K070455 · Zoll Medical Corporation, World Wide Headquarters · MKJ · Apr 26, 2007 · Cardiovascular
Device Facts
Record ID
K070455
Device Name
ZOLL E SERIES WITH SEE-THRU CPR
Applicant
Zoll Medical Corporation, World Wide Headquarters
Product Code
MKJ · Cardiovascular
Decision Date
Apr 26, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation. Note: This device must not be connected to internal pacemaker electrodes. The purposes of pacing include: Resuscitation from standstill or bradycardia of any etiology: Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes. As a standby when standstill or bradycardia might be expected: Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phiebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing. Suppression of tachycardia: Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia. Pediatric Pacing: Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL. pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended. This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP). ECG monitoring is performed by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles. SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions. EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients. 12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias. NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients. The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients. The CPR monitoring function is not intended for use on patients under 8 years of age. Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse. This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats. The E Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient. They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol. The E Series products must be prescribed for use by a physician or medical advisor of an emergency response team. Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by: Unconsciousness, Absence of breathing, Absence of pulse.
Device Story
ZOLL E Series is a multi-parameter monitor/defibrillator/pacemaker for emergency/clinical use. Inputs: ECG (via leads/paddles/pads), SpO2, EtCO2, NIBP, and chest compression data. Operation: Manual or semiautomatic (AED) defibrillation; external pacing; vital sign monitoring; Real CPR Help provides audio/visual feedback on compression rate/depth. Output: Defibrillation shocks, pacing pulses, real-time physiological data, and CPR performance metrics. Used by emergency personnel/physicians to treat cardiac arrest, bradycardia, and tachycardia. Benefits: Rapid resuscitation, non-invasive pacing alternative to transvenous methods, and improved CPR quality via real-time feedback. Healthcare providers use output to guide resuscitation protocols and clinical interventions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Multi-parameter monitor/defibrillator/pacemaker. Energy: DC defibrillator (up to 200J). Sensing: ECG, SpO2, EtCO2, NIBP. Connectivity: 3/5 lead cables, MFE pads, paddles. Software: Real CPR Help feedback algorithm. Form factor: Portable emergency unit.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring cardiac monitoring, external pacing, or defibrillation. Specific populations: cardiac arrest victims (unconscious, apneic, pulseless); patients with bradycardia, standstill, or tachycardia; patients requiring vital sign monitoring (ECG, SpO2, EtCO2, NIBP). CPR monitoring indicated for patients >= 8 years. Contraindications: Do not connect to internal pacemaker electrodes.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Related Devices
K062177 — ZOLL E SERIES WITH REAL CPR HELP · ZOLL Medical Corporation · Dec 13, 2006
K072923 — ZOLL E SERIES DEFIBRILLATOR · Zoll Medical Corporation, World Wide Headquarters · Apr 24, 2008
K060559 — ZOLL R SERIES DEFIBRILLATOR · ZOLL Medical Corporation · Aug 17, 2006
K111594 — ZOLL E SERIES · Zoll Medical Corporation, World Wide Headquarters · Aug 17, 2011
K091265 — ZOLL E SERIES WITH CPRREADYCHARGE · Zoll Medical Corporation, World Wide Headquarters · Dec 23, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
K070455
# External Pacemaker (Pacer Version Only)
APR 2 6 2007
# Intended Use — Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
## Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
## As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phiebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
#### Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
## Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL. pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
# Monitor
## Intended -- Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is performed by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
## (continued on next page)
{1}------------------------------------------------
# Indications for Use (continued from previous page)
# Intended -- Use Real CPR Help™
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
{2}------------------------------------------------
Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is positioned to the right of the text, which is arranged in a circular pattern around the logo. The text reads "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 6 2007
Zoll Medical Corporation c/o Eileen Boyle 269 Mill Rd Chelmsford, Massachusetts 01824
Re: K070455
Trade/Device Name: E Series with See-Thru CPR Regulation Number: 21 CFR 870.5310 Regulation Name: Regulatory Class: Class III Product Code: MKJ Dated: March 26, 2007 Received: March 26, 2007
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Boyle
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276- (see bottom for #s). Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
B. Zunimana for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):
#### Device Name: ZOLL E Series
# Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
### Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- . Unconsciousness
- . Absence of breathing
- . Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
# Intended Use - Semiautomatic Operation (AED)
The E Series products are designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
The E Series products must be prescribed for use by a physician or medical advisor of an emergency response team.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by:
- Unconsciousness ◆
- Absence of breathing .
- Absence of pulse.
#### (continued on next page)
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Blamming
(División Sign-Off)
Division of Cardiovascular Devices
510(k) Number K070455
Page 1 of 2
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.