DR. HAYS BITE GUARD

K014079 · Inventive Resources, Inc. · MQC · Feb 22, 2002 · DE

Device Facts

Record IDK014079
Device NameDR. HAYS BITE GUARD
ApplicantInventive Resources, Inc.
Product CodeMQC · DE
Decision DateFeb 22, 2002
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

Protecting against teeth grinding, bruxism and jaw clenching. Short-term pain relief from muscle spasm pain due to occlusal interference. For the prevention of chronic tension and temporal mandibular joint (TMJ) syndrome that is caused by chronic jaw clenching of the posterior mandibular and maxillary teeth by the temporalis muscle. The device is custom made for the individual.

Device Story

Dr. Hays Bite Guard is a custom-molded, horseshoe-shaped dental tray; constructed from polycarbonate (Lexan) and thermoplastic agent (Elvax). Supplied non-sterile to healthcare practitioners; practitioners heat, mold, and trim device to fit individual patient anatomy. Used by patients at home to manage occlusal disorders, bruxism, and jaw clenching. Device acts as physical barrier/cushion between mandibular and maxillary teeth; prevents clenching-induced muscle spasms and TMJ syndrome. Patient cleans device with soap and water. Benefits include reduction of dental wear, relief of muscle pain, and prevention of chronic tension associated with TMJ disorders.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on design, material composition, and intended use comparisons to legally marketed predicate devices.

Technological Characteristics

Horseshoe-shaped dental tray; materials: Lexan (polycarbonate) and Elvax (thermoplastic agent). Custom-molded to patient anatomy by clinician. Non-sterile; reusable by single patient. Cleaning via soap and water. No electronic components or software.

Indications for Use

Indicated for patients requiring protection against teeth grinding, bruxism, and jaw clenching; provides short-term pain relief from muscle spasms due to occlusal interference; prevents chronic tension and TMJ syndrome caused by chronic clenching of posterior teeth by the temporalis muscle. Custom-made for individual patients.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K014079 # 10(k) Premarket Notification Summary of Safety and Effectiveness Information Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807,92 ## 1. Device Name: FEB 2 2 2002 Trade Name: Dr. Hays Bite Guard Common Name: Night Guard Classification Name: Device, Jaw Repositioning #### 2. Establishment Name & Registration Number: Name: Inventive Resource, Inc. Number: Pending #### 3. Classification: Unclassified Device Product Code(s): LQZ Device Class: unclassified Classification Panel: Anesthesia Devices Panel & Dental Device Panel #### Contact Person: 4. Mr. John A. Paoluccio Inventive Resources, Inc. 5038 Salida Blvd. Salida, CA 95368 209.545.2616 - 209.545.3533 - fax ### Submission Correspondent: Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 925.356.2640 / 925.356.2654 FAX #### ડ. Description of the Device: The Dr. Hays Bite Guard is designed to assist in the treatment of occlusal disorders resulting in teeth grinding, jaw clenching, and mandibular muscle spasm. The device is comprised of a thin horseshoe shaped plastic (polycarbonate, Lexan) tray filled with a thermoplastic agent (Elvax). The device is supplied to health care practioners to heat, mold, and otherwise custom fit and trim each device to the patient. Materials: the Dr. Hays Bite Guard is made of Lexan and Elvax, the same materials as the Snore Guard device, cleared under K882303. The device is supplied Non-Sterile. The device is packaged in "clean only" condition but is free of manufacturing debris and/or residues. Components are inspected after processing to evaluate and document the removal of manufacturing residue and debris. However, it is recommended that the device be removed from its shipping and packing materials and washed with warm soap and water before use. The device may not be autoclaved. {2}------------------------------------------------ The Dr. Hays Bite Guard is non-sterile and must be cleaned and sanitized before first and all Sanitizing. subsequent uses. Wash the device thoroughly with hot water using a typical commercial detergent or antibacterial soap. Rinse completely in running warm tap water. Allow to air dry and return the device to its case. See Appendix I for complete instructions. Packaging. Materials used in the packaging of the device are typical injection molded plastic boxes. Carton and shippers are made from appropriate paper based products. Storage. The device should be stored clean and dry. The location should be at the bedside or bathroom on in locations where ambient temperatures do not exceed 130 degrees F. #### 6. Comparison to Predicate Device(s): Dr. Hays Bite Guard may be directly contrasted with the following equivalent devices: - SnoreGuard™K882303. . - Best Bite Decluder™ K980953 . - . NTI Clenching Suppression System K981546 ### Summary Basis for Equivalence and Comparison Table: Based on the available information concerning the referenced comparison devices are similar in that: The devices have the same intended use and indications for use. The devices are made of the same basic materials. The devices have a similar design, form and function. The referenced comparison devices demonstrate that the Dr. Hays Bite Guard is substantially equivalent . The table on the following page compares the available legally marketed devices. | Comparison Table: | | | | | |-------------------|--|--|--|--| |-------------------|--|--|--|--| | FEATURE | Dr. Hays Bite Guard | Snore Guard | Best Bite | NTI System | SE? | |----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|---------------|---------------|-----| | Indications for<br>Use: | Protecting against teeth grinding,<br>bruxism and jaw clenching.<br>Short-term pain relief from muscle<br>spasm pain due to occlusal<br>interference.<br>For the prevention of chronic tension<br>and temporal mandibular joint<br>(TMJ) syndrome that is caused by<br>chronic jaw clenching of the<br>posterior mandibular and maxillary<br>teeth by the temporalis muscle. The<br>device is custom made for the<br>individual. | Snoring - sometimes used<br>for Bruxism | Same | Same | Yes | | Materials: | Lexan & Elvax | Same | Thermoplastic | Thermoplastic | Yes | | Design: | Custom molded mouth guard | Same | Same | Same | Yes | | Prescription<br>Device: | Yes | Same | Same | Same | Yes | | Reusable: | Yes - Single patient | Yes | Yes | Yes | Yes | | Method of<br>Disinfection: | Warm water, soap & toothbrush | Same | Same | Same | Yes | | K-Number: | Pending | K882303 | K980953 | K981546 | Yes | {3}------------------------------------------------ Image /page/2/Picture/2 description: The image is a black and white seal for the Department of Health and Human Services, USA. The seal is circular, with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other, with flowing lines beneath them. JUL 1 7 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Inventive Resources, Incorporated C/O Mr. David W. Schlerf Buckman Company 200 Gregory Lane, Suite C-100 Pleasant Hill, California 94523-3389 Re: K014079 Trade/Device Name: Dr. Hayes Bite Guard Regulation Number: None Regulation Name: None Regulatory Class: Unclassified Product Code: MQC Dated: November 9, 2001 Received: December 11, 2001 Dear Mr. Schlerf: This letter corrects our substantially equivalent letter of February 22, 2002. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Schlerf Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 http://www.fda.gov/cdrh/organiz.html#OC for OC organization structure). Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Suite Y. Michael DMD. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ K014079 Page 1 of 1 510(k) Number K014079 Dr. Hays Bite Guard Device Name: Indications For Use: - Protecting against teeth grinding, bruxism and jaw clenching. . - Short-term pain relief from muscle spasm pain due to occlusal . interference. - For the prevention of chronic tension and temporal mandibular . joint (TMJ) syndrome that is caused by chronic jaw clenching of the posterior mandibular and maxillary teeth by the temporalis muscle. The device is custom made for the individual. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) | | <img alt="Signature" src="signature.png"/> | |---------------------------------------------------------------------|--------------------------------------------| | (Division Sign-Off) | Susan Runar | | Division of Dental, Infection Control, and General Hospital Devices | | | F10(k) Number | 1501409 | | Prescription Use | <span style="text-decoration: underline;">✓</span> | OR | Over-The-Counter Use | |----------------------|----------------------------------------------------|----|--------------------------| | (Per 21 CFR 801.109) | | | (Optional format 1-2-96) | bite.doc 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...