ARGYLE AQUA-SEAL ALL PURPOSE AUTOTRANSFUSION CHEST DRAINAGE UNIT AND OPTIONAL ARGYLE AQUA-SEAL EVACUATING AUTOTRANSFUSIO
K970857 · Davis & Geck, Inc. · CAC · Nov 5, 1997 · Anesthesiology
Device Facts
Record ID
K970857
Device Name
ARGYLE AQUA-SEAL ALL PURPOSE AUTOTRANSFUSION CHEST DRAINAGE UNIT AND OPTIONAL ARGYLE AQUA-SEAL EVACUATING AUTOTRANSFUSIO
Applicant
Davis & Geck, Inc.
Product Code
CAC · Anesthesiology
Decision Date
Nov 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory is a sterile, single-use accessory to the Argyle® Aqua-Seal® All Purpose Autotransfusion CDU and is intended to be used for the purpose of transferring the lost blood recovered by the CDU into a conventional blood bag. The blood can subsequently be re-infused back into the patient using a conventional pressure cuff or by gravity means.
Device Story
Accessory to Argyle Aqua-Seal Autotransfusion Chest Drainage Unit (CDU); facilitates transfer of recovered post-operative blood from CDU to conventional blood bag. Device consists of blood bag within air-tight vessel; input line connects to CDU infusion tube assembly. Hospital vacuum applied to vessel suction port creates pressure differential; bag expands, drawing blood from CDU into bag. Blood subsequently re-infused via pressure cuff or gravity. Used in clinical settings; operated by healthcare professionals. Benefits patient by enabling autologous blood transfusion, reducing need for donor blood.
Clinical Evidence
Bench testing only. Device met ANSI/AAMI standards for autologous devices. Biocompatibility testing performed per ISO 10993. Performance testing conducted under simulated conditions of use.
Technological Characteristics
Sterile, single-use accessory. Materials: PVC blood bag, air-tight vessel. Principle: vacuum-assisted fluid transfer. Connectivity: connects to CDU infusion tube assembly and hospital vacuum source. Complies with ANSI/AAMI standards for autologous devices and ISO 10993 biocompatibility requirements.
Indications for Use
Indicated for evacuation and collection of blood/air from mediastinal and pleural cavities in post-operative and trauma patients; collection and reinfusion of autologous blood; prevention of fluid/air re-accumulation; and facilitation of lung re-expansion and restoration of normal breathing dynamics.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
Atrium Self-Sealing ATS bag
Reference Devices
Argyle® Aqua-Seal® Continuous Autotransfusion Chest Drainage Unit
Atrium Multipurpose Chest Drainage Unit
Related Devices
K963850 — PLEUR-EVAC SAHARA PLUS MODEL S-1150 CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM · Deknatel, Inc. · Apr 22, 1997
K963098 — PLEUR-EVAC SHARA PLUS MODEL S-2150/CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM(S-100) · Deknatel, Inc. · Nov 8, 1996
K022489 — SUMMIT MEDICAL LTD CELLTRANS POSTOPERATIVE AUTOTRANSFUSION SYSTEM, MODELS ABT4510, ABT4512 & ABT4514 · Summit Medical , Ltd. · Oct 9, 2002
K140197 — PLEUR-EVAC PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM · Teleflexmedical, Inc. · Oct 21, 2014
K120953 — PLEUR-EVAC SAHARA PLUS CONTINOUS REINFUSION AUTOTRANSFUSION SYSTEM · Tlelflex Medical, Inc. · Dec 10, 2012
Submission Summary (Full Text)
{0}
K410851
# ATTACHMENT # 1.
510 (k) Summary of Safety and Efficacy - Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory
NOV - 5 1997
{1}
Submitted by: Sherwood - Davis & Geck
1915 Olive St.
St. Louis, MO 63103
(314) 621-7788
Contact: Vanada Johnson
Date Prepared: 4 March, 1997
The Argyle® Aqua-Seal® All Purpose Autotransfusion Chest Drainage Unit (CDU) and Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory (Evacuating Accessory) is classified as an “Auto-transfusion Device”, Class II (performance standards) under section 868.5830 of the Code of Federal Regulations (CFR).
The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory is a sterile, single-use accessory to the Argyle® Aqua-Seal® All Purpose Autotransfusion CDU and is intended to be used for the purpose of transferring the lost blood recovered by the CDU into a conventional blood bag. The blood can subsequently be re-infused back into the patient using a conventional pressure cuff or by gravity means.
The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory consists of a conventional blood bag contained in an air-tight vessel. The input line from the bag passes through a sealed opening located at the top of the vessel. This line attaches to the spike port “Y” connector of the infusion tube assembly of the CDU. Transfer of blood from the CDU to the bag commences when a vacuum line is removed from the CDU and attached to the suction port of the Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory. The pressure differential across the surface of the bag compels the bag to expand and thereby drawing blood from the CDU into the bag.
The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory is substantially equivalent to the Atrium Self-Sealing ATS bag in that:
1. Each device allows for the batch transfer of a patient’s lost post-operative blood from the CDU into an external container bag.
2. Each device operates on the principle that fluid will rush to fill a vacuum. The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory utilizes hospital vacuum while the Atrium Self-Sealing ATS bag creates its own vacuum by means of a spring.
3. Each device is designed for single use only.
The differences between the Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory and the Atrium Self-Sealing ATS bag are:
1. The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory operates on hospital vacuum while the Atrium Self-Sealing ATS bag utilizes the stored potential energy of a fully compressed metal spring.
{2}
2. The Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory is a conventional PVC blood bag. The Atrium Self-Sealing ATS bag is thick PVC and inflexible because of the pressure plates and spring built within. As such the Atrium Self-Sealing ATS bag will not fit in a conventional pressure cuff.
In addition, the Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory has undergone testing to ensure that the device meets the applicable blood handling requirements outlined in the American National Standards Institute (ANSI) / Association for the Advancement of Medical Instrumentation (AAMI) Standard for Autologous devices. Biocompatibility testing in accordance with ISO 10993 requirements has been done, as has performance testing under simulated conditions of use. Results of all testing demonstrate the safety and effectiveness of the Argyle® Aqua-Seal® Evacuating Autotransfusion Accessory for its intended use.
The Argyle® Aqua-Seal® All Purpose Autotransfusion Chest Drainage Unit is also substantially equivalent to the Argyle® Aqua-Seal® Continuous Autotransfusion Chest Drainage Unit and to the Atrium Multipurpose Chest Drainage Unit in design, function and appearance. The only differences between the two are:
1. The Argyle® Aqua-Seal® All Purpose Autotransfusion CDU incorporates the automatic negative pressure relief valve with a manual override feature whereas the Argyle® Aqua-Seal® Continuous Autotransfusion CDU and the Atrium Multipurpose CDU utilize a strictly manual negative pressure relief valve.
2. While all three CDU’s incorporate shut-off valves to prevent the loss of the water seal, the shut-off valves on the Argyle® Aqua-Seal® All Purpose Autotransfusion CDU and the Atrium Multipurpose CDU are flow sensing. Therefore, at low flow rates associated with the autotransfusion process, the valve “shuttles” and remains open to allow for improved blood collection times.
The Argyle® Aqua-Seal® All Purpose Autotransfusion Chest Drainage Unit with the two new features has undergone testing to ensure that the device meets the applicable blood handling requirements outlined in the American National Standards Institute (ANSI) / Association for the Advancement of Medical Instrumentation (AAMI) Standard for Autologous devices. Biocompatibility testing in accordance with ISO 10993 requirements has been done, as has performance testing under simulated conditions of use. Results of all testing demonstrate the safety and effectiveness of the Argyle® Aqua-Seal® All Purpose Autotransfusion CDU for its intended use.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Mr. Vanada Johnson
Regulatory Affairs Specialist
Sherwood Davis & Geck
444 McDonnell Boulevard
Hazelwood, Montana 63042-2516
NoV - 5 1997
Re: K970857
Argyle® Aqua-Seal All Purpose Autotransfusion Chest Drainage Unit
Regulatory Class: II (Two)
Product Code: CAC
Dated: August 6, 1997
Received: August 7, 1997
Dear Mr. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Mr. Vanada Johnson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
868.5850 - CAC II - Autotransfusion Apparatus
510(k) Number (if known): K 970857
Device Name: S 001
Indications For Use:
1. Evacuation and collection of blood and/or air from the mediastinal and pleural cavity in post-operative and trauma situations.
2. Collection and reinfusion of autologous blood from the mediastinal and pleural cavity in post-operative and trauma situations.
3. Prevention of fluid and/or air re-accumulation in the mediastinal and pleural cavity.
4. Facilitation of complete lung re-expansion and restoration of normal breathing dynamics.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970857
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.