K193068 · Airos Medical, Inc. · JOW · Dec 4, 2019 · Cardiovascular
Device Facts
Record ID
K193068
Device Name
AIROS 8 Sequential Compression Device
Applicant
Airos Medical, Inc.
Product Code
JOW · Cardiovascular
Decision Date
Dec 4, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AIROS 8 Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use.
Device Story
AIROS 8 is a gradient pneumatic compression system for treating venous or lymphatic disorders. It consists of a digitally-controlled pump unit and 8-chambered garments. The device generates compressed air, delivered via tubing to the garment, which inflates chambers sequentially from distal to proximal. This cycle increases blood flow and encourages extracellular fluid clearance. Used in home or hospital settings, the device is operated by patients or clinicians based on a physician's prescription. The provider sets therapy parameters; the device applies pressure cyclically for specified durations. By facilitating fluid movement, the device aids in managing edema and venous conditions.
Clinical Evidence
Bench testing only. Performance testing included alarm, LED/LCD, cycle time, pressure accuracy, therapy time, therapeutic performance, garment integrity, pull, and garment printing tests. Conformance to ISO 14971 (risk management) and ISO 10993 (biocompatibility: cytotoxicity, irritation, and hypersensitivity) was demonstrated.
Technological Characteristics
Digitally-controlled pneumatic compression system. Components: electrically generated compressed air source, tubing, 8-chambered garments. Operates via cyclic pressure application. Standards: ISO 14971:2007, ISO 10993-5, ISO 10993-10. Connectivity: not specified.
Indications for Use
Indicated for patients with lymphedema, venous stasis ulcers, venous insufficiency, or peripheral edema.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
K172779 — AIROS 8 Sequential Compression Device · Airos Medical, Inc. · Jun 22, 2018
K223193 — AIROS 8P Sequential Compression Device · Airos Medical, Inc. · Dec 9, 2022
K223195 — AIROS 6P Sequential Compression Device · Airos Medical, Inc. · Dec 9, 2022
K172770 — AIROS 6 Sequential Compression Device · Airos Medical, Inc. · Jun 22, 2018
K193069 — AIROS 6 Sequential Compression Device · Airos Medical, Inc. · Dec 4, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2019
AIROS Medical, Inc. Darren Behuniak Vice President, Operations and Marketing 2501 Monroe Blvd. Suite 1200 Audubon, Pennsylvania 19403
Re: K193068
Trade/Device Name: AIROS 8 Sequential Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: November 1, 2019 Received: November 4, 2019
Dear Darren Behuniak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K193068
Device Name AIROS 8 Sequential Compression Device
Indications for Use (Describe)
The AIROS 8 Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions:
- Lymphedema
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a stylized "A" inside, and the words "AIROS MEDICAL" to the right of the hexagon. The "A" inside the hexagon is green, while the hexagon itself is gray. The words "AIROS MEDICAL" are also gray, with "AIROS" in a larger font than "MEDICAL".
# 510(k) SUMMARY
#### November 1, 2019
# Submitter:
AIROS Medical, Inc. 2501 Monroe Blvd, Suite 1200 Audubon, PA 19403
### Contact Person:
Darren Behuniak, VP Operations & Marketing Email: dbehuniak@airosmedical.com Phone: 866-991-6956
#### Common Classification & Proprietary Names:
Trade Name: AIROS 8 Sequential Compression Device Common Names: Sequential Compression Device
#### Classification
The classification name, 21 CFR Part and Paragraph number, product code and classification of the AIROS 8 Sequential Compression Device.
| Classification Name | 21 CFR<br>Section | Product<br>Code | Class |
|-----------------------------|-------------------|-----------------|-------|
| Compressible Limb<br>Sleeve | 870.5800 | JOW | II |
### Predicate Device:
The AIROS 8 Sequential Compression Device is substantially equivalent to the following.
| Predicate Device | Manufacturer | 510(k)# |
|---------------------------------------|---------------------|---------|
| AIROS 8 Sequential Compression Device | AIROS Medical, Inc. | K172779 |
www.airosmedical.com 1.866.991.6956
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon shape with a green "A" inside, and the words "AIROS" and "MEDICAL" to the right of the hexagon. The word "AIROS" is in a larger font than the word "MEDICAL".
# Device Description
The AIROS 8 Sequential Compression Device is a gradient pneumatic compression device. The device is used for treatment and management of venous or lymphatic disorders. The application of gradient sequential compression increases blood flow and encourages extracellular fluid clearance.
The AIROS 8 system consists of the device and 8-chambered garments. The device provides cycles of compressed air and sequentially inflates the garments from distal to proximal.
# Intended Use:
The AIROS & Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions:
- Lymphedema ●
- Venous stasis ulcers
- Venous insufficiency
- Peripheral edema
The device is safe for both home and hospital use.
#### Technological Characteristics:
The manufacturer believes that the technological characteristics of the AIROS 8 are substantially similar to those of the predicate device.
The AIROS 8 has similar components to its predicate device and has similar operating principles. The digitally-controlled device consists of an electrically generated source of compressed air, tubing to convey the pressurized air to the sleeve, and like the predicate, pressure is applied cyclically for a specified period of time, according to the physician's prescription.
ww.airosmedical.com 866.991.6956
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with a stylized "A" inside, with the word "AIROS" to the right of the hexagon. Below the word "AIROS" is the word "MEDICAL" in a smaller font. The hexagon is gray, and the "A" is green.
# Functional Performance Testing
Testing was performed and to ensure that the system meets its specifications. The manufacturer believes that the technological characteristics of the AIROS 8 are substantially equivalent to those of the predicate device. The functional performance testing includes the following tests:
| Test Description |
|---------------------------------|
| Alarm Testing |
| LED/LCD Testing |
| Cycle Time Testing |
| Pressure Accuracy Testing |
| Therapy Time Testing |
| Therapeutic Performance Testing |
| Garment Integrity Testing |
| Pull Testing |
| Garment Printing Testing |
# Standards
The AIROS 8 conforms to the following standards:
- ISO 14971:2007 Medical Devices Application of Risk Management to Medical Devices ●
- ISO 10993-5 Biological evaluation of medical devices Part 1: Biological evaluation of ● medical devices -- Part 5: Tests for in vitro cytotoxicity
- o ISO 10993-10 Biological evaluation of medical devices -Part 10:Tests for irritation and delayed-type hypersensitivity
### Statement of Substantial Equivalence
The AIROS 8 is substantially equivalent in technology, function, operating parameters, and indicated use to the predicate device that is currently commercially available and in distribution. The system does not raise any new risks when compared to the predicate product.
### Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, AIROS Medical, Inc., believes that the AIROS 8 is substantially equivalent to the predicate device.
w.airosmedical.com
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.