DenTek Fresh Protect Dental Guard

K202974 · Medtech Products, Inc. · OBR · Dec 22, 2020 · DE

Device Facts

Record IDK202974
Device NameDenTek Fresh Protect Dental Guard
ApplicantMedtech Products, Inc.
Product CodeOBR · DE
Decision DateDec 22, 2020
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The DenTek™ Fresh Protect™ Dental Guard is an over-the-counter (OTC) device that is intended to be used by lay people for protection against the effects of nighttime teeth grinding. The DenTek™ Fresh Protect™ Dental Guard is indicated for protection against bruxism or nighttime teeth grinding. It is intended to reduce damage to the teeth and reduce the noise associated with bruxing or grinding.

Device Story

The DenTek Fresh Protect Dental Guard is an over-the-counter, ready-to-wear, one-size-fits-all dental guard for adult use at home. Constructed of ethylene/vinyl acetate copolymer via injection molding, the device features a full-occlusion design worn on the lower teeth. It acts as a physical barrier to maintain separation between upper and lower teeth during sleep. By preventing direct tooth-to-tooth contact, the guard reduces noise and damage caused by bruxism. The device is intended for short-term use (1-3 days) and is discarded after its useful life. Users self-operate the device by placing it in the oral cavity; no clinician involvement is required. The device provides a passive mechanical benefit by cushioning the teeth.

Clinical Evidence

No clinical data. Bench testing only: separation test, fit test, and 3-day durability wear test. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (sensitization and oral mucosa irritation).

Technological Characteristics

Ethylene/vinyl acetate copolymer; injection molded; full-occlusion, one-size-fits-all form factor; dimensions 39.3 mm x 58 mm x 8.35 mm; weight 0.79 grams; non-sterile; home-use; passive mechanical barrier.

Indications for Use

Indicated for adults 18 and older for protection against bruxism or nighttime teeth grinding; intended to reduce tooth damage and noise associated with grinding.

Regulatory Classification

Identification

To protect the teeth and reduce damage caused by bruxing or nighttime grinding and prevention of the noise associated with bruxing and grinding.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The text reads "U.S. FOOD & DRUG ADMINISTRATION". December 22, 2020 Vincent Argiro Director, Regulatory Affairs Medtech Products, Inc. 660 White Plains Rd. Tarrytown, New York 10591 Re: K202974 Trade/Device Name: DenTek Fresh Protect Dental Guard Regulatory Class: Unclassified Product Code: OBR Dated: September 29, 2020 Received: September 30, 2020 Dear Vincent Argiro: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Michael E. Adiodha. M.ChE. Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K202974 Device Name DenTek™ Fresh Protect™ Dental Guard Indications for Use (Describe) The DenTek™ Fresh Protect™ Dental Guard is indicated for protection against bruxism or nighttime teeth grinding. It is intended to reduce damage to the teeth and reduce the noise associated with bruxing or grinding. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Medtech. The word "MEDTECH" is in large, bold, blue letters. Underneath the word is a blue line, and below that is the text "a PrestigeBrands company", with "PrestigeBrands" in red. # 510(k) Summary # DenTek™ Fresh Protect™ Dental Guard K202974 K # #### 1. Submitter | Name & Address: | Medtech Products Inc.<br>660 White Plains Road<br>Tarrytown, NY, 10591 | |-----------------|------------------------------------------------------------------------| | Contact: | Vincent Argiro, RAC | | Title: | Director, Regulatory Affairs | | Phone number: | (914) 524-8721 | | Email: | vargiro@prestigebrands.com | - 2. Date Prepared December 21, 2020 ### 3. Device Identification | Trade/Proprietary Name: | DenTek™ Fresh Protect™ Dental Guard | |-------------------------|-----------------------------------------------------------------------------------| | Common/Usual Name: | Over-the-Counter Dental Guard | | Classification Name: | Mouthguard, Over-the Counter | | Regulation Number: | N/A | | Product Code: | OBR | | Device Class: | Unclassified | | Classification Panel: | Division of Anesthesiology, General Surgery, Infection Control and Dental Devices | ### 4. Legally Marketed Predicate Device(s) | Predicate Device: | DenTek™ Ready-Fit™ Disposable Dental Guard (K151149), Product Code | |-------------------|--------------------------------------------------------------------| | | OBR | {4}------------------------------------------------ #### 5. Device Description The DenTek™ Fresh Protect™ Dental Guard is an over-the-counter (OTC) device to be used by lay people for protection against the effects of nighttime teeth grinding. The guard is constructed of an ethylene/vinyl acetate copolymer and utilizes a ready-to-wear, one-size-fits-all design. The fully occlusive guard is worn on the lower teeth, maintaining separation between the upper and lower teeth, thereby preventing the noise and damage associated with teeth grinding. #### 6. Intended Use The DenTek™ Fresh Protect™ Dental Guard is an over-the-counter (OTC) device that is intended to be used by lay people for protection against the effects of nighttime teeth grinding. The DenTek™ Fresh Protect™ Dental Guard bears the following indications for use statement: The DenTek™ Fresh Protect™ Dental Guard is indicated for protection aqainst bruxism or nighttime teeth grinding. It is intended to reduce damage to the teeth and reduce the noise associated with bruxing or grinding. The DenTek™ Fresh Protect™ Dental Guard intended use statement is identical to that of the predicate device. ### 7. Substantial Equivalence Discussion The following table compares the DenTek™ Fresh Protect™ Dental Guard to the chosen predicate device with respect to intended use and technological characteristics. This comparison of the devices provides detailed information demonstrating the basis for the determination of substantial equivalence. Any minor differences are discussed in the narrative below the table. | | Proposed Device | Predicate Device | Differences | |-----------------------------|-------------------------------------------|-------------------------------------|-------------| | Trade Name /<br>Device Name | DenTek™ Fresh<br>Protect™ Dental<br>Guard | DenTek™ Ready-<br>Fit™ Dental Guard | N/A | | 510(k) Number | TBD | K151149 | N/A | | Date Cleared | 90 days from date<br>of receipt by FDA | 09/30/2015 | N/A | | Original applicant | Medtech Products<br>Inc. | DenTek Oral Care,<br>Inc. | N/A | | REGULATORY CLASSIFICATION | | | | | Regulatory Class | Unclassified | Unclassified | None | {5}------------------------------------------------ | Name of Generic<br>Device Type | Mouthguard,<br>Over-the-Counter | Mouthguard, Over-<br>the-Counter | None | |---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | Regulation | N/A | N/A | None | | Product Code | OBR | OBR | None | | Applicable<br>Performance<br>Standards or<br>Special Controls | None specified by<br>FDA for Product<br>Code OBR | None specified by<br>FDA for Product<br>Code OBR | None | | DEVICE DESCRIPTION – SUBSTANTIAL EQUIVALENCE COMPARATORS | | | | | Intended Use | Keep upper and<br>lower teeth<br>separated during<br>sleep. | Keep upper and<br>lower teeth<br>separated during<br>sleep. | None | | OTC or Rx | OTC | OTC | None | | Indications for Use | For protection<br>against bruxism or<br>nighttime teeth<br>grinding. It is<br>intended to reduce<br>damage to the<br>teeth and reduce<br>the noise<br>associated with<br>bruxing or grinding. | Ready-Fit<br>Disposable Dental<br>Guard is indicated<br>for protection for<br>nighttime teeth<br>grinding or bruxism.<br>It is intended to<br>reduce damage to<br>teeth by cushioning<br>them and keeping<br>them apart during<br>grinding. | None | | Target Population | Adults 18 and older | Adults 18 and older | None | | Duration of Use | 1-3 days | 1 day | Minor (testing conducted to<br>confirm durability to 3 days, see<br>Section 18 - Bench Testing) | | Technological<br>Characteristics | Flexible guard used<br>as a barrier<br>between teeth. | Flexible guard used<br>as a barrier between<br>teeth. | None | | DEVICE DESCRIPTION - DESIGN FEATURES | | | | | Material(s) | Ethylene/vinyl<br>acetate copolymer) | Ethylene/methyl<br>acrylate copolymer) | Minor (biocompatibility testing<br>confirmed the DenTek™ Fresh<br>Protect™ resin is appropriate for<br>this intended use) | | Design /<br>Presentation | Full occlusion,<br>ready-to-wear<br>dental guard | Posterior occlusion,<br>ready-to-wear<br>dental guard | Minor (the full occlusion design is<br>common in OTC dental guards) | | Method of | Injection molding | Injection molding | None | {6}------------------------------------------------ | Manufacture | | | | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------| | Method of Cleaning | Rinse with water<br>before first use; if<br>using for more<br>than one night,<br>wash between uses<br>and inspect for<br>signs of wear. | The device is meant<br>to be discarded<br>after one night's<br>use. No instructions<br>for cleaning are<br>included. | Minor (the basic cleaning<br>instruction provided are similar<br>to those seen on other OTC<br>dental guards) | | Fit | One size fits all | One size fits all | None | | Dimensions (L x W x<br>H) | 39.3 mm x 58 mm x<br>8.35 mm | 38 mm x 39 mm x<br>19 mm | Minor (similar to the predicate) | | Weight | 0.79 grams (0.03<br>ounces) | 1.13 grams (0.04<br>ounces) | Minor (similar to the predicate) | | Accessories | None | None | None | | Sterile Device | No | No | None | | Use environment | Home | Home | None | | Anatomical site of<br>use | Oral Cavity | Oral Cavity | None | The table above identifies a few minor differences between the proposed device and the cited predicate. While the contact material for both devices is an ethylene-based copolymer, the exact resins are not identical; the appopriateness of the subject device's resin for use in a dental guard was demonstrated through biocompatability testing of the finished guards, which is provided in this application. Also, while both guards are labeled as "disposable" due to the short term intended use of each individual guard, the specific durations of use per the products' labeling differs: 1 day for the predicate and 1-3 days for the proposed device; this difference is minor because the DenTek™ Fresh Protect™ Dental Guard was shown adequately durable in a 3-day use simulation, the report for which is provided with this application. Another minor difference is that the predicate utilizes a posterior-occlusion design where the subject device – like many other OTC dental guards (refer to K091660 and K083400) - utilizes a full-occlusion design. Finally, though not identical, the measurements of the proposed guard are similar to the cited predicate (i.e., the resting bite radius is wider than the predicate because the predicate has a narrow strap between the bite pads that does not retain its in-use width) with a lower overall weight as intended to deliver an ultra-light guard. In accordance with section 513(i)(1)(A) of the FDCA, a device is substantially equivalent (SE) when it has the same intended use and technological characteristics as a legally marketed predicate device. As demonstrated in this traditional 510(k), any differences between the subject device and the cited predicate do not raise different questions of safety or effectiveness and this application establishes that the device is as safe and effective as the predicates. {7}------------------------------------------------ #### 8. Non-Clinical Performance Data The following bench testing was conducted to confirm the performance of the DenTek™ Fresh Protect™ Dental Guard: - Separation test. Simulation study demonstrating the finished guard's ability to keep the ● teeth separated when the jaw is clenched. - . Fit test. Simulation study demonstrating the finished guard's ability to accommodate a range of arch widths without heating or adjustments. - . Wear test. Simulation study demonstrating the finished guard's durability for the intended use duration of 1-3 days. The non-clinical testing of DenTek™ Fresh Protect™ Dental Guard demonstrates the device's performance in providing teeth separation, accommodating different mouth sizes, and lasting the expected use life, which supports its substantial equivalence to the cited predicate. Additionally, the following tests for biocompatibility were conducted on both the DenTek™ Fresh Protect ™ Dental Guard and the provided forming tray: - In Vitro Cytotoxicity Assay (Elution method) in accordance with ISO 10993-5; ● - . Guinea Pig Maximization Test (Sensitization) in accordance with ISO 10093-10, Section 7.5 and Annex E; - . Oral Mucosa Irritation Test in accordance with ISO 10993-10 Annex B.3. These studies demonstrated that the DenTek™ Fresh Protect ™ Dental Guard is not cytotoxic, is not a contact skin sensitizer, and is not irritating to the buccal mucosa, which further supports the device's substantial equivalence to the predicate. ### 9. Clinical Performance Data There are no differences in intended use and technological characteristics between the DenTek™ Fresh Protect™ Dental Guard and the predicate that necessitate conducting a clinical trial. ### 10. Statement of Substantial Equivalence As demonstrated in this application, the proposed device, the DenTek™ Fresh Protect™ Dental Guard, has the same intended use as the identified predicate device, the DenTek™ Ready-Fit™ Dental Guard (K151149), and employs the same basic technological characteristics; any differences between the proposed device and the predicates are minor and do not constitute different technological characteristics. The relevant information on biocompatibility and the performance testing confirm the DenTek™ Fresh Protect™ Dental Guard fulfills its intended use as safely and effectively as the legally marketed predicate device. The DenTek Fresh Protect™ Dental Guard is therefore substantially equivalent to the cited predicate.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...