The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw. The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1: [Table 1 data omitted for brevity]. All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are only intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility.
Device Story
Patient-specific dental abutment; fabricated from Titanium alloy (ASTM F136). Device connects to dental implants via included prosthetic screw; provides foundation for cement-retained single or multi-unit crowns/copings. Designed via CAD/CAM software; manufactured at Elos Medtech approved milling facility. Used by dental clinicians in clinical settings. Device allows up to 30° angulation; provides restorative flexibility for various implant platforms. Delivered non-sterile; requires steam sterilization by clinician before placement. Benefits include customized fit for specific patient anatomy and implant systems.
Clinical Evidence
Bench testing only. Evidence includes engineering/dimensional analysis for compatibility, fatigue testing per ISO 14801 (following FDA guidance for root-form endosseous dental implants), and biocompatibility cytotoxicity testing per ISO 10993-5. No clinical data provided.
Technological Characteristics
Material: Titanium Alloy 6Al-4V ELI (ASTM F136). Design: Patient-specific, indexed connection (flat top and conical). Dimensions: Diameter Ø3.0–Ø6.0 mm, max angulation 30°, min wall thickness 0.4 mm. Connectivity: CAD/CAM digital design workflow. Sterilization: Steam sterilization (non-sterile delivery).
Indications for Use
Indicated for patients requiring single or multiple tooth prosthetic restorations supported by endosseous dental implants. Compatible with specific Nobel Biocare and Straumann implant platforms.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 26, 2019
Elos Medtech Pinol A/S Tina Poulsen Head of Compliance Engvei 33 DK-3330 Goerloese DENMARK
Re: K190299
Trade/Device Name: Elos Accurate® Customized Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 28, 2019 Received: May 28, 2019
Dear Tina Poulsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# INDICATIONS FOR USE
# 510(k) Number: K190299 Device Name: Elos Accurate® Customized Abutment
### Indications for Use
The Elos Accurate® Customized Abutments are intents in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.
The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:
### Table 1.
| Elos Accurate Customized<br>Abutment - Model Type | Platform compatibility | Platform diameter<br>[mm] | Implant Body diameter<br>[mm] |
|---------------------------------------------------|-------------------------|---------------------------|-------------------------------|
| AB-NBR35 | Nobel Replace NP | 3.5 | 3.5 |
| AB-NBA30 | Nobel CC 3.0 | 3 | 3 |
| AB-NBA43 | Nobel CC RP | 3.9 | 4.3 & 5 |
| AB-NBA60 | Nobel CC WP | 5.1 | 5.5 |
| AB-SBO33 | Straumann Bone Level NC | 3.3 | 3.3 |
| AB-SBO41 | Straumann Bone Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are only intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary Elos Accurate® Customized Abutment June 20, 2019
This summary of 510(k) information is being submitted in accordance with the requirements of 21 CFR § 807.92.
| I. | Company: | Elos Medtech Pinol A/S<br>Engvej 33<br>DK-3330 Goerloese<br>Denmark |
|------|-------------------------|------------------------------------------------------------------------------------------------------------------------|
| | Contacts: | Tina Friis Poulsen<br>Head of Compliance<br>Tel: +45 27 77 11 17<br>E-mail:tina.poulsen@elosmedtech.com |
| | | Magnus Lynge Bergholt<br>Regulatory Affairs Manager<br>Tel: +45 48 21 64 58<br>E-mail: magnus.jacobsen@elosmedtech.com |
| II. | Proprietary Trade Name: | Elos Accurate® Customized Abutment |
| III. | Classification Name: | Endosseous Dental Implant Abutment |
| IV. | Classification: | Class II, 21 CFR 872.3630 |
| V. | Product Code(s): | NHA |
### VI. Identification of Legally Marketed Devices:
The design features, materials and Indications for Use of the subject devices are substantially equivalent to the predicate devices noted below.
# Primary Predicate Device:
- K171799 / SE 01/15/2018 Elos Accurate® Customized Abutment .
## Reference Devices:
- K023113 / SE 09/26/2002 NOBEL REPLACE TIUNITE ENDOSSEOUS ● IMPLANT
- K083550 / SE 02/26/2009 - STRAUMANN DENTAL IMPLANT SYSTEM
- K102436 / SE 12/09/2010 NOBEL ACTIVE 3.0 ●
- . K071370 / SE 08/03/2007 – NOBEL ACTIVE INTERNAL CONNECTION IMPLANT
- K133731 / SE 05/08/2014 NOBEL ACTIVE WIDE PLATFORM (WP) ●
{4}------------------------------------------------
### VII. Product Description:
The Elos Accurate® Customized Abutment is a patient specific abutment intended for attaching to dental implants in order to provide basis for single- or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to the implant using the included Elos Prosthetic Screw and attached to the crown/coping manually by cementation. The Elos Accurate® Customized Abutment consists of an Abutment Blank used in fabricating of a full patient-specific abutment in Titanium alloy per ASTM F136. The Abutment Blank used in creation of the Elos Accurate® Customized Abutment has a pre-manufactured connection interface that fits directly to a pre-specified dental implant. The customized shape of the abutment is intended to be manufactured by an Elos Medtech approved milling facility. The Elos Accurate® Customized Abutment is delivered nonsterile and the final restoration including corresponding Elos Prosthetic Screw is intended to be sterilized at the dental clinic before it is placed in the patient. The Elos Accurate® Customized Abutment provides clinicians and laboratories with a prosthetic device that can be used in definitive (permanent) single- or multi restorations.
### VIII. Indications for Use:
The Elos Accurate® Customized Abutments are intended for attaching to dental implants in order to provide basis for single or multiple tooth prosthetic restorations. The Elos Accurate® Customized Abutment will be attached to a dental implant using the included Elos Prosthetic screw.
The Elos Accurate® Customized Abutments are compatible with the implant systems listed in table 1:
| Elos Accurate Customized<br>Abutment - Model Type | Platform<br>compatibility | Platform<br>diameter<br>[mm] | Implant Body<br>diameter<br>[mm] |
|---------------------------------------------------|----------------------------|------------------------------|----------------------------------|
| AB-NBR35 | Nobel Replace NP | 3.5 | 3.5 |
| AB-NBA30 | Nobel CC 3.0 | 3 | 3 |
| AB-NBA43 | Nobel CC RP | 3.9 | 4.3 & 5 |
| AB-NBA60 | Nobel CC WP | 5.1 | 5.5 |
| AB-SBO33 | Straumann Bone<br>Level NC | 3.3 | 3.3 |
| AB-SBO41 | Straumann Bone<br>Level RC | 4.1 & 4.8 | 4.1 & 4.8 |
| Table | 1. |
|-------|----|
All digitally designed CAD/CAM customizations for the Elos Accurate® Customized Abutments are only intended to be sent and manufactured at a FDA registered Elos Medtech approved milling facility.
{5}------------------------------------------------
### IX. Summary of the Technological Characteristics:
The subject devices provide additional restorative options for connection to existing implant platforms. The subject devices has similar Indications for Use, intended use, designs, sizes and configurations, materials, and principles of operation as the primary predicate device Elos Accurate® Customized Abutment (K171799 / SE 01/15/2018). In order to determine nominal dimensions and tolerances of the Elos Accurate® Customized Abutment products, measuring- and dimensional analyses of original manufacturers' components (abutments, implants & abutment screws) have been made.
Comparing to the primary predicate device, the specific language (wording) of the Indications for Use Statements are not identical. However, the subject device and corresponding primary predicates are all abutments intended for use with endosseous dental implants for support of single-tooth or multiple-tooth restorations in the upper or lower arch, and therefore, have the same intended use. Also, for the subject device, all digitally designed abutments are intended to be manufactured at an Elos Medtech approved milling facility which is equivalent to the predicate device. The implant system compatibility of the subject device is extended to include compatibility other implant platforms. All differences between the subject- and predicate devices are further specified in the 4 bullets below. The 4 bullets have been substantiated by engineering and dimensional analysis of original manufactures' components (abutments, implants & abutment screws) for determination of compatibility (bullet 1), fatigue testing (bullet 1, 2 and 3) and additional biocompatibility cytotoxicity testing (bullet 4).
- . The implant system compatibility of the subject device is extended to include Nobel Replace NP, Nobel CC; 3.0, RP and WP and Straumann Bone Level NC and RC platforms.
- The subject device allows an angulation up to 30° compared to the primary predicate device that allows an angulation up to 20°.
- . Minimum wall thickness is 0.4mm for subject device compared to 0.5mm for the primary predicate
- . Subject devices includes a non-coated prosthetic screw
{6}------------------------------------------------
| Element of Comparison | Indications for Use | | | | |
|--------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------------------|-------------------------------------|--|
| Subject Device<br>Elos Accurate® Customized<br>Abutment | The Elos Accurate® Customized Abutments are intended for attaching to<br>dental implants in order to provide basis for single or multiple tooth<br>prosthetic restorations. The Elos Accurate® Customized Abutment will be<br>attached to a dental implant using the included Elos Prosthetic screw.<br>The Elos Accurate® Customized Abutments are compatible with the implant<br>systems listed in table 1: | | | | |
| | Table 1.<br>Elos Accurate<br>Customized Abutment<br>- Model Type | Platform<br>compatibility | Platform<br>diameter<br>[mm] | Implant<br>Body<br>diameter<br>[mm] | |
| | AB-NBR35 | Nobel Replace<br>NP | 3.5 | 3.5 | |
| | AB-NBA30 | Nobel CC 3.0 | 3 | 3 | |
| | AB-NBA43 | Nobel CC RP | 3.9 | 4.3 & 5 | |
| | AB-NBA60 | Nobel CC WP | 5.1 | 5.5 | |
| | AB-SBO33 | Straumann Bone<br>Level NC | 3.3 | 3.3 | |
| | AB-SBO41 | Straumann Bone<br>Level RC | 4.1 & 4.8 | 4.1 & 4.8 | |
| | All digitally designed CAD/CAM customizations for the Elos Accurate®<br>Customized Abutments are only intended to be sent and manufactured at a<br>FDA registered Elos Medtech approved milling facility. | | | | |
| Primary Predicate Device<br>(K171799)<br>Elos Accurate® Customized<br>Abutment | The Elos Accurate® Customized Abutments are intended for attaching to<br>dental implants in order to provide basis for single or multiple tooth<br>prosthetic restorations. The Elos Accurate® Customized Abutment will be<br>attached to a dental implant using the included Elos Prosthetic screw.<br>The Elos Accurate® Customized Abutments are compatible with the<br>following implant systems: | | | | |
| | Ref. No. | Platform compatibility | | Implant<br>diameter | |
| | AB-BRA411213-US | Nobel Biocare® / Brånemark® RP | | 3.75 mm & 4<br>mm | |
| | AB-BRA351213-US | Nobel Biocare® / Brånemark® NP | | 3.3 mm | |
| | AB-BRA511213-US | Nobel Biocare® / Brånemark® WP | | 5 mm | |
| | All digitally designed Elos Accurate® Customized Abutments are intended to<br>be manufactured at an Elos Medtech approved milling facility. | | | | |
{7}------------------------------------------------
| Element of Comparison | Subject Device<br>Elos Accurate® Customized Abutment | Primary Predicate Device (K171799)<br>Elos Accurate® Customized Abutment |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FDA Classification & Product Code | 21 CFR 872.3630: NHA<br>Endosseous Dental Implant Abutment | 21 CFR 872.3630: NHA<br>Endosseous Dental Implant Abutment |
| Base Materials | Elos Accurate® Customized Abutments:<br>Titanium Alloy 6Al-4V ELI, medical grade 5<br>Elos Prosthetic screws:<br>Titanium Alloy 6Al-4V ELI, medical grade 5 | Elos Accurate® Customized Abutments:<br>Titanium Alloy 6Al-4V ELI, medical grade 5<br>Elos Prosthetic screws:<br>Titanium Alloy 6Al-4V ELI, medical grade 5 |
| Surface Finish | Elos Accurate® Customized Abutment:<br>Non-coated<br>Elos Abutment screws:<br>Non-coated or MediCarb (DLC) | Elos Accurate® Customized Abutment:<br>Non-coated<br>Elos Abutment screws:<br>MediCarb (DLC) |
| Implant interface | Indexed | Indexed |
| Connection type | Flat top and conical | Flat top |
| Abutment diameter | Ø3.0 - Ø6.0 mm | Ø3.5 - Ø5.1 mm |
| Prosthesis attachment | Cement-retained | Cement-retained |
| Restoration | Single-unit &Multi-unit | Single-unit &Multi-unit |
| Abutment Design Matrix | Up to 30° | Up to 20° |
| Sterility | Provided non-sterile | Provided non-sterile |
| Sterilization method | Steam sterilization | Steam sterilization |
| Digital CAD Systems | 510(k) cleared CAD software | 510(k) cleared CAD software |
| Abutment Design Matrix | Minimum wall thickness: 0.4 mm<br>Maximum post height: 13 mm<br>Maximum angulation: 30°<br>Maximum diameter: 12 mm<br>Minimum post height: 4mm<br>Minimum gingiva height: 0.5mm<br>Maximum gingiva height: 5mm | Minimum wall thickness: 0.5 mm<br>Maximum post height: 13 mm<br>Maximum angulation: 20°<br>Maximum diameter: 12 mm<br>Minimum post height: 4mm<br>Minimum gingiva height: 0.5mm<br>Maximum gingiva height: not stated |
{8}------------------------------------------------
The data included in this submission demonstrate substantial equivalence to the predicate device and/or reference device listed above.
Overall, the subject device has the following substantial equivalencies to the predicate device:
- . has the same intended use,
- uses the same operating principle,
- incorporates the same basic design, ●
- incorporates the same or very similar materials, and ●
- . is to be sterilized using the same processes.
#### X. Discussion of the Non-Clinical Testing:
Non clinical testing data submitted included:
- engineering and dimensional analysis of original manufactures' components (abutments, implants & abutment screws) for determination of compatibility.
- fatigue testing per ISO 14801 according to FDA guidance for Industry and FDA Staff "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments" dated May 12, 2004.
- biocompatibility testing for cytotoxicity according to ISO 10993-5. Tests included covered: ●
- non-coated prosthetic screw representative for subject device -
- -Medicarb coated prosthetic screw representative for subject devices (test of primary predicate device (K171799))
- -Elos accurate® Customized Abutment representative for subject devices (test of primary predicate device (K171799))
As the primary predicate device were tested according to ISO 17665-1 & ISO 17665-2, demonstrating a SAL of 10th no additional testing were necessary on the subject device.
#### XI. Conclusions:
Based on the test results and additional supporting documentation provided in this premarket notification, the subject devices demonstrated substantial equivalence to the previously listed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.