HUNTLEIGH FP5000 SYSTEM MODEL FP5000

K965153 · Huntleigh Healthcare, Inc. · JOW · Jan 26, 1998 · Cardiovascular

Device Facts

Record IDK965153
Device NameHUNTLEIGH FP5000 SYSTEM MODEL FP5000
ApplicantHuntleigh Healthcare, Inc.
Product CodeJOW · Cardiovascular
Decision DateJan 26, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

1. Prevention of deep vein thrombosis 2. Enhancement of venous & arterial circulation 3. Prevention of venous stasis 4. Assist healing of cutaneous ulcers 5. Reduction of acute or chronic edema 6. Reduction of lower limb pain due to surgery or trauma 7. Reduction of compartmental pressures

Device Story

External pneumatic compression system; consists of pump and pair of foot wrap garments. Pump supplies compressed air to garments; garments inflate to compress venous plantar plexus of foot. Compression moves blood out of leg. Used in clinical settings to prevent DVT and treat venous/arterial conditions of lower limbs. Benefits patient by improving venous outflow velocity, reducing edema, and assisting in ulcer healing.

Clinical Evidence

Bench testing only. Laboratory evaluation on healthy subjects compared venous outflow velocity at the common femoral vein between the FP5000 System and predicate devices (AV5000 and PlexiPulse). Results documented comparable performance.

Technological Characteristics

External pneumatic compression system. Components: pump and foot wrap garments. Energy source: pneumatic (compressed air). Compliance: UL listing, European Medical Directive for electromagnetic compatibility.

Indications for Use

Indicated for patients requiring prevention of deep vein thrombosis, enhancement of venous/arterial circulation, prevention of venous stasis, assistance in healing cutaneous ulcers, reduction of acute/chronic edema, reduction of lower limb pain due to surgery/trauma, and reduction of compartmental pressures. Prescription use only.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} K965153 # 510(k) Summary ## FP5000 System JAN 26 1998 The FP5000 System is an external pneumatic compression system comprised of a pump which supplies compressed air to a pair of foot wrap garments. The garments alternately inflate to provide compression to the venous plantar plexus of the foot. Venous plantar plexus compression move blood out of the leg, and may be used to prevent deep vein thrombosis (DVT), and as part of the treatment regimen for venous and arterial conditions of the lower limbs. The system is designed to provide therapy comparable to existing marketed devices such as the AV5000 system (Kendall) and the PlexiPulse system (NuTech/KCI). There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. Laboratory evaluation on healthy subjects documented that the venous outflow velocity, as measured at the common femoral vein, is comparable between the FP5000 System and the two predicate systems. There are no performance standards established for this category of medical device (Class II - Compressible Limb Sleeves). Voluntary standards met include UL listing and compliance with the European Medical Directive for electromagnetic compatibility. The FP5000 System raises no new safety or effectiveness concerns. 000010 {1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Audrey Witko Director of Corporate Affairs Huntleigh Healthcare Inc. 227 Route 33 East Manalapan, NJ 07726 Re: K965153 Huntleigh FP5000 System Model FP5000 Regulatory Class: II (Two) Product Code: JOW Dated: October 30, 1997 Received: October 31, 1997 Dear Ms. Witko: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. JAN 26 1998 {2} Page 2 - Ms. Audrey Witko This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html." Sincerely yours, ![img-0.jpeg](img-0.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 510(k) Number (if known): K965153 Device Name: Huntleigh FP5000 System ## Indications For Use: 1. Prevention of deep vein thrombosis 2. Enhancement of venous & arterial circulation 3. Prevention of venous stasis 4. Assist healing of cutaneous ulcers 5. Reduction of acute or chronic edema 6. Reduction of lower limb pain due to surgery or trauma 7. Reduction of compartmental pressures (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number K965153 Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) 000005
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...