K062177 · ZOLL Medical Corporation · MKJ · Dec 13, 2006 · Cardiovascular
Device Facts
Record ID
K062177
Device Name
ZOLL E SERIES WITH REAL CPR HELP
Applicant
ZOLL Medical Corporation
Product Code
MKJ · Cardiovascular
Decision Date
Dec 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL E Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The device is light weight and easy to carry for transport. The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned.
Device Story
ZOLL E Series is a portable external defibrillator/monitor for critical care and transport; operated by qualified medical personnel. Device inputs include ECG signals (via leads/paddles), SpO2, EtCO2, NIBP, and chest compression data (via CPR-D-padz sensor). Device processes inputs to provide real-time CPR feedback (compression rate/depth) via voice/visual prompts and metronome; performs ECG rhythm analysis (Advisory Algorithm) to identify shockable rhythms; delivers defibrillation shocks or noninvasive pacing. Output includes vital sign monitoring, CPR performance metrics, and therapy delivery. Healthcare providers use real-time feedback to adjust CPR technique and monitor patient status, facilitating timely resuscitation and stabilization.
Clinical Evidence
Bench testing only. Performance testing verified that the Real CPR Help feature and software modifications meet functional requirements and performance specifications. Safety testing confirmed compliance with recognized industry standards. No clinical data presented.
Technological Characteristics
Portable defibrillator/monitor. Materials: disposable pre-gelled MFE pads. Sensing: ECG leads, SpO2, EtCO2, NIBP, and CPR sensor (compression depth/rate). Energy: DC defibrillation (up to 200J) and noninvasive pacing. Connectivity: data printing/recording. Software: Advisory Algorithm for rhythm analysis and CPR feedback logic. Sterilization: N/A (disposable pads).
Indications for Use
Indicated for cardiac arrest victims (unconscious, apneic, pulseless) for defibrillation (manual/semiautomatic) and synchronized cardioversion. Indicated for temporary external cardiac pacing (bradycardia/standby). Indicated for multi-parameter monitoring (ECG, SpO2, EtCO2, NIBP) in adult, pediatric, and neonatal patients. CPR monitoring indicated for adults (not for patients <8 years).
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten number, which appears to be "1062177". The numbers are written in a cursive style with varying stroke thicknesses. The image also contains the partial text "OU Medical Corporation" at the bottom left.
ZOLL Me
Worldwide Headquarters
269 Mill Road
Chelmsford, MA 01824-4105
U.S.A.
978 421-9655
978 421-0025 Main Fax
# 510(k) Summary:
Submitter's Name and Address:
ZOLL Medical Corporation 269 Mill Road Chelmsford, MA 01824-4105 (978) 421-9655
Contact Person:
Sean Reynolds
(978) 421-9655, Ext. 9386
Date Summary Prepared:
December 11, 2006
Device:
ZOLL E Series with Real CPR Help™
Classification:
Automatic External Defibrillators: Class III (21 CFR 870.5310)
Cardiopulmonary Resuscitation Aid: Class III (21 CFR 870.5200)
Description:
The ZOLL E Series with Real CPR Help™ provides 'real-time' user CPR assistance when used in conjunction with ZOLL CPR-D-padz™ Multi-function Electrodes. CPR feedback is provided via ZOLL's unique sensor assembly that relays user compression data to the ZOLL E Series. The ZOLL E Series with Real CPR Help™ provides a metronome to encourage rescuers to perform CPR at the AHA/ECR recommended rate of 100 compressions per minute (CPM). The ZOLL E Series additionally provides chest compression performance feedback to the user through displayed symbols, text messages and voice prompts.
The ZOLL E Series External Defibrillator is indicated for the defibrillation, Noninvasive Transcutaneous Pacing and multi-parameter monitoring of patient vital signs, including: ECG Monitoring, Pulse Oximetry, end tidal CO2, 12-Lead ECG monitoring, non-invasive blood pressure measurement, CPR performance and data printing and recording for resting patients in critical care and transport. The ZOLL E Series is intended for use by qualified medical personnel who are trained and authorized to respond to medical emergencies, to facilitate the ability
Image /page/0/Picture/20 description: The image shows the word "ZOLL" in large, bold letters at the top. Below the word, the phrase "Advancing Resuscitation. Today." is written in a smaller font. The image appears to be a logo or advertisement for a company or product related to resuscitation.
DEC 1 3 2006
{1}------------------------------------------------
to monitor and assess the physiological characteristics of the indicated patients in a critical care environment. The device is light weight and easy to carry for transport.
The device is capable of providing Basic Life-Saving (BLS) personnel the option of analyzing a patient's ECG signal via the Advisory feature on the device The Advisory Algorithm will determine if the acquired heart rhythm is shockable or non-shockable and will prompt the end-user to provide therapy in the event of a shock advised determination. The user will be prompted to re-assess the patient in the event a no shock advised determination is returned.
Indications for Use:
# Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
The E Series products must be prescribed for use by a physician or medical advisor of an emergecy response team.
# Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- . Unconsciousness
- . Absence of breathing, and
- . Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricualr fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed).
# Intended Use - Semiautomatic Operation (AED)
The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.
{2}------------------------------------------------
# Intended Use - CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
# External Pacemaker (Pacer Version Only)
## Intended Use - Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
# Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, quinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
### As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, druq toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous therapy that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
# Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
# Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
{3}------------------------------------------------
# Monitor
## Intended-Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is performed by connecting the patient to the unit via the 3 or 5 lead patient cable. MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (EtCO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
## Substantial Equivalence:
The features and functions of the ZOLL E Series with Real CPR Help™ are identical to those of the E Series Defibrillator, with the exception of the incorporation of Real CPR Help™. The characteristics, features and functions of the Real CPR Help™ are substantially equivalent to other ZOLL Defibrillators equipped with CPR Feedback.
# Comparison of Technological Characteristics
The ZOLL E Series with Real CPR Help™ acquires compression depth and rate signals from the CPR sensor incorporated in the currently marketed CPR-D.padz™ Multi-function Electrodes. The acquired depth and rate signals are used to provide CPR compression performance feedback to the user through displayed symbols, text messages and voice prompts which are equivalent to other ZOLL defibrillators with user CPR feedback.
### Performance Testing:
Extensive performance testing ensures that the ZOLL E Series with Real CPR Help™ performs as well as the indicated predicate devices and meets all of its functional requirements and performance specifications. Safety testing assures the device complies with applicable sections of recognized industry and safety standards.
Functional testing of the device's features and functions was conducted to ensure that the modifications to the software did not degrade or impact other product features, functions or performance specifications.
### Conclusion
Performance and safety testing of the ZOLL E Series with Real CPR Help™ demonstrates that its features and functions are substantially equivalent to that of the indicated commercially distributed device with regard to performance, safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three lines forming its body and wings. The seal is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA".
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 3 2006
ZOLL Medical Corporation c/o Sean Reynolds Regulatory Affairs Engineer 269 Mill Road Chelmsford, MA 01824-4105
Re: K062177
Trade/Device Name: ZOLL E Series Defibrillator with Real CPR Help Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III Product Code: MKJ, LIX Dated: November 13, 2006 Received: November 29, 2006
Dear Mr. Reynolds:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Reynolds
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
B.Zimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# SECTION 4 - INDICATIONS FOR USE
510(k) Number (if known): K062177
#### ZOLL E Series Device Name:
## Defibrillator Function
The E Series products contain a DC defibrillator capable of delivering up to 200 joules of energy. It may be used in synchronized mode to perform synchronized cardioversion by using the R-wave of the patient's ECG as a timing reference. The unit uses paddles or disposable, pre-gelled, MFE Pads for defibrillation.
The E Series products must be prescribed for use by a physician or medical advisor of an emergecy response team.
### Intended Use - Manual Operation
Use of the E Series products in the manual mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation as indicated by these three conditions:
- . Unconsciousness
- . Absence of breathing, and
- Absence of pulse.
This product should be used only by qualified medical personnel for converting ventricular fibrillation and rapid ventricular tachycardia to sinus rhythm or other cardiac rhythms capable of producing hemodynamically significant heart beats.
In manual mode, the E Series unit may also be used for synchronized cardioversion to terminate atrial fibrillation (AF) or ventricular tachycardias (VT) by using the R-wave of the patient's ECG as a timing reference. A qualified physician must decide when synchronized cardioversion is appropriate.
The advisory function should be used to confirm ventricualr fibrillation and wide complex ventricular tachycardia (greater than 150 beats per minute) in patients meeting the three conditions indicating lack of circulation (previously listed).
# Intended Use - Semiautomatic Operation (AED)
The E Series AED unit is designed for use by emergency care personnel who have completed training and certification requirements applicable to the use of a defibrillator where the device operator controls delivery of shocks to the patient.
# (continued on next page)
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| All man concerner |
| (Division Sign-Off) |
|---------------------|
|---------------------|
Division of Cardiovascular Devices
| 510(k) Number | k062177 |
|---------------|---------|
|---------------|---------|
{7}------------------------------------------------
### Indications for Use (continued from previous page)
They are specifically designed for use in early defibrillation programs where the delivery of a defibrillator shock during resuscitation involving CPR, transportation, and definitive care are incorporated into a medically-approved patient care protocol.
Use of the device in the Semiautomatic mode for defibrillation is indicated on victims of cardiac arrest where there is apparent lack of circulation.
### Intended Use - CPR Monitoring
The CPR monitoring function provides visual and audio feedback designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of 1.5 to 2 inches (3.8 to 5.0 cm) for adult patients.
The CPR monitoring function is not intended for use on patients under 8 years of age.
## External Pacemaker (Pacer Version Only)
### Intended Use — Pacemaker
This product may be used for temporary external cardiac pacing in conscious or unconscious patients as an alternative to endocardial stimulation.
Note: This device must not be connected to internal pacemaker electrodes.
The purposes of pacing include:
### Resuscitation from standstill or bradycardia of any etiology:
Noninvasive pacing has been used for resuscitation from cardiac standstill, reflex vagal standstill, drug induced standstill (due to procainamide, guinidine, digitalis, b- blockers, verapamil, etc.) and unexpected circulatory arrest (due to anesthesia, surgery, angiography, and other therapeutic or diagnostic procedures). It has also been used for temporary acceleration of bradycardia in Stokes-Adams disease and sick-sinus syndrome. It is safer, more reliable, and more rapidly applied in an emergency than endocardial or other temporary electrodes.
### As a standby when standstill or bradycardia might be expected:
Noninvasive pacing may be useful as a standby when cardiac arrest or symptomatic bradycardia might be expected due to acute myocardial infarction, drug toxicity, anesthesia or surgery. It is also useful as a temporary treatment in patients awaiting pacemaker implants or the introduction of transvenous therapy. In standby pacing applications, noninvasive pacing may provide an alternative to transvenous that avoids the risks of displacement, infection, hemorrhage, embolization, perforation, phlebitis and mechanical or electrical stimulation of ventricular tachycardia or fibrillation associated with endocardial pacing.
### Suppression of tachycardia:
Increased heart rates in response to external pacing often suppress ventricular ectopic activity and may prevent tachycardia.
### Pediatric Pacing
Pacing can be performed on pediatric patients weighing 33lbs / 15kg or less using special ZOLL pediatric MFE Pads. Prolonged pacing (in excess of 30 minutes), particularly in neonates, could cause burns. Periodic inspection of the underlying skin is recommended.
### (continued on next page)
{8}------------------------------------------------
### Monitor
### Intended-Use Multi-parameter Monitoring
This product may be used for monitoring various patient vital signs, including: electrocardiogram (ECG), Pulse Oximetry (SpO2), End Tidal CO2, 12-Lead ECG, and Non-Invasive Blood Pressure (NIBP).
ECG monitoring is performed by connecting the patient to the 3 or 5 lead patient cable, MFE Pads, or through the paddles.
SpO2 monitoring is indicated for detecting arterial oxygen saturation of blood and pulse rate for adult, pediatric and neonatal patients who are well or poorly perfusing, during both no motion and patient motion conditions.
EtCO2 monitoring is indicated for the continuous measurement of end tidal carbon dioxide (ECO2) and respiration rate for adult, pediatric and neonatal patients.
12 Lead ECG analysis is indicated for the diagnosis and treatment of adult and pediatric patients with acute myocardial infarction or other cardiac arrhythmias.
NIBP monitoring is indicated for the measurement of arterial blood pressure for resting adult, pediatric, and neonatal patients.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.