K203355 · Institut Straumann AG · NHA · Feb 12, 2021 · Dental
Device Facts
Record ID
K203355
Device Name
Straumann TLX Novaloc and Cementable Abutments
Applicant
Institut Straumann AG
Product Code
NHA · Dental
Decision Date
Feb 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TLX Novaloc Abutments: The Straumann® Retentive System is indicated for the attachment of full or partial dentures on Straumann dental implants (NT, RT, and WT). TLX Cementable Abutments: Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Final abutments may be placed into occlusion for implants with sufficient primary stability and with appropriate occlusal loading to restore chewing function or for implants that are fully osseointegrated.
Device Story
Straumann TLX Novaloc and Cementable Abutments are prosthetic components for dental implants. Novaloc abutments support full-arch detachable over-dentures via snap-on fixtures; cementable abutments support crowns and bridges. Both types connect to Straumann TLX implants (NT, RT, WT platforms) using a TorcFit interface. Manufactured from titanium alloys (Ti-6Al-4V or Ti-6Al-7Nb), they are provided non-sterile for end-user moist heat sterilization. Clinicians select appropriate heights and angulations (0° or 15°) to restore chewing function. The device facilitates prosthetic rehabilitation by providing a stable interface between the endosseous implant and the final dental restoration.
Clinical Evidence
Bench testing only. Dynamic fatigue and static strength testing conducted per ISO 14801 and FDA guidance for endosseous dental abutments. Biocompatibility evaluated per ISO 10993-1. Sterilization validated per ISO 17665-1/2.
Technological Characteristics
Materials: TAV (Ti-6Al-4V) or TAN (Ti-6Al-7Nb). Interface: TorcFit. Surface: TiN coating (Novaloc coronal) and laser marking. Sterilization: Moist heat (steam) end-user. Form factor: Straight and angulated (15°) abutments for NT, RT, WT platforms.
Indications for Use
Indicated for patients requiring full or partial denture attachment or prosthetic rehabilitation (crowns/bridges) on Straumann dental implants (NT, RT, WT).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Straumann® Retentive System Novaloc TiN Abutments (K193046)
MRI Compatibility for Existing Straumann Dental Implant Systems (K190662)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA, LLC 60 Minuteman Road Andover, Massachusetts 01801
Re: K203355
Trade/Device Name: Straumann TLX Novaloc and Cementable Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: November 12, 2020 Received: November 16, 2020
Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew Steen Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
#### K203355
Device Name:
Straumann TLX Novaloc and Cementable Abutments
Indications for Use (Describe)
#### TLX Novaloc Abutments
The Straumann® Retentive System is indicated for the attachment of full or partial dentures on Straumann dental implants (NT, RT, and WT).
#### TLX Cementable Abutments
Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Final abutments may be placed into occlusion for implants with sufficient primary stability and with appropriate occlusal loading to restore chewing function or for implants that are fully osseointegrated.
Type of Use (Select one or both, as applicable) �Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED
#### This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
## Straumann TLX Novaloc and Cementable Abutments
510(k) Summary
#### 5 510(k) Summary
### 5.1 Submitter's Contact Information
| Submitter: | Straumann USA, LLC<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No.: 1222315 Owner/Operator No.: 9005052 |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| | On the behalf of: |
| | Institut Straumann AG<br>Peter Merian-Weg 12<br>4052 Basel, Switzerland |
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 |
| Prepared By &<br>Alternate Contact: | Laure Kuhner<br>Regulatory Affairs Specialist<br>Institut Straumann AG<br>Phone Number: +41 61 965 1389 |
Date of Submission: February 4, 2021
### 5.2 Name of the Device
| Trade Names: | Straumann TLX Novaloc and Cementable Abutments |
|------------------------|------------------------------------------------|
| Common Name: | Endosseous Dental Implant Abutment |
| Classification Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | 21 CFR 872.3630 |
| Device Classification: | Class II |
| Product Code(s): | NHA |
| Classification Panel: | Dental devices |
| Proprietary Name: | Straumann TLX Novaloc and Cementable Abutments |
{4}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
#### 510(k) Summary
#### 5.3 Predicate Device(s)
Primary Predicate:
- K193046 Straumann® Retentive System Novaloc TiN Abutments .
Reference Devices:
- K113283 Narrow Neck CrossFit (NNC) Cementable Abutments .
- K200586 Straumann TLX Implant System .
- . K190662 - MRI Compatibility for Existing Straumann Dental Implant Systems
#### 5.4 Device Description
#### Straumann TLX Novaloc Abutments
The subject TLX Novaloc Abutments are intended to be placed onto Straumann TLX implants to provide support for full arch detachable restorations (over-denture). The coronal portion of the subject abutments is similar to the primary predicate devices, whereas the apical portion of the abutments has the appropriate implant-to-abutment interface geometry for each of the platforms of the Straumann TLX Implant System (NT, RT and WT). The Novaloc Abutments are manufactured from TAV (Ti-6Al-4V, Titanium-Aluminum-Vanadium). The restoration is connected to the Novaloc Abutment through a snap-on fixture provided by a negative shape of Novaloc snapon fixture embedded into the final restoration. The snap-on feature is TiN coated. The subject TLX Novaloc Abutments are provided in straight and angulated versions, in different heights. The TLX Novaloc Abutments are provided non-sterile with instructions for end user sterilization.
#### Straumann TLX Cementable Abutments
The subject TLX Cementable Abutments are intended to provide support for prosthetic reconstructions such as crowns and bridges and are used with cemented restorations. They are made of titanium alloy (Ti-6Al-7Nb or TAN) and are attached to the implant with a basal screw. To allow for more flexibility they are offered in a straight and in an angulated version for each of the three platforms (NT, RT and WT). The TLX Cementable Abutments are provided non-sterile with instructions for end user sterilization.
{5}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
510(k) Summary
#### 5.5 Indications for Use
#### Straumann TLX Novaloc Abutments
The Straumann® Retentive System is indicated for the attachment of full or partial dentures on Straumann dental implants (NT, RT, and WT).
#### Straumann TLX Cementable Abutments
Prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. Final abutments may be placed into occlusion for implants with sufficient primary stability and with appropriate occlusal loading to restore chewing function or for implants that are fully osseointegrated.
#### 5.6 Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in Table 1 and Table 2.
{6}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
### 510(k) Summary
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K203355 | K193046<br>Straumann® Retentive System -<br>Novaloc TiN Abutments | K200586<br>Straumann TLX Implant System<br>(TLX Variobase for Crown) |
| Indications for<br>Use | The Straumann® Retentive System is<br>indicated for the attachment of full or<br>partial dentures on Straumann dental<br>implants (NT, RT, and WT). | The Straumann® Retentive System is<br>indicated for the attachment of full or<br>partial dentures on Straumann dental<br>implants. | Straumann® Variobase® prosthetic components directly<br>connected to the endosseous dental implant are intended for<br>use as an aid in prosthetic rehabilitations. The prosthetic<br>restoration (crowns) can be cemented onto the Straumann®<br>Variobase® prosthetic components. A temporary restoration<br>can be used prior to the insertion of the final components to<br>maintain, stabilize and shape the soft tissue during the<br>healing phase; they must be placed out of occlusion. Final<br>abutments and restorations may be placed into occlusion<br>when the implant is fully osseointegrated. All digitally<br>designed copings and/or crowns for use with the Straumann®<br>Variobase® Abutment system are intended to be sent to<br>Straumann for manufacture at a validated milling center. |
| Material | TAV (Ti-6Al-4V) | TAV (Ti-6Al-4V) | TAN (Ti-6Al-7Nb) |
| Surface<br>Treatment | Coronal portion: TiN coating<br>Apical portion: Laser marking | Coronal portion: TiN coating<br>Apical portion: Laser marking | Laser marking |
| Implant to<br>Abutment<br>Connection | TorcFit | CrossFit and synOcta | TorcFit |
| Prosthetic<br>platform | NT, RT and WT | NC, RC, RN and WN | NT, RT and WT |
| Gingival<br>Height | Straight: 1, 2, 3, 4, 5, and 6 mm<br>Angled: 2, 3, 4, 5, and 6 mm | Straight: 1, 2, 3, 4, 5, and 6 mm<br>Angled: 2, 3, 4, 5, and 6 mm | 5.5 mm, 6 mm and 6,5 mm |
| Angulation | 0° and 15° | 0° and 15° | 0° |
| Sterilization<br>method | Moist heat end user sterilization | Moist heat end user sterilization | Moist heat end user sterilization |
| FEATURE | PROPOSED DEVICE | REFERENCE DEVICE | REFERENCE DEVICE |
| K Number | K203355 | K113283<br>Narrow Neck CrossFit (NNC) Cementable<br>Abutments | K200586<br>Straumann TLX Implant System |
| Indications for<br>Use | Prosthetic components directly connected<br>to the endosseous dental implant are<br>intended for use as an aid in prosthetic<br>rehabilitations. Final abutments may be<br>placed into occlusion for implants with<br>sufficient primary stability and with<br>appropriate occlusal loading to restore<br>chewing function or for implants that are<br>fully osseointegrated. | Abutments are used in connection with the<br>prosthetic restoration of Straumann dental<br>implants. Abutments are intended to be<br>placed into dental implants to provide support<br>for prosthetic reconstructions such as crowns<br>and bridges.<br><br>Narrow Neck CrossFit Cementable<br>Abutments are indicated for cement-retained<br>single tooth and bridge restorations. | TLX Dental Implant:<br>Straumann TLX Implants are suitable for endosteal<br>implantation in the upper and lower jaws and for the<br>functional and esthetic oral rehabilitation of edentulous and<br>partially edentulous patients. TLX Implants can be placed<br>with immediate function on single-tooth and multi-unit<br>restorations when good primary stability is achieved and<br>with appropriate occlusal loading to restore chewing<br>function. The prosthetic restorations are connected to the<br>implants through the corresponding abutment<br>components.<br>TLX Variobase for Crown:<br>Straumann Variobase prosthetic components directly<br>connected to the endosseous dental implant are intended<br>for use as an aid in prosthetic rehabilitations. The<br>prosthetic restoration (crowns) can be cemented onto the<br>Straumann Variobase prosthetic components. A temporary<br>restoration can be used prior to the insertion of the final<br>components to maintain, stabilize and shape the soft tissue<br>during the healing phase; they must be placed out of<br>occlusion. Final abutments and restorations may be placed<br>into occlusion when the implant is fully osseointegrated. All<br>digitally designed copings and/or crowns for use with the<br>Straumann Variobase Abutment system are intended to be<br>sent to Straumann for manufacture at a validated milling<br>center. |
| Material | TAN (Ti-6Al-7Nb) | Titanium, Grade 4 | TAN (Ti-6Al-7Nb) |
| Surface<br>Treatment | Laser marking | Laser marking | Laser marking |
| Implant to<br>Abutment<br>Connection | TorcFit | CrossFit | TorcFit |
| Prosthetic<br>platform | NT, RT and WT | NNC | NT, RT and WT |
| FEATURE | PROPOSED DEVICE | REFERENCE DEVICE | REFERENCE DEVICE |
| K Number | K203355 | K113283<br>Narrow Neck CrossFit (NNC) Cementable<br>Abutments | K200586<br>Straumann TLX Implant System |
| Chimney<br>height | Straight: 5,7 mm and 6 mm<br>Angled: 6 mm | Straight: 5,7 mm<br>Angled: 5,9 mm | 5.5 mm, 6 mm and 6,5 mm |
| Angulation | 0° and 15° | 0° and 15° | 0° |
| Sterilization<br>method | Moist heat end user sterilization | Moist heat end user sterilization | Moist heat end user sterilization |
Table 1 – Comparison of subject TLX Novaloc Abutments versus primary predicate and reference devices
{7}------------------------------------------------
## Straumann TLX Novaloc and Cementable Abutments
### 510(k) Summary
{8}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
510(k) Summary
Table 2 – Comparison of subject TLX Cementable Abutments versus reference devices
{9}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
510(k) Summary
#### 5.7 Performance Testing
#### Bench Testing 5.7.1
Dynamic fatigue and static strength tests were conducted according to the FDA guidance document "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" and ISO 14801, "Dentistry - Implants - Dynamic fatigue test for endosseous dental implants" and demonstrated that the Straumann TLX Novaloc and Cementable Abutments are equivalent to the primary predicate and reference devices.
#### 5.7.2 Biocompatibility Testing
Biological assessment has been performed according to ISO 10993-1:2009 "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" and to the FDA Guidance document "Use of International Standard ISO 10993-
1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff", Document issued on: June 16, 2016" for each of the subject devices.
The subject devices have equivalent nature of body contact duration, material formulation and sterilization methods compared to the primary predicate and reference devices, therefore, no new issues regarding biocompatibility were raised.
#### 5.7.3 Sterilization Validation
The subject Straumann TLX Novaloc and Cementable Abutments are provided non-sterile and need to be sterilized by moist heat (steam) by the end-user. The recommended sterilization method has been validated according to ISO 17665-1 and ISO 17665-2 and to applicable recommendations in the FDA guidance document "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015". There are no changes to the sterilization procedures or processes from those of the primary predicate and reference devices.
{10}------------------------------------------------
### Straumann TLX Novaloc and Cementable Abutments
510(k) Summary
### 5.8 Conclusion
The documentation submitted in this premarket notification demonstrates the Straumann TLX Novaloc and Cementable Abutments are substantially equivalent to the primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.