MEDI-TRACE COMBINATION DEFIBRILLATION & ECG ELECTRODE FOR LAERDAL
Device Facts
| Record ID | K955923 |
|---|---|
| Device Name | MEDI-TRACE COMBINATION DEFIBRILLATION & ECG ELECTRODE FOR LAERDAL |
| Applicant | Graphic Controls Corp. |
| Product Code | MKJ · Cardiovascular |
| Decision Date | Oct 24, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.5310 |
| Device Class | Class 3 |
| Attributes | Therapeutic |
Indications for Use
Intended for use in defibrillation procedures. This device also allows for ECG monitoring. There are no population restrictions for the device except as noted in the directions for use.
Device Story
Medi-Trace 1110L is a pre-gelled, self-adhesive electrode; used for defibrillation and ECG monitoring. Components: cable attachment, vinyl insulating layer, metal foil, conductive adhesive hydrogel, pressure-sensitive adhesive ring, release liner. Packaged in pairs in heat-sealed pouches. Used by clinicians in clinical settings; provides electrical interface between defibrillator/monitor and patient skin. Facilitates energy delivery for defibrillation and signal acquisition for ECG monitoring.
Clinical Evidence
Bench testing only. Device met or exceeded ANSI/AAMI DF39-1993 specifications for electrical testing and labeling. Biocompatibility testing (ISO-10993) confirmed device is non-irritating, non-cytotoxic, and non-sensitizing. Accelerated shelf-life studies substantiated claimed shelf life.
Technological Characteristics
Materials: vinyl insulating layer, metal foil, conductive adhesive hydrogel, pressure-sensitive adhesive ring. Standards: ANSI/AAMI DF39-1993 (electrical/labeling), ISO-10993 (biocompatibility). Form factor: self-adhesive, pre-gelled electrode. Energy: passive electrical conductor for defibrillation/ECG.
Indications for Use
Indicated for patients requiring defibrillation procedures and ECG monitoring. No specific age or gender restrictions.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
- Laerdal Heartstart® Defibrillation Electrode
Related Devices
- K980857 — MEDI-TRACE 1310P COMBINATION DEFIBRILLATION, MONITORING & PACING ELECTRODE · Graphic Controls Corp. · May 12, 1998
- K983287 — MEDI-TRACE 1610C COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE · Graphic Controls Corp. · Nov 5, 1998
- K955882 — MEDI-TRACE COMBINATION DEFIBRILLATION & ECG ELECTRODE FOR HEWLETT PACKARD DEFIBRILLATORS · Graphic Controls Corp. · Oct 24, 1996
- K972970 — MEDI-TRACE 1510M COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE · Graphic Controls Corp. · Aug 27, 1997
- K041883 — SKINTACT MULTIFUNCTION ELECTRODES, MODEL DF 21 · Leonhard Lang GmbH · Sep 7, 2004