K203750 · Institut Straumann AG · NHA · Oct 9, 2021 · Dental
Device Facts
Record ID
K203750
Device Name
Straumann CARES P-Series CAD/CAM System
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Oct 9, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Straumann CARES P-Series CAD/CAM System is intended to be used for the design and fabrication of patient-specific dental restorations. The dental restorations are intended to restore dental implants via a Ti-base style abutment (Straumann tradename Variobase).
Device Story
System integrates digital dentistry workflow components: intra-oral/extra-oral scanner files, CAD software, CAM software, 3D printer, post-curing unit, and restorative acrylic resin. User scans teeth/models; CARES Visual CAD software designs restoration geometry; CAM module converts design to layer slices for 3D printing. Printer produces restoration using Straumann P pro Crown & Bridge resin. Restorations are cemented to titanium-alloy Variobase abutments, then affixed to dental implants. Used in dental laboratories by technicians. Output allows restoration of edentulous sites; benefits patient via temporary implant-borne prosthetic.
Clinical Evidence
Bench testing only. Dynamic fatigue and static strength testing per FDA guidance for root-form endosseous dental implants/abutments. Biocompatibility data leveraged from reference device K200039. Software verification/validation confirmed design limitations. MRI simulations performed for MR Conditional status.
Technological Characteristics
System components: CAD software, CAM software, 3D printer, post-curing unit. Materials: Ti6Al7Nb (TAN) Variobase; P pro Crown & Bridge acrylic resin. Connectivity: Digital file transfer (STL). Software: Closed CAD system with encrypted design library. Sterilization: Non-sterile, autoclave required prior to implantation.
Indications for Use
Indicated for design and fabrication of single or multiple-unit implant-borne prosthetics for restoration of partially or fully edentulous mandibles and maxillae. Used to design/fabricate 3D printed coping, crown, and bridge restorations cemented onto Straumann Variobase Abutments affixed to Straumann Dental Implant System implants via basal screw. Straumann P pro Crown & Bridge material with Variobase is indicated for temporary (up to 180 days) dental restoration.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
October 9, 2021
Institut Straumann AG % Jennifer Jackson Director of Regulatory Affairs and Ouality Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01801
Re: K203750
Trade/Device Name: Straumann CARES P-Series CAD/CAM System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA. PNP Dated: September 15, 2021 Received: September 16, 2021
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen for Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K203750
Device Name
Straumann CARES P-Series CAD/CAM System
Indications for Use (Describe)
The Straumann CARES P-Series CAD/CAM System is indicated for the design and fabrication of single or multiple-unit implant-borne prosthetics for the restoration of partially or fully edentulous mandibles. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, a restorative acrylic resin, 3D printers, post-curing unit and associated accessories. The system is used to design and fabricate CAD/CAM 3D printed coping, crown and bridge restorations to be cemented onto Straumann® Variobase® Abutments, that are affixed to the endosseous dental implants of the Straumann® Dental Implant System using a basal screw.
The Straumann P pro Crown & Bridge material in combination with the Straumann Variobase is indicated for temporary (up to 180 days) dental restoration of a Straumann dental implant.
Type of Use (Select one or both, as applicable) 図Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Straumann CARES P-Series CAD/CAM System
510(k) Summary
# 510(k) Summary
### Submitter's Contact Information
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315<br>Owner/Operator No.: 9005052 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | On behalf of: |
| | Institut Straumann AG<br>Peter Merian Weg 12<br>CH-4002 Basel, Switzerland |
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs and Quality<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 |
| Prepared By: | Gordon Dodds<br>Manager Design Control QM<br>Etkon GmbH |
| Date of Submission: | September 30, 2021 |
### Name of the Device
| Trade Names: | Straumann CARES P-Series CAD/CAM System |
|------------------------|-----------------------------------------|
| Common Name: | Endosseous dental implant abutment |
| Classification Name: | Endosseous dental implant abutment |
| Regulation Number: | 21 CFR 872.3630 |
| Device Classification: | II |
| Product Code(s): | NHA, PNP |
| Classification Panel: | Dental |
{4}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
510(k) Summary
#### Predicate and Reference Device(s)
Primary Predicate:
- K171649 CARES M-Series CADCAM System .
Reference Devices:
- K200039 P pro Crown & Bridge (DeltaMed GmbH) .
- K120822 Straumann CARES Variobase® Abutment NNC, RN, WN, NC, RC ●
- K173968 CARES Variobase® for Bridge/Bar Cylindrical (Institut Straumann AG) .
- K173379 Variobase® AS (Institut Straumann AG) .
- K190662 MR I Compatibility for Existing Straumann Dental Implant Systems (Institut . Straumann AG)
- . K190082 – Straumann® BLX Variobase® Abutment (Institut Straumann AG)
- . K190040 – Straumann BLX Line Extension - New Abutments (Institut Straumann AG)
- K100247 DETAX Implantlink Semi Classic cement .
#### Device Description
The Straumann CARES P-Series CAD/CAM System is intended for the design and fabrication of dental restorations by dental laboratories by means of a digital workflow for 3D printing.
The Straumann CARES P-Series CAD/CAM System employs optical impression files that document the topoqraphical characteristics of teeth, traditional dental impressions, or stone models. The Straumann CARES Visual CAD software then allows the desired restorations. The CAM software converts the digital restoration design into the sequential slice geometries needed to 3D print a restoration or multiple restorations in a print job. The printing file is transferred to the Straumann P-Series 3D printer; where it is decoded into the defined printing slices and the user initiates the printing operation. The Straumann P-Series 3D printer will then print the designed restoration using the Straumann P Pro Crown & Bridge resin.
The user of a Straumann CARES P-Series CAD/CAM System can design dental implant borne restorations using Straumann Variobases as the connecting interface to the implant. By this, the user will create two-piece abutments with the 3D printed part being the upper part of the implant borne restoration.
{5}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
#### 510(k) Summary
The Straumann Variobases that can be restored via the Straumann CARES P-Series CAD/CAM System include the following models and platforms:
- . RC (Reqular CrossFit®)
- . NC (Narrow CrossFit®)
- . RN (Regular Neck)
- . WN (Wide Neck)
- RB (Regular Base) .
- RB/WB (Regular Base/Wide Base) .
The Straumann® Variobases are pre-manufactured (stock) items, sometimes referred to as "Tibases", made from a titanium-aluminum-niobium (Ti6Al7Nb, TAN) allov. They are the bottom half of the two-piece abutment.
The 3D printed polymeric portion of the dental restoration will have a connecting interface to the coronal aspect of the Variobase, so that it can be cemented to that connecting interface. The 3D printed items feature a screw channel hole so that the cemented two-piece abutment can be affixed via a basal screw to the dental implant. The dental restoration is the top half of the twopiece abutment.
| Type | Implant-borne | Image |
|-------------|-------------------------------------------------|------------------------------------------------|
| Single unit | Using previously cleared<br>Variobase Abutments | Image: dental implant<br>Image: dental implant |
| Multi-unit | Using previously cleared<br>Variobase Abutments | Image: dental implant in mouth |
The different types of restorations are described in Table 1.
Table 1 – Types of restorations
{6}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
#### 510(k) Summary
The digital workflow using the Straumann CARES P-Series CAD/CAM System includes the use of the following products:
#### Dental Scanner(s)
The Straumann CARES P-Series CAD/CAM system can accept files generated using the following devices (note that these are not subject devices to this submission):
- . Dental Wings Intra-Oral Scanner, DWIO
- . Dental Wings 3-Series & 7-Series desktop scanners (extra-oral)
The dental scanner takes optical impressions that record the topographical characteristics of teeth, traditional dental impressions, or stone models. This includes the location and orientation of dental implants or abutments when a Scanbody is employed during the scan.
### CARES Visual CAD Software
The CARES Visual software is a dental CAD application that allows the user to digitally design dental restorations, based on information that was acquired by a dental scanner. As a result of the design process and the indication and the Straumann P pro Crown & Bridge design dimension limits, a three-dimensional geometry is created. The use of Straumann provided digital device models (interface library) assures the accuracy of the interfaces between the designed restoration and the Variobase® being restored.
#### CAM Module
The CAM interface module converts the digital three-dimensional restoration geometry into the layer geometries for the 3D printer.
#### Straumann P-Series 3D printer
The 3D printer receives the CAM file from the CAM software. The user loads the CAM file into the Straumann P-Series 3D printer and initiates the 3D printing process.
#### Centrifuge
A centrifuging process is carried out to remove non polymerized resin from the surface of the printed dental restoration, without the use of a solvent, before the curing process.
{7}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
510(k) Summary
#### Post Curing
After the centrifyge process, the printed restorations need to undergo a final light-induced polymerization process, so that the device achieves its tested physical and mechanical characteristics.
#### 3D printing resin
The Straumann P pro Crown & Bridge resin is the raw material used to create the 3D printed restorations. The material is available in three different color shades, A1, A2 and A3. The Straumann P pro Crown & Bridge material has received 510(k) clearance through K200039, as a temporary restoration for crowns and bridges (CFR 872.3770, Product Code EBG). The Straumann P pro Crown & Bridge resin device in this submission, while of the same material composition, is for use as an endosseous dental implant abutment (CFR 872.3630, Product Code NHA). The P pro Crown & Bridge design specifications for crowns and bridges are:
| Restoration type | Parameter | Value |
|------------------|-----------------------------------------------------------------------------------|----------------|
| Crown | Minimum wall thickness | 1.0 mm |
| | Maximum angulation | 30° |
| | Minimum post-height of Variobase<br>combined with P pro Crown& Bridge<br>crown | 4.6 mm |
| | Minimum diameter | 3.8 mm |
| Bridge | Minimum wall thickness | 1.0 mm |
| | Maximum angulation | 30° |
| | Minimum post-height of Variobase<br>combined with P pro Crown& Bridge<br>abutment | 4.6 mm |
| | Minimum diameter | 3.8 mm |
| | Maximum number of units | 6 |
| | Maximum number of pontics | 2 |
| | Minimum cross-section of connector<br>anterior/posterior | 12 mm²/ 14 mm² |
Table 2 – Design parameters for 3D printed restorations from Straumann P pro Crown & Bridge
{8}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
510(k) Summary
#### Intended Use
The Straumann CARES P-Series CAD/CAM System is intended to be used for the design and fabrication of patient-specific dental restorations. The dental restorations are intended to restore dental implants via a Ti-base style abutment (Straumann tradename Variobase).
#### Indications for Use
The Straumann CARES P-Series CAD/CAM System is indicated for the design and fabrication of single or multiple-unit implant-borne prosthetics for the restoration of partially or fully edentulous mandibles and maxillae. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners or Extra-Oral Scanners, CAD software, CAM software, a restorative acrylic resin, 3D printers, post-curing unit and associated accessories. The system is used to design and fabricate CAD/CAM 3D printed coping, crown and bridge restorations to be cemented onto Straumann® Variobase® Abutments that are affixed to the endosseous dental implants of the Straumann® Dental Implant System using a basal screw.
The Straumann P pro Crown & Bridge material in combination with the Straumann Variobase is indicated for temporary (up to 180 days) dental restoration of a Straumann dental implant.
#### Technoloqical Characteristics
The technological characteristics of the subject devices are compared to the primary predicate device in Table 3.
{9}------------------------------------------------
# Straumann CARES P-Series CAD/CAM System
### 510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Subject Device | K171649 |
| Indications for<br>Use | Straumann CARES P-Series CAD/CAM System<br>The Straumann CARES P-Series CAD/CAM<br>System is indicated for the design and fabrication of<br>single or multiple-unit implant-borne prosthetics for<br>the restoration of partially or fully edentulous<br>mandibles and maxillae. The system integrates<br>multiple components of the digital dentistry<br>workflow: scan files from Intra-Oral Scanners or<br>Extra-Oral Scanners, CAD software, CAM software,<br>a restorative acrylic resin, 3D printers, post-curing<br>unit and associated accessories. The system is<br>used to design and fabricate CAD/CAM 3D printed<br>coping, crown and bridge restorations to be<br>cemented onto Straumann® Variobase® Abutments<br>that are affixed to the endosseous dental implants<br>of the Straumann® Dental Implant System using a<br>basal screw.<br>The Straumann P pro Crown & Bridge material in<br>combination with the Straumann Variobase is<br>indicated for temporary (up to 180 days) dental<br>restoration of a Straumann dental implant. | Straumann CARES M-Series CAD/CAM System<br>The Straumann CARES M-Series CAD/CAM System<br>is indicated for the design and fabrication of single or<br>multiple-unit implant-borne prosthetics for the<br>restoration of partially or fully edentulous mandibles<br>and maxillae. The system integrates multiple<br>components of the digital dentistry workflow: scan<br>files from Intra-Oral Scanners or Extra-Oral<br>Scanners, CAD software, CAM software, restoration<br>material blanks, milling machines and associated<br>tooling and accessories. The system is used to<br>design and fabricate CAD/CAM milled coping, crown<br>and bridge restorations to be cemented onto<br>Straumann® Variobase® Abutments, as well as<br>milled abutments to be affixed to the endosseous<br>dental implants of the Straumann® Dental Implant<br>System using a basal screw. |
| Source of Input<br>Files | Intra-Oral Scanner<br>Bench-top Scanners | Intra-Oral Scanner<br>Bench-top Scanners |
| Bench Scanner<br>Control | Yes | Yes |
| Implant<br>Detection | Yes, using Scanbodies | Yes, using Scanbodies |
| Design<br>Environment | Straumann CARES Visual:<br>Closed CAD System facilitating the design of<br>restorations used in conjunction with the devices of<br>the Straumann Dental Implant System (SDIS). | Straumann CARES Visual:<br>Closed CAD System facilitating the design of<br>restorations used in conjunction with the devices of<br>the Straumann Dental Implant System (SDIS). |
| Restoration<br>Types<br>Supported | Single unit restorations and bridges for Straumann<br>Variobases | Single unit restorations and bridges for Straumann<br>Variobases |
| Supported<br>Variobases | Straumann Variobase Abutments for NC, RC, RN<br>and WN implant-to-abutment interfaces (K120822).<br>Straumann Variobase Abutments for Bridges and<br>Bars for NC, RC, RN and WN implant-to-abutment<br>interfaces (K173968)<br>Copings for Straumann Screw-Retained Abutments<br>(K133421).<br>Straumann Variobase AS (K173379)<br>Straumann Variobase BLX (K190082)<br>Straumann BLX Varibase for Bar & Bridge<br>(K190040) | Straumann Variobase Abutments for NC, RC, NNC,<br>RN and WN implant-to-abutment interfaces<br>(K120822).<br>Straumann Variobase Abutments for Bridges and<br>Bars for NC, RC, NNC, RN and WN implant-to-<br>abutment interfaces (K151157)<br>Copings for Straumann Screw-Retained Abutments<br>(K133421). |
| Supported<br>Restorative<br>Materials | Straumann P pro Crown & Bridge (K200039) | Zirconia and glass ceramic milling blanks |
| Restoration<br>Sizes | Single crown up to 6-unit bridge | Single crown up to 16-unit bridge |
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE |
| K Number | Subject Device | K171649 |
| Interface to<br>Variobase | 3D printed by the system | Milled by the system |
| CAD to Local<br>Production<br>Transfer | Straumann CARES Visual output STL is converted<br>to printing layer format using Netfabb software and<br>transferred to printer | Straumann CARES Visual output geometry is<br>converted to a milling tool path sequence using<br>Moduleworks software and transferred to mill |
| Supported<br>manufacturing<br>equipment | Straumann CARES P-Series 3D printers | Straumann CARES M-Series mills |
| Fabrication<br>Workflow | 3D printing (additive) | Milling (subtractive) |
| Compatible<br>Implants | Straumann Bone Level implants having the NC,<br>RC, RB and RB/WB implant-to-abutment interface<br>geometries.<br>Straumann Tissue Level implants having the RN,<br>and WN implant-to-abutment interface geometries. | Straumann Bone Level implants having the NC and<br>RC implant-to-abutment interface geometries.<br>Straumann Tissue Level implants having the NNC,<br>RN, and WN implant-to-abutment interface<br>geometries. |
| Abutment<br>Platform<br>Diameters | 3.8 mm - 7.0 mm | 3.8 mm - 7.0 mm |
| Material of<br>Variobase | Titanium alloy<br>(Ti-6Al-7Nb, TAN) | Titanium alloy<br>(Ti-6Al-7Nb, TAN) |
| Material of<br>restoration | Acrylic - P Pro Crown & Bridge | ZrO2 |
| Lifetime of<br>Restoration | Temporary (up to 180 days) acrylic material | Permanent zirconia materials |
| Construction | Two-piece abutment base with bonded restoration<br>produced via the Straumann CAD/CAM System<br>digital workflow with a 3D printed crown from P pro<br>Crown & Bridge material. | Two-piece abutment base with bonded restoration<br>produced via the Straumann CAD/CAM System<br>digital workflow via local milling by the Straumann M-<br>Series milling machine. |
| Maximum<br>Angulation | 30° controlled in design software | 30° controlled in design software |
| Minimum post-<br>height | 4.6 mm | 4.55 mm Zolid SHT<br>4.2 mm Zolid HT |
| Minimum wall<br>thickness | 1.0 mm | 0.95 mm Zolid SHT<br>0.6 mm Zolid HT |
| Maximum<br>number of<br>pontics | 2 pontics | 1 pontic - Zolid SHT<br>2 pontics - Zolid HT |
| Maximum<br>number of units | 6 units | 3 units - Zolid SHT<br>16 units - Zolid HT |
| Sterility | Provided non-sterile - terminally sterilized via<br>autoclave prior to implantation. | Provided non-sterile - terminally sterilized via<br>autoclave prior to implantation. |
{10}------------------------------------------------
# Straumann CARES P-Series CAD/CAM System
### 510(k) Summary
#### Table 3 – Comparison of subject device versus primary predicate device
{11}------------------------------------------------
### Straumann CARES P-Series CAD/CAM System
510(k) Summary
#### Feature Comparison:
Indications for Use - The differences between the subject and predicate device are due to the different manufacturing processes and materials. The subject device features an additive 3D printing CAD/CAM procedure while the primary predicate device features a subtractive milling CAD/CAM process. The proposed device uses a 510(k) cleared acrylic resin, Straumann P pro Crown & Bridge (K200039) for temporary use, while the primary predicate uses zirconia milling disks or glass ceramic blocks for permanent use. The intended use for both, production of a crown or bridge on Ti-bases, is substantially equivalent as an endosseous dental implant abutment (CFR 872.3630, Product Code NHA).
Design Specifications – There are differences in the minimum abutment post height and minimum wall thickness between the subject and predicate devices. Fatique testing of the subject and predicate devices has assured that the subject and predicate device values for these are sufficient for their intended use. The maximum number of units and maximum number of pontics is specified by the material manufacturer and depends on the material. The values used for the subject and predicate devices correspond to the values specified by the respective material manufacturer. The Design Specifications are thus substantially equivalent.
#### Performance Testing
Dynamic fatigue and static strength tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and demonstrated that the Straumann Variobases restored with P pro Crown & Bridge constructs are equivalent to the primary predicate and reference devices.
The subject device materials are identical to reference device materials. Biocompatibility tests of the P pro Crown & Bridge material in K200039 (including cytotoxicity, sensitization, irritation) were passed successfully. Therefore, no new issues regarding biocompatibility were raised.
The sterilization process for the Straumann Variobase® was validated according to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015".
Software verification and validation testing was provided for the subject abutment design library to demonstrate effective use with the Straumann CARES Visual CAD software. Software
{12}------------------------------------------------
## Straumann CARES P-Series CAD/CAM System
#### 510(k) Summary
verification and validation testing was conducted to demonstrate that the restrictions prevent design of the top half component of the two-piece abutment outside of the allowable design limitations, including screenshots under user verification testing. In addition, the encrypted abutment design library was validated to demonstrate that the established design limitations and specifications are locked and cannot be modified within the abutment design library. Non-clinical testing and MRI simulations were performed to evaluate the dental implant system offered by Institut Straumann AG. Non-clinical testing demonstrates that these products are MR Conditional.
Verification of the 3D printer system showed that the design input matched the output, that the process is r…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.