Straumann Bone Level NC Angled Screw Retained Abutments

K141871 · Straumann USA, LLC · NHA · Aug 7, 2014 · Dental

Device Facts

Record IDK141871
Device NameStraumann Bone Level NC Angled Screw Retained Abutments
ApplicantStraumann USA, LLC
Product CodeNHA · Dental
Decision DateAug 7, 2014
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 872.3630
Device ClassClass 2

Indications for Use

The Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations.

Device Story

Straumann Screw Retained Abutments are dental components placed into Straumann dental implants to support prosthetic reconstructions (crowns, bridges, bars). Devices are one-piece titanium alloy components manufactured via precision machining. Used in dental clinics by dentists/prosthodontists to restore chewing function in patients. Output is a mechanical interface providing structural support for dental prosthetics. Benefit is restoration of oral function.

Clinical Evidence

Bench testing only. Dynamic fatigue testing performed consistent with FDA guidance and ISO 14801 to demonstrate mechanical integrity.

Technological Characteristics

Material: Ti-6Al-7Nb alloy (ISO 5832-11). Construction: One-piece precision machined. Connection: Narrow CrossFit (NC) engaging. Angulations: 17°, 30°. Sterilization: Non-sterile, intended for moist heat (autoclave) terminal sterilization.

Indications for Use

Indicated for patients requiring prosthetic restoration (crowns, bridges, bars) supported by Straumann Dental Implant System implants to restore chewing function.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 7, 2014 Straumann USA, Limited Liability Company Mr. Christopher Klaczyk Director of Regulatory Affairs & Clinical Research 60 Minuteman Road Andover, MA 01810 Re: K141871 Trade/Device Name: Straumann® Screw Retained Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: July 9, 2014 Received: July 11, 2014 Dear Mr. Klaczyk: This letter corrects our substantially equivalent letter of August 7, 2014. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Mr. Klaczyk Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Mary S. Runner -S Erin I. Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## 4. Indications For Use 510(k) Number (if known): K141871 Straumann® Screw Retained Abutments Device Name: Indications for Use: The Straumann® Screw Retained Abutments are indicated to be placed into the implants of the Straumann® Dental Implant System to provide support for prosthetic reconstructions such as crowns, bridges and bars. The final processed devices have the purpose of restoring chewing function. Straumann® Screw Retained Abutments are indicated for screw-retained restorations. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sheena A. Green-S 2014.08.07 13:44:29-04'00' {4}------------------------------------------------ ## 5. 510(k) Summary | Submitter: | Straumann USA, LLC (on behalf of Institut Straumann AG)<br>60 Minuteman Road<br>Andover, MA 01810<br>Registration No .: 1222315 Owner/Operator No .: 9005052 | | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--| | Contact Person: | Christopher Klaczyk<br>Director of Regulatory Affairs and Clinical Research<br>Tel.: (978) 747-2575 | | | Date Prepared: | July 9, 2014 | | | Product Code(s): | NHA (21 CFR 872.3630) | | | Device Class: | (21 CFR 872.3630)<br>II | | | Classification Panel: | Dental | | | Classification Name: | Endosseous dental implant abutments (21 CFR 872.3630) | | | Proprietary Name: | Straumann® Screw Retained Abutments | | | Predicate Device(s): | Straumann® Magellan Abutment System (K133421)<br>● | | | Device Description: | The proposed Straumann® NC Angled Screw Retained<br>Abutments are a line extension of the NC Straight and RC<br>Straight and Angled Screw Retained Abutments cleared to<br>market per premarket notification submission K133421. | | | | The Straumann® NC Angled Screw Retained Abutments<br>include 17° and 30° abutments. This is consistent with the RC<br>Angled Screw Retained Abutments cleared to market per<br>premarket notification submission K133421. | | | Intended Use: | The Straumann® Screw Retained Abutments are indicated to be<br>placed into the implants of the Straumann® Dental Implant<br>System to provide support for prosthetic reconstructions such as<br>crowns, bridges and bars. | | | | The final processed devices have the purpose of restoring<br>chewing function. | | | | Straumann® Screw Retained Abutments are indicated for<br>screw-retained restorations. | | | Materials: | The subject devices are produced from titanium-6aluminum-<br>7niobium alloy (TAN) conforming to ISO 5832-11. This is the<br>same material as for the predicate abutments cleared to market<br>per premarket notification submission K133421. | | | Technological<br>Characteristics: | The proposed Straumann® NC Angled Screw Retained<br>Abutments are manufactured using precision machining<br>systems from solid material (i.e. one-piece construction). All<br>technological characteristics of the subject devices are the same<br>as for devices as shown in the table below. | | {5}------------------------------------------------ | Feature | Subject Devices<br>NC Angled Screw Retained<br>Abutments | Predicate Devices<br>Magellan Abutment System<br>K133421 | |-----------------------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------| | Implant-to-Abutment<br>Connection | Narrow CrossFit (NC) | Narrow CrossFit (NC)<br>Regular CrossFit (RC) | | Interface Type | Engaging | Engaging | | Platform Diameter(s) | NC: Ø4.6 mm | NC: Ø3.5 mm, Ø4.6 mm<br>RC: Ø4.6 mm | | Abutment Angulation(s) | NC: 17°, 30° | NC: 0°<br>RC: 0°, 17°, 30° | | Gingival Height(s) | NC Angled: 2.5 mm, 4.0 mm | NC Straight: 1.0, 2.5 and 4.0 mm<br>RC Straight: 1.0, 2.5 and 4.0 mm<br>RC Angled: 2.5 and 4.0 mm | | Orientation of Angulation<br>to Engagement Features | NC: Type A (45°), Type B (0°) | NC: N/A<br>RC: Type A (45°), Type B (0°) | | Material | Ti-6Al-7Nb titanium alloy | Ti-6Al-7Nb titanium alloy | | Primary Package | Medical grade polyethylene blister<br>with a sealing lid | Medical grade polyethylene blister<br>with a sealing lid | | Sterilization | Non-sterile; intended for terminal<br>sterilization via moist heat<br>(autoclave) | Non-sterile; intended for terminal<br>sterilization via moist heat<br>(autoclave) | Per Guidance for Industry and FDA Staff - Class II Special Performance Data: Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments dated May 12, 2004, the substantial equivalence of the subject device(s) are satisfactorily addressed via bench studies. Dynamic fatigue test data consistent with FDA guidance and ISO 14801 have been referenced in support of this submission. - Based upon our assessment of the design and applicable Conclusions: performance data, the subject devices have been determined to be substantially equivalent to the identified predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...