AIROS 6P Sequential Compression Device

K223195 · Airos Medical, Inc. · JOW · Dec 9, 2022 · Cardiovascular

Device Facts

Record IDK223195
Device NameAIROS 6P Sequential Compression Device
ApplicantAiros Medical, Inc.
Product CodeJOW · Cardiovascular
Decision DateDec 9, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AIROS 6P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use.

Device Story

AIROS 6P is a pneumatic compression device for managing venous or lymphatic disorders. It consists of a control unit and inflatable compression garments for upper/lower extremities and trunk. The device supplies cycles of pressurized air to garments, inflating them from distal to proximal areas to increase blood flow and clear extracellular fluid. It features two treatment modes: Gradient and Peristaltic. Operated by patients or clinicians in home or hospital settings, the device is digitally controlled to manage treatment options, pressure, and sequences per physician prescription. By providing cyclic compression, it aids in edema reduction and ulcer management, potentially improving patient outcomes in lymphatic and venous conditions.

Clinical Evidence

Bench testing only. Functional performance testing included error indicator, noise, cycle time accuracy, pressure time accuracy, therapeutic performance, therapy time accuracy, and garment integrity testing.

Technological Characteristics

Digitally controlled pneumatic compression device. Electrically generated compressed air source with tubing to sleeves. Gradient and Peristaltic operating modes. Cyclic pressure application. Complies with ISO 14971, ISO 10993-1, ES 60601-1, IEC 60601-1-2, IEC 61000-3-2, IEC 61000-3-3, IEC 60601-1-6, and IEC 60601-1-11.

Indications for Use

Indicated for patients with lymphedema, venous stasis ulcers, venous insufficiency, or peripheral edema. Safe for use in home and hospital settings.

Regulatory Classification

Identification

A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text. December 9, 2022 AIROS Medical, Inc Darren Behuniak Vice President, Operations and Marketing 2501 Monroe Blvd. Suite 1200 Audubon, Pennsylvania 19403 Re: K223195 Trade/Device Name: AIROS 6P Sequential Compression Device Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible limb sleeve Regulatory Class: Class II Product Code: JOW Dated: October 12, 2022 Received: October 13, 2022 Dear Darren Behuniak: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Eric E. Richardson --S 2022.12.09 16:20:07 -05'00' Nicole Gillette for Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K223195 Device Name AIROS 6P Sequential Compression Device Indications for Use (Describe) The AIROS 6P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a logo for AIROS Medical. The logo consists of a gray hexagon with a green letter "A" inside of it. To the right of the hexagon is the word "AIROS" in gray, with the word "MEDICAL" underneath it in a smaller font. # 510(k) Summary October 12, 2022 ### Submitter: AIROS Medical, Inc. 2501 Monroe Blvd, Suite 1200 Audubon, PA 19403 # Contact Person: Darren Behuniak, VP Operations & Marketing Email: dbehuniak(@airosmedical.com Phone: 866-991-6956 ### Common Classification & Proprietary Names Trade Name: AIROS 6P Sequential Compression Device Common Name: Sequential Compression Device #### Classification | Classification Name | 21 CFR Regulation | Product Code | Class | |-----------------------------|-------------------|--------------|-------| | Compressible Limb<br>Sleeve | 870.5800 | JOW | II | #### Primary Predicate Device and Reference Device The AIROS 6P Sequential Compression Device is substantially equivalent to the primary predicate. AIROS 6 Sequential Device, and the reference device, Mego Afek Lympha Press Optimal. | | Primary Predicate | Reference Device | |--------------------------|------------------------------------------|-----------------------------------| | | AIROS 6 Sequential<br>Compression Device | Mego Afek Lympha Press<br>Optimal | | 510(k) | K193069 | K182003 | | 21 CFR Regulation Number | 870.5800 | 870.5800 | | Product Code | JOW | JOW | | Classification | II | II | # Device Description AIROS 6P Sequential Compression Device is a pneumatic compression device used for treatment and management of venous or lymphatic disorders. The application of compression is effective by increasing blood flow and encouraging extracellular fluid clearance. The device consists of the mechanical device that is used to set the treatment options and supplies cycles of air to the compression garments. There are two primary treatment modes, Gradient Mode and Peristaltic Mode. The air is supplied at adjustable pressures and sequences and inflates the www.airosmedical.com 1.866.991.6956 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a gray hexagon with a green letter A inside of it. To the right of the hexagon is the word "AIROS" in gray, with the word "MEDICAL" underneath it. compression garments from the distal to proximal areas of the body. The compression garments are supplied in various sizes for the upper and lower extremity areas of the body. # Indication for Use The AIROS 6P Sequential Compression Device utilizes gradient pneumatic compression, which is intended for treatment of patients with the following conditions: - Lymphedema - Venous stasis ulcers - Venous insufficiency - Peripheral edema The device is safe for both home and hospital use. #### Technological Characteristics The manufacturer believes that the technological characteristics of the AIROS 6P are substantially equivalent to those of the primary predicate and reference devices. The AIROS 6P has the same therapy modes as the AIROS 6 predicate with the exception that the AIROS 6P also includes the Peristaltic mode which is substantially equivalent to the reference device, Mego Afek Lympha Press Optimal. There have also been some component changes with the 60 that do not affect safety or efficacy of the device. The digitally controlled device consists of an electrically generated source of compressed air, tubing to convey the pressurized air to the sleeves, and like the primary predicate and reference devices, pressure is applied cyclically for a specified period of time, according to the physician's prescription. Two primary operating modes, Gradient and Peristaltic are available to the user and multiple size garments are available for the upper extremities, lower extremities, and trunk of the body. #### Functional Performance Testing Testing was performed and to ensure that the system meets its specifications. The manufacturer believes that the technological characteristics of the AIROS 6P are substantially equivalent to those of the primary predicate and reference devices. The functional performance testing includes the following tests: - Error Indicator Testing ● - Noise Testing - Cycle Time Accuracy ● - Pressure Time Accuracy - Therapeutic Performance - Therapy Time Accuracy - . Garment Integrity #### Standards www.airosmedical.com 1.866.991.6956 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for AIROS Medical. The logo consists of a hexagon with the letter A inside, and the words "AIROS MEDICAL" to the right of the hexagon. The hexagon is gray, the letter A is green, and the words "AIROS MEDICAL" are gray. - ISO 14971:2019 Medical Devices Application of Risk Management to Medical . Devices - ISO 10993-1 2018 Biological evaluation of medical devices Part 1: Evaluation and ● testing within a risk management process - ISTA 3A:2018 Packaged Products for Parcel Delivery System Shipment ● - ES 60601-1: 2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 -● Medical electrical equipment- Part 1: General requirements for basic safety and essential performance - . IEC 60601-1-2: 2014 + AMD1:2020 CSV - Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral Standard: Electromagnetic disturbances- Requirements and tests - IEC 61000-3-2: 2018 A1: 2020 EMC Part 3-2: Limits Limits for harmonic current . emissions (Equipment input current 16A per phase) - IEC 61000-3-3: 2013 A1: 2017 EMC - Part 3-3: Limits - Limitation of voltage changes, voltage fluctuations and flicker in public low-voltage supply systems for equipment with rated current ≤ 16 A per phase and not subject to conditional connections - IEC 60601-1-6: 2010/AMD1: 2013 Medical electrical equipment- Part 1-6: General requirements for basic safety and essential performance- Collateral Standard: Usability - IEC 60601-1-11: 2015 Medical electrical equipment- Part 1-11: General requirements . for basic safety and essential performance- Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment # Statement of Substantial Equivalence The AIROS 6P is substantially equivalent in technology, function, operating parameters, and indications to the AIROS 6P primary predicate device and the Lympha Press Optimal reference device. There are no new risks introduced with the AIROS 6P. # Conclusion In accordance with the Federal Food, Drug and Cosmetic Act and 21 CFR Part 807, and based on the information provided in this pre-market notification, AIROS Medical, Inc., believes that the AIROS 6P is substantially equivalent to the primary predicate and reference devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...