The S-Plant Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller S-Plant Dental Implants (03.4, 3.6, 3.8, 4.2, 4.7, 5.2 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger S-Plant Dental Implants (Ø6.0, 7.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading. Dual abutments are intended for use with a dental implant to prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.
Device Story
S-Plant Dental Implant System consists of endosseous dental implants (fixtures), abutments, and screws. Fixtures are screw-type, made of Grade 4 CP titanium (ASTM F67) with S.L.A. surface treatment. Abutments and screws are made of Ti-6Al-4V ELI alloy (ASTM F136). Used by dental surgeons in clinical settings to replace tooth roots and provide foundation for prosthetic restorations (crowns, bridges, overdentures). Fixtures are surgically placed into jawbone; abutments connect to fixtures via tapered joint to support prosthetics. Healing abutments shape soft tissue; cover screws protect internal threads during osseointegration. System supports both immediate and delayed loading protocols depending on implant size and clinical stability. Benefits include restoration of chewing function and esthetics for edentulous patients.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993 series), sterilization validation (ISO 11137), shelf-life (ASTM F1980), and surface characterization (EDS/SEM). MRI safety evaluated via scientific rationale and literature review per FDA guidance.
Technological Characteristics
Materials: CP Ti Grade 4 (ASTM F67) for fixtures; Ti-6Al-4V ELI (ASTM F136) for abutments/screws. Surface: S.L.A. (Sand-blasted, Large-grit, Acid-etched). Connection: Tapered conical hex. Sterilization: Gamma radiation (fixtures). Form factor: Screw-type endosseous implants with various diameters/lengths.
Indications for Use
Indicated for patients requiring single or multiple unit prosthetic restoration in upper/lower jaw arches. Smaller implants (3.4-5.2mm) allow two-stage or single-stage (immediate loading) surgery. Larger implants (6.0-7.0mm) indicated for molar region with delayed loading. Contraindications not specified.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
DIO UF HSA Internal Sub-merged Implant System (K122519)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 18, 2023
IDIS Co., Ltd. % Joyce Kwon CEO Provision Consulting Group, Inc. 13925 City Center Dr., Suite 200 Chino Hills, California 91709
Re: K231429
Trade/Device Name: S-Plant Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous dental implant Regulatory Class: Class II Product Code: DZE, NHA Dated: November 13, 2023 Received: November 14, 2023
Dear Joyce Kwon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Blitzer
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K231429
Device Name S-Plant Dental Implant System
#### Indications for Use (Describe)
The S-Plant Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller S-Plant Dental Implants (03.4, 3.6, 3.8, 4.2, 4.7, 5.2 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger S-Plant Dental Implants (Ø6.0, 7.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.
Dual abutments are intended for use with a dental implant to prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### 510(k) Submitter
Aiden Lee IDIS Co, Ltd. #1314~ #1319, 122, Jomaru-ro, 385 beon gil, Bucheon-si, Gyeunggi-do, 14556, Republic of Korea
#### Official Correspondent / Contact Person
Joyce Kwon, CEO/Principal Consultant Provision Consulting Group, Inc. 13925 City Center Dr. Suite 200 Chino Hills, CA 91709 +1-909-493-3276 joyce@provisionfda.com
#### Date Prepared: 12/18/2023
#### Device Information
- Trade Name: S-Plant Dental Implant System
- Common Name: Endosseous Dental Implant
- Classification Name: Endosseous dental implant
- Regulation Number: 21 CFR 872.3640
- Device Class: Class II
- Primary Product Code: DZE
- Secondary Product Code: NHA ●
#### Predicate Devices
Fixture:
- DIO UF HSA Internal Sub-merged Implant System (K122519) ●
#### Reference Device
Fixture
- DIO UF(II) Narrow Implant System (K161987) .
#### Abutment:
- Angled Abutment- Osstem Implant Angled Abutment (K182091) ●
- Dual Abutment- Osstem Implant Dual Abutment (K182091) ●
- Healing Abutment- Hiossen Healing Abutment (K203360)
#### Accessories:
- Cover Screw- Dentium Implantium SuperLine Abutments Cover Screw (K141457) ●
- Abutment Screw- Osstem Implant Abutment Screw (K182091) ●
{5}------------------------------------------------
# Prior Submission Information
N/A
#### Indications for Use
The S-Plant Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple units' prosthetic attachment to restore a patient's chewing function. The smaller S-Plant Dental Implants (03.4, 3.6, 3.8, 4.2, 4.7, 5.2 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading when good primary stability is achieved with appropriate occlusal loading. The larger S-Plant Dental Implants (Ø6.0, 7.0 mm) can be placed with a conventional two stage surgical process with an option for transmucosal healing and are indicated for the molar region with delayed loading.
Dual abutments are intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. Healing abutments are used to make a natural soft tissue shape before setting up prosthetics and removing cover screw after osseointegration. Cover Screws are used to protect the internal portion of the implant, preventing soft tissue growth into the implant, facilitating provisional restorations when necessary, and enabling the transition to final restoration components once osseointegration is complete.
#### Device Description
The S-Plant Dental Implant System is comprised of dental implants, superstructures, instruments for prosthetics and surgical instruments. The S-Plant Dental Implant System is specially designed for use in dental implant surgery. A successfully osseointegrated implant will achieve a firm implant when surgically implanted under controlled conditions. They are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations.
The S-Plant Dental Implant System fixtures are made of commercial pure titanium, grade 4 (ASTM F67) which have a S.L.A (Sand blasted large grit acid etched) treated surface and supplied sterile (gamma radiation). These fixtures can be used one stage surgery method or two stage surgery method. And that are surgically inserted into the upper and/or lower jawbone. The fixtures replace tooth roots providing a stable foundation for restorations. Geometrically, the implant is screw tvpe. An abutment is connected to the implant through a tapered joint.
The Abutment made of Ti-6AL-4V ELI alloy (ASTM F136) is intended for use as an aid in single or multiple-unit prosthetic restorations. It consists of Healing Abutment, Dual Abutment, and Abutment Screws. All abutments are supplied non-sterile and autoclaved by the end user.
The Cover Screw made of Ti-6Al-4V ELI alloy (ASTM F136), is an essential component in dental implant procedures. This device safeguards the internal threads of dental implant fixtures during the healing phase, ensuring a sterile environment for successful osseointegration.
{6}------------------------------------------------
The dimensions of subject devices are as following:
| No. | Device Name | Dimensions | | | | | | | | | | | | | | | | | | | | | |
|-----|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| 1. | S-Plant Fixture | Ø 3.4 x 8.5, 10.0, 11.5, 13.0, 15.0 mm (L)<br>Ø 3.6 x 8.5, 10.0, 11.5, 13.0, 15.0 mm (L)<br>Ø 3.8 x 7.0, 8.5, 10.0, 11.5, 13.0, 15.0 mm (L)<br>Ø 4.2 x 7.0, 8.5, 10.0, 11.5, 13.0 mm (L)<br>Ø 4.7 x 7.0, 8.5, 10.0, 11.5, 13.0 mm (L)<br>Ø 5.2 x 7.0, 8.5, 10.0, 11.5 mm (L)<br>Ø 6.0 x 7.0, 8.5, 10.0, 11.5 mm (L)<br>Ø 7.0 x 7.0, 8.5 mm (L) | | | | | | | | | | | | | | | | | | | | | |
| 2. | Dual Abutment | D(Ø) G/H Post 4 1,2,3,4,5 5.5,7 4.5 1,2,3,4,5 5.5,7 4.5 1,2,3,4,5 5.5,7 5 1,2,3,4,5 4,5.5,7 6 1,2,3,4,5 4,5.5,7 7 1,2,3,4,5 5.5 | | | | | | | | | | | | | | | | | | | | | |
| 3. | Healing Abutment | Ø 4.0 x 3.0, 4.0, 5.0, 6.0, 7.0 mm (L)<br>Ø 4.5 x 3.0, 4.0, 5.0, 6.0, 7.0 mm (L)<br>Ø 5.0 x 3.0, 4.0, 5.0, 6.0, 7.0 mm (L)<br>Ø 6.0 x 3.0, 4.0, 5.0, 6.0, 7.0 mm (L)<br>Ø 7.0 x 3.0, 4.0, 5.0, 6.0, 7.0 mm (L) | | | | | | | | | | | | | | | | | | | | | |
| 4. | Abutment Screw | Ø 2.25 x 10.2 mm (L)<br>Ø 2.5 x 9.0 mm (L) | | | | | | | | | | | | | | | | | | | | | |
| 5. | Cover Screw | Ø 3.60 x 6.5 mm (L) | | | | | | | | | | | | | | | | | | | | | |
### Substantial Equivalent Comparison Chart with Primary Predicate Device and Reference Device for Fixture
| | Subject Device | Primary Predicate Device | Reference Device |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | S-Plant Implant System<br>Fixture | DIO UF HSA Internal<br>Sub-merged Implant System | DIO UF(II) Narrow<br>Implant System |
| 510(k)<br>Number | K231429 | K122519 | K161987 |
| Manufacturer | IDIS | DIO Corporation | DIO Corporation |
| Classification<br>Regulation | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 |
| Primary<br>Product Code | DZE | DZE | DZE |
| Intended Use | Functional and esthetic<br>rehabilitation of the<br>edentulous mandible or<br>maxilla | Functional and esthetic<br>rehabilitation of the<br>edentulous mandible or<br>maxilla | Functional and<br>esthetic<br>rehabilitation of the<br>edentulous<br>mandible or maxilla |
| Indications | The S-Plant Dental Implant<br>System is indicated for surgical<br>placement in the upper and<br>lower jaw arches, to provide a<br>root form means for single or<br>multiple units' prosthetic<br>attachment to restore a patient's<br>chewing function. The smaller<br>S-Plant Dental Implants (Ø3.4,<br>3.6, 3.8, 4.2, 4.7, 5.2 mm) can<br>be placed with a conventional<br>two stage surgical process with<br>an option for transmucosal<br>healing or they can be placed in<br>a single stage surgical process<br>for immediate loading when<br>good primary stability is<br>achieved with appropriate<br>occlusal loading. The larger<br>S-Plant Dental Implants<br>(Ø6.0, 7.0 mm) can be placed<br>with a conventional two stage<br>surgical process with an option<br>for transmucosal healing and are<br>indicated for the molar region<br>with delayed loading. Dual<br>abutments are intended for use<br>with a dental implant to provide<br>support for prosthetic<br>restorations such as crowns,<br>bridges, or overdentures.<br>Healing abutments are used to<br>make a natural soft tissue shape<br>before setting up prosthetics and<br>removing cover screw after<br>osseointegration. Cover Screws<br>are used to protect the internal<br>portion of the implant,<br>preventing soft tissue growth<br>into the implant, facilitating<br>provisional restorations when<br>necessary, and enabling the<br>transition to final restoration<br>components once<br>osseointegration is complete. | The DIO UF HSA Internal<br>Sub-Merged Implant Systems<br>indicated for surgical<br>placement in the upper and<br>lower jaw arches, to provide a<br>root form means for single or<br>multiple units' prosthetic<br>attachment to restore a<br>patient's chewing function.<br>The smaller The DIO UF<br>HSA Internal Sub-Merged<br>(03.8~ Ø5.5) implants can be<br>placed with a conventional<br>two stage surgical process<br>with an option for<br>transmucosal healing or they<br>can be placed in a single stage<br>surgical process for<br>immediate loading when good<br>primary stability is achieved<br>with appropriate occlusal<br>loading. The larger (Ø6.0 ~<br>Ø7.0) implants can be placed<br>with a conventional two stage<br>surgical process with an<br>option for transmucosal<br>healing and are indicated for<br>the molar region with delayed<br>loading. | The UF(II) Narrow<br>Implant System is<br>intended for two-<br>stage surgical<br>procedures in the<br>following situations<br>and with the<br>following clinical<br>protocols:<br>- The intended use<br>for the 3.0mm,<br>3.3mm diameter<br>UF(II) Narrow<br>Implant is limited<br>to the replacement<br>of maxillary lateral<br>incisors and<br>mandibular<br>incisors.<br>- Immediate<br>placement in<br>extraction sites and<br>in situations with a<br>partially or<br>completely healed<br>alveolar ridge.<br>- It is intended for<br>delayed loading. |
| Prosthetic<br>Interface<br>Connection | Tapered conical hex | Tapered conical hex | Tapered conical<br>hex |
| Fixture<br>Diameter and<br>Length | Diameters 3.4 mm; 3.6 mm<br>Each in lengths of 8.5, 10.0,<br>11.5, 13.0, 15.0 mm<br>Diameter 3.8 mm<br>In lengths of 7.0, 8.5, 10.0,<br>11.5, 13.0, 15.0 mm<br>Diameters 4.2 mm; 4.7 mm;<br>Each in lengths of 7.0, 8.5,<br>10.0, 11.5, 13.0 mm<br>Diameter 5.2 mm; 6.0 mm;<br>Each in lengths of 7.0, 8.5,<br>10.0, 11.5 mm<br>Diameter 7.0 mm<br>in lengths of 7.0, 8.5 mm | Fixture Diameter:<br>3.8, 4.0, 4.5, 5.0, 5.5, 6.0, 7.0<br>Fixture Length:<br>7.0, 8.5, 10, 11.5, 13.0, 15.0,<br>16.0 | Fixture Diameter:<br>3.0, 3.3<br>Fixture Length:<br>8.5, 10, 11.5, 13.0,<br>15.0 |
| Implant<br>Material | CP Ti Grade 4<br>(ASTM F67) | CP Ti Grade 4<br>(ASTM F67) | CP Ti Grade 4<br>(ASTM F67) |
| Implant<br>Endosseous<br>Surface | S.L.A. | S.L.A. | S.L.A. |
| Implant<br>Placement | Bone Level | Bone Level | Bone Level |
| Sterilization | Gamma Sterilization | Gamma Sterilization | Gamma<br>Sterilization |
| Shelf Life | 8 years | 5 years | 5 years |
{7}------------------------------------------------
{8}------------------------------------------------
#### Substantial Equivalence Discussion
- 1. General Indications:
Both the subject device and the predicate device share similar overarching indications for use. They are both designed for surgical placement in the upper and lower jaw arches, with the primary goal of providing a root form means for prosthetic attachment to restore a patient's chewing function.
- 2. Implant Sizes and Surgical Procedures:
Both systems offer a range of implant sizes to accommodate various clinical scenarios. The smaller implants of both systems (03.4 ~ Ø5.5 mm) can be placed using a conventional twostage surgical process, with the option for transmucosal healing. Additionally, both systems provide the flexibility of a single-stage surgical process for immediate loading when there is good primary stability, with appropriate occlusal loading.
{9}------------------------------------------------
- 3. Molar Region and Delayed Loading:
The larger implants in both systems (Ø6.0 ~ Ø7.0 mm) are specifically indicated for the molar region and involve a conventional two-stage surgical process with an option for transmucosal healing, accompanied by delayed loading. This demonstrates a parallel approach to addressing the unique requirements of the molar region.
#### Discussion:
The comparisons highlight the substantial equivalence between the S-Plant Dental Implant System and the DIO UF HSA Internal Sub-Merged Implant System. The similarities in their indications for use, implant sizes, and surgical procedures demonstrate that the differences between the two devices are not substantial enough to raise any safety or efficacy concerns.
The minor variations in implant dimensions and specific procedural details are well within the acceptable range for dental implant systems, considering the diverse clinical needs and preferences of practitioners. Both systems provide suitable options for immediate loading and delayed loading based on clinical conditions, indicating flexibility in accommodating various patient cases.
The S-Plant Dental Implant System is similar to the predicate device (K122519) in terms of its design, dimensions, material, surface treatment, intended use, and technological characteristics, The S-Plant Dental Implant System also includes diameters (3.4mm and 3.6mm) that are not covered by the predicate device, but they fall within the range between the diameter of the reference device (K161987) and the diameter of the predicate device. Therefore, upon comparing the subject device with both the predicate and reference devices, there have been no new substantial equivalence concerns identified for the S-Plant Dental Implant System in terms of safety and effectiveness.
| | Subject Device | Reference Device |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
| Type | Cover Screw | Cover Screw |
| Secondary<br>Product<br>Code | NHA | NHA |
| K Number | K231429 | K141457 |
| Manufacturer | IDIS | Dentium Company Limited |
| Description | Cover Screws are used to protect the<br>internal portion of the implant,<br>preventing soft tissue growth into the<br>implant, facilitating provisional<br>restorations when necessary, and | Cover Screws are used to protect the<br>internal portion of the implant. |
#### Substantial Equivalent Comparison Chart with Reference Device for Cover Screw
{10}------------------------------------------------
| | enabling the transition to final<br>restoration components once<br>osseointegration is complete. | |
|----------|--------------------------------------------------------------------------------------------------|------------------------------------------|
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) |
| | Diameter<br>(mm) | 3.6 |
Substantial Equivalent Comparison Discussion: Proposed Cover Screw and the predicated cover screw have common in design, function, indications for use, material, and dimensions; therefore, the proposed Cover Screw is substantially equivalent to the predicated cover screw (K141457).
#### Substantial Equivalent Comparison Chart with Reference Device for Abutments
| | Subject Device | Reference Device |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Abutment Type | Dual Abutment | Dual Abutment |
| Secondary<br>Product Code | NHA | NHA |
| K Number | K231429 | K182091 |
| Manufacturer | IDIS | Osstem Implant Co., Ltd. |
| Design | Image: Subject Device Design | Image: Reference Device Design |
| Indications for<br>Use | Dual abutments are intended for<br>use with a dental implant to<br>provide support for prosthetic<br>restorations such as crowns,<br>bridges, or overdentures. | The Osstem Abutment System is<br>intended for use with a dental<br>implant to provide support for<br>prosthetic restorations such as<br>crowns, bridges, or overdentures. |
| Principle of<br>Operation | Using making for general cement-<br>type prosthesis. | Using making for general cement-<br>type prosthesis. |
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) |
#### 1. Dual Abutment
{11}------------------------------------------------
| Diameter (mm) | D(Ø) | G/H | Post | D(Ø) | G/H | Post |
|---------------|------|-----------|---------|------|---------------|---------|
| | 4 | 1,2,3,4,5 | 5.5,7 | 4 | 1,2,3,4,5,6,7 | 5.5,7 |
| | 4.5 | 1,2,3,4,5 | 5.5,7 | 4.6 | 6,7 | 5.5,7 |
| | 4.5 | 1,2,3,4,5 | 5.5,7 | 5 | 6,7 | 4,5.5,7 |
| | 5 | 1,2,3,4,5 | 4,5.5,7 | 6 | 6,7 | 4,5.5,7 |
| | 6 | 1,2,3,4,5 | 4,5.5,7 | 7 | 1,2,3,4,5,6,7 | 4 |
| | 7 | 1,2,3,4,5 | 5.5 | 7 | 6,7 | 5.5 |
Substantial Equivalent Comparison Discussion: Proposed Dual Abutment and the Reference Device Dual Abutment have common in design, functions for use, material, and dimensions; therefore, the proposed Dual Abutment is substantially equivalent to the predicated Dual Abutment (K182091).
#### 2. Healing Abutment
| | Subject Device | Reference Device |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Abutment Type | Healing Abutment | Healing Abutment |
| Secondary Product Code | NHA | NHA |
| K Number | K231429 | K203360 |
| Manufacturer | IDIS | Hiossen, Inc. |
| Design | Image: Subject Device Design | Image: Reference Device Design |
| Indications for Use | Healing abutments are used to<br>make a natural soft tissue shape<br>before setting up prosthetics and<br>removing cover screw after<br>osseointegration. | Used to make a natural soft tissue<br>shape before setting up prosthetics<br>and removing cover screw after<br>osseointegration. |
| Material | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) | Pure Titanium Grade 4 (ASTM F67) |
| Diameters (mm) | 4.0, 4.5, 5.0, 6.0, 7.0 | 4.0 ~ 8.0 |
| Lengths | 3.0~7.0 | 3.0~7.0 |
{12}------------------------------------------------
Substantial Equivalent Comparison Discussion: Proposed Healing Abutment and the Reference Device Healing Abutment have common in design, function, indications for use, material, and dimensions; the proposed Healing Abutment is substantially equivalent to the reference device Healing Abutment (K203360).
| | Subject Device | Reference Device |
|---------------------------|-----------------------------------------------------------------|-----------------------------------------------------------------|
| Type | Abutment Screw | Abutment Screw |
| Secondary Product<br>Code | NHA | NHA |
| K Number | K231429 | K182091 |
| Manufacturer | IDIS | Osstem Implant Co., Ltd. |
| Description | Abutment Screw is used to<br>connect an abutment to the fixture | Abutment Screw is used to connect<br>an abutment to the fixture |
| Material | Titanium Alloy<br>(Ti-6A1-4V, ASTM F136) | Titanium Alloy<br>(Ti-6Al-4V, ASTM F136) |
| Diameter (mm) | 2.25, 2.5 | 2, 2.05 |
| Length (mm) | 9.0, 10.2 | 7.5, 9.6 |
### 3. Abutment Screw
#### Substantial Equivalence Discussion for Abutment Screw
The Proposed Abutment Screw and the reference device Abutment Screw share similar indications for use and material composition. Although the diameter and length of the screws differ slightly, the gap range is minimal, measuring less than 1-1.5mm. Consequently, this variance does not give rise to any concerns about the safety and effectiveness of the Proposed Abutment Screw. To sum up, the Proposed Abutment Screw can be deemed substantially equivalent to the reference device Abutment Screw (K182091).
#### Non-Clinical Test Data
The subject device was tested to evaluate its performance as below.
- ISO 7405:2018, Dentistry Evaluation of biocompatibility of medical devices used in . dentistrv
- ISO 10993-1:2018, Biological evaluation of medical devices Part 1: Evaluation and . testing
- ISO 10993-3:2014, Biological evaluation of medical devices Part 3: Tests for ● genotoxicity, carcinogenicity and reproductive toxicity
- . ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-6:2016, Biological evaluation of medical devices Part…