Resin for Temporary Crown & Bridge

K180657 · Dentis Co., Ltd. · EBG · Dec 7, 2018 · Dental

Device Facts

Record IDK180657
Device NameResin for Temporary Crown & Bridge
ApplicantDentis Co., Ltd.
Product CodeEBG · Dental
Decision DateDec 7, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3770
Device ClassClass 2

Indications for Use

Resin for Temporary Crown & Bridge is indicated for the fabrication of temporary dental restorations in conjunction with extra-oral curing light equipment. Duration is less than 30 days in oral environment.

Device Story

Liquid photo-curable resin (methacrylate/urethane acrylate oligomer with inorganic filler) used for 3D printing temporary dental restorations; requires ZENITH 3D printer (405nm UV light source). Printer layers resin at 100, 50, or 16um thickness; resolution 0.1-0.03mm on x, y axis. Used in dental clinics/labs by professionals to create crowns and bridges. Curing process polymerizes liquid into hardened material. Provides temporary dental function for patients; biocompatible per ISO 10993.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (EN ISO 10993-3, -5, -10, -11) and mechanical performance testing (compressive/tensile strength, hardness, water sorption, solubility) compared against ISO 10477 standards.

Technological Characteristics

Methacrylate/urethane acrylate oligomer resin with 0.01-0.1wt% inorganic filler. Photo-curable at 405nm UV. Form factor: liquid in 1000ml HDPE bottle. Process: 3D printing (layer-by-layer curing). Biocompatibility: ISO 10993 compliant. Sterilization: non-sterile.

Indications for Use

Indicated for fabrication of temporary dental restorations for patients requiring short-term (less than 30 days) oral prosthetic treatment.

Regulatory Classification

Identification

A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. December 7, 2018 Dentis Co., Ltd. % Hwi Joon Park Consultant Allura Medical Solution Inc. 8141 Lampson Ave. #4 Garden Grove, California 92841 Re: K180657 Trade/Device Name: Resin for Temporary Crown & Bridge Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary Crown and Bridge Resin Regulatory Class: Class II Product Code: EBG Dated: September 13, 2018 Received: September 14, 2018 Dear Hwi Joon Park: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Image /page/1/Figure/5 description: The image shows a digital signature. The signature is from Mary S. Runner -S3. The date of the signature is 2018.12.07, and the time is 10:21:47 -05'00'. For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K180657 Device Name Resin for Temporary Crown & Bridge Indications for Use (Describe) Resin for Temporary Crown & Bridge is indication of temporary dental restorations in conjunction with extra-oral curing light equipment. Duration is less than 30 days in oral environment. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><span> </span><span style="margin-left: 5px;">Prescription Use (Part 21 CFR 801 Subpart D)</span></div> | <div style="display:flex; align-items:center;"><span> </span><span style="margin-left: 5px;">Over-The-Counter Use (21 CFR 801 Subpart C)</span></div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ - ## 510(k) Summary K180657 | Submitter | Dentis Co., Ltd.<br>99, Seongseoseo-ro, Dalseo-gu, Daegu, 42718 Republic of Korea<br>Phone. +82-53-583-2804<br>Fax. +82-53-283-2806 | |-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | HWI JOON PARK<br>8141 Lampson Ave. #4, Garden Grove, CA 92841, USA<br>Phone: +1-972-800-0044<br>Fax: +1-210-899-0079 | | Submission Date | Mar 08, 2018 | | Trade / Proprietary<br>name | Resin for Temporary Crown & Bridge | | Common / Usual<br>Name | Crown and Bridge, Temporary, Resin | | Classification Name<br>Classification Code<br>Regulatory Class<br>Regulation Number | Temporary Crown and Bridge Resin<br>EBG<br>Class II 872.3770 | | Predicate Device | DeltaMed GMBH<br>e-Dent Temporary Resin and Extra-Oral Curing System<br>(K102776, Feb 18, 2011 - Primary Predicate) | | Description of<br>Device | The Resin for Temporary Crown & Bridge is made by Methacrylate<br>Oligomer based on the Urethane Acrylate Oligomer with 0.01~0.1wt%<br>inorganic filler. It has stored in a brown 1000ml of HDPE bottle. It contains<br>materials with colors of A2 based on the shade guide.<br>This Product is a liquid photo-curable material that is polymerized by UV laser<br>at 405nm. It can be used to make a tooth model with a photo-curable polymer<br>that is cured by ultraviolet light. The liquid UV curing resin is cured at a<br>specific wavelength (405nm) by the photo-initiator contained in the resin.<br>Curing in a 3D printer is related to the conditions of the printer equipment, and<br>is typically 0.1 to 0.010mm in thickness, and is output at a resolution of 0.1 to<br>0.03 mm on the x, y axis. This device should use ZENITH 3D Printer<br>equipment using UV light source of Dentis Co., Ltd., and it is possible to<br>produce three dimensional printed matter by curing lamination step by step a<br>thicknesses of 100, 50 and 16um. | | Indication for Use | Resin for Temporary Crown & Bridge is indicated for the fabrication of<br>temporary dental restorations in conjunction with extra-oral curing light<br>equipment. Duration is less than 30 days in oral environment. | | Comparison of<br>Technological<br>Characteristics with<br>the Predicate Devices | Indication for Use<br>The Resin for Temporary Crown & Bridge and the predicate device have<br>similar indications for use statements. Limitation on the duration of the<br>subject device was intended to ensure the performance of subject device by<br>narrowing down the allowed terms of its prosthetic use. Such difference is<br>not critical to the intended use of the subject device.<br>Therefore it doesn't affect the safety and effectiveness of the device when<br>used as labeled. | | | Technological Characteristics<br>The subject device is a polymer that changes its properties when exposed<br>to UV light. It conforms into a hardened polymeric material through a<br>process called curing. Temporary dental restoration using resin is<br>technologically useful applications for dental treatment. | | | Above technological characteristics of the subject device is identical to<br>the predicate device. | | | Materials<br>The subject device is composed of the each raw material whose safety were<br>evaluated by suppliers and the results were indicated on MSDS. Moreover,<br>the Resin for Temporary Crown & Bridge was evaluated for biocompatibility<br>based on ISO 10993. | | | Clinical Tests<br>The Resin for Temporary Crown & Bridge and the cited predicate device were<br>not conducted clinical tests to determine substantial equivalence. | | | Non-clinical Tests<br>Non-clinical testing was performed in order to validate the product against<br>the company's specified design requirement.<br>The Resin for Temporary Crown & Bridge in all variants were tested with<br>respect to biocompatibility according to EN ISO 10993-3, EN ISO 10993-5,<br>EN ISO 10993-10 and EN ISO 10993-11. The laboratory certified that the<br>insolubility is in compliance with the requirements of the standard. There is<br>no evidence that effects hazardous to the patient will arise by leachable<br>ingredients/contaminants.<br>The composition of the Resin for Temporary Crown & Bridge exhibits<br>sufficient strengths and performances in all intraoral conditions and will<br>sufficiently resist compressive & tensile loads, hardness, water sorption and<br>solubility. | | | Sterilization<br>The Resin for Temporary Crown & Bridge and the cited predicate device are<br>provided non-sterile. | | | Performance Specification<br>Performance testing confirmed the Resin for Temporary Crown & Bridge<br>demonstrated equivalent or better performance to the predicate device or<br>acceptance criteria referred to ISO 10477. | | Conclusion | It is adequate for the intended use and it shows that there are no<br>significant differences between the Resin for Temporary Crown &<br>Bridge and the predicate device currently being marketed that would<br>adversely affect the use of the product. Any differences in<br>technological characteristics do not raise new issues of safety or<br>efficacy.<br>Therefore, conclusions drawn from testing demonstrate that the<br>Resin for Temporary Crown & Bridge is substantially equivalent in performance to predicate device. | {4}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...