TPH Spectra+ Universal Composite Restorative

K162107 · Dentsply Sirona · EBF · Sep 27, 2016 · Dental

Device Facts

Record IDK162107
Device NameTPH Spectra+ Universal Composite Restorative
ApplicantDentsply Sirona
Product CodeEBF · Dental
Decision DateSep 27, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3690
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

TPH Spectra® + Universal Composite Restorative is indicated for direct or indirect replacement of missing or deficient tooth structure in primary and permanent anterior and posterior teeth, e.g., 1. Direct anterior and posterior restorations (including occlusal surfaces) 2. Core build-ups 3. Splinting 4. Indirect restorations including inlays, onlays

Device Story

Visible light-cured, radiopaque, methacrylate-based composite restorative; used for anterior/posterior tooth restoration and reshaping. Input: composite material dispensed via multi-use syringes or pre-dosed Compules® Tips (using DENTSPLY Compules® Tips Gun). Operation: clinician applies material to tooth following adhesive/cement bonding; light-cured to harden. Output: permanent dental restoration. Benefit: high strength, low wear, and aesthetic tooth-like shade/fluorescence for functional/cosmetic repair. Used in dental clinics by dentists.

Clinical Evidence

No clinical performance data was submitted. Evidence consists of in-vitro bench testing including radiopacity, fracture toughness, fluorescence, depth of cure, flexural strength/modulus, water sorption/solubility, localized wear loss, and sensitivity to ambient light, evaluated against ISO 4049:2009 and internal standards.

Technological Characteristics

Methacrylate-based resin, photo initiator, silanated inorganic filler, and pigments. Visible light-cured. Radiopaque. Packaged in syringes or Compules® Tips. Complies with ISO 4049:2009 (Dentistry - Polymer-based restorative) and ISO 10993-1/ISO 7405 for biocompatibility.

Indications for Use

Indicated for direct or indirect replacement of missing or deficient tooth structure in primary and permanent anterior and posterior teeth, including direct restorations, core build-ups, splinting, and indirect restorations (inlays, onlays).

Regulatory Classification

Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 27, 2016 Dentsply Sirona Helen Lewis Director Corporate Regulatory Affairs 221 W Philadelphia St, Suite 60 York, Pennsylvania 17404 Re: K162107 Trade/Device Name: Tph Spectra+ Universal Composite Restorative Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: July 28, 2016 Received: July 29, 2016 Dear Helen Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Image /page/1/Picture/7 description: The image shows a signature that reads "Susan Runno DDS, MA". The signature is written in cursive with a black ink. The letters are connected and flow smoothly together. Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) Device Name TPH Spectra® + Universal Composite Restorative Indications for Use (Describe) TPH Spectra® + Universal Composite Restorative is indirect replacement of missing or deficient tooth structure in primary and permanent anterior and posterior teeth, e.g., 1. Direct anterior and posterior restorations (including occlusal surfaces) - 2. Core build-ups - 3. Splinting - 4. Indirect restorations including inlays, onlays Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404 Image /page/3/Picture/1 description: The image shows the Dentsply Sirona logo. The logo consists of a stylized leaf-like shape on the left and the words "Dentsply Sirona" on the right. The leaf shape is gray, and the text is a darker gray. The text is stacked, with "Dentsply" on top of "Sirona". #### 510(k) SUMMARY for TPH Spectra® + Universal Composite Restorative Submitter Information: Dentsply Sirona 221 West Philadelphia Street Suite 60 York, PA 17404 | Contact Person: | Helen Lewis | |-------------------|--------------| | Telephone Number: | 717-487-1332 | | Fax Number: | 717-849-4343 | Date Prepared: July 28, 2016 ## Device Name: - Proprietary Name: ● - . Classification Name: - CFR Number: . - Device Class: . - Product Code: ● Predicate Device: TPH Spectra® + Universal Composite Restorative Tooth Shade Resin Material 872.3690 II EBF | Predicate Device Name | 510(k) | Company Name | |---------------------------------------------|---------|--------------| | Filtek™ Supreme Ultra Universal Restorative | K083610 | 3M ESPE | ## Description of Device: TPH Spectra® + Universal Composite Restorative is a visible light cured, radiopaque, composite restorative for anterior and posterior restorations and cosmetic and functional reshaping of primary and permanent teeth. The new device contains methacrylate-based resin, photo initiator, silanated inorganic filler and pigments. Available shades include opaque dentin shades, regular body shades and translucent enamel shades. This restorative provides high strength and low wear for durability. TPH Spectra® + Universal Composite Restorative is applied to the tooth following use of a methacrylate-based dental adhesive and/or a cement, such as manufactured by DENSTPLY, which permanently bonds the restoration to the tooth structure. TPH Spectra® + Universal Composite Restorative is packaged in traditional multi-use syringes, for outside-the-mouth dispensing onto a mixing pad, and in pre-dosed Compules® Tips for intraoral dispensing. The restorative composite in Compules® Tips is dispensed using a DENTSPLY Compules® Tips Gun. {4}------------------------------------------------ ## Indications for Use: TPH Spectra® + Universal Composite Restorative is indicated for direct or indirect replacement of missing or deficient tooth structure in primary and permanent anterior and posterior teeth, e.g., - 1. Direct anterior and posterior restorations (including occlusal surfaces) - 2. Core build-ups - 3. Splinting - 4. Indirect restorations including inlays, onlays #### Substantial Equivalence: ## Technological Characteristics: Information provided in this 510(k) submission shows that TPH Spectra® + Universal Composite Restorative supports substantial equivalence when compared to the predicate device, Filtek "" Supreme Ultra Universal Restorative (K083610) in terms of intended use, indications for use, composition, physical properties and technological characteristics. {5}------------------------------------------------ | Element | Predicate Device<br>Filtek™ Supreme Ultra Universal Restorative | Proposed Device<br>TPH Spectra® + Universal Composite<br>Restorative | Difference | |-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | K083610 | To be assigned | | | Indications<br>for use | Direct anterior and posterior restorations<br>• (including occlusal surfaces)<br>• Core build-ups<br>• Splinting<br>• Indirect restorations including inlays,<br>onlays and veneers | Direct or indirect replacement of missing<br>or deficient tooth structure in primary and<br>permanent anterior and posterior teeth,<br>e.g.<br>1. Direct anterior and posterior<br>restorations (including occlusal<br>surfaces)<br>2. Core build-ups<br>3. Splinting<br>4. Indirect restorations including inlays,<br>onlays and veneers | "Direct or indirect replacement of<br>missing or deficient tooth structure<br>in primary and permanent anterior<br>and posterior teeth" is a description<br>of the indications in general.<br>Indications of the proposed device<br>are the same as the predicate device.<br>According to ISO 4049, the physical<br>property data support that the<br>proposed device meets the<br>requirements primarily for the direct<br>and indirect restorations. | | Composition<br>of Materials | Methacrylate-based resin, photo initiator,<br>fillers and pigments. | Methacrylate-based resin, photo initiator,<br>fillers and pigments. | Chemically similar to the predicate<br>device. | | Physical<br>Properties | Radiopacity: 2.10 (0.0) mmAl<br>Fracture Toughness: $1.64 (0.09) MPa.m^{1/2}$<br>Fluorescence: similar to a natural tooth.<br>Depth of Cure: 2.60 (0.02) mm<br>Flexural Strength: 131 (14) MPa<br>Flexural Modulus: 10.6 (0.6) MPa<br>Water Sorption: 33.1 (2.1) µg/mm³<br>Water Solubility: 1.0 (0.7) µg/mm³<br>Localized Wear Loss: 0.87 (0.24) x10⁻² mm³<br>Sensitivity to Ambient Light: 135 s<br>Shade and Color Stability: Stable/Pass<br>Stain Resistance: 3.30 (1.06) | Radiopacity: 2.23 (0.08) mmAl<br>Fracture Toughness: $1.71 (0.16) MPa.m^{1/2}$<br>Fluorescence: similar to a natural tooth.<br>Depth of Cure: 2.12 (0.02) mm<br>Flexural Strength: 117 (10) MPa<br>Flexural Modulus: 11.6 (0.8) MPa<br>Water Sorption: 19.3 (1.9) µg/mm³<br>Water Solubility: 0 µg/mm³<br>Localized Wear Loss: 0.47 (0.15) x10⁻² mm³<br>Sensitivity to Ambient Light: 180 s<br>Shade and Color Stability: Stable/Pass<br>Stain Resistance: 1.69 (0.25) | Physical property data support<br>substantial equivalence of the<br>proposed device when compared to<br>the predicate device. | Table 5.1 Similarities and Differences between the proposed and the predicate devices {6}------------------------------------------------ ## Biocompatibility Testing: An evaluation of biocompatibility was performed for the TPH Spectra® + Universal Composite Restorative in accordance with ISO 10993-1(Biological Evaluation of Medical Devices- Part1: Evaluation and Testing) and ISO 7405 (Dentistry - Evaluation of Biocompatibility of Medical Devices used in Dentistry). TPH Spectra® + Universal Composite Restorative has been demonstrated as biocompatible for its intended use. ## Physical Properties: In-vitro bench tests were performed on the TPH Spectra® + Universal Composite Restorative including Radiopacity, Fracture Toughness, Fluorescence, Depth of Cure, Flexural Strength, Flexural Modulus, Water Sorption and Solubility, Localized Wear Loss, and Sensitivity to Ambient Light. The compressive strength was no evaluated due to the unbonded, free standing failure mode of the test which is clinically irrelevant under real clinical situations. The results indicated that the TPH Spectra® + Universal Composite Restorative meets or exceeds the requirements of ISO 4049: 2009 (Dentistry -- Polymer-based restorative ) or DENTSPLY internal standards and supports substantial equivalence when compared to the predicate devices on physical properties. ## Clinical Performance Data: No clinical performance data was submitted. ## Risk Analysis The risk analysis of TPH Spectra® + Universal Composite Restorative was conducted by a design Failure Mode and Effects Analysis (FMEA). As designed, the residual risk of the TPH Spectra® + Universal Composite Restorative is considered acceptable. The benefits of the product are considered to outweigh the risks outlined in the risk analysis. ## Conclusion Regarding Substantial Equivalence: TPH Spectra® + Universal Composite Restorative is a composite which is intended to be used for anterior and posterior restorations and cosmetic and functional reshaping of primary and permanent teeth. The TPH Spectra® + Universal Composite Restorative has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicate Filtek™ Supreme Ultra Universal Restorative cleared under premarket notification K083610. Test data to verify the performance of the TPH Spectra® + Universal Composite Restorative has been provided for Radiopacity, Fracture Toughness, Fluorescence, Depth of Cure, Flexural Strength, Flexural Modulus, Water Sorption and Solubility, Localized Wear Loss, and Sensitivity to Ambient Light. Combined with the design and intended use comparison with the predicate device, we have determined that the proposed device TPH Spectra® + Universal Composite Restorative supports substantial equivalence when compared to predicate device Filtek™ Supreme Ultra Universal Restorative (K083610).
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...