K200586 · Institut Straumann AG · DZE · Aug 20, 2020 · Dental
Device Facts
Record ID
K200586
Device Name
Straumann TLX Implant System
Applicant
Institut Straumann AG
Product Code
DZE · Dental
Decision Date
Aug 20, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Straumann TLX Implants are suitable for endosteal implantation in the upper and lower jaws and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. TLX Implants can be placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components. Straumann Closure Caps and Healing Caps are indicated to be placed in the patient's mouth at the end of the implant placement to protect the inner configuration of the implant and to form, maintain and stabilize the soft tissue during the healing process. Closure caps and healing caps should be used only with suitable implant connections. They have a maximum duration of usage of 6 months. TLX Temporary Abutments can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. TLX Temporary Abutments have a maximum duration of usage of 180 days. Straumann Variobase prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (crowns) can be cemented onto the Straumann Variobase prosthetic components. A temporary restoration can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center. The Straumann CARES Abutments TAN are indicated for single tooth replacement and multiple tooth restorations. The prosthetic restoration can be cemented. CARES Screw-retained Bridges and Bars (SRBB) are indicated for use as bars and bridges that attach to implants to provide support for prosthetic reconstructions such as bridges and overdentures. The final processed products have the purpose of restoring function. Straumann CARES Screw-retained Bridges and Bars are indicated for Screw-retained restorations. Straumann CARES Screw-retained Bridges and Bars are designed to interface with the Bone Level (BL), Tissue Level (TL), BLX implants and TLX implants of the Straumann Dental Implant System (SDIS).
Device Story
System comprises fully tapered endosseous dental implants (Roxolid material, SLActive surface) and prosthetic components (closure/healing caps, temporary abutments, Variobase, CARES abutments, screw-retained bridges/bars). Implants placed in clinic by dental professionals; provide foundation for prosthetic restorations. Digital workflow utilizes Straumann CARES Visual software for CAD design of copings/crowns/bridges, manufactured at validated milling centers. Components connect via TorcFit conical interface. Temporary components stabilize soft tissue during healing; final restorations provide chewing function. Benefits include immediate function capability and customized prosthetic fit.
Clinical Evidence
Bench testing only. Dynamic fatigue testing conducted per FDA guidance in saline and air. Surface area and pull-out force testing performed on Ø3.75 x 6 mm implants. Biocompatibility assessed per ISO 10993-1. Sterilization validated per ISO 11137 and ISO 17665.
Indicated for edentulous and partially edentulous patients requiring endosteal dental implants in upper or lower jaws for functional and esthetic oral rehabilitation. Supports single-tooth and multi-unit restorations, including immediate function when primary stability is achieved. Includes components for soft tissue healing, temporary abutments, and prosthetic support (crowns, bridges, bars).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2020
Institut Straumann AG % Jennifer Jackson Director, Regulatory Affairs Straumann USA. LLC 60 Minuteman Road Andover, Massachusetts 01810
Re: K200586
Trade/Device Name: Straumann TLX Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: July 21, 2020 Received: July 22, 2020
### Dear Jennifer Jackson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K200586
#### Device Name Straumann TLX Implant System
#### Indications for Use (Describe)
#### TLX Dental Implant:
Straumann TLX Implants are suitable for endosteal implantation in the upper and for the functional and esthetic oral rehabilitation of edentulous patients. TLX Implants can be placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.
#### TLX Closure Caps and Healing Caps:
Straumann Closure Caps and Healing Caps are indicated to be placed in the patient's mouth at the end of the implant placement to protect the inner configuration of the implant and stabilize the soft tissue during the healing process. Closure caps and healing caps should be used only with suitable implant connections. They have a maximum duration of usage of 6 months.
#### TLX Temporary Abutment:
TLX Temporary Abutments can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. TLX Temporary Abutments have a maximum duration of usage of 180 days.
#### TLX Variobase for Crown:
Straumann Variobase prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (crowns) can be cemented onto the Straumann Variobase prosthetic components. A temporary restoration can be used prior to the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseontegrated. All digitally designed copings and/or crowns for use with the Straumann Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
#### TLX CARES Abutment TAN:
The Straumann CARES Abutments TAN are indicated for single tooth replacement and multiple tooth restorations. The prosthetic restoration can be cemented.
#### TLX Screw-retained Bridges and Bars:
CARES Screw-retained Bridges and Bars (SRBB) are indicated for use as bars and bridges that attach to implants to provide support for prosthetic reconstructions such as bridges and overdentures. The final processed products have the purpose of restoring chewing function. Straumann CARES Screw-retained Bridges and Bars are indicated for Screw-retained restorations. Straumann CARES Screw-retained Bridges and Bars are designed to interface with the Bone Level (BL), Tissue Level (TL), BLX implants of the Straumann Dental Implant System (SDIS).
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
Straumann® TLX Implant System
Substantial Equivalence Discussion
#### 510(k) Summary 5
#### 5.1 Submitter's Contact Information
| Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810 | | |
|-------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------|---------|
| Registration No.: | 1222315 | Owner/Operator No.: | 9005052 |
| | On the behalf of:<br>Institut Straumann AG<br>Peter Merian weg 12<br>4052 Basel, Switzerland | | |
| Contact Person: | Jennifer M. Jackson, MS<br>Director of Regulatory Affairs<br>Phone Number: +1-978-747-2509<br>Fax Number: +1-978-747-0023 | | |
| Prepared By: | Renate Reiss<br>Regulatory Affairs and Compliance Manager<br>Institut Straumann AG<br>Phone number: +41 61 965 1260 | | |
| Date Prepared: | August 19, 2020 | | |
#### 5.2 Name of the Device
| Trade Names: | Straumann TLX Implant System |
|------------------------|------------------------------------------------------------|
| Common Name: | Endosseous dental implant |
| Classification Name: | Endosseous dental implant |
| Regulation Number: | §872.3640 |
| Device Classification: | II |
| Product Code(s): | Primary product code – DZE<br>Secondary product code – NHA |
{6}------------------------------------------------
### Straumann® TLX Implant System
Substantial Equivalence Discussion
#### Predicate Device(s) 5.3
Primary Predicate:
- K173961 Straumann BLX Implant System .
Reference Devices:
- K181703 Straumann® BLX Line Extension .
- K171784 Straumann Dental implant system .
- . K190082 – Straumann BLX Variobase Abutment
- K190040 BLX Line Extension New Abutments .
- K132844 Straumann CARES Bone Level Screw-retained Bars, STRAUMANN CARES . Bone Level Screw-retained Bridges
- K112280 Straumann CARES Screw-retained Bridge Titanium, STRAUMANN CARES . Dolder Bar Titanium
- K101465 Straumann CARES Bridge; STRAUMANN CARES Dolder Bar .
- . K190097 – Straumann CARES Screw-retained Bridges and Bars
- K190662 MRI Compatibility for Existing Straumann Dental Implant Systems ●
- K150938 Straumann Dental Implant System Roxolid SLA Implants .
- K163194 Neodent Implant System GM Line .
- . K172798 – Straumann CARES Abutments CoCr
#### Device Description 5.4
### TLX Dental Implant:
The TLX Dental Implant are fully tapered implants manufactured utilizing the Roxolid material and are finished with SLActive surface. The connection is identified as conical fitting with Torx style engaging feature. TLX implants are presented with 3 prosthetic platforms as listed below:
- . NT (Narrow TorcFit)
- RT (Regular TorcFit) .
- WT (Wide TorcFit) .
The internal connection is identical for all prosthetic platforms, implant diameters, and implant lengths.
{7}------------------------------------------------
### Straumann® TLX Implant System
Substantial Equivalence Discussion
#### TLX Closure Caps and Healing Caps:
The closure caps are screwed into the implant to protect the inner configuration and shoulder of the implant during the healing phase in cases of submucosal) healing protocols. The healing caps are screwed into the implant to protect the inner configuration of the implant in cases of transmucosal healing protocols. They are placed out of occlusion and do not support a prosthetic restoration. Closure caps and healing caps are used during the healing phase only.
The TLX Closure Caps and Healing Caps are manufactured from Titanium Grade 4 and are laser marked with NT, RT or WT for identification purposes. They are provided sterile and are available in different heights and diameters.
#### TLX Temporary Abutment
TLX Temporary Abutments can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion.
TLX Temporary Abutments have a maximum duration of usage of 180 days.
The TLX Temporary Abutments are manufactured from TAN and consist of a coronal section, a platform and a connection part. The abutments are provided non-sterile with instructions for end user sterilization. The Temporary Abutments are seated in the implant with a basal screw which is also manufactured from TAN and are laser marked with NT, RT or WT for identification purposes. The Basal screw is delivered with the abutment. The TLX Temporary Abutments are available for Crown and Bridge/Bar restorations.
### TLX Variobase for Crown
The TLX Variobase for Crown incorporates the implant to abutment connection (TorcFit) and is available for each of the three implant diameter platforms (NT, RT & WT) with a different abutment chimney height and prosthetic platform diameter. The TLX Variobase Abutments for Crown are titanium bases to be used as the lower part of two-piece abutments. The upper part of the two-piece abutment is a CAD/CAM designed and manufactured restoration. These components, which once assembled together and placed with the corresponding basal screw, constitute the final medical device.
TLX Variobase for Crown will be marketed as stand-alone component or through the CARES® X-Stream workflow. In the latter the prosthetic restoration is designed though CARES® Visual
{8}------------------------------------------------
### Straumann® TLX Implant System
Substantial Equivalence Discussion
software (Digital CARES workflow) and manufactured in a validated Straumann milling center. The prosthetic restoration is then shipped together with the TLX Variobase for Crown and the Basal screw.
All digitally designed copings and/or crowns for use with the TLX Variobase for Crown are intended to be sent to Straumann for manufacture at a validated milling center.
The TLX Variobase for Crown is provided non-sterile with instructions for end user steam sterilization.
### Prosthetic Restoration Design and Materials
The following materials are available within the digital workflow for the manufacturing of prosthetic restorations:
Final restorations:
- . zerion® LT
- . zerion® ML
- zerion® UTML ●
- IPS e.max CAD .
- coron® .
Temporary restoration:
- polycon® ae .
### TLX CARES Abutment TAN
The TLX CARES Abutments TAN are packed and delivered with the Basel screw. Both are manufactured from TAN (titanium-aluminum-niobium alloy/ Ti-6AI-7Nb).
TLX CARES Abutments TAN are intended to be placed into Straumann implants to provide support for prosthetic reconstructions such as crowns and bridges.
The final abutment, fabricated from a pre-milled blank, is designed to allow for individual customization regarding function and esthetics.
The pre-milled blank incorporates the pre-milled implant to abutment connection (TorcFit) and has a cylindrical body with enough material volume to create a wide range of geometries for the final abutment.
The TLX CARES Abutments TAN is available for each of the three implant platforms.
{9}------------------------------------------------
### Straumann® TLX Implant System
Substantial Equivalence Discussion
Inside the abutment, a screw channel provides access to the internal thread feature of the implant, such that the component can be firmly attached while providing fit between screw, abutment and implant.
The design of the customized abutment must be made using the validated Straumann CARES Visual software (Digital CARES workflow).
Finally, the design file is transferred digitally to a Straumann validated milling center.
The TLX CARES Abutment TAN is provided non-sterile with instructions for end user steam sterilization.
### TLX Screw-retained Bridges and Bars:
The Straumann CARES Screw-retained Bridges and Bars, also referred to as SRBB are packed and delivered with the corresponding basal screws.
SRBB devices are manufacture from either
- . Titanium Grade 4 or
- . Cobalt chromium (also referred to as CoCr (or coron).
- . the Basal Screw is manufactured from TAN (titanium-aluminum-niobium alloy/ Ti-6Al-7Nb).
SRBB are used for the restoration of Straumann dental implants with different endosteal diameters, lengths and platforms.
- CARES bars are to be combined with an overdenture to treat edentulous cases. .
- . CARES fixed bars are superstructures for the direct application with dental resin and prefabricated teeth to treat edentulous cases.
- . CARES Screw-retained Bridges are intended to be directly veneered with dental veneering ceramics.
The purpose of this premarket notification is to expand the currently cleared abutment-toimplant interfaces to include the TLX implant system.
The SRBB devices are available for each of the three prosthetic platforms (NT, RT, WT).
The design of the SRBB devices must be made using the validated Straumann CARES Visual software (Digital CARES workflow).
Finally, the design file is transferred digitally to a Straumann validated milling center.
{10}------------------------------------------------
# Straumann® TLX Implant System
Substantial Equivalence Discussion
The TLX SRBB are provided non-sterile with instructions for end user steam sterilization.
#### TLX Basal screw
The Basal screw is used to seat the temporary abutments, the TLX Variobase Abutments or the TLX Screw-retained Bridges and Bars to the dental implant and can be also be used during lab procedures to fix lab prosthetic parts on implant analogs. They are provided along the prosthetic components, but they are also provided as standalone screws. The TLX basal screws is manufactured from TAN.
#### Indications for Use 5.5
#### TLX Dental Implant:
Straumann TLX Implants are suitable for endosteal implantation in the upper and lower jaws and for the functional and esthetic oral rehabilitation of edentulous and partially edentulous patients. TLX Implants can be placed with immediate function on single-tooth and multi-unit restorations when good primary stability is achieved and with appropriate occlusal loading to restore chewing function. The prosthetic restorations are connected to the implants through the corresponding abutment components.
#### TLX Closure Caps and Healing Caps:
Straumann Closure Caps and Healing Caps are indicated to be placed in the patient's mouth at the end of the implant placement to protect the inner configuration of the implant and to form, maintain and stabilize the soft tissue during the healing process. Closure caps and healing caps should be used only with suitable implant connections. They have a maximum duration of usage of 6 months.
#### TLX Temporary Abutment:
TLX Temporary Abutments can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they may not be placed into occlusion. TLX Temporary Abutments have a maximum duration of usage of 180 days.
#### TLX Variobase for Crown:
Straumann Variobase prosthetic components directly connected to the endosseous dental implant are intended for use as an aid in prosthetic rehabilitations. The prosthetic restoration (crowns) can be cemented onto the Straumann Variobase prosthetic components. A temporary
{11}------------------------------------------------
# Straumann® TLX Implant System
### Substantial Equivalence Discussion
restoration can be used prior to the insertion of the final components to maintain, stabilize and shape the soft tissue during the healing phase; they must be placed out of occlusion. Final abutments and restorations may be placed into occlusion when the implant is fully osseointegrated. All digitally designed copings and/or crowns for use with the Straumann Variobase Abutment system are intended to be sent to Straumann for manufacture at a validated milling center.
### TLX CARES Abutment TAN:
The Straumann CARES Abutments TAN are indicated for single tooth replacement and multiple tooth restorations. The prosthetic restoration can be cemented.
### TLX Screw-retained Bridges and Bars:
CARES Screw-retained Bridges and Bars (SRBB) are indicated for use as bars and bridges that attach to implants to provide support for prosthetic reconstructions such as bridges and overdentures. The final processed products have the purpose of restoring function. Straumann CARES Screw-retained Bridges and Bars are indicated for Screw-retained restorations. Straumann CARES Screw-retained Bridges and Bars are designed to interface with the Bone Level (BL), Tissue Level (TL), BLX implants and TLX implants of the Straumann Dental Implant System (SDIS).
#### Technological Characteristics 5.6
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in Table 1 through Table 6. Regarding the technological characteristics of the Implants described in Table 1, the following describes the relevant equivalence discussion:
- K150938 was introduced for comparison between the subject Ø3.75 x 6 mm and the . reference Ø4.1 x 6 mm implant. In order to address the difference in the diameter/thread design, surface area comparison as well as pull-out strength were conducted to demonstrate equivalence.
- K163194 was introduced for comparison between the subject Ø5.0 x 18 mm and the . reference Ø5.0 x 18 mm implant.
- K171784 and K150938 were introduced for comparison of the implant neck shape. .
{12}------------------------------------------------
# Straumann® TLX Implant System
# Substantial Equivalence Discussion
| | Proposed Device | Primary Predicate<br>Device | Reference Device | Reference Device | Reference Device | Reference Device |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | K200586<br>Straumann TLX Implant<br>System | K173961<br>Straumann BLX<br>Implant System | K181703<br>Straumann® BLX<br>Line Extension | K171784<br>Straumann Dental<br>implant system | K150938<br>Roxolid SLA Implants | K163194<br>Neodent Implant<br>System - GM Line |
| Indications for<br>Use | Straumann TLX Implants<br>are suitable for endosteal<br>implantation in the upper<br>and lower jaws and for<br>the functional and<br>esthetic oral rehabilitation<br>of edentulous and<br>partially edentulous<br>patients. TLX Implants<br>can be placed with<br>immediate function on<br>single-tooth and multi-unit<br>restorations when good<br>primary stability is<br>achieved and with<br>appropriate occlusal<br>loading to restore<br>chewing function. The<br>prosthetic restorations are<br>connected to the implants<br>through the<br>corresponding abutment<br>components. | Straumann® BLX<br>Implants are suitable for<br>endosteal implantation<br>in the upper and lower<br>jaw and for the<br>functional and esthetic<br>oral rehabilitation of<br>edentulous and partially<br>edentulous patients.<br>BLX Implants can be<br>placed with immediate<br>function on single-tooth<br>applications when good<br>primary stability is<br>achieved and with<br>appropriate occlusal<br>loading to restore<br>chewing function. The<br>prosthetic restorations<br>are connected to the<br>implants through the<br>corresponding abutment<br>components. | Straumann® BLX<br>Implants are suitable<br>for endosteal<br>implantation in the<br>upper and lower jaw<br>and for the functional<br>and esthetic oral<br>rehabilitation of<br>edentulous and<br>partially edentulous<br>patients. BLX Implants<br>can be placed with<br>immediate function on<br>single-tooth, bar and<br>bridge applications<br>when good primary<br>stability is achieved<br>and with appropriate<br>occlusal loading to<br>restore chewing<br>function. The<br>prosthetic restorations<br>are connected to the<br>implants through the<br>corresponding<br>abutment components. | Straumann® Dental<br>implants are indicated for<br>oral endosteal<br>implantation in the upper<br>and lower jaw arches and<br>for the functional and<br>aesthetic oral<br>rehabilitation of<br>edentulous and partially<br>dentate patients.<br>Straumann® Dental<br>implants are also<br>indicated for immediate or<br>early implantation<br>following extraction or loss<br>of natural teeth. Implants<br>can be placed with<br>immediate function on<br>single-tooth and/or<br>multiple-tooth applications<br>when good primary<br>stability is achieved and<br>with appropriate occlusal<br>loading to restore chewing<br>function. The prosthetic<br>restorations used are<br>single crowns, bridges<br>and partial or full<br>dentures, which are<br>connected to the implants<br>through the corresponding<br>components (abutments). | Straumann® Dental<br>implants are indicated for<br>oral endosteal<br>implantation in the upper<br>and lower jaw arches and<br>for the functional and<br>aesthetic oral<br>rehabilitation of<br>edentulous and partially<br>dentate patients.<br>Straumann® Dental<br>implants are also<br>indicated for immediate or<br>early implantation<br>following extraction or<br>loss of natural teeth.<br>Implants can be placed<br>with immediate function<br>on single-tooth and/or<br>multiple-tooth applications<br>when good primary<br>stability is achieved and<br>with appropriate occlusal<br>loading to restore<br>chewing function. The<br>prosthetic restorations<br>used are single crowns,<br>bridges and partial or full<br>dentures, which are<br>connected to the implants<br>through the<br>corresponding<br>components (abutments). | The Neodent Implant<br>System is intended to be<br>surgically placed in the<br>bone of the upper or<br>lower jaw to provide<br>support for prosthetic<br>devices such as artificial<br>teeth, to restore chewing<br>function. It may be used<br>with single-stage or two-<br>stage procedures, for<br>single or multiple unit<br>restorations, and may be<br>loaded immediately<br>when good primary<br>stability is achieved and<br>with appropriate occlusal<br>loading. |
| Material | Titanium-13 Zirconium<br>alloy (Roxolid®) | Titanium-13 Zirconium<br>alloy (Roxolid®) | Titanium-13 Zirconium<br>alloy (Roxolid®) | Titanium-13 Zirconium<br>alloy<br>(Roxolid®) | Titanium-13 Zirconium<br>alloy<br>(Roxolid®) | Titanium grade 4 |
| | Proposed Device | Primary Predicate<br>Device | Reference Device | Reference Device | Reference Device | Reference Device |
| Feature | K200586<br>Straumann TLX Implant<br>System | K173961<br>Straumann BLX<br>Implant System | K181703<br>Straumann® BLX<br>Line Extension | K171784<br>Straumann Dental<br>implant system | K150938<br>Roxolid SLA Implants | K163194<br>Neodent Implant<br>System - GM Line |
| Surface<br>Treatment | Hydrophilic SLActive® | Hydrophilic SLActive® | Hydrophilic SLActive® | Hydrophilic SLActive® | SLA | Neoporos<br>Acqua |
| Implant to<br>Abutment<br>Connection | TorcFit<br>(with conical fitting) | TorcFit<br>(with conical fitting) | TorcFit<br>(with conical fitting) | Narrow CrossFit (NC)<br>Regular CrossFit (RC)<br>Regular Neck (RN)<br>Wide Neck (WN) | Narrow CrossFit (NC)<br>Regular CrossFit (RC)<br>Narrow Neck CrossFit<br>(NNC)<br>Regular Neck (RN)<br>Wide Neck (WN) | GM interface; 16° Morse<br>taper with anti-rotational<br>features. |
| Implant<br>Diameter | Ø3.75, 4.0, 4.5, 5.0, 5.5,<br>and 6.5 mm | Ø4.5, 5.5, and 6.5 mm | Ø3.75 mm | Ø3.3, 4.1, and 4.8 mm | Ø3.3, 4.1, and 4.8 mm | Ø3.5 to 5.0 mm |
| Implant Length | Ø3.75, 4.0, 4.5, 5.0 mm:<br>6, 8, 10, 12, 14, 16 and<br>18 mm<br>Ø5.5 and 6.5 mm:<br>6, 8, 10, 12 mm | 6 to 18 mm | 8 to 18 mm | 6 to 18 mm | 6, 8, 10, 12, 14, 16 mm | 8 to 18 mm |
| Implant Design | Tapered body | Tapered body | Tapered body | Parallel wall and bone<br>level tapered (BLT) | Parallel wall and bone<br>level tapered (BLT) | Titamax<br>Helix<br>Drive |
| Implant neck | Tulip shape | n/a (bone level) | n/a (bone level) | Tulip shape<br>n/a (bone level) | Tulip shape<br>n/a (bone level) | n/a (bone level) |
| Prosthetic<br>platforms | NT, RT, and WT | RB and WB | RB and WB | RN, WN, RC, and NC | NC, RC, NNC, RN, and<br>WN | GM interface |
| Thread Pitch | 1.7, 2.0, 2.1, 2.2, 2.5, 2.6,<br>and 2.8 mm | 2.0 to 2.8 mm | 1.7 to 2.6 mm | 0.8 and 1.25 mm | 0.8 and 1.25 mm…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.