Inclusive Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System EV

K191222 · Prismatik Dentalcraft, Inc. · NHA · Aug 5, 2019 · Dental

Device Facts

Record IDK191222
Device NameInclusive Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System EV
ApplicantPrismatik Dentalcraft, Inc.
Product CodeNHA · Dental
Decision DateAug 5, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Inclusive® Titanium Abutments are premanufactured prosthetic components connected to endosseous dental implants in the edentulous or partially edentulous maxilla or mandible to provide support for cement-retained or screw-retained prosthetic restorations. All digitally designed abutments for use with Inclusive Titanium Abutments for CAD/CAM are intended to be sent to a Prismatik Dentalcraft validated milling center for manufacture. Compatible Implant System: Dentsply Implants Astra Tech Implant System® EV

Device Story

Inclusive® Titanium Abutments are premanufactured prosthetic components for endosseous dental implants. Input: patient-specific clinical data for CAD/CAM design. Transformation: clinician prescribes custom abutment; validated milling center manufactures device using CAD/CAM. Output: titanium abutments (stock or custom) and titanium screws. Used in dental clinics; operated by clinicians. Output used by clinicians to support cement-retained or screw-retained crowns, bridges, or zirconia copings. Benefits: restores chewing function in edentulous/partially edentulous patients. Device includes titanium blanks, 4.5mmH/6mmH abutments, esthetic abutments, and screws.

Clinical Evidence

No clinical data included. Bench testing only: sterilization validation (ISO 17665-1), biocompatibility (ISO 10993-5/12), reverse engineering of OEM implant bodies/abutments, and static compression/compression fatigue testing (ISO 14801).

Technological Characteristics

Materials: Titanium alloy (ASTM F136/Ti-6Al-4V ELI). Design: Prefabricated precision interface, taper with keyway anti-rotation feature. Dimensions: Compatible with Astra Tech OsseoSpeed EV implants (3.0S-5.4S). Connectivity: CAD/CAM workflow via validated milling center. Sterilization: Non-sterile, end-user sterilized per ISO 17665-1.

Indications for Use

Indicated for patients with edentulous or partially edentulous maxilla or mandible requiring support for cement-retained or screw-retained prosthetic restorations connected to endosseous dental implants.

Regulatory Classification

Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 5, 2019 Prismatik Dentalcraft, Inc. % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130 Re: K191222 Trade/Device Name: Inclusive® Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System® EV Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 6, 2019 Received: May 7, 2019 Dear Kevin Thomas: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas Nandkumar, Ph.D. Acting Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K191222 #### Device Name Inclusive® Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System® EV #### Indications for Use (Describe) Inclusive® Titanium Abutments are premanufactured prosthetic components connected to endosseous dental implants in the edentulous or partially edentulous maxilla or mandible to provide support for cement-retained or screw-retained prosthetic restorations. All digitally designed abutments for use with Inclusive Titanium Abutments for CAD/CAM are intended to be sent to a Prismatik Dentalcraft validated milling center for manufacture. Compatible Implant System: Dentsply Implants Astra Tech Implant System® EV | Compatible Implant Fixtures | Implant Body Diameter | Implant Platform Diameter | |-----------------------------|-----------------------|---------------------------| | OsseoSpeed® EV 3.0S | 3.0 mm | 3.0 mm | | OsseoSpeed® EV 3.6S | 3.6 mm | 3.6 mm | | OsseoSpeed® EV 4.2S | 4.2 mm | 4.2 mm | | OsseoSpeed® EV 4.2C | 3.6 mm | 4.2 mm | | OsseoSpeed® EV 4.8S | 4.8 mm | 4.8 mm | | OsseoSpeed® EV 4.8C | 4.2 mm | 4.8 mm | | OsseoSpeed® EV 5.4S | 5.4 mm | 5.4 mm | | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------|--| |-------------------------------------------------|--| > Prescription Use (Part 21 CFR 801 Subpart D) |__ Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary – K191222 # Inclusive® Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System® EV # Prismatik Dentalcraft, Inc. August 5, 2019 #### ADMINISTRATIVE INFORMATION | Manufacturer Name | Prismatik Dentalcraft, Inc.<br>2212 Dupont Drive, Suite P<br>Irvine, CA 92612<br>Telephone: +1 949-440-2629<br>Fax: +1 978-313-0850 | |---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Official Contact | Herbert Schoenhoefer, Director of RA/QA<br>So Hyun Park, Regulatory Affairs Specialist | | Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: kthomas@paxmed.com<br>flarson@paxmed.com | #### DEVICE NAME AND CLASSIFICATION | Trade/Proprietary Name | Inclusive® Titanium Abutments compatible with: | |------------------------|-------------------------------------------------| | | Dentsply Implants Astra Tech Implant System® EV | | Common Name | Dental implant abutment | | Regulation Number | 21 CFR 872.3630 | | Regulation Name | Endosseous dental implant abutment | | Regulatory Class | Class II | | Product Code | NHA | | Classification Panel | Dental Products Panel | | Reviewing Branch | Dental Devices Branch | #### PREDICATE DEVICE INFORMATION Primary Predicate K170044, Inclusive® Titanium Abutments compatible with: MegaGen AnyRidge® Implant System, Prismatik Dentalcraft, Inc. Reference Devices K120414, OsseoSpeed™ Plus, Astra Tech AB K073217, Inclusive™ Abutment for Zimmer, 31 and Nobel Biocare Implants, Inclusive Dental Solutions K083192, Inclusive Titanium Abutment Blanks, Inclusive Dental Solutions K160979, Inclusive Abutments, Prismatik Dentalcraft, Inc. {5}------------------------------------------------ ## INDICATIONS FOR USE STATEMENT Inclusive® Titanium Abutments are premanufactured prosthetic components connected to endosseous dental implants in the edentulous or partially edentulous maxilla or mandible to provide support for cement-retained or screw-retained prosthetic restorations. All digitally designed abutments for use with Inclusive Titanium Abutments for CAD/CAM are intended to be sent to a Prismatik Dentalcraft validated milling center for manufacture. Compatible Implant System: Dentsply Implants Astra Tech Implant System® EV | Compatible Implant Fixtures | Implant Body Diameter | Implant Platform Diameter | |-----------------------------|-----------------------|---------------------------| | OsseoSpeed® EV 3.0S | 3.0 mm | 3.0 mm | | OsseoSpeed® EV 3.6S | 3.6 mm | 3.6 mm | | OsseoSpeed® EV 4.2S | 4.2 mm | 4.2 mm | | OsseoSpeed® EV 4.2C | 3.6 mm | 4.2 mm | | OsseoSpeed® EV 4.8S | 4.8 mm | 4.8 mm | | OsseoSpeed® EV 4.8C | 4.2 mm | 4.8 mm | | OsseoSpeed® EV 5.4S | 5.4 mm | 5.4 mm | ## SUBJECT DEVICE DESCRIPTION Inclusive® Titanium Abutments compatible with: Dentsply Implants Astra Tech Implant System EV, which are the subject of this submission, are designed and fabricated to be compatible exclusively with Dentsply Implants Astra Tech Implant System EV, and have a prefabricated, precision interface (implant/abutment connection) compatible with the primary predicate device connection. The proposed product line includes stock abutments and abutments intended for fabrication of patient-specific custom abutments using CAD/CAM technology. Each patient-specific abutment is individually prescribed by the clinician and manufactured by an authorized milling center validated by Prismatik Dentalcraft. The subject devices are manufactured from titanium alloy conforming to ASTM F136, and have a taper followed by a keyway connection feature that prevents rotation of the abutment. The abutments are compatible exclusively with the Astra Tech OsseoSpeed EV implant line (cleared under K120414) in the following diameters: 3.0S, 3.6S, 4.2S, 4.2C, 4.8S, 4.8C, 5.4S mm. The subject device includes the following components: Titanium Abutment Blanks, intended to fabricate all-titanium, patientspecific abutments using CAD/CAM technology; Titanium Abutment 4.5mmH and Titanium Abutment 6mmH, intended to be used for support of CAD/CAM fabricated crowns/bridges or zirconia copings; Titanium Esthetic Abutments, prefabricated abutments intended to be connected directly to an endosseous implant for retention of a prosthesis in straight and 15° angled designs; and Titanium Screws, indicated for the temporary or long-term retention of the abutments to the compatible dental implant fixtures. All subject device components are provided non-sterile. Titanium Abutment 4.5mmH are two-piece abutments. The crowns/bridges or zirconia copings produced at the validated milling center compose the second part of the two piece abutment. ## PERFORMANCE DATA Non-clinical data submitted to demonstrate substantial equivalence included: sterilization validation according to ISO 17665-1, and referenced from K073217, and K160979; biocompatibility {6}------------------------------------------------ according to ISO 10993-5 and ISO 10993-12, and referenced from K160979; reverse engineering of OEM implant bodies, abutments, and abutment screws to confirm compatibility; and static compression and compression fatigue testing according to ISO 14801. No clinical data were included in this submission. ## EQUIVALENCE TO MARKETED DEVICES The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference device listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device K170044, and the reference device K120414. The subject device is substantially equivalent to the primary predicate device K170044 in intended use, material, design, and performance. The subject device and the primary predicate device K170044 each include prefabricated, precision interface (implant/abutments that are manufactured from titanium alloy conforming to ASTM F136. The subject device and the primary predicate device K 170044 each include stock abutments and abutments intended to be used for fabrication of patientspecific custom abutments using CAD/CAM technology. Each patient-specific abutment is individually prescribed by the clinician and manufactured by an authorized milling center validated by Prismatik Dentalcraft. The purpose of including the reference device K120414 is to support the indication for use of the subject device with the compatible Astra Tech OsseoSpeed® EV implants 3.0S, 3.6S, 4.2S, 4.2C, 4.8S, 4.8C, and 5.4S. The reference devices K073217, K083192, and K160979 are included to reference prior sterilization validation information; K160979 also is included to reference prior biocompatibility information. The Indications for Use Statement (IFUS) for the subject device is substantially equivalent to that of the primary predicate K170044. Slight differences in language of the subject device and primary predicate device Indications for Use statements do not affect the intended use as endosseous dental implant abutments for the support of a prosthesis to restore chewing function. Minor differences in language of the subject device and primary predicate device IFUS are related to the specific device names and the compatible OEM implant lines. None of these minor differences impact safety or effectiveness because both IFUS express equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording. Similarly, the differences between the subject device IFUS and that of the reference device are related to the specific device names and design features, and the compatible implant lines. None of these minor differences impact safety or effectiveness because all IFUS express equivalent intended use to facilitate dental prosthetic restorations, and the indications are expressed equivalently using different specific wording. The subject device designs are substantially equivalent to the corresponding designs of the primary predicate device K170044, including titanium blank abutments, titanium base abutments (engaging and non-engaging), and esthetic abutments. The subject device and the primary predicate device K170044 are for single-unit or multi-unit restorations, are for cement-retained prostheses, have internal implant interface connections, and are made of Ti-6Al-4V alloy (abutment screws), and the titanium base abutments are to be used with zirconia superstructures. {7}------------------------------------------------ The subject device includes designs for implant platforms ranging from 3.0 mm to 5.4 mm. The primary predicate K170044 includes implant platform sizes of 3.5 mm. Substantial equivalence of the range of platform sizes of the subject device are supported by the reference device K120414. The subject device includes designs for abutment angulation up to 20°; this angulation is within the range of abutment angulation of the primary predicate device K170044 and is the same range as the reference device K120414. The difference in compatible OEM implant designs between the subject device and the primary predicate device K17044 are mitigated by dimensional analysis and reverse engineering analysis regarding specific critical dimensions. When the subject device is used according to its labeling, this difference does not impact safety or effectiveness. The subject device is to be sterilized by the end-user, the same as the primary predicate device K170044. Sterilization validation for the subject device was performed according to ISO 17665-1. This validated sterilization method is the same as that for the primary predicate device K170044, and the reference devices K073217, K083192, and K160979. Substantial equivalence of the subject device components in terms of biocompatibility is supported by the fact that materials are identical in formulation, processing, component interactions, and storage conditions to the predicate devices in K170044. The biocompatibility testing included in the previous 510(k) submissions, K160979 and K170044, were referenced as the confirmatory biocompatibility testing of the subject device. Mechanical performance testing was performed according to ISO 14801. Worst-case constructs were subjected to static compression and compression fatigue testing. The fatigue limit data demonstrated that the subject device has sufficient strength for its intended use. ## CONCLUSION The subject device, the primary predicate device, and the reference device have the same intended use, have similar technological characteristics, and are materials. The subject device, the primary predicate, and reference device encompass the same range of physical dimensions, are packaged in similar materials, and are to be sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. {8}------------------------------------------------ # Table of Substantial Equivalence | | Subject Device | Primary Predicate Device | Reference Device | |--------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Comparison | Inclusive® Titanium Abutments compatible with:<br>Dentsply Implants Astra Tech Implant System® EV<br>Prismatik Dentalcraft, Inc. | K170044<br>Inclusive® Titanium Abutments compatible with:<br>MegaGen AnyRidge® Implant System<br>Prismatik Dentalcraft, Inc. | K120414<br>OsseoSpeed™ Plus<br>Astra Tech AB | | Indications for Use Statement | Inclusive® Titanium Abutments are premanufactured prosthetic<br>components connected to endosseous dental implants in the<br>edentulous or partially edentulous maxilla or mandible to provide<br>support for cement-retained or screw-retained prosthetic<br>restorations.<br>All digitally designed abutments for use with Inclusive Titanium<br>Abutments for CAD/CAM are intended to be sent to a Prismatik<br>Dentalcraft validated milling center for manufacture.<br>Compatible Implant System: Dentsply Implants Astra Tech Implant<br>System® EV | Inclusive® Titanium Abutments are premanufactured prosthetic<br>components connected to endosseous dental implants in the<br>edentulous or partially edentulous maxilla or mandible to provide<br>support for cement-retained or screw-retained prosthetic<br>restorations.<br>All digitally designed abutments for use with Inclusive<br>Abutments for CAD/CAM are intended to be sent to a Prismatik<br>Dentalcraft validated milling center for manufacture.<br>Compatible Implant System: MegaGen AnyRidge® Implant<br>System<br>Implant Diameter (mm) 3.5, 4.0, 4.4, 4.9, 5.4, 5.9, 6.4, 6.9, 7.4,<br>7.9, 8.4<br>Platform Diameter (mm) 3.5 and 5.0 | [Implant portion of IFUS not applicable to subject device]<br>Abutments:<br>Astra Tech Implant System Plus abutments are intended to be<br>used in conjunction with Astra Tech Implant System Plus in fully<br>edentulous or partially edentulous maxillary and/or mandibular<br>arches to provide support for crowns, bridges or overdentures.<br>Atlantis Abutments:<br>The Atlantis™ Abutment is intended for use with an endosseous<br>implant to support a prosthetic device in a partially or completely<br>edentulous; patient. It is intended for use to support single and<br>multiple tooth prostheses, in the mandible or maxilla. The<br>prosthesis can be cemented, screw retained or friction fit to the<br>abutment. The abutment screw is intended to secure the abutment<br>to the endosseous implant.<br>The Atlantis™ Crown Abutment in Zirconia is intended for use<br>with an endosseous; implant to function as a substructure that also<br>serves as the final restoration, in partially or completely<br>edentulous; patients. The prosthesis is screw retained. The<br>abutment screw is intended to secure the crown abutment to the<br>endosseous implant. | | Reason for Predicate /<br>Reference Device | Not applicable | IFUS; Abutment designs | Compatible implant fixtures | | Summary | | | | | Compatible Implant Diameters | 3.0, 3.6, 4.2, 4.8, 5.4 mm | 3.5, 4.0, 4.4, 4.9, 5.4, 5.9, 6.4, 6.9, 7.4, 7.9, 8.4 mm | 3.0, 3.6, 4.2, 4.8, 5.4 mm | | Compatible Astra Tech<br>OsseoSpeed™ EV Implants | 3.0S, 3.6S, 4.2S, 4.2C, 4.8S, 4.8C, 5.4S | Not applicable | 3.0S, 3.6S, 4.2S, 4.2C, 4.8S, 4.8C, 5.4S | | Abutment Platform Sizes | 3.0 mm - 5.4 mm | 3.5 mm - 5.0 mm | 3.0 mm - 5.4 mm | | Abutment Margin Height | 0.5 mm - 6.0 mm | 0.5 mm - 6.5 mm | 1.0 mm - 7.0 mm | | Abutment Angle | 0° - 20° | 0° - 30° | 0° - 30° | | Prosthesis Attachment | Cement-retained, Screw-retained | Cement-retained, Screw-retained | Cement-retained, Screw-retained | | Restoration | Single-unit, Multi-unit | Single-unit, Multi-unit | Single-unit, Multi-unit | | Abutment Designs | | | | | Titanium Abutment Blank | Provided in 9.4 mm diameter;<br>to machine a patient-specific, engaging, one-piece<br>abutment | Provided in 9.4 mm diameter;<br>to machine a patient-specific one-piece abutment | Not applicable | | Prosthetic Post Height | 4.0 mm minimum | 4.0 minimum | | | Margin Height | 0.5 mm minimum; 6 mm maximum | 0.5 mm minimum; 6 mm maximum | 0.5 mm minimum; 6 mm maximum | | | Subject Device | Primary Predicate Device | Reference Device | | Comparison | Inclusive® Titanium Abutments compatible with:<br>Dentsply Implants Astra Tech Implant System® EV<br>Prismatik Dentalcraft, Inc. | K170044<br>Inclusive® Titanium Abutments compatible with:<br>MegaGen AnyRidge® Implant System<br>Prismatik Dentalcraft, Inc. | K120414<br>OsseoSpeed™ Plus<br>Astra Tech AB | | Angulation | 20° maximum | 30° maximum | | | Prosthesis Attachment | Cement-retained, Screw-retained | Cement-retained, Screw-retained | | | Restoration Types | Single-unit, Multi-unit | Single-unit, Multi-unit | | | Titanium Abutment Engaging | Engaging, two-piece abutment | Engaging, two-piece abutment | | | Prosthetic Post Height | 4 mm, 5.5 mm | 4 mm, 5.5 mm | | | Margin Height | 0.5 mm minimum, 6 mm maximum | 1 mm | | | Angulation | 0°…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...