prosthetic use. Such difference is not critical<br>to the intended use of the subject device.<br>Therefore it doesn't affect the safety and<br>effectiveness of the device when used as<br>labeled. | Fabricating temporary restorations<br>1. Taking an impression<br>1) Take an impression using alginate or silicone<br>impression material. Alternatively, temporary<br>can be made using a laboratory-prepared<br>thermoforming matrix.<br>2) Block out preparation undercuts.<br>2. Teeth preparation<br>1) Block out preparation undercuts.<br>2) Clean teeth preparation with water spray and<br>lightly dry.<br>3) Lightly lubricate separating medium such as<br>Vaseline on prepared teeth or resin core.<br>3. Dispensing<br>1) Before dispensing the first time from a new<br>syringe or after a long interval in between use,<br>bleed the pastes from the syringe.<br>2) Remove the cap of the syringe by rotating 1/4<br>turn anti-clockwise. Attach a Mixing Tip by<br>aligning the V shaped notch on the rim of the<br>mixing tip with the V shaped notch between the<br>syringe barrels. Push firmly to attach the mixing<br>tip. Then rotate the colored collar of the mixing<br>tip 1/4 turn clockwise.<br>3) Discard first cm extruded before use.<br>4) Extrude material directly into the impression.<br>NOTE:<br>The used mixing tip can be left on the syringe to<br>serve as a storage cap until next use. For each<br>application, place a new mixing tip.<br>4. Fabricating a temporary restoration<br>Fabricate a temporary crown and bridge<br>according to timing showed below. Time<br>indicate after start of mixing paste.<br>Option1 Fabricating in the mouth | 기 {7}------------------------------------------------ | • Color difference may occur if shaking is insufficient.<br>• Avoid cross contamination with other products.<br>• Before each printing process, gently stir the remaining material in the reservoir.<br>Before each printing process, check there is no polymerized area or visual evidence of contamination. If such, replace material.<br>• Replace material in the reservoir according the additive manufacturing system instructions.<br>3. Cleaning of the printed objects<br>Carefully remove the printed restoration(s) from the printing platform.<br>Do not remove the support structures under the object. For Cleaning, place the restoration(s) in a glass with Isopropanol solution (>96%), then place the glass in an ultrasonic water bath for 2min. Dry parts with compressed air.<br>Repeat the cleaning procedure in a clean Isopropanol solution (>96%), then place in an ultrasonic water bath for 2 min. Important that the second rinse is done in a clean Isopropanol solution.<br>After cleaning, check if the surface is still shiny, wet or sticky, which would indicate there is still Isopropanol or residual monomers on the printed object. If such, please repeat the cleaning procedure until restoration(s) are completely clean.<br>Note:<br>• Do not rinse the restoration in an Isopropanol solution for more than 5 min.<br>• Do not wipe the restoration(s) surface.<br>• After cleaning, check the restoration for:<br>o Inhomogeneous color distribution linked to the printing process<br>o Deformation | 0'00"~<br>1'00" Place TEMPSMART into the impression and reposition into the mouth.<br>1'00"~<br>2'00" Setting in the mouth.<br>Up to<br>2'00" Remove the impression from the teeth.<br>If necessary, tack cure with a light curing unit (GC G-Light) for 2seconds each tooth unit on the teeth.<br>Remove the temporary restoration from the mouth and final cure with light curing unit for 5seconds each tooth unit.<br>After<br>final Completely remove the oxygen inhibition layer from the surface of the temporary restoration with alcohol.<br>cure<br>Option2 Fabricating on the model<br>0'00"~<br>1'00" Place TEMPSMART into the impression and reposition on the model.<br>1'00"~<br>3'00" Setting on the model.<br>Up to<br>3'00" Remove the impression from the model.<br>Remove the impression from the model and final cure with an extraoral light curing oven for 1minute.<br>Remove the temporary restoration from the model.<br>After<br>final Completely remove the oxygen inhibition layer from surface of the temporary restoration with alcohol.<br>cure<br>Caution!<br>When using a light curing unit on teeth, do not irradiate the temporary restoration more than 2 seconds for each surface because of high polymerization temperature.<br>Note: | |------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | o Defects like cracks or holes<br>4. Post curing<br>Perform a final polymerization using<br>Labolight DUO. For use of other<br>devices, we refer to respective<br>instructions manufacturing systems<br>which support GC Temp PRINT. This list<br>can be downloaded from GC website -<br>www.gceurope.com<br><br>For final polymerization with Labolight<br>DUO, cure for 3min on opposite from<br>support side.<br>Remove supports with a nipper and a<br>carbide bur.<br>Apply light for another 3 min from the<br>other side to obtain final homogeneous<br>polymerization. When using other<br>manufacturers' light curing devices,<br>refer to the respective manufacturer's<br>instructions for use.<br>Note:<br>• Make sure the Isopropanol solution is<br>completely removed prior to post curing.<br>• Do not apply light curing longer than<br>instructed.<br>• Do not use chair side hand-held type<br>curing light for post curing.<br>• Do not store the printed restoration(s)<br>in direct sunlight.<br>5. Adjustment - Finishing - Gloss<br>1. Carefully remove the remaining<br>supports with a carbide bur. Too much<br>pressure could break the restoration(s).<br>2. Shape and finish in the usual way.<br>3. When necessary, minor fit<br>adjustments can be undertaken using<br>UNIFAST III.<br>4. If required, modify shape by using a<br>resin material such as G-aenial Flo X,<br>G-aenial Universal Flo, UNIFAST III or<br>grind to obtain proper occlusion.<br>5. Apply a coating agent OPTIGLAZE<br>color following the manufacturer's<br>instructions for use.<br>Note: | 1) The above processing times are applicable to<br>products which have been stored and<br>processed at a temperature of 23° C / 74° F and<br>at 50% relative humidity. Higher temperatures<br>shorten these times, while lower temperatures<br>extend them. The size of the restoration also<br>has an impact on the setting times: as the size<br>of the workpiece increases, the setting<br>accelerates.<br>2) When light irradiating to the temporary<br>bridge, irradiate the light onto the occlusal<br>surface.<br>3) When final curing the temporary bridge with a<br>light curing unit, irradiate the light by one tooth<br>from both ends to of prevent deformation.<br><br>5. Finishing<br>1) Remove the excess material using a fine<br>carbide bur.<br>2) Check occlusion and if necessary, correct<br>shape by using a resin material (TEMPSMART,<br>G-ænial Flo, G-ænial Universal Flo, UNIFAST<br>Trad) or grind to obtain proper occlusion.<br>3) Polish the surface of the temporary<br>restoration using silicon polisher and goat's hair<br>wheel. Alternatively, apply a coating agent<br>(OPTIGLAZE).<br><br>Caution!<br>Do not breathe in polishing dust. Use personal<br>protective devices such as facemask and safety<br>glasses.<br><br>Note:<br>1) When using a resin material (TEMPSMART,<br>G-ænial Flo, G-ænial Universal Flo, UNIFAST<br>Trad) for custom-shaping and shape correction,<br>there usually is no need for a bonding agent.<br>But completely remove saliva contaminated part<br>using a carbide bur. And if necessary, apply a<br>ceramic primer (CERAMIC PRIMER II) to the<br>surface of the temporary restoration.<br>2) When using TEMPSMART for custom-<br>shaping and shape correction, light cure the<br>mixing paste with a light curing unit for 10<br>seconds. | {8}------------------------------------------------ D : : . T {9}------------------------------------------------ | | · Do not use GC Temp PRINT as an<br>add on material.<br>When using a resin material such as G-<br>aenial Flo X, G-aenial Universal Flo or<br>UNIFAST III for custom-shaping and<br>shape correction, there is no<br>requirement for prior application of a<br>bonding agent.<br>6. Cementing<br>Cement the completed temporary<br>restoration using FREEGENOL or<br>equivalent temporary cement. | | 6. Cementing<br>Cement using temporary cement.<br>Note:<br>Do not use eugenol cement because it will<br>prevent polymerization of resin.<br>Repairing / Modifying of temporary restorations<br>1) Abrade the surface of the temporary<br>restorations using carbide bur.<br>2) If necessary, apply a ceramic primer<br>(CERAMIC PRIMER II) and dry with air.<br>3) Place a resin material (TEMPSMART, G-<br>ænial Flo, G-ænial Universal Flo, UNIFAST<br>Trad) and cure according to manufacturer's<br>instructions.<br>Note:<br>Completely remove saliva contaminated part<br>using a carbide bur. | |-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological<br>Characteristics<br>and Mode of<br>action | The curing mechanism is<br>polymerization of uncured methacrylate<br>ester monomers photo initiator. It is<br>fabricated by using Additive<br>manufacturing system (3D printer). | The subject device is a polymer that changes its<br>properties when exposed to UV light. It conforms<br>into a hardened polymeric material through a<br>process called curing. Temporary dental<br>restoration using resin is technologically useful<br>applications for dental treatment. | The curing mechanism is polymerization of<br>uncured methacrylate ester monomers by<br>"mixing initiator(s) and activator(s)" and/or<br>"photo initiator." It is fabricated manually. | # 10. Conclusion Based on similarities in intended use, mode of action, chemical composition, and performance testing, CC Temp PRINT is substantially equivalent to the selected predicate devices. :
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...