Apex Locator
K231990 · Cefla S.C. · LQY · Feb 13, 2024 · DE
Device Facts
| Record ID | K231990 |
| Device Name | Apex Locator |
| Applicant | Cefla S.C. |
| Product Code | LQY · DE |
| Decision Date | Feb 13, 2024 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
Indications for Use
APEX LOCATOR is a device intended for locating the apex of the root canal during endodontic treatments.
Device Story
Apex Locator is an accessory for Cefla dental units used to estimate root canal apex position during endodontic procedures. Device measures electrical impedance of dental tissue, which varies by density and composition; utilizes 400Hz and 8kHz frequencies. Input consists of electrical signals from dental tissue via file clip and lip hook; output is displayed on the dental unit interface. Operated by qualified dental personnel in clinics or dental offices. Provides real-time feedback to assist clinicians in determining working length during root canal preparation, potentially improving treatment accuracy and patient outcomes.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), biocompatibility (ISO 10993-1), and reprocessing validation.
Technological Characteristics
Impedance-based apex locator; measures electrical impedance at 400Hz and 8kHz. Nominal measuring voltage <200mV AC. Powered by Cefla dental unit. Components include file clips, lip hooks, and external wiring. Biocompatible per ISO 10993-1. Software documentation provided per FDA guidance for basic level of concern.
Indications for Use
Indicated for patients undergoing endodontic treatments requiring root canal apex localization.
Predicate Devices
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
February 13, 2024
Cefla S.C. Daidone Simona Regulatory Affairs Via Selice Provinciale N.23/A Imola, Bologna 40026 ITALY
Re: K231990
Trade/Device Name: Apex Locator Regulatory Class: Unclassified Product Code: LQY Dated: January 17, 2024 Received: January 17, 2024
Dear Daidone Simona:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{2}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}------------------------------------------------
Sincerely,
# Michael E. Adjodha -S
Michael E. Adjodha, MChE, RAC, CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231990
Device Name APEX LOCATOR
Indications for Use (Describe)
APEX LOCATOR is a device intended for locating the apex of the root canal during endodontic treatments.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{5}------------------------------------------------
# K231990
# 510(k) SUMMARY, AS REQUIRED BY CFR 807.92
| Submitter's Name: | CEFLA S.C. |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Via Selice Provinciale 23/a<br>Imola, BO 40026 ITALY<br>Tel. +39 0542 653111<br>Fax +39 0542 653444 |
| Establishment<br>Registration<br>Number: | 3006610845 |
| Summary<br>Preparation Date: | February, 12th, 2024 |
| Contact Person: | Simona Daidone, Regulatory Affairs |
| Telephone<br>Number: | +39 0542 653441 |
| Email: | regulatory@cefla.it |
| Trade/Device<br>name: | Apex Locator |
| Common or Usual<br>Name: | Dental Root Apex Locator |
| Classification<br>Name: | Classification Name: Unclassified<br>Device Class: Unclassified<br>Product Code: LQY |
| Description: | APEX LOCATOR is a device used in dentistry to estimate the position of the root apex of teeth during root canal preparation. It is used connected with<br>the Cefla dental units. Its principle of operation relies on measuring the electrical impedance of dental tissue, which varies according to tissue density<br>and composition.<br>The APEX LOCATOR is an accessory device for the Cefla dental units used to estimate the position of the root apex of teeth during root canal preparation.<br>The kit is composed of the following elements: |
{6}------------------------------------------------
# CEFLA S.C. 510(k) Premarket Notification
| • | file clip (2 pieces) | | |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------|----------------|---------------------------|
| • | lip hook (4 piece) | | |
| • | external wiring (1 piece) | | |
| Indication for Use: | APEX LOCATOR is a device intended for locating the apex of the root canal during endodontic treatments. | | |
| Identification of<br>Predicate Device: | CEFLA S.C. will refer to the following predicate device (1): | | |
| | Proprietary Name: Guilin Woodpecker Medical Instrument | | |
| | Classification Name: Unclassified | | |
| | Registered Establishment Name: 3005581016 | | |
| | Guilin Woodpecker Medical Instrument Co., Ltd. | | |
| | Information Industrial Park, Guilin National High-Tech Zone | | |
| | Guilin City, CN 541004Owner/Operator: W&H | | |
| | Establishment Operations: Manufacturer | | |
| | 510 (k): K181087 | | |
| | Device Name: APEX LOCATOR DPEX III | | |
| | Applicant: Guilin Woodpecker Medical Instrument | | |
| Comparison of<br>technological<br>characteristics | Characteristics | Subject Device | Predicate Device |
| | Device Trade Name | Apex Locator | Apex Locator, DPEX III |
| | 510K Applicant | Cefla S.C. | GUILIN WOODPECKER MEDICAL |
| technological<br>characteristics<br>with the<br>predicate and<br>reference<br>devices: | Device Trade Name | Apex Locator | Apex Locator, DPEX III | |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|--|
| | 510K Applicant | Cefla S.C. | GUILIN WOODPECKER MEDICAL<br>INSTRUMENT CO., LTD | |
| | 510(K) Number | - | K181087 | |
| | Regulation Number | Unclassified | Unclassified | |
| | Product Code | LQY | LQY | |
| | Classification Name | Root Apex Locator | Root Apex Locator | |
| | OTC or Prescription | Prescription Use | Prescription Use | |
| | Medical Specialty | Dental | Dental | |
| | Indications for Use | APEX LOCATOR is a device intended<br>for locating the apex of the root<br>canal during endodontic treatments. | DPEX III is a microprocessor-<br>controlled device used for locating<br>the apex of root canal. | |
| | Electrical Power | Dental Unit power supply | Rechargeable lithium battery(3.7V) | |
| | Adapter | Not applicable | Input: AC100-240V 50/60Hz<br>Output: DC5V 1A | |
| | Display | On the dental unit | Custom colorized 4.5" LCD | |
| | Measuring Voltage | Nominal – doesn't exceed 200mV<br>AC | Nominal - doesn't exceed 200mV<br>AC | |
| | Frequencies used<br>for measurements | 400Hz & 8kHz | 400Hz & 8kHz | |
| | Method of<br>Calculation | The ratio of impedance at two<br>frequencies | The ratio of impedance at two<br>frequencies | |
| | Accessory | Measuring wire, File clip, Lip hook,<br>Touch probe | Measuring wire, File clip, Lip hook,<br>Touch probe, Adapter, Tester | |
| | Operating<br>Environment | Hospital environment, clinics or<br>dental offices by qualified dental<br>personnel | Hospital environment, clinics or<br>dental offices by qualified dental<br>personnel | |
| | Button | On the dental unit | Three push buttons:<br>1. ON/OFF<br>2. Sound adjustment<br>3. Demonstration function | |
| | Sound Indication | On the dental unit | Piezo transducer with sound level<br>control (high, medium, low, mute) | |
| | Automatic Turn off<br>function | Not present | The device turns off automatically<br>after 5 minutes of the idle state. | |
| | Calibration | Not required | Not required | |
| | Biocompatibility | Complies with ISO10993-1 | Complies with ISO10993-1 | |
| Non-clinical<br>Performance<br>Testing: | | Electrical safety Test was conducted and performed in accordance with IEC 60601-1. | | |
| | | Electromagnetic compatibility Test was conducted and performed in accordance with IEC 60601-1-2. | | |
| | | Usability test was conducted in accordance to IEC 60601-1-6. | | |
| | Basic level software documentation per Content of Premarket Submissions for Device Software Functions | | | |
| | | Evaluation of biocompatibility is based on: | | |
| | ISO 10993-1 "Biological evaluation of medical devices - Part 1: Evaluation and Testing"; | | | |
| | Validation of the reprocessing according "FDA guidance Reprocessing Medical Devices Health Care Settings: Validation Methods and Labeling." | | | |
| | | Performance tests have been performed according to internal protocol and method. | | |
| Clinical Testing: | | Clinical performance testing was not conducted. | | |
| | | | | |
| Conclusion: | | CEFLA S.C. considers the Apex Locator to be substantially equivalent to the predicate device listed above. This conclusion is based on the similarities in | | |
| | | intended use, principle of operation, functional design, and established medical use. Differences between the devices shown in the comparison section | | |
| | | above are minor and do not have any negative effect on substantial equivalence. | | |
{7}------------------------------------------------
# CEFLA S.C. 510(k) Premarket Notification
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------